Automatic Disposable Biopsy Needle: Automatic Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid, lymph nodes, breast and muscle. It is not intended for use in bone. Semi-automatic Disposable Biopsy Needle: Semi-Automatic Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid, lymph nodes, breast and muscle. It is not intended for use in bone. Coaxial Biopsy Needle: Coaxial Biopsy Needle is intended for use as a guiding needle in obtaining core biopsy samples from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid, lymph nodes, breast and muscle. It is not intended for use in bone.
Device Story
Hand-operated, non-electronic surgical instruments for soft tissue biopsy. Automatic and semi-automatic needles extract tissue samples via spring-loaded firing mechanisms; coaxial needles guide biopsy needle insertion under imaging (ultrasound, X-ray, CT). Used in clinical settings by physicians. Automatic needle: inner rod enters lesion, outer tube fires to cut and capture sample in groove. Semi-automatic needle: manual advancement of inner needle, spring-loaded outer cannula fires to cut. Optional swab component assists in specimen removal. Single-use, sterile. Output is a tissue core sample for pathological examination. Benefits include minimal surrounding tissue damage during extraction.
Clinical Evidence
No human clinical trials. Evidence consists of in vivo animal biopsy sampling performance comparison against predicate devices (K133948 and K171953) to validate expanded indications, and bench testing for the new swab component (functionality, sterility, shelf-life, EO/ECH residuals).
Technological Characteristics
Hand-operated, non-electronic, single-use, sterile (EO) biopsy needles. Materials meet ISO 10993 biocompatibility standards. Coaxial needles include trocar tip stylets. Swab component added for specimen handling. Standards: ISO 11607-1/2 (packaging), ASTM F 1980 (accelerated aging), ISO 10993-7 (EO/ECH residuals).
Indications for Use
Indicated for obtaining core biopsy samples from soft tissues (liver, kidney, prostate, spleen, lung, thyroid, lymph nodes, breast, muscle) in patients requiring biopsy for soft tissue abnormalities. Not indicated for use in bone.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
January 17, 2024
Canyon Medical Inc. Mingzhi Jin, RA Building 3, Phase 2 Accelerator, No. 11 Yaogu Avenue, Jiangbei New Area. Jiangsu Province Nanjing, 210032 China
Re: K233031
Trade/Device Name: M Biopsy /SureCore Automatic Disposable Biopsy Needle. M Biopsy /SureCore Semi-automatic Disposable Biopsy Needle, M Biopsy /SureAim Coaxial Biopsy Needle Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-urology biopsy instrument Regulatory Class: Class II Product Code: KNW Dated: September 25, 2023 Received: September 25, 2023
Dear Mingzhi Jin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Jessica Carr -S
Jessica Carr, PhD Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
#### 510(k) Number (if known) K233031
#### Device Name
M·Biopsy /SureCore Automatic Disposable Biopsy Needle, M·Biopsy /SureCore Semi-automatic Disposable Biopsy Needle, M·Biopsy /SureAim Coaxial Biopsy Needle
Indications for Use (Describe)
Automatic Disposable Biopsy Needle:
Automatic Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid, lymph nodes, breast and muscle. It is not intended for use in bone.
Semi-automatic Disposable Biopsy Needle:
Semi-Automatic Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid, lymph nodes, breast and muscle. It is not intended for use in bone.
Coaxial Biopsy Needle:
Coaxial Biopsy Needle is intended for use as a guiding needle in obtaining core biopsy samples from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid, lymph nodes, breast and muscle. It is not intended for use in bone.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a logo for Canyon. The logo features two intertwined shapes resembling hearts or stylized letter 'C's. One shape is colored in blue, while the other is in yellow. Below the intertwined shapes, the word 'CANYON' is written in a stylized font, with the 'Y' colored in yellow to match the upper shape.
# 1. 510(k) Owner
Canyon Medical Inc.
