FMS VUE™ Fluid Management and Tissue Debridement System
K233675 · DePuy Mitek, Inc. · HRX · Feb 26, 2024 · Orthopedic
Device Facts
Record ID
K233675
Device Name
FMS VUE™ Fluid Management and Tissue Debridement System
Applicant
DePuy Mitek, Inc.
Product Code
HRX · Orthopedic
Decision Date
Feb 26, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.1100
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The FMS VUE™ Fluid Management and Tissue Debridement System is intended to provide controlled fluid distention and suction, controlled cutting, shaving and abrading of bone and tissue for use in arthroscopic surgery of the shoulder, knee, ankle, elbow, wrist and hip joints.
Device Story
Arthroscopic pump system; provides controlled fluid distention and suction; integrates shaver console for cutting, burring, shaving, abrading bone/soft tissue. Used in OR by surgeons during arthroscopic procedures. System comprises irrigation pump (joint pressure control), suction pump (saline/waste flow control), and shaver console. Accessories include foot pedal, remote hand control, shaver handpieces, tubing sets, blades, burrs. Software/microelectronics execute control algorithms to maintain set pump pressure. Hardware updates include component replacements and new rear panel connectors; software update adds shaver oscillation profile algorithm. Benefits include optimal surgical field visibility and controlled tissue debridement.
Clinical Evidence
No clinical data. Bench testing only. Performance verified via functional testing (pressure transducer, pressure security, DUO/SOLO mode performance, RFID, accessory compatibility, backflow prevention, microbial/viral ingress) and software V&V.
Technological Characteristics
System includes console, irrigation/suction peristaltic pumps, tubing pinch valve, display panel. Operates in DUO (inflow/outflow) or SOLO (inflow) modes. Software-controlled microelectronics. Complies with IEC 60601-1 (safety) and IEC 60601-1-2 (EMC). Software development conforms to IEC 62304.
Indications for Use
Indicated for patients undergoing arthroscopic surgery of the shoulder, knee, ankle, elbow, wrist, or hip joints requiring controlled fluid distention, suction, and tissue/bone debridement. Contraindicated for non-arthroscopic procedures, use with gas distension media, or patients where arthroscopy is contraindicated.
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
Predicate Devices
FMS VUE Fluid Management and Tissue Debridement System (K130169)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
February 26, 2024
DePuy Mitek, Inc. Pol Fort Grebol Project Manager Regulatory Affairs 325 Paramount Drive Raynham, Massachusetts 02767
Re: K233675
Trade/Device Name: FMS VUE™ Fluid Management and Tissue Debridement System Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: November 16, 2023 Received: January 23, 2024
Dear Pol Fort Grebol:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jesse Muir -S Digitally signed by Jesse Muir
Jesse Muir, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
Enclosure
{3}------------------------------------------------
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 07/31/2026 See PRA Statement below.
Submission Number (if known)
K233675
Device Name
FMS VUE™ Fluid Management and Tissue Debridement System
Indications for Use (Describe)
The FMS VUE™ Fluid Management and Tissue Debridement System is intended to provide controlled fluid distention and suction, controlled cutting, shaving and abrading of bone and tissue for use in arthroscopic surgery of the shoulder, knee, ankle, elbow, wrist and hip joints.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/2 description: The image shows the logo for DePuy Synthes, which is a company that specializes in orthopedics. The logo features a red symbol on the left, followed by the company name in gray. Below the company name, it says "THE ORTHOPAEDICS COMPANY OF Johnson & Johnson" in a smaller font. The words "Johnson & Johnson" are in red.
## 510(k) Summary
| Submitter's Name<br>and Address | DePuy Mitek, Inc.<br>325 Paramount Drive<br>Raynham, MA 02767, USA |
|---------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Pol Fort Grèbol<br>Project Manager Regulatory Affairs<br>DePuy Synthes<br>a Johnson & Johnson company<br>Phone: +34 672 672 579<br>Email: pfortgre@its.jnj.com |
| Date Prepared | February 22, 2023 |
| Proprietary Name | FMS VUE™ Fluid Management and Tissue Debridement System |
| Common or Usual<br>Name | Arthroscopic Pump, Tubing Sets and Accessories |
| Regulation Number | 21 CFR § 888.1100 (Arthroscope) |
| Regulatory Class | Class II |
| Product Code | HRX |
| Predicate Device | The FMS VUE Fluid Management and Tissue Debridement System<br>(K130169)<br>No reference devices were used in this submission. |
| Device Description<br>Summary | The FMS VUE Fluid Management and Tissue Debridement System is an<br>arthroscopic pump system designed to provide optimal visibility of the<br>surgical field by computer-controlled fluid regulation of intra-articular<br>pressure, and flow during arthroscopic procedures. The system integrates a<br>tissue debridement for controlled cutting, burring, shaving and abrading of<br>bone and soft tissue.<br>The FMS VUE Fluid Management and Tissue Debridement System consists<br>of irrigation pump, which controls joint pressure, and the suction pump, which<br>controls the flow of saline and waste from the joint.<br>The FMS VUE Fluid Management and Tissue Debridement System includes<br>an integrated shaver console, intended to provide controlled cutting, burring,<br>shaving and abrading of bone and soft tissue.<br>The FMS VUE Fluid Management and Tissue Debridement System is<br>designed to work with the existing FMS VUE platform accessories including:<br>foot pedal, remote hand control, shaver handpieces, FMS connect cable,<br>tubing sets, blades, and burrs. |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a logo for DePuy Synthes, a company that specializes in orthopedics. The logo features a red symbol on the left, followed by the company name in gray. Below the company name, it says "THE ORTHOPAEDICS COMPANY OF Johnson & Johnson" in a smaller font. The symbol on the left is a stylized representation of a bone or joint.
