+ACG is a Clinical Decision Support device that only matches data to Hospital-defined protocols to generate Alarms and Advisories. +ACG uses data from multiple sources including medical devices and healthcare information systems. Medical Facilities use their clinical practices, protocols, and policies to define clinically relevant alarms and advisory criteria. Talis then configures Alarms and Advisories within +ACG to be used by the Medical Facility's clinical physicians or appropriate medical staff under the direction of physicians. +ACG is intended to support the Medical Facility's efforts to improve compliance to their patient care protocols and achievement of their quality initiatives. It is not intended to replace clinicians' judgement, but rather to assist clinicians in making timely, informed decisions. +ACG is not intended to be used for secondary monitoring. +ACG Alarms are not intended to be relied upon in deciding to take immediate clinical action. +ACG only works with Talis EMR Products.
Device Story
Clinical decision support system; integrates data from medical devices and healthcare information systems. Matches patient data against hospital-defined protocols to trigger alarms/advisories. Used in medical facilities by clinicians; assists in informed decision-making. Does not perform secondary monitoring; not for immediate clinical action. Operates exclusively with Talis EMR software. Provides notifications to healthcare professionals to support protocol compliance.
Clinical Evidence
No clinical data. Bench testing only, including software verification/validation, usability testing, and cybersecurity assessment.
Technological Characteristics
Software-based alarm notification system. Operates on Talis EMR platform. Connectivity via workstation or web browser. Standards: ANSI AAMI IEC 62304 (software lifecycle), ANSI AAMI IEC 62366-1 (usability), ISO 60601-1-8 (alarm systems), ISO 27001 (cybersecurity).
Indications for Use
Indicated for use by clinical physicians or appropriate medical staff under the direction of physicians in medical facilities to support compliance with patient care protocols and quality initiatives via alarm and advisory notifications based on hospital-defined criteria.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
April 1, 2024
Talis Clinical, LLC William Murphy Vice President Regulatory & Quality 650 Mondial Parkway Streetsboro, Ohio 44241
Re: K233133
Trade/Device Name: Talis EMR with +ACG Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac monitor (including cardiotachometer and rate alarm) Regulatory Class: Class II Product Code: MWI Dated: February 23, 2024 Received: February 26, 2024
Dear William Murphy:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jennifer W. Shih -S
Jennifer Shih Kozen Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K233133
Device Name
Talis EMR with +ACG
Indications for Use (Describe)
+ACG is a Clinical Decision Support device that only matches data to Hospital-defined protocols to generate Alarms and Advisories. +ACG uses data from multiple sources including medical devices and healthcare information systems.
Medical Facilities use their clinical protocols, and policies to define clinically relevant alarms and advisory criteria. Talis then configures Alarms and Advisories within +ACG to be used by the Medical Facility < clinical physicians or appropriate medical staff under the direction of physicians.
+ACG is intended to support the Medical Facility's efforts to improve compliance to their patient care protocols and achievement of their quality initiatives.
lt is not intended to replace clinicians' judgement, but rather to assist clinicians in making timely, informed decisions.
+ACG is not intended to be used for secondary monitoring.
+ACG Alarms are not intended to be relied upon in deciding to take immediate clinical action.
+ACG only works with Talis EMR Products.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|-----------------------------------------------|
| ☑ Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) Summary | | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------|--------------|
| I. SUBMITTER | | | |
| Date Prepared | March 15, 2024 | | |
| Submitter/Owner | Talis Clinical LLC<br>650 Mondial Parkway<br>Streetsboro, OH 44241 USA<br>Phone: 1-234-284-2400 | | |
| Key Contact | William Murphy<br>VP Regulatory & Quality<br>william.murphy@getinge.com | | |
| 510(k) Submission Type | This is a traditional 510(k) for multi-function software as a<br>medical device. | | |
| II. DEVICE | | | |
| Trade Name | Talis EMR with +ACG | | |
| Common Name | Alarm Notification System | | |
| Classification Name | Panel & Name: Cardiovascular (OHT2)<br>Subpart & Division: Cardiac Electrophysiology, Diagnostics, and<br>Monitoring Devices (DHT2A)<br>21 CFR 870.2300 Regulatory Class: Class II<br>Product Code: MWI | | |
| III. PREDICATE DEVICE | | | |
| Predicate Device | 510(k) No. | Company Name<br>Device Name | Product Code |
| | K213335 | Capsule Technologies, SAS / Capsule Tech Inc.<br>Capsule Surveillance System | MWI |
| The Talis EMR with +ACG is substantially equivalent to the legally marketed predicate Capsule<br>Technologies, SAS / Capsule Tech Inc.'s Capsule Surveillance System (K213335). | | | |
#### IV. DEVICE DESCRIPTION
### Talis EMR with +ACG - description of device per 21 CFR 807.92(a)(4)
Talis EMR software is an Electronic Medical Record intended to support patient care documentation compliant with patient care record and billing requirements. Talis EMR leverages a non-device MDDS to acquire, transmit, store, and display electronic patient information available from multiple electronic sources including other EMRs and medical devices. Talis EMR provides no clinical recommendations, performs no assessment of patient conditions, and uses no proprietary algorithms. Display or access to patient data is not intended to influence the judgement of the healthcare professional user. Talis EMR software functions are either non-device software functions or software functions for which FDA has determined to exercise enforcement discretion.