# 2. Submission Correspondent
Submitter: Ms. Mingzhi Jin, RA
E-mail: imz@canyonmedical.com.cn
Telephone number: +86 13275719503
Address: Building 3, Phase 2 Accelerator, No. 11 Yaogu Avenue, Jiangbei New Area, 210032, Nanjing, Jiangsu Province People's Republic of China
# 3. Date of Preparation: January 13, 2024
## 4. Subject Device(s) Information
| Common Name: | Automatic Disposable Biopsy Needle<br>Semi-automatic Disposable Biopsy Needle<br>Coaxial Biopsy Needle |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name: | M·Biopsy /SureCore Automatic Disposable Biopsy Needle<br>M·Biopsy /SureCore Semi-automatic Disposable Biopsy Needle<br>M·Biopsy /SureAim Coaxial Biopsy Needle |
| Regulation Number: | 21 CFR 876.1075 |
| Classification Name: | Instrument, Biopsy |
| Device Class: | II |
| Product Code: | KNW |
| Review Panel: | Gastroenterology/Urology |
#### 5. Predicate Device(s) Information
| Subject device | Predicate device - K222865 |
|-----------------------------------------|-----------------------------------------|
| Automatic Disposable Biopsy Needle | Automatic Disposable Biopsy Needle |
| Semi-automatic Disposable Biopsy Needle | Semi-automatic Disposable Biopsy Needle |
| Coaxial Biopsy Needle | Coaxial Biopsy Needle |
# 6. Device Description
Automatic Disposable Biopsy Needle, Semi-automatic Disposable Biopsy Needle and Coaxial Biopsy Needle are hand-operated, non-electronic, surgical instruments.
Automatic Disposable Biopsy Needle is designed for the automatic extraction of a specimen from soft tissues, while causing minimal surrounding tissue damage, for
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo with the word "CANYON" in bold, sans-serif font. The "C" and "Y" are in yellow, while the "A", "N", "O", and "N" are in blue. Above the word "CANYON" is a stylized heart shape, with the left side in blue and the right side in yellow. The logo appears to be for an organization or company named Canyon.
tissue pathological examination/ testing. Automatic Disposable Biopsy Needle is first loaded and then inserted into the edge of the target tissue. Then, the inner needle rod is threaded into the target lesion (automatic firing), then, the outer needle tube is fired to push forward, and the tissue sample is cut through the relative movement of the outer needle tube and the inner needle rod of the biopsy needle, later, the tissue sample is cut off and stored in the sampling groove. Finally, specimen was removed after withdrawing the biopsy needle.
Semi-automatic Disposable Biopsy Needle is designed for the extraction of a specimen from soft tissues, while causing minimal surrounding tissue damage, for tissue pathological examination/testing. The semi-automated biopsy needle requires manual advancement of the inner needle to expose the specimen notch. With pressure on its plunger, a spring action rapidly advances the outer needle (cutting cannula) over the specimen notch of the inner needle.
Coaxial Biopsy Needle is used with biopsy needles to guide the insertion of biopsy needle into the soft tissue under imaging control (ultrasound, X-ray, CT, etc.). It is supplied with trocar tip stylet with or without blunt tip needle.
All of these devices are sterile with a Sterility Assurance Level (SAL) of 10-6, nonpyrogenic and single-use devices. The ultrasound, X-ray, CT and other equipments are used to guide the puncture and sampling. These devices cannot be used under MRI.
#### 7. Indications for Use
Automatic Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid, lymph nodes, breast and muscle. It is not intended for use in bone.
Semi-Automatic Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid, lymph nodes, breast and muscle. It is not intended for use in bone.
Coaxial Biopsy Needle is intended for use as a guiding needle in obtaining core biopsy samples from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid, lymph nodes, breast and muscle. It is not intended for use in bone.
#### 8. Purpose of Submission
The purpose of this submission is to introduce some modifications to the products
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image is a logo with a stylized heart shape. The top left portion of the heart is blue, while the top right portion is yellow. Below the heart shape, the word "CANYON" is written in blue, with the "Y" in yellow. The logo is simple and modern, with a focus on color and shape.
base on K222865.
1) Extend indications;
2) Add a component (swab) which can be used optionally to help take the biopsy specimen for further diagnosis preparation to the Automatic and Semi-automatic Disposable Biopsy Needle;
3) Add some models within previous specifications (K222865). The specifications of the new models are within the previously approved specification range (Automatic Disposable Biopsy Needle are available in specification range from 12G to 20G; Semiautomatic Disposable Biopsy Needle are available in specification range from 14G to 20G; Coaxial Biopsy Needle are available in specification range from 13G to 19G).