| The FMS VUE Fluid Management and Tissue Debridement System contains updates in the hardware, which include replaced near obsolescence components and new real panel connectors.<br>The FMS VUE Fluid Management and Tissue Debridement System software modification includes an additional shaver oscillation profile algorithm. | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use /<br>Indications for Use | The FMS VUE Fluid Management and Tissue Debridement System is intended to provide controlled fluid distention and suction, controlled cutting, burring, shaving and abrading of bone and tissue for use in arthroscopic surgery of the shoulder, knee, ankle, elbow, wrist and hip joints. | | |
| Indications for Use<br>Comparison | The proposed device has the same indications for use as the predicate device. | | |
| Comparison of<br>Technological<br>Characteristics with<br>the Predicate Device | Device comparison demonstrated that the subject FMS VUE Fluid Management and Tissue Debridement System is substantially equivalent to the previously cleared FMS VUE Fluid Management and Tissue Debridement System (K130169) regarding intended use, indications for use, technological characteristics (design features and performance) as well as operating principle. | | |
| | Characteristic | Subject Device<br>FMS VUE Fluid<br>Management and<br>Tissue Debridement<br>System | Predicate Device<br>FMS VUE Fluid<br>Management and<br>Tissue Debridement<br>System (K130169) |
| | Product Code | Same as the predicate device. | HRX |
| | Intended<br>Use/Indications for Use<br>Statement | Same as the predicate device. | The FMS VUE Fluid<br>Management and<br>Tissue Debridement<br>System is intended to<br>provide controlled fluid<br>distention and suction,<br>controlled cutting,<br>burring, shaving and<br>abrading of bone and<br>tissue for use in<br>arthroscopic surgery of<br>the shoulder, knee,<br>ankle, elbow, wrist and<br>hip joints. |
| | Contraindications | Same as the predicate device. | The FMS VUE System<br>is contraindicated in<br>any non-arthroscopic<br>surgical procedure and<br>is for use with saline<br>irrigation only. The<br>system is not |
| | | appropriate for patients<br>for whom an<br>arthroscopic procedure<br>is contraindicated for<br>any reason. | |
| | | • DO NOT use the<br>FMS VUE in<br>instances where<br>capsular integrity is<br>suspect. | |
| | | • DO NOT use the<br>FMS VUE with a<br>gas distension<br>medium. Use only<br>with sterile<br>irrigation solution<br>for distension of<br>the operative site. | |
| | | • DO NOT use the<br>FMS VUE to<br>administer drugs,<br>IV-fluids, blood or<br>blood substitutes. | |
| Principles of Operation | Same as the predicate<br>device. | The device is<br>controlled by software<br>and microelectronics<br>that execute control<br>algorithms that ensure<br>the set pump pressure<br>is being achieved. | |
| Energy Source | Same as the predicate<br>device. | Supply voltage<br>provided to the console<br>is that of a typical<br>business or hospital<br>environment | |
| Pump Mechanism | Same as the predicate<br>device. | Two peristaltic pumps:<br>1 inflow and 1 outflow | |
| Functional Mode | Same as the predicate<br>device. | • DUO Mode<br>(Inflow/Outflow) | |
| | | • SOLO Mode<br>(Inflow only) | |
| Roller Wheel | Same as the predicate<br>device. | The device has<br>internally mounted<br>roller wheels that<br>interface with<br>disposable tubing sets. | |
| Non-clinical and<br>Clinical<br>Performance Data | Design | Same as the predicate device. | Pump consists of the following main components:<br>Console housing Power supply Two peristaltic pumps Tubing pinch valve Display panel |
| | Software Development | Same as the predicate device. | IEC 62304 |
| | Electrical Safety and<br>Electromagnetic<br>Compatibility | Same as the predicate device. | IEC 60601-1<br>IEC 60601-1-2 |
| | The core functionality of the predicate FMS VUE Fluid Management and Tissue Debridement System K130169, as cleared per K130169, remains unchanged. Similarly, the indications for use and intended use remain unchanged.<br>The changes described in this premarket notification are related to hardware and software upgrades. | | |
| | Design verification and validation testing were performed on the subject FMS VUE Fluid Management and Tissue Debridement System as a result of the risk analysis and product requirements.<br>Proposed FMS VUE Fluid Management and Tissue Debridement System testing included functional performance, software verification and validation and testing per IEC 60601-1 and IEC 60601-1-2 standards. Results of functional performance testing were compared to the predicate device and have demonstrated that the proposed device is suitable for its intended use. | | |
