DePuy Mitek FMS VUE Fluid Management and Tissue Debridement System is intended to provide controlled fluid distention and suction, controlled cutting, burring, shaving and abrading of bone and tissue for use in arthroscopic surgery of the shoulder, knee, ankle, elbow, wrist and hip joints. The FMS CONNECT Interface Cables are accessories to the DePuy Mitek FMS Fluid Management and Tissue Debridement Systems; they connect competitive shavers to FMS arthroscopy pumps.
Device Story
FMS VUE is a fluid management and tissue debridement system for arthroscopic surgery. System includes a pump, tube sets, and optional accessories (foot pedal, remote control, handpiece, interface cable). Device uses two peristaltic roller pumps to provide irrigation (inflow) and aspiration (outflow) to joint cavities. Software-controlled pumps automatically manage fluid, intra-articular pressure, and shaver/cannula suction based on surgeon-adjusted settings (mmHg). System supports two modes: DUO (irrigation, shaver control, suction) and SOLO (irrigation, shaver control). Integrated debridement function enables controlled shaving and burring of tissue and bone. FMS Connect Interface Cable allows integration of competitive shaver systems by sending signals to the FMS VUE pump. Used in clinical settings by surgeons to maintain joint distention and perform tissue resection, potentially improving surgical visualization and procedural efficiency.
Clinical Evidence
No clinical studies were required. Safety and efficacy were demonstrated through bench testing, including electrical, software, and performance verification, and validation testing to ensure user needs were met.
Technological Characteristics
System utilizes two peristaltic roller pumps for fluid management. Features include software-controlled pressure and flow regulation, integrated debridement (shaving/burring), and interface cables for competitive shavers. Complies with EN 60601-1 (safety), EN 60601-1-2 (EMC), and IEC 62304 (software lifecycle).
Indications for Use
Indicated for patients undergoing arthroscopic surgery of the shoulder, knee, ankle, elbow, wrist, or hip joints requiring fluid distention, suction, or tissue/bone debridement (cutting, burring, shaving, abrading).
Regulatory Classification
Identification
An arthroscope is an electrically powered endoscope intended to make visible the interior of a joint. The arthroscope and accessories also is intended to perform surgery within a joint.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for DePuy Mitek, a Johnson & Johnson company. The logo features a black circle followed by the company name in bold, stylized font. Below the company name is the text "a Johnson & Johnson company" in a smaller font. The text "K130169" is located below the company information.
510(k) SUMMARY
MAY 0 1 2013
## Date Prepared January 22, 2013
| Submitter's Name and Address: | DePuy Mitek, Inc.<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767 |
|-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person | Susan Kagan<br>Project Manager, Regulatory Affairs<br>DePuy Mitek, Inc.<br>a Johnson & Johnson company<br>325 Paramount Drive<br>Raynham, MA 02767, USA<br>Telephone: 508-880-8097<br>Facsimile: 508-977-6955<br>e-mail: skagan@its.jnj.com |
| Name of Medical Device | Classification Name: Arthroscope: 21 CFR 888.1100<br>Common: Pump<br>Trade Name: FMS VUE Fluid Management & Tissue Debridement System<br>FMS Connect Interface Cable<br>FDA Classification: II<br>FDA product code: HRX |
| Predicate Device(s) | The proposed FMS VUE Fluid Management & Tissue Debridement System, and FMS Connect Interface Cable are substantially equivalent to:<br>• K954465 FMS DUO (November 9, 1995)<br>• K002040 FMS SOLO (September 11, 2000) |
| Device Description | The FMS VUE is DePuy Mitek's next generation of Fluid Management and Tissue Debridement System. The system consists of a pump, tube sets, and the following optional accessories Foot Pedal, Remote Control, Handpiece and FMS Connect Interface Cable. |
.
{1}------------------------------------------------
2002 : 0 YAM
As with its predicate devices, FMS DUO+ and FMS SOLO, this system utilizes pump technology designed to provide a range of arthroscopic surgical treatments including soft tissue ablation, contouring, cutting and coagulation and temperature control.
The FMS VUE provides irrigation (inflow) and aspiration (outflow) of fluid to and from joint cavities during arthroscopic procedures by way of 2 peristaltic roller pumps. Both roller pumps are software controlled to automatically manage fluid and intra-articular joint pressure based on default settings or adjusted setting chosen by the surgeon. Additionally, the pumps automatically manage shaver and cannula suction. By controlling both flow and pressure setting, the FMS VUE Fluid Management & Tissue Debridement System accurately requlates pressure and flow in the joint.
The FMS VUE can be used in either the DUO mode (default mode) which provides irrigation, shaver control and suction or the SOLO mode which provides irrigation and shaver control but not suction.
Like its predicates, the FMS VUE also incorporates an integrated debridement function for controlled shaving and burring of tissue and bone.