## 9. Technical Characteristics
The subject devices are substantially equivalent to the predicate devices in terms of intended use and technological characteristics. The differences between the subject devices and predicate devices do not affect the basic design principle, usage, effectiveness and safety of the subject device.
The technological characteristics of the subject devices are identical to those of predicate devices, with the exception of the following design changes:
1) More soft tissues types for indication - breast and muscle
2) A new component swab which can be used optionally to help take the biopsy specimen for further diagnosis preparation
3) Add some models within previous specifications (K222865)
Equivalence has been identified as follows:
| Subject device | Predicate device - K222865 |
|-----------------------------------------|-----------------------------------------|
| Automatic Disposable Biopsy Needle | Automatic Disposable Biopsy Needle |
| Semi-automatic Disposable Biopsy Needle | Semi-automatic Disposable Biopsy Needle |
| Coaxial Biopsy Needle | Coaxial Biopsy Needle |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo for Canyon. The logo features two stylized shapes resembling intertwined hearts or loops, one in blue and the other in yellow. Below the shapes, the word "CANYON" is written in a sans-serif font, with the "Y" colored in yellow to match the loop above it. The overall design is simple and modern.
# Table 1. Comparison of the Automatic Disposable Biopsy Needle to the predicate device.
| Item | Subject Device-Canyon | Predicate Device-Canyon | |
|---------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device | Automatic Disposable<br>Biopsy Needle | K222865 Automatic<br>Disposable Biopsy Needle | Comments |
| Classification | Class II | Class II | Same |
| Product Code | KNW | KNW | Same |
| Regulation Number | 21 CFR 876.1075 | 21 CFR 876.1075 | Same |
| Regulation Name | Gastroenterology-<br>Urology Biopsy Instrument | Gastroenterology-<br>Urology Biopsy Instrument | Same |
| Indications for<br>Use | Intended for use in<br>obtaining biopsies from soft<br>tissues such as liver,<br>kidney, prostate, spleen,<br>lung, thyroid, lymph nodes,<br>breast and muscle. It is not<br>intended for use in bone. | Intended for use in obtaining<br>biopsies from soft tissues<br>such as liver, kidney,<br>prostate, spleen, lung,<br>thyroid and lymph nodes. It<br>is not intended for use in<br>bone. | This change is to add new<br>indications (breast and<br>muscle). As for new<br>indications, Canyon<br>conducted sampling<br>performance comparison<br>with previously chosen<br>predicate in K222865 to<br>ensure the biopsy needles<br>can be used on the new<br>types of soft tissues. |
| Target<br>Population | Individuals requiring<br>biopsy for sampling of<br>soft tissue abnormalities | Individuals requiring<br>biopsy for sampling of<br>soft tissue abnormalities | Same |
| Operation<br>Mechanics | Single hand automatic<br>activation | Single hand automatic<br>activation | Same |
| Gauge | 12G, 14G, 16G, 18G, 20G | 12G, 14G, 16G, 18G, 20G | Same |
| Length (mm) | 100, 130, 160, 200, 250 | 100, 130, 160, 200, 250 | Same |
| Penetration<br>Depth (mm) | 11, 22 | 11, 22 | Same |
| Slot Size (mm) | 18 | 18 | Same |
| Sterilization | EO Sterilization | EO Sterilization | Same |
| Single Use | Yes | Yes | Same |
| Patient-<br>contacting<br>Information | Externally communicating<br>device, in contact with the<br>patient for a limited<br>duration (no more than 24<br>hours) | Externally communicating<br>device, in contact with the<br>patient for a limited duration<br>(no more than 24 hours) | Same |
| Biocompatibility | Biocompatible according to<br>ISO 10993 applicable parts | Biocompatible according to<br>ISO 10993 applicable parts | Same |
| Item | Subject Device-Canyon | Predicate Device-Canyon | Comments |
| Device | Semi-automatic Disposable<br>Biopsy Needle | K222865 Semi-automatic<br>Disposable Biopsy Needle | |
| Classification | Class II | Class II | Same |
| Product Code | KNW | KNW | Same |
| Regulation Name | Gastroenterology-Urology Biopsy Instrument | Gastroenterology-Urology Biopsy Instrument | Same |
| Regulation Number | 21 CFR 876.1075 | 21 CFR 876.1075 | Same |