| | <b>Functional Performance Testing</b><br>Functional performance testing was conducted to ensure the device design meets user needs in the environment of use: | | |
| | Pressure Transducer Test Pressure Security Test (Low Pressure Fault Checks and High Pressure Fault Checks) DUO (Inflow/Outflow) Performance SOLO (Inflow) Performance RFID Capability Functional Testing with existing electrical accessories: Foot Pedal Remote Hand Control | | |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for DePuy Synthes, which is a company that specializes in orthopedics. The logo consists of a red symbol on the left, followed by the company name in gray text. Below the company name, there is a tagline that reads "THE ORTHOPAEDICS COMPANY OF Johnson & Johnson" in a smaller font size.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows the DePuy Synthes logo. The logo features a red abstract symbol on the left, followed by the words "DePuy Synthes" in gray. Below the company name, it says "THE ORTHOPAEDICS COMPANY OF Johnson & Johnson" in a smaller font size, with "Johnson & Johnson" in red.
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for DePuy Synthes, a company that specializes in orthopedics. The logo features a stylized red symbol on the left, followed by the company name in gray. Below the company name, it says "THE ORTHOPAEDICS COMPANY OF Johnson-Johnson" in a smaller font.
| | Ο Shaver Handpieces Ο FMS Connect Cable • Backflow-Prevention Valve Testing (Dye Leak Testing) • Microbial and Viral Ingress Testing |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <b>Software Verification and Validation</b> |
| | Software verification and validation testing was conducted, and documentation was provided as recommended by Guidance for Industry and FDA Staff "Content of Premarket Submission for Device Software Functions" issued on June 14, 2023. Based on the cited FDA guidance document, the most appropriate documentation level for the subject FMS VUE software is Basic Documentation. The software development process for the subject FMS VUE Fluid Management and Tissue Debridement System software conforms to all clauses of IEC 62304 standard. |
| | <b>Electrical Safety and Electromagnetic Compatibility (EMC)</b> |
| | Electrical safety and EMC testing were conducted on the predicate FMS VUE Fluid Management and Tissue Debridement System and results provided as part of the original 510(k) submission, K130169. The subject FMS VUE Fluid Management and Tissue Debridement System complies with the IEC 60601-1 standard for safety and the IEC 60601-1-2 standard for EMC. |
| | <b>Animal Study</b> |
| | No animal testing was conducted to determine the safety and effectiveness of the FMS VUE Fluid Management and Tissue Debridement System. |
| | <b>Clinical Studies</b> |
| | No human clinical testing was conducted to determine the safety and effectiveness of the FMS VUE Fluid Management and Tissue Debridement System. |
| <b>Substantial<br/>Equivalence</b> | The determination of substantial equivalence for this device was based on a detailed device description, performance data, and conformance to consensus and voluntary standards.<br>The testing included in this submission demonstrates that: Any differences in technological characteristics of the predicate devices do not raise any new questions of safety or effectiveness. The proposed devices are at least as safe and effective as the predicate devices. |
| <b>Conclusion</b> | The intended use, design principles, and fundamental technology of the subject FMS VUE Fluid Management and Tissue Debridement System remains unchanged and therefore is substantially equivalent to the predicate device – the FMS VUE Fluid Management and Tissue Debridement System (K130169). The proposed hardware and software modifications do not raise new questions of safety or effectiveness of the subject FMS VUE Fluid Management and Tissue Debridement System. The results of the non-clinical |
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows the logo for DePuy Synthes, a company that specializes in orthopedics. The logo consists of a red graphic to the left of the company name, which is written in gray. Below the company name is the text "THE ORTHOPAEDICS COMPANY OF Johnson & Johnson."
| demonstrates that proposed FMS VUE performs as intended in the specified |
|--------------------------------------------------------------------------|
| use conditions. |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.