M Biopsy /SureCore Automatic Disposable Biopsy Needle, M Biopsy /SureCore Semi-Automatic Disposable Biopsy Needle, M Biopsy /SureAim Coaxial Biopsy Needle
Applicant
Canyon Medical, Inc.
Product Code
KNW · Gastroenterology, Urology
Decision Date
Apr 3, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 876.1075
Device Class
Class 2
Indications for Use
Automatic Disposable Biopsy Needle: Automatic Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid and lymph nodes. It is not intended for use in bone. Semi-automatic Disposable Biopsy Needle: Semi-Automatic Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid and lymph nodes. It is not intended for use in bone. Coaxial Biopsy Needle: Coaxial Biopsy Needle is intended for use as a guiding needle in obtaining core biopsy samples from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid and lymph nodes. It is not intended for use in bone.
Device Story
Hand-operated, non-electronic surgical instruments for soft tissue biopsy. Automatic needle: spring-loaded mechanism for automatic firing and tissue cutting. Semi-automatic needle: manual advancement of inner needle; spring-loaded outer cannula for cutting. Coaxial needle: guiding cannula for biopsy needle insertion under imaging (ultrasound, X-ray, CT). Used in clinical settings by physicians. Devices are single-use, sterile, and non-MRI compatible. Output is a tissue specimen for pathological examination. Benefits include minimally invasive tissue sampling for diagnosis.
Clinical Evidence
No clinical data. Bench testing only. Performance evaluated via comparative testing against predicates (puncture force, stiffness, breakage resistance, corrosion resistance, joint strength, heavy metal content) and compliance with ISO 9626, ISO 10993, ISO 11607, and ASTM F 1980 standards.
Technological Characteristics
Hand-operated, non-electronic, single-use biopsy needles. Materials meet ISO 10993 biocompatibility requirements. EO sterilization (SAL 10^-6). Automatic/semi-automatic spring-actuated cutting mechanisms. Gauge range 12G-20G. Not MRI compatible.
Indications for Use
Indicated for patients requiring core biopsy sampling of soft tissue abnormalities (liver, kidney, prostate, spleen, lung, thyroid, lymph nodes). Not for use in bone.
Regulatory Classification
Identification
A gastroenterology-urology biopsy instrument is a device used to remove, by cutting or aspiration, a specimen of tissue for microscopic examination. This generic type of device includes the biopsy punch, gastrointestinal mechanical biopsy instrument, suction biopsy instrument, gastro-urology biopsy needle and needle set, and nonelectric biopsy forceps. This section does not apply to biopsy instruments that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font.
April 3, 2023
Canyon Medical Inc. Huiqi Jiang RA Building 3, Phase 2 Accelerator, No. 11 Yaogu Avenue Jiangbei New Area Nanjing, Jiangsu Province 210032 China
Re: K222865
Trade/Device Name: M Biopsy /SureCore Automatic Disposable Biopsy Needle, M Biopsy /SureCore Semi-Automatic Disposable Biopsy Needle, M Biopsy /SureAim Coaxial Biopsy Needle Regulation Number: 21 CFR 876.1075 Regulation Name: Gastroenterology-Urology Biopsy Instrument Regulatory Class: Class II Product Code: KNW Dated: March 3, 2023 Received: March 3, 2023
Dear Huiqi Jiang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-
542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
| Mark | Digitally signed by |
|----------|-------------------------------------|
| Trumbore | Mark Trumbore -S |
| -S | Date: 2023.04.03<br>11:20:56-04'00' |
Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) K222865
Device Name
Automatic Disposable Biopsy Needle Semi-automatic Disposable Biopsy Needle Coaxial Biopsy Needle
Indications for Use (Describe)
Automatic Disposable Biopsy Needle:
Automatic Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid and lymph nodes. It is not intended for use in bone.
Semi-automatic Disposable Biopsy Needle:
Semi-Automatic Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid and lymph nodes. It is not intended for use in bone.