+ACG software is an alarm notification system which may be configured to notify health care professionals when alarm or advisory conditions are identified based on patient data matching the
{4}------------------------------------------------
Medical Facility defined alarm or advisory condition criteria. +ACG alarm/advisory notifications are limited to alarm or advisory conditions defined by the Medical Facility.
Medical Facilities use their clinical practices, protocols, and policies to define clinically relevant alarms and advisory criteria. Talis then configures Alarms and Advisories within +ACG to be used by the Medical Facility's clinical physicians or appropriate medical staff under the direction of physicians. +ACG is intended to support the Medical Facility's efforts to improve compliance to their patient care protocols and achievement of their quality initiatives.
+ACG software only works with Talis EMR software.
#### INDICATIONS FOR USE V.
# Intended Use as required per 21 CFR 807.92(a)(5)
+ACG is a Clinical Decision Support device that only matches data to Hospital-defined protocols to generate Alarms and Advisories. +ACG uses data from multiple sources including medical devices and healthcare information systems.
Medical Facilities use their clinical practices, protocols, and policies to define clinically relevant alarms and advisory criteria. Talis then configures Alarms and Advisories within +ACG to be used by the Medical Facility's clinical physicians or appropriate medical staff under the direction of physicians. +ACG is intended to support the Medical Facility's efforts to improve compliance to their patient care protocols and achievement of their quality initiatives.
It is not intended to replace clinicians' judgement, but rather to assist clinicians in making timely, informed decisions.
+ACG is not intended to be used for secondary monitoring.
+ACG Alarms are not intended to be relied upon in deciding to take immediate clinical action.
+ACG only works with Talis EMR Products.
| Comparison of Intended Uses for Subject Device and Predicate | |
|--------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | Indications for Use/Intended Use |
| Talis EMR with +ACG<br>Subject Device | +ACG is a Clinical Decision Support device that only matches<br>data to Hospital-defined protocols to generate Alarms and<br>Advisories. +ACG uses data from multiple sources including<br>medical devices and healthcare information systems.<br><br>Medical Facilities use their clinical practices, protocols, and<br>policies to define clinically relevant alarms and advisory criteria.<br>Talis then configures Alarms and Advisories within +ACG to be<br>used by the Medical Facility's clinical physicians or appropriate<br>medical staff under the direction of physicians.<br>+ACG is intended to support the Medical Facility's efforts to<br>improve compliance to their patient care protocols and<br>achievement of their quality initiatives.<br><br>It is not intended to replace clinicians' judgement, but rather to<br>assist clinicians in making timely, informed decisions.<br><br>+ACG is not intended to be used for secondary monitoring.<br>+ACG Alarms are not intended to be relied upon in deciding to<br>take immediate clinical action.<br>+ACG only works with Talis EMR Products. |
| K213335 | Capsule Surveillance is a clinical decision support device that<br>integrates, analyzes, and displays data from multiple sources<br>including medical devices and healthcare information systems. It |
{5}------------------------------------------------
| Capsule Technologies, SAS /<br>Capsule Tech Inc<br>Capsule Surveillance System<br>Predicate Device | uses standardized rules that are based on customers approved<br>clinical practices, protocols, and policies to create clinically<br>relevant alerts in health care facilities when used by clinical<br>physicians or appropriate medical staff under the direction of<br>physicians. It is not intended to replace clinicians' judgment, but<br>rather to assist clinicians in making timely, informed, higher<br>quality decisions.<br>Capsule Surveillance may be configured for secondary<br>monitoring and alerting intended to be relied upon in deciding to<br>take immediate clinical action.<br>Capsule Surveillance may also be configured for remote display<br>of physiological data and alerts not intended to be relied upon in<br>deciding to take immediate clinical action. |
|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
# VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
| Similarities | |
|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item of Comparison | Description/Rationale |
| Target Populations | Both devices are intended to be used by trained clinicians. |
| Environment of Use | Both devices are intended to be used in medical facility settings. |
| Matrix and Detailed Views | Both devices allow for matrix or simultaneous multi-patient<br>display or access to detailed views of a single patient's data. |
| Data Input | Both devices use input from multiple electronic sources including<br>EMRs and medical devices. |
| Methods for Accessing the<br>Application | Both devices allow access via a workstation or remotely via a web<br>browser |
| Not intended to be relied upon in<br>deciding to take immediate<br>clinical action. | Both devices may be configured for remote display of<br>physiological data and alerts not intended to be relied upon in<br>deciding to take immediate clinical action. |
| Differences | |
| Item of Comparison | Description/Rationale |
| Secondary Monitoring | Subject device is not intended for secondary monitoring |
