K233347 · Cryosurgery, Inc. · GEH · Dec 4, 2023 · General, Plastic Surgery
Device Facts
Record ID
K233347
Device Name
Verruca-Freeze® H
Applicant
Cryosurgery, Inc.
Product Code
GEH · General, Plastic Surgery
Decision Date
Dec 4, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Verruca-Freeze® H cryosurgical system is used for treatment of benign skin lesions.
Device Story
Verruca-Freeze® H is a cryosurgical system for treating benign skin lesions. The device consists of a pressurized aerosol canister containing a mixture of dimethyl ether, propane, and isobutane, and non-sterile, single-use polyurethane foam swabs. The operator (physician) sprays the cryogen onto a foam swab, which is then applied directly to the target lesion for 15-40 seconds. The device operates via evaporative cooling, reaching temperatures of -50°C to induce extracellular and intracellular ice formation, resulting in tissue destruction. The procedure is simple, with a typical thaw time of 20-40 seconds. The device is intended for prescription use in a clinical setting.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical performance testing comparing output temperature and effectiveness to the predicate device, including visual confirmation of ice ball formation on a neoprene pad.
Technological Characteristics
Pressurized aerosol canister containing dimethyl ether (95%), propane (2%), and isobutane (3%). Applicators are polyurethane foam swabs. Operates via evaporative cooling to reach -50°C. Non-sterile, single-use components. 2-year shelf life.
Indications for Use
Indicated for the treatment of actinic keratosis, genital warts, lentigo, molluscum contagiosum, seborrheic keratosis, skin tags, verruca plantaris, verruca vulgaris, and verruca plana.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. Next to that is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 4, 2023
CryoSurgery, Inc. % Rachel Paul Lead Consultant QA&RA Emergo Europe Consulting Prinsessegracht 20 The Hague, 2514AP Netherlands
Re: K233347
Trade/Device Name: Verruca-Freeze® H Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit And Accessories Regulatory Class: Class II Product Code: GEH Dated: September 29, 2023 Received: September 29, 2023
Dear Rachel Paul:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the OS regulation reguires device manufacturers to review and approve changes to device design and production (21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to https://www.fda.gov/medical-device-safety/medical-device-reporting-mdr-howreport-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and -CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatorythe DICE assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
| Mark Trumbore -S | Digitally signed by Mark Trumbore -S<br>Date: 2023.12.04 08:24:35 -05'00' |
|------------------|---------------------------------------------------------------------------|
|------------------|---------------------------------------------------------------------------|
| Mark Trumbore, Ph.D. |
|--------------------------------------------------------|
| Assistant Director |
| DHT4A: Division of General Surgery Devices |
| OHT4: Office of Surgical and Infection Control Devices |
| Office of Product Evaluation and Quality |
| Center for Devices and Radiological Health |
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known)
K233347
Device Name
Verruca-Freeze® H
Indications for Use (Describe)
The Verruca-Freeze® H cryosurgical system is indicated for use in the following skin lesions:
- · Actinic Keratosis
- · Genital Warts
- Lentigo
- Molluscum Contaqiosum
- Seborrheic Keratosis
- Skin Tags
- · Verruca Plantaris
- · Verruca Vulgaris • Verruca Plana
- Verruca Plana
Type of Use (Select one or both, as applicable)
ر Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
The following information is provided in accordance with 21 CFR 807.92 for the Premarket 510(k) Summary:
# 1. Submitter Information
| Company: | Ronald A. McDow, MD |
|------------------------|-------------------------------------------------------------------|
| | Medical Director |
| | CryoSurgery, Inc. |
| | 5829 Old Harding Pike |
| | Nashville, TN 37205 USA |
| | Telephone: (615) 354-0414 |
| | Fax: (615) 354-0466 |
| | rmcdow@comcast.net |
| Contact: | Rachel Paul |
| | Lead Consultant QA&RA |
| | Emergo Europe Consulting |
| | Prinsessegracht 20 |
| | The Hague, 2514AP The Netherlands |
| | Telephone: + 31 (0) 70 345 85 70/Direct: + 31<br>(0) 70 850 82 49 |
| | Fax: |
| | LST.AUS.ProjectManagement@ul.com |
| Date Summary Prepared: | November 27, 2023 |
| 2. Name of the Device | |
| Trade Name: | Verruca-Freeze® H |
| Common Name: | Unit, Cryosurgical, Accessories |
| Classification Name: | General & Plastic Surgery |
| Review Panel: | General & Plastic Surgery (SU) |
| Regulation: | 878.4350 |
| Class: | Class II |
| Product Code: | GEH |
{4}------------------------------------------------
# 3. Equivalence Claimed to Predicate Device
The Verruca-Freeze® H is equivalent to the CryoDose H Or Similar (K161337), manufactured by Nuance Medical, LLC.
# 4. Device Description
The Verruca-Freeze® H cryosurgical system is used for treatment of benign skin lesions.
