Genesys Spine Sacroiliac Joint Fusion System with Navigation
Applicant
Genesys Spine
Product Code
OLO · Neurology
Decision Date
Mar 29, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Genesys Spine Sacroillac Joint Fusion System with Navigation is intended to be used with the Genesys Spine Sacroiliac Joint Fusion System to assist the surgeon in precisely locating anatomical structures in SIros and SIros O procedures, in which the use of stereotactic surgery may be appropriate, and where reference to anatomical structures, such as the pelvis or vertebra, can be identified relative to acquired imaging (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or image data based models of the anatomy. Navigation instruments are intended to be used with the Medtronic O-arm® Imaging System and StealthStation® Navigation System.
Device Story
System consists of navigation-enabled surgical instruments designed to interface with Medtronic O-arm® Imaging and StealthStation® Navigation systems. Instruments assist surgeons in precisely locating anatomical structures during open or minimally invasive sacroiliac joint fusion procedures. Input data includes patient anatomical images (CT, MR, 2D/3D fluoroscopy). Instruments feature mechanical interfaces for Medtronic NavLock trackers, allowing real-time tracking of instrument position relative to patient anatomy on the navigation workstation. Output is visual guidance on the navigation display, enabling accurate placement of Genesys Spine cannulated, self-tapping titanium (Ti-6AL-4V ELI) implants. Used in clinical/OR settings by surgeons. Benefits include improved precision in implant placement and reduced micromotion to facilitate bony fusion.
Clinical Evidence
Bench testing only. Testing verified functionality and compatibility with Medtronic O-arm and StealthStation systems, including registration accuracy, instrument-to-NavLock connection rigidity, and mechanical compatibility with existing Genesys Spine implants.
Technological Characteristics
Instruments manufactured from implant-grade titanium (Ti-6AL-4V ELI). System utilizes stereotactic navigation principles via integration with Medtronic O-arm® and StealthStation® platforms. Features cannulated, self-tapping design for implant delivery. Connectivity is achieved through mechanical interface with NavLock trackers for optical/electromagnetic tracking within the navigation environment.
Indications for Use
Indicated for patients undergoing SIros or SIros O sacroiliac joint fusion procedures where stereotactic surgery is appropriate and anatomical structures (pelvis/vertebra) can be referenced via CT, MR, or fluoroscopic imaging.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
Genesys Spine Sacroiliac Joint Fusion System (K191748)
Medtronic Navigated Manual Reusable Instruments for use with th (K161210)
Medtronic Navigated StealthStation System with Synergy Cranial (K150216)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date March 29, 2024. The text is in a simple, sans-serif font. The date is written out in full, with the month, day, and year clearly visible. The text is horizontally aligned and evenly spaced.
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the text "FDA U.S. FOOD & DRUG ADMINISTRATION" on the right. The symbol on the left is a stylized image of a staff entwined with a serpent, which is a common symbol for medicine and healthcare. The text on the right is written in a clear, sans-serif font, with "FDA" in a blue square.
Genesys Spine Derek Southard Chief Technical Officer 1250 S. Capital of Texas Highway Building 3, Suite 600 Austin, Texas 78746
Re: K233595
Trade/Device Name: Genesys Spine Sacroiliac Joint Fusion System with Navigation Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: February 28, 2024 Received: February 28, 2024
Dear Derek Southard:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
#### K233595 - Derek Southard
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Image /page/1/Picture/7 description: The image contains the name "Shumaya Ali-S" in a large, sans-serif font. The text is black against a white background. The letters are evenly spaced and clearly legible.
Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
### Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
Submission Number (if known)
K233595
Device Name
Genesys Spine Sacroiliac Joint Fusion System with Navigation
Indications for Use (Describe)
The Genesys Spine Sacroillac Joint Fusion System with Navigation is intended to be used with the Genesys Spine Sacroiliac Joint Fusion System to assist the surgeon in precisely locating anatomical structures in SIros and SIros O procedures, in which the use of stereotactic surgery may be appropriate, and where reference to anatomical structures, such as the pelvis or vertebra, can be identified relative to acquired imaging (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or image data based models of the anatomy. Navigation instruments are intended to be used with the Medtronic O-arm® Imaging System and StealthStation® Navigation System.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) Summary | Prepared |
|----------------|----------|
| | 21 CFR |
Contact Details
# R 807.92(a)(1)
T
| Applicant Name | Genesys Spine | | | | |
|------------------------------------|-----------------------------------------------------------------------------------------|--|--|--------------|--|
| Applicant Address | 1250 S. Capital of Texas Highway Building 3, Suite 600 Austin TX 78746<br>United States | | | | |
| Applicant Contact Telephone | 512-381-7080 | | | | |
| Applicant Contact | Mr. Southard Derek | | | | |
| Applicant Contact Email | derek.southard@genesysspine.com | | | | |
| Device Name | 21 CFR 807.92(a)(2) | | | | |
| Device Trade Name | Genesys Spine Sacroiliac Joint Fusion System with Navigation | | | | |
| Common Name | Stereotaxic instrument | | | | |
| Classification Name | Orthopedic Stereotaxic Instrument | | | | |
| Regulation Number | 882.4560 | | | | |
| Product Code | OLO | | | | |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) | | | | |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | | | Product Code | |
| K172268 | iFuse Implant System - iFuse Navigation | | | OLO | |
| K191748 | Genesys Spine Sacroiliac Joint Fusion System | | | OUR | |
| K161210 | Medtronic Navigated Manual Reusable Instruments for use with th | | | OLO | |
| K150216 | Medtronic Navigated StealthStation System with Synergy Cranial | | | HAW | |
| Device Description Summary | | | | | |
| 21 CFR 807.92(a)(4) | | | | | |
{4}------------------------------------------------
#### GENERAL OVERVIEW
The Genesys Spine Sacrolliac Joint Fusion with Navigation consist of instruments compatible with the Medtronic O-arm® Imaging System and StealthStation™ Navigation System to assist the surqeon in precisely locating anatomical structures in either open or minimally invasive procedures. The Medtronic O-arm® Imaging System and StealthStation™ Navigation System is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the spine or pelvis, can be identified relative to images of the anatomy.
The Genesys Spine Sacrolliac Joint Fusion System consists of implants designed to secure the sacroiliac joint and minimize micromotion to enable bony fusion. All Implants are cannulated and self-tapping; they are offered with different diameters and lengths to accommodate variations in patient anatomy and surgeon preference. The Slros OTM Sacrolliac Joint Fusion System implants are manufactured from implant grade titanium (Ti-6AL-4V ELI).
#### PRODUCT SPECIFICATIONS AND DIMENSIONS
Product and engineering specifications, and engineering drawings for all system specific instruments are provided as attached.
#### SURGICAL TECHNIQUE
A detailed surgical technique is available as attached. The surgical technique for instrumentation use, anatomical location of use, various user interfaces, how the devices and how the devices interact with the patient.
## Intended Use/Indications for Use
### 21 CFR 807.92(a)(5)
The Genesys Spine Sacrolliac Joint Fusion System with Navigation is intended to be used with the Genesys Spine Sacroiliac Joint Fusion System to assist the surgeon in precisely locating anatomical structures in which the use of stereotactic surqery may be appropriate, and where reference to anatomical structures, such as the identified relative to acquired imaging (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or image data based models of the anatomy. Navigation instruments are intended to be used with the Medtronic O-arm® Imaging System and StealthStation® Navigation System.
### Indications for Use Comparison
### 21 CFR 807.92(a)(5)
The Genesys Spine Sacrolliac Joint Fusion System with Navigation instruments are intended to be used with the Genesys Spine Sacrollac Joint Fusion System (implants and existing instruments are designed to interface with the already-cleared Medtronic StealthStation® Navigation System to assist surgeons in precisely locating anatomical structures for placement of Genesys Spine SJ Joint Fusion implants These inilar designs as the predicate instruments and incorporate the additional design features based on the predicate and reference devices to enable navigation and use with the SteathStation Navl ock trackers.
The Navigation instrument set is intended to be used with the Medtronic StealthStation using the Medtronic O-arm imaging system. The instrument modification detailed in this submission the technological characteristics of either the existing Genesys Spine Sacroiliac Joint Fusion Implants, instruments, or the Medtronic O-arm® and StealthStation® systems.
{5}------------------------------------------------
The primary predicate device for this 510(k) is the iFuse Navigation (K172268). The Genesys Spine Sacrolliar Joint Fusion System with Navigation is similar to the cited predicate in regard to instrument design, device description, intended use/indications for use, technological characteristics (operating principle, design, materials, manufacturing, etc.) and performance testing.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Testing was completed to ensure the functionality and compatibility with the Medtronic o-arm and StealthStation Systems. The performance testing completed:
1. Registration - Registration testing was performed to ensure that the instruments can be registered to the StealthStation®.
2. Accuracy - Accuracy testing was completed for comparison to the reference instruments.
3. Rigidity - Rigidity testing evaluated the connection between the Medtronic NavLock and the subject instruments.
4. Compatibility with Genesys Spine Sacroiliac Fusion System Implants - Compatibility with the Implant System has been evaluated to ensure that the isubject instruments are compatible with the Implant System.
#### Not Applicable
The intended use of the subject Genesys Spine Sacroiliac Fusion System with Navigation instruments, as part of the Genesys Spine Sacroiliac Fusion Implant System, is substantially equivalent to the predicate instruments. The verification and validation results support substantial equivalence of the subject instruments compared to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.