| Indications for<br>Use | Intended for use in obtaining<br>biopsies from soft tissues<br>such as liver, kidney,<br>prostate, spleen, lung,<br>thyroid, lymph nodes, breast<br>and muscle. It is not<br>intended for use in bone. | Intended for use in obtaining<br>biopsies from soft tissues<br>such as liver, kidney,<br>prostate, spleen, lung,<br>thyroid and lymph nodes. It<br>is not intended for use in bone. | This change is to add new<br>indications (breast and<br>muscle). As for new<br>indications, Canyon<br>conducted sampling<br>performance comparison<br>with previously chosen<br>predicate in K222865 to<br>ensure the biopsy needles<br>can be used on the new<br>types of soft tissues. |
| Target<br>Population | Individuals requiring<br>biopsy for sampling of<br>soft tissue abnormalities | Individuals requiring<br>biopsy for sampling of<br>soft tissue abnormalities | Same |
| Operation<br>Mechanics | Single hand semi-automatic<br>activation | Single hand semi-automatic<br>activation | Same |
| Gauge | 14G, 16G, 18G, 20G | 14G, 16G, 18G, 20G | Same |
| Length (mm) | 60, 100, 130, 160, 200, 250,<br>300 | 60, 100, 130, 160, 200, 250,<br>300 | Same |
| Penetration<br>Depth (mm) | 10, 20 | 10, 20 | Same |
| Sterilization | EO Sterilization | EO Sterilization | Same |
| Single Use | Yes | Yes | Same |
| Patient-<br>contacting<br>Information | Externally communicating<br>device, in contact with the<br>patient for a limited duration<br>(no more than 24 hours) | Externally communicating<br>device, in contact with the<br>patient for a limited duration<br>(no more than 24 hours) | Same |
| Biocompatibility | Biocompatible according to<br>ISO 10993 applicable parts | Biocompatible according to<br>ISO 10993 applicable parts | Same |
| Item | Subject Device-Canyon | Predicate Device-Canyon | |
| Device | Coaxial Biopsy Needle | K222865 Coaxial Biopsy<br>Needle | Comments |
| Classification | Class II | Class II | Same |
| Product Code | KNW | KNW | Same |
| Regulation<br>Number | 21 CFR 876.1075 | 21 CFR 876.1075 | Same |
| Regulation<br>Name | Gastroenterology-<br>Urology Biopsy Instrument | Gastroenterology-<br>Urology Biopsy Instrument | Same |
| Indications for<br>Use | Intended for use as a<br>guiding needle in obtaining<br>core biopsy samples from<br>soft tissues such as liver,<br>kidney, prostate, spleen,<br>lung, thyroid and lymph<br>nodes, breast and muscle. It<br>is not intended for use in<br>bone | Intended for use as a<br>guiding needle in obtaining<br>core biopsy samples from<br>soft tissues such as liver,<br>kidney, prostate, spleen,<br>lung, thyroid and lymph<br>nodes. It is not intended for<br>use in bone. | This change is to add new<br>indications (breast and<br>muscle). As for new<br>indications, Canyon<br>conducted sampling<br>performance comparison<br>with previously chosen<br>predicate in K222865 to<br>ensure the biopsy needles<br>can be used on the new<br>types of soft tissues. |
| Operation<br>Mechanics | Single hand automatic<br>activation | Single hand automatic<br>activation | Same |
| Target<br>Population | Individuals requiring<br>biopsy for sampling of<br>soft tissue abnormalities | Individuals requiring<br>biopsy for sampling of<br>soft tissue abnormalities | Same |
| Gauge | 13G, 14G, 15G, 16G, 17G,<br>18G, 19G | 13G, 15G, 17G, 19G | This change is to add new<br>gauges, the gauges to be<br>added are all within the<br>previous cleared<br>specifications of K222865. |
| Sterilization | EO Sterilization | EO Sterilization | Same |
| Single use | Yes | Yes | Same |
| Patient<br>contacting | Externally communicating<br>device, in contact with the<br>patient for a limited duration<br>(no more than 24 hours) | Externally communicating<br>device, in contact with the<br>patient for a limited duration<br>(no more than 24 hours) | Same |
| Biocompatibility | Biocompatible according to<br>ISO 10993 applicable parts | Biocompatible according to<br>ISO 10993 applicable parts | Same |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo for Canyon. The logo features two intertwined shapes resembling hearts or stylized letter 'C's. One shape is colored blue, and the other is yellow. Below the shapes, the word 'CANYON' is written in a sans-serif font, with the 'Y' in yellow to match the color of the shape above it.