| Intended to be relied upon in<br>deciding to take immediate action | Subject device is not intended to be relied upon in deciding to<br>take immediate clinical action. |
| Clinical Surveillance | Subject device does not detect the onset of emergent, potentially<br>actionable conditions and inform the care team. |
| Patient Monitoring | Subject device is not integrated with a predictive scoring system<br>that calculates a score quickly at the bedside (combining certain<br>vitals and clinician observations) to help in identifying patients at<br>risk of deterioration. |
| Default Alarms or Advisories | Subject device does not support any clinical alarm or advisory<br>condition unless the medical facility identifies and defines the<br>alarm or advisory condition and messages to the HCP. |
| Clinical Surveillance | Subject device does not monitor patient conditions |
{6}------------------------------------------------
| Patient Monitoring | Subject device does not monitor, assess, diagnose, or predict<br>patient conditions. Software is limited to match patient data to<br>Medical Facility defined patient care protocols (guidances) |
|--------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended to be relied upon in<br>deciding to take immediate action | Subject device provides no alarms / advisories not understood by<br>HCP authorized to perform and trained to Medical Facility patient<br>care protocols. |
# Substantial Equivalence Summary
Both devices provide clinical decision support. Both may be configured to support Medical Facility defined alarms / advisories condition criteria. Both have the ability to send alarm notification signals to HCP as identified by the Medical Facility. Operational and technological characteristics form the basis for the determination of substantial equivalence of the Talis EMR with +ACG software with the legally marketed predicate device (K213335). Talis EMR with +ACG is substantially equivalent to the predicate device.
#### VI. Performance Data
# Non-Clinical Tests – Harmonized Standards
Talis EMR with +ACG has passed all safety tests for demonstrated compliance with the harmonized standards below.
| Standard | FDA<br>Recognition # | Title |
|--------------------------------------------------------------------|----------------------|------------------------------------------------------------------------------------------------------------|
| ANSI AAMI IEC<br>62304:2006/A1:2016 | 13-79 | Medical device software ">– Software life cycle processes<br>[Including Amendment 1 (2016)] |
| ANSI AAMI IEC<br>62366-<br>1:2015+AMDI:2020<br>(Consolidated Text) | 5-129 | Medical devices – Part 1 Application of usability engineering<br>to medical devices, including Amendment 1 |
# Non-Clinical Tests - International Recognized Standard
Talis +ACG Alarm Notification System testing confirmed that the alarm and advisory signals comply with the international standard below.
| Standard | Title |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 60601-1-8 Edition 2.2 2020-<br>07 | Medical electrical equipment – Part 1-8: General requirements for<br>basic safety and essential performance – Collateral standard:<br>General requirements, tests, and guidance for alarm systems in<br>medical electrical equipment and medical electrical systems. |
Talis EMR with +ACG cybersecurity processes and controls demonstrate compliance with the recognized standard below, as certified by TUV SUD and assessed by 3th party vulnerability and penetration testing. MedISAO membership also supports identification of newly identified vulnerabilities.
| Standard | Title |
|-----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 27001: 2017 | Information technology – Security techniques – Information<br>security management systems – Requirements (ISO/IEC<br>27001:2013 including Cor 1:2014 and Cor 2:2015) Released<br>2017-06 |
{7}------------------------------------------------
Talis Hub hardware which functions as a non-device MDDS was tested to all applicable sections of the standard below, as certified by UL.
| Standard | Title |
|-------------------------|-----------------------------------------------------------------------------------------------------------|
| IEC 60601-1 Edition 3.1 | Medical electrical equipment – Part 1: General requirements for<br>basic safety and essential performance |
Human factors and usability testing has been completed, as well as certification to IEC 62304 for software lifecycle and IEC 62366. No new issues of safety or effectiveness as compared to the predicate are introduced as a result of using this device.
# Clinical Studies
Talis EMR with +ACG, like the predicate device, did not require clinical trials.
FDA recognized standards, FDA guidance documents, harmonized standards, verification and validation, software validation, usability validation, and risk management activities have taken place for the Talis EMR with +ACG.
Based upon the design, intended use, indications for use, classification, usability, and safety testing the Talis EMR with +ACG is substantially equivalent to the listed predicate device.
#### VII. CONCLUSIONS
The results of the substantial equivalence assessment, taken together with non-clinical bench testing, electrical safety and electromagnetic compatibility, software verification and validation, human factors and usability testing demonstrate that Talis EMR with +ACG does not raise new questions of safety and effectiveness when compared to the predicate device, performs as intended, and has performance characteristics that are substantially equivalent to the Capsule Surveillance System predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.