Coaxial Biopsy Needle:
Coaxial Biopsy Needle is intended for use as a guiding needle in obtaining core biopsy samples from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid and lymph nodes. It is not intended for use in bone.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
{4}------------------------------------------------
# 1. 510(k) Submitter Information:
Submitter Name: Canyon Medical Inc.
Address: Building 3, Phase 2 Accelerator, No. 11 Yaogu Avenue, Jiangbei New Area, 210032, Nanjing, Jiangsu Province, People's Republic of China
# Submission Correspondent: Ms. Huiqi Jiang, RA
E-mail: jhq@canyonmedical.com.cn
Telephone: +86-13241196637
# 2. Date of Preparation: January 24, 2023
| Common Name: | Automatic Disposable Biopsy Needle<br>Semi-automatic Disposable Biopsy Needle<br>Coaxial Biopsy Needle |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name: | M·Biopsy /SureCore Automatic Disposable Biopsy Needle<br>M·Biopsy /SureCore Semi-automatic Disposable Biopsy Needle<br>M·Biopsy /SureAim Coaxial Biopsy Needle |
| Regulation Number: | 21 CFR 876.1075 |
| Classification Name: | Instrument, Biopsy |
| Device Class: | II |
| Product Code: | KNW |
| Review Panel: | Gastroenterology/Urology |
# 3. Subject Device(s) Information:
# 4. Predicate Device(s) Information:
| Subject device | Predicate device |
|--------------------------------------------|--------------------------------------------------------------|
| Automatic Disposable<br>Biopsy Needle | K133948 BARD® MAX-CORE® Disposable Core Biopsy<br>Instrument |
| Semi-automatic Disposable<br>Biopsy Needle | K171953 BARD® MISSION® Disposable Core Biopsy<br>Instrument |
| Coaxial Biopsy Needle | K171953 BARD® TRUGUIDE® Disposable Coaxial Biopsy<br>Needle |
# 5. Device Description
Automatic Disposable Biopsy Needle, Semi-automatic Disposable Biopsy Needle and Coaxial Biopsy Needle are hand-operated, non-electronic, surgical instruments.
Automatic Disposable Biopsy Needle is designed for the automatic extraction of
{5}------------------------------------------------
### 510(K) Summarv
a specimen from soft tissues, while causing minimal surrounding tissue damage, for tissue pathological examination/ testing. Automatic Disposable Biopsy Needle is first loaded and then inserted into the edge of the target tissue. Then, the inner needle rod is threaded into the target lesion (automatic firing), then, the outer needle tube is fired to push forward, and the tissue sample is cut through the relative movement of the outer needle tube and the inner needle rod of the biopsy needle, later, the tissue sample is cut off and stored in the sampling groove. Finally, specimen was removed after withdrawing the biopsy needle.
Semi-automatic Disposable Biopsy Needle is designed for the extraction of a specimen from soft tissues, while causing minimal surrounding tissue damage, for tissue pathological examination/testing. The semi-automated biopsy needle requires manual advancement of the inner needle to expose the specimen notch. With pressure on its plunger, a spring action rapidly advances the outer needle (cutting cannula) over the specimen notch of the inner needle.
Coaxial Biopsy Needle is used with biopsy needles to quide the insertion of biopsy needle into the soft tissue under imaging control (ultrasound, X-ray, CT, etc.). It is supplied with trocar tip stylet with or without blunt tip needle.
All of these devices are sterile with a Sterility Assurance Level (SAL) of 10-6, nonpyrogenic and single-use devices. The ultrasound, X-ray, CT and other equipments are used to guide the puncture and sampling. These devices can't be used under MRI.
## 6. Indications for Use
Automatic Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid and lymph nodes. lt is not intended for use in bone.
Semi-Automatic Disposable Biopsy Needle is intended for use in obtaining biopsies from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid and lymph nodes. It is not intended for use in bone.