The FMS Connect Interface Cable allows the use of competitive shaver systems to interface with the FMS VUE. When the FMS Connect device is attached to the cable of a competitors hand piece and the hand piece is running, the FMS Connect sends a signal to the FMS VUE.
#### Indications for Use
DePuy Mitek FMS VUE Fluid Management and Tissue Debridement System is intended to provide controlled fluid distention and suction, controlled cutting, burring, shaving and abrading of bone and tissue for use in arthroscopic surgery of the shoulder, knee, ankle, elbow, wrist and hip joints.
The FMS CONNECT Interface Cables are accessories to the DePuy Mitek FMS Fluid Management and Tissue Debridement Systems; they connect competitive shavers to FMS arthroscopy pumps.
{2}------------------------------------------------
### Comparison to Predicate Device
This submission is intended to demonstrate that the FMS VUE and FMS Connect Interface Cable are substantially equivalent to their legally marketed devices: FMS DUO+ (K954465) and FMS SOLO (K002040).
The FMS VUE Fluid Management & Tissue Debridement System and FMS Connect Interface Cable have been carefully compared to legally marketed devices with respect to intended use, essential components and material, performance specifications and technology characteristics. They have the same Indications for Use and technology characteristics.
- Enhancements on the FMS VUE include: .
- o Incorporating DUO+ and SOLO mode functionality in one pump system
- o Measurement displayed in mmHg
- FMS Connect Interface Cable has a different mode of connecting to completive shavers and is powered for a wall source rather that the pump.
#### Safety and Performance
Verification of the FMS VUE Fluid Management & Tissue Debridement System and accessories includes electrical, software and performance tests to show that the device meets its product specifications over a range of operating conditions. Validation testing for the FMS VUE includes testing to show the device meets user needs. Verification testing conform to the following Standards and Guidance documents:
| Standard/<br>Guidance | Description |
|-----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| EN 60601-<br>1:2005 | Medical electrical equipment -- Part 1: General<br>requirements for safety |
| EN60601-1-<br>2:2007 | Medical electrical equipment -- Part 1-2: General<br>requirements for safety - Collateral standard:<br>Electromagnetic compatibility - Requirements and tests. |
| IEC<br>62304:2006 | Medical device software-software life cycle process |
| | Guidance for the Content of Premarket Submission for Software<br>Contained in Medical Devices: 2005 |
| | General Principles of Software Validation; Final Guidance for Industry and<br>FDA Staff: 2002 |
| | Guidance Off-The-Shelf Software Use in Medical Devices: 1999 |
#### TABLE 1 Standards & Guidance Documents
{3}------------------------------------------------
Clinical No clinical studies are required to demonstrate safety and efficacy of Testing the device in support of an application for premarket clearance. The FMS VUE Fluid Management & Tissue Debridement System and FMS Connect Interface Cables do not differ from the predicate device in fundamental scientific technology or intended use.
Safety and performance testing have been executed to validate the Conclusion performance and safety of the devices. It has been demonstrated that these device modifications will not affect safety and effectiveness of the subject devices. Results of performance and safety testing have demonstrated that the modified device is suitable for its intended use.
> Based on the indications for use, fundamental scientific technology, and comparison to the predicate devices, the FMS VUE Fluid Management & Tissue Debridement System, and FMS Connect Interface Cable are shown to be substantially equivalent to the predicate devices under the Federal Food, Drug and Cosmetic Act.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, with three wavy lines representing snakes intertwined around a staff.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
May 1. 2013
DePuy Mitek, Inc. % Ms. Susan Kagan Project Manager, Regulatory Affairs 325 Paramount Drive Raynham, Massachusetts 02767
Re: K130169
Trade/Device Name: FMS VUE Fluid Management & Tissue Debridement System Regulation Number: 21 CFR 888.1100 Regulation Name: Arthroscope Regulatory Class: Class II Product Code: HRX Dated: January 22, 2013 Received: February 07, 2013
Dear Ms. Kagan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 - Ms. Susan Kagan
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours, FOR
# Peter DERRumm -S
Mark N. Melkerson Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### Indications for Use
510(k) Number (if known): K130169
FMS VUE™ Fluid Management and Tissue Debridement System Device Name:
Indications for Use:
DePuy Mitek FMS VUE Fluid Management and Tissue Debridement System is intended to provide controlled fluid distention and suction, controlled cutting, burring, shaving and abrading of bone and tissue for use in arthroscopic surgery of the shoulder, knee, ankle, elbow, wrist and hip joints.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Joshua C.
Nipper -S
(Division Sign-Off)
For
Division of Surgical Devices
510(k) Number_K130169
Page 1 of 1
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): K130169
FMS™ Connect Interface Cable Device Name:
Indications for Use:
The FMS CONNECT Interface Cables are accessories to the DePuy Mitek FMS Fluid Management and Tissue Debridement Systems; they connect competitive shavers to FMS arthroscopy pumps.
Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Joshua C.
Nipper -S
For
(Division Sign-Off)
Division of Surgical Devices
K130169 510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.