510(k) Summary (K233031)
# Table 2. Comparison of the Semi-automatic Disposable Biopsy Needle to the predicate device.
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows a logo for Canyon. The logo consists of two intertwined shapes, one in blue and one in yellow, forming a heart-like design. Below the shapes, the word "CANYON" is written, with the "Y" in yellow to match the color of the shape above it. The overall design is simple and modern.
Table 3. Comparison of the Coaxial Biopsy Needle to the predicate device.
The technological characteristics of the subject device are identical to those of predicate device. The subject device has the same basic design as the predicate device. The comparison between the subject and predicate devices is based on the following:
- Similar indications for use
- Same material types that meet ISO 10993 biocompatibility requirements
- · Same sterilization methods
- · Same fundamental technology
There is no significant risk raised by the difference.
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows a logo for Canyon. The logo features two intertwined shapes resembling hearts or stylized letter 'C's. One shape is blue, and the other is yellow. Below the shapes, the word "CANYON" is written in a sans-serif font, with the 'O' in yellow, matching the color of the heart shape above it.
# 10. Summary of Non-Clinical Testing
Based on the purpose of this submission and as for the targeted design changes, we explicate and support with following:
## 10.1 As for indication extension
We conducted in vivo biopsy sampling from animal to compare the sampling performance of the subject devices and comparator devices (K133948 and K171953) , and the results indicated no substantial difference.
## 10.2 As for the addition of swab
Summary of non-clinical and performance bench testing was performed to evaluate the performance and functionality of the subject device against requirement specification. The subject device has been subjected to compliance testing according to, by FDA, recognized consensus standards ISO 11607-1, ISO 11607-2, ASTM F 1980 etc.
## 1) Biocompatibility
Since this component does not contact with the patient, therefore, biocompatibility test is unnecessary.
# 2) Package Validation and Transport
The swab is packaged in the previous package of Automatic Disposable Biopsy Needle and Semi-automatic Disposable Biopsy Needle without any other modification, therefore, no more test is carried out.
#### 3) Sterility test
In order to ensure the sterility, the sterility test is carried out and the result meets the requirement.
## 4) Swab performance and shelf life
Overall requirement: the swab shall can help take the soft tissue
Performance 1: the tissue strip should be successfully removed from the sampling tank, and there should be no impurities such as dander (exfoliation of the swab) in the tissue strip.
Performance 2: after helping take soft tissue, the swab head should be complete, no broken, and no residual tissue strip on the swab head.
Performance 3: shelf life: 5 years
Accelerated aging was used to simulate the storage of 5 years, then the physical performance tests and chemical performance tests were performed on the accelerated aged samples. The test results demonstrate that the aged samples complied with the pre-determined acceptance criteria.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows a logo for Canyon. The logo features two interlocking shapes, one in blue and one in yellow, that resemble stylized hearts or loops. Below the shapes, the word "CANYON" is written in a sans-serif font, with the "Y" colored yellow to match the yellow shape above. The overall design is simple and modern, with a focus on the interlocking shapes and the contrasting colors.
#### EO/ECH Residual 5)
Examination of the EO and ECH residuals have been conducted in accordance with ISO 10993-7:2008+Amd.1:2019 to evaluate whether the sterilized proposed device comply with the above selected allowable limits, and the results meet the requirements. EO/ECH residual testing is based on the whole product of biopsy needle and the swab.
# 10.3 As for the addition of product models
Since the models to be added are all within the previous cleared specifications of K222865. Therefore, as this change, no more tests are needed.
## 11. Conclusion
The conclusions drawn from the comparison and analysis above demonstrate that the proposed subject devices are as safe, as effective, and performs as well as the legally marketed predicate devices and raises no new questions of safety or effectiveness.
The differences between both devices are insignificant in terms of safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.