Coaxial Biopsy Needle is intended for use as a guiding needle in obtaining core biopsy samples from soft tissues such as liver, kidney, prostate, spleen, lung, thyroid and lymph nodes. It is not intended for use in bone.
## 7. Technical Characteristics
The subject devices are substantially equivalent to the predicate devices in terms of intended use and technological characteristics. The differences between the subject
{6}------------------------------------------------
devices and predicate devices do not affect the basic design principle, usage, effectiveness and safety of the subject device. Equivalence has been identified as
follows:
| Subject device | Predicate device |
|--------------------------------------------|--------------------------------------------------------------|
| Automatic Disposable<br>Biopsy Needle | K133948 BARD® MAX-CORE® Disposable Core Biopsy<br>Instrument |
| Semi-automatic Disposable<br>Biopsy Needle | K171953 BARD® MISSION® Disposable Core Biopsy<br>Instrument |
| Coaxial Biopsy Needle | K171953 BARD® TRUGUIDE® Disposable Coaxial Biopsy<br>Needle |
{7}------------------------------------------------
| Item | Subject Device-Canyon | Predicate Device-Bard |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| Device | Automatic Disposable Biopsy<br>Needle | K133948 BARD® MAX-CORE®<br>Disposable Core Biopsy<br>Instrument |
| Classification | Class II | Same as subject device |
| Product Code | KNW | Same as subject device |
| Regulation Number | 21 CFR 876.1075 | Same as subject device |
| Regulation Name | Gastroenterology-<br>Urology Biopsy Instrument | Same as subject device |
| Indications for Use | Intended for use in obtaining<br>biopsies from soft tissues such as<br>liver, kidney, prostate, spleen,<br>lung, thyroid and lymph nodes. It is<br>not intended for use in bone. | Same as subject device |
| Target Population | Individuals requiring<br>biopsy for sampling of<br>soft tissue abnormalities | Same as subject device |
| Operation Mechanics | Single hand automatic activation | Same as subject device |
| Gauge | 12G, 14G, 16G, 18G, 20G | 14G, 16G, 18G, 20G |
| Length (mm) | 100, 130, 160, 200, 250 | 100, 160, 200, 250 |
| Penetration<br>Depth<br>(mm) | 11, 22 | Same as subject device |
| Slot Size (mm) | 18 | 18, 19 |
| Sterilization | EO Sterilization | Same as subject device |
| Single Use | Yes | Yes |
| Patient-contacting<br>Infomarion | Externally communicating device,<br>in contact with the patient for a<br>limited duration (no more than 24<br>hours) | Same as subject device |
| Biocompatibility | Biocompatible according to ISO<br>10993 applicable parts | Biocompatible according to ISO<br>10993 applicable parts |
| Item | Subject Device-Canyon | Predicate Device-Bard |
| Device | Semi-automatic Disposable Biopsy<br>Needle | K171953 BARD® MISSION®<br>Disposable Core Biopsy<br>Instrument |
| Classification | Class II | Same as subject device |
| Product Code | KNW | Same as subject device |
| Regulation<br>Number | 21 CFR 876.1075 | Same as subject device |
| Regulation Name | Gastroenterology-<br>Urology Biopsy Instrument | Same as subject device |
| Indications for<br>Use | Intended for use in obtaining biopsies<br>from soft tissues such as liver, kidney,<br>prostate, spleen, lung, thyroid and lymph<br>nodes. It is not intended for use in bone. | Same as subject device |
| Target Population | Individuals requiring<br>biopsy for sampling of<br>soft tissue abnormalities | Same as subject device |
| Operation<br>Mechanics | Single hand semi-automatic activation | Same as subject device |
| Gauge | 14G, 16G, 18G, 20G | 14G, 16G, 18G, 20G |
| Length (mm) | 60, 100, 130, 160, 200, 250, 300 | 60, 100, 160, 200, 250 |
| Penetration Depth<br>(mm) | 10, 20 | Same as subject device |
| Sterilization | EO Sterilization | Same as subject device |
| Single Use | Yes | Yes |
| Patient-contacting<br>Information | Externally communicating device, in<br>contact with the patient for a limited<br>duration (no more than 24 hours) | Same as subject device |
| Biocompatibility | Biocompatible according to ISO 10993<br>applicable parts | Biocompatible according to<br>ISO 10993 applicable parts |
Table 1. Comparison of the Automatic Disposable Biopsy Needle to the predicate device.
