K172268 · SI-BONE, Inc. · OLO · Oct 31, 2017 · Neurology
Device Facts
Record ID
K172268
Device Name
iFuse Implant System- iFuse Navigation
Applicant
SI-BONE, Inc.
Product Code
OLO · Neurology
Decision Date
Oct 31, 2017
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
iFuse-Navigation instruments are intended to be used with the iFuse Implant System to assist the surgeon in precisely locating anatomical structures in iFuse procedures, in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the pelvis or vertebra can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy. iFuse-Navigation instruments are intended to be used with the Medtronic StealthStation System.
Device Story
iFuse-Navigation instruments are reusable manual surgical tools designed to interface with the Medtronic StealthStation Navigation System. They assist surgeons in precisely locating anatomical structures during iFuse procedures. The system uses pre-acquired patient images (CT, MR, or fluoroscopic) to provide dynamic, multi-plane 3D/2D graphical representations of instrument and implant placement. The device includes a mounting bracket for TeraTracker connection, enabling compatibility with O-arm imaging. Surgeons use the real-time navigation output to guide implant placement, improving accuracy in complex or minimally invasive procedures. The device is intended for use in clinical settings by trained surgeons.
Clinical Evidence
Bench testing only. Testing included registration verification, accuracy comparison against reference instruments, rigidity testing of the TeraTracker connection, and compatibility evaluation with the iFuse Implant System.
Technological Characteristics
Reusable manual surgical instruments constructed of stainless steel, PEEK, and titanium alloy. Features a mounting bracket for TeraTracker integration. Designed for use with Medtronic StealthStation navigation and O-arm imaging systems. No electronic components; passive navigation interface.
Indications for Use
Indicated for patients undergoing iFuse procedures where stereotactic surgery is appropriate and anatomical structures (pelvis or vertebra) can be referenced against CT, MR, or fluoroscopic images/models.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
October 31, 2017
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
SI-BONE, Inc. Roxanne Dubois VP, Regulatory and Quality Assurance 3055 Olin Avenue, Suite 2200 San Jose, California 95128
Re: K172268
Trade/Device Name: iFuse Implant System- iFuse Navigation Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: July 26, 2017 Received: July 27, 2017
Dear Roxanne Dubois:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevicesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely,
Katherine D. Kavlock -S
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172268
## Device Name
iFuse Implant System - iFuse-Navigation
## Indications for Use (Describe)
iFuse-Navigation instruments are intended to be used with the iFuse Implant System to assist the surgeon in precisely locating anatomical structures in iFuse procedures, in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the pelvis or vertebra can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy. iFuse-Navigation instruments are intended to be used with the Medtronic StealthStation System.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16"></rect> <path d="M2,2 L14,14 M14,2 L2,14" stroke="black" stroke-width="2"></path> </svg> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> | <div> <span> <svg height="16" width="16"> <rect fill="white" height="16" stroke="black" stroke-width="1" width="16"></rect> </svg> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY - iFuse Implant System®
#### I. SUBMITTER
| SI-BONE, Inc. | |
|-----------------|-----------------------------------------------------------|
| Address: | 3055 Olin Avenue, Suite 2200, San Jose, CA 95128 |
| Phone: | 408-207-0700; Fax: 408-557-8312 |
| Mobile: | 408-828-5019; Office: 408-207-0700 x2236 |
| Contact Person: | Roxanne Dubois, VP, Regulatory and Quality, SI-BONE, Inc. |
| Email: | rdubois@si-bone.com |
| Date Prepared: | July 26, 2017 |
#### II. DEVICE
| Name of Device: | iFuse Implant System - iFuse-Navigation |
|-----------------------|-----------------------------------------|
| Common or Usual Name: | Orthopedic Stereotaxic Instrument |
| Regulation Numbers: | 21 CFR 882.4560- Stereotaxic instrument |
### Product Codes: OLO; Stereotaxic Instrument
### PREDICATE DEVICES III.
| Primary Predicate: | iFuse Implant System by SI-BONE: K131405, K162733 |
|-----------------------|-----------------------------------------------------|
| Reference Predicates: | -K161210, Medtronic Reusable Instruments Compatible |
| | with the STEALTHSTATION® System, |
| | -K150216, Medtronic Navigated StealthStation System |
| | with Synergy Cranial Software |
| | - K162921, Orthofix Navigated Instrument System |
#### IV. DEVICE DESCRIPTION
The iFuse-Navigation instrument set is comprised of reusable manual surgical instruments specifically designed for use with the iFuse Implant System. 8 These instruments are designed to interface with the already-cleared Medtronic StealthStation® Navigation System to assist surgeons in precisely locating anatomical structures for placement of iFuse implants. This surgical imaging technology provides visualization for complex and MIS procedures and confirms the accuracy of advanced surgical procedures. Use of these navigation systems provides the surgeon access to dynamic, graphical representation of multi-plane 3D images (and 2D images) providing indication of instrument and implant placement.
{4}------------------------------------------------
### V. INDICATIONS FOR USE
iFuse-Navigation instruments are intended to be used with the iFuse Implant System to assist the surgeon in precisely locating anatomical structures in iFuse procedures, in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure, such as the pelvis or vertebra can be identified relative to the acquired image (CT, MR, 2D fluoroscopic image or 3D fluoroscopic image reconstruction) and/or an image data based model of the anatomy. iFuse-Navigation instruments are intended to be used with the Medtronic StealthStation System.
### TECHNOLOGICAL CHARACTERISTICS COMPARED TO THE PREDICATE VI.
The iFuse Navigation instruments are intended to be used with the iFuse Implant System® (implants and instruments). The iFuse-Navigation instruments are designed to interface with the already-cleared Medtronic StealthStation® Navigation System to assist surgeons in precisely locating anatomical structures for placement of iFuse implants. These instruments have similar designs as the predicate instruments and incorporate the additional design features based on the reference devices to enable navigation and use with the StealthStation TeraTrackers. Like the predicate devices, the subject iFuse-Navigation instruments are made of stainless steel, PEEK and titanium alloy.
The iFuse-Navigation set is intended to be used with the Medtronic StealthStation using an O-arm imaging system instead of a C-arm imaging system. Based on that, the subject device includes a mounting bracket to enable the connection of a TeraTracker™.
The instrument modifications detailed in this submission have no impact on the technological characteristics of either the existing iFuse Implant System implants, instruments, or the StealthStation®.
# VII. PERFORMANCE DATA
Testing was completed to ensure the functionality and compatibility with the identified Medtronic products. The following table summarizes the performance testing completed:
| Test | Description |
|---------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Registration | Registration testing was performed to ensure that the instruments can be registered to the StealthStation.® |
| Accuracy | Accuracy testing was completed for comparison to the reference instruments. |
| Rigidity | Rigidity testing evaluated the connection between the TeraTracker and the instruments. |
| Compatibility<br>with the iFuse<br>Implant System | Compatibility with the iFuse Implant System has been evaluated to ensure that the iFuse-Navigation instruments are compatible with the iFuse Implant System. |
{5}------------------------------------------------
# VIII. CONCLUSIONS
The intended use of the subject iFuse-Navigation instruments, as part of the iFuse Implant System, is substantially equivalent to the intended use of the predicate instruments. The verification and validation results support substantial equivalence of the iFuse- Navigation instruments compared to the predicate and reference devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.