The device consists of:
- Canister filled with a liquid mixture of the compressed gases dimethyl ether, propane, and isobutane
- . Polyurethane foam swabs
- . Detailed instructions including illustrated descriptions
The main component of the device is the pressurized aerosol canister containing cryogen spray. The canister is used with foam swabs, which are non-sterile, single-use applicators that are disposed after each use. The foam swabs allow the cryogen to be localized to a target lesion.
The one- to two-minute procedure begins by spraying cryogen onto a foam swab, which is then placed directly onto the lesion and held for 15-40 seconds. Upon removal of the swab from the skin, the frozen tissue appears white and begins thawing. The typical thaw time is 20-40 seconds.
This treatment methodology is simple and has been an accepted practice used by physicians for decades. It utilizes a standard cryogen composition profile to freeze lesions and destroy the underlying tissue through evaporative cooling. On the cellular level, both extracellular and intracellular ice formation are occurring.
| Substantial Equivalence Table | | |
|-------------------------------|-------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|
| Characteristic | Proposed Device: Verruca-<br>Freeze® H | Primary Predicate Device:<br>CryoDose H Or Similar<br>(K161337) |
| Intended Use | The Verruca-Freeze H<br>cryosurgical system is used<br>for treatment of benign skin<br>lesions. | The CryoDose H is used for<br>treatment of benign skin<br>lesions. |
## 5. Substantial Equivalence Comparison
{5}------------------------------------------------
| Indication for Use | The Verruca-Freeze H<br>cryosurgical system is<br>indicated for use in the<br>treatment of the following<br>skin lesions: Actinic<br>Keratosis, Genital Warts,<br>Lentigo, Molluscum<br>Contagiosum, Seborrheic<br>Keratosis, Skin Tags, Verruca<br>Plantaris, Verruca Vulgaris,<br>and Verruca Plana | The CryoDose H indications<br>for use as follows: CryoDose<br>H is indicated for use in the<br>treatment of the following:<br>actinic keratosis, genital<br>warts, lentigo, molluscum<br>contagiosum, seborrheic<br>keratosis, skin tags, verruca<br>plantaris, verruca vulgaris,<br>verruca plana |
|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | GEH | GEH |
| Regulation Number | 878.4350 | 878.4350 |
| Classification Name | Unit, Cryosurgical,<br>Accessories | Unit, Cryosurgical,<br>Accessories |
| Mechanism of Action | Application of extreme cold<br>causing tissue destruction | Application of extreme cold<br>causing tissue destruction |
| Cryogen Gas Material | Dimethyl Ether (95%),<br>Propane (2%) and Isobutane<br>(3%) | Dimethyl Ether (95%),<br>Propane (2%) and Isobutane<br>(3%) |
| Freeze Time | 15-40 seconds | 15-40 seconds |
| Applicators | Foam swabs | Foam swabs |
| Sterile | No | No |
| OTC or Rx | Rx | Rx |
| Design | Aerosol-filled canister and<br>parts within the canister<br>(valve, actuator, cover cap) | Aerosol-filled canister and<br>parts within the canister<br>(valve, actuator, cover cap) |
| Freezing temperature | -50°C | -50°C |
| Shelf life | 2-year | 2-year |
## 6. Additional Information
#### 6.1 Substantial Equivalence Discussion
The technological characteristics of Verruca-Freeze® H, including materials, design, and energy source, are identical to the predicate device. The table as above compares Verruca-Freeze® H to the predicate device with respect to indications for use, principles of operation, technological characteristics, materials, and performance, and forms the basis for the determination of substantial equivalence.
Other elements of Verruca-Freeze® H not mentioned in the table that have the same technical characteristics as the predicate device include:
{6}------------------------------------------------
- Canister ●
- . Valve
- Actuator ●
- Cover Cap
Verruca-Freeze® H has the same technological characteristics as the predicate device. Both devices use the same design, energy source, materials, and other features. When the cryogen is dispensed, it saturates the foam swab is then applied directly to the desired area of treatment. Subsequently, the skin lesion is frozen and destroyed. Verruca-Freeze® H does not raise any new questions of safety or effectiveness as compared to the predicate device.
It is concluded that Verruca-Freeze® H is substantially equivalent to the predicate device.
#### 6.2 Non-Clinical Performance Data
Performance testing was conducted to determine the output temperature and effectiveness of Verruca-Freeze® H as compared to the predicate device. The output temperature was measured in the cryogen-saturated foam-tipped applicators followed by a visual confirmation of an "ice ball" on a neoprene pad. The tests determined that Verruca-Freeze® H cryogen and applicators perform equivalently to the cryogen and applicators of the predicate device. Further, the tests demonstrated that the device is capable of reaching the minimum required temperature to destroy the underlying tissue of a skin lesion.
#### 6.3 Clinical Performance Data
There was no clinical testing required to support the medical device as the indications for use are equivalent to those of the predicate device. These types of devices have been on the market for decades and their clinical safety and efficacy has been established. The nonclinical testing detailed in this submission supports the substantial equivalence of the device.
#### 7. Substantial Equivalence Conclusion
Based on the device characteristics and testing, the Verruca-Freeze® H is substantially equivalent to the predicate device and raises no new questions related to safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.