{8}------------------------------------------------
Table 2. Comparison of the Semi-automatic Disposable Biopsy Needle to the predicate device.
{9}------------------------------------------------
| Item | Subject Device-Canyon | Predicate Device-Bard |
|---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------|
| Device | Coaxial Biopsy Needle | K171953 BARD® TRUGUIDE®<br>Disposable Coaxial Biopsy Needle |
| Classification | Class II | Same as subject device |
| Product Code | KNW | Same as subject device |
| Regulation Number | 21 CFR 876.1075 | Same as subject device |
| Regulation Name | Gastroenterology-<br>Urology Biopsy Instrument | Same as subject device |
| Indications for Use | Intended for use as a guiding needle<br>in obtaining core biopsy samples from<br>soft tissues such as liver, kidney,<br>prostate, spleen, lung, thyroid and<br>lymph nodes. It is not intended for use<br>in bone | Same as subject device |
| Operation Mechanics | Single hand automatic activation | Same as subject device |
| Target Population | Individuals requiring<br>biopsy for sampling of<br>soft tissue abnormalities | Same as subject device |
| Gauge | 13G, 15G, 17G, 19G | 11G, 13G, 15G, 17G, 19G |
| Sterilization | EO Sterilization | Same as subject device |
| Single use | Yes | Yes |
| Patient contacting | Externally communicating device, in<br>contact with the patient for a limited<br>duration (no more than 24 hours) | Same as subject device |
| Biocompatibility | Biocompatible according to ISO<br>10993 applicable parts | Biocompatible according to ISO<br>10993 applicable parts |
# Table 3. Comparison of the Coaxial Biopsy Needle to the predicate device.
The technological characteristics of the subject device are identical to those of predicate device. The subject device has the same basic design as the predicate device. The comparison between the subject and predicate devices is based on the following:
- · Same indications for use
- Same material types that meet ISO 10993 biocompatibility requirements
- Same sterilization methods
- Same fundamental technology
There is no significant risk raised by the difference.
{10}------------------------------------------------
# 8. Summary of Non-Clinical Testing
Summary of non-clinical and performance bench testing was performed to evaluate the performance and functionality of the subject device aqainst requirement specification. The subject device has been subjected to compliance testing according to, by FDA, recognized consensus standards ISO 9626, ISO 10993 series standards, ISO 11607-1, ISO 11607-2, ASTM F 1980 etc. Results from testing performed confirm the design requirement.
- 1) Biocompatibility
The subject devices are externally communicating medical devices with tissue less than 24h which means limited duration. Therefore, as per "Table A.1: Biocompatibility Evaluation Endpoints" of FDA quidance, following items shall be carried out.
| No. | Test Item | Standard |
|-----|----------------------------|---------------------------------------------|
| 1 | Cytotoxicity | ISO10993-5:2009 |
| 2 | Maximization Sensitization | ISO10993-10:2010 |
| 3 | Intracutaneous Reactivity | ISO10993-10:2010 |
| 4 | Acute Systemic Toxicity | ISO10993-11:2017 |
| 5 | Pyrogen Test | ISO10993-11:2017/ USP General Chapter <151> |
| 6 | Hemolysis | ISO 10993-4:2017 |
- 2) Package Validation
The primary package of the proposed device is intended to maintain the sterility of the product during its claimed shelf life. The integrity performance as per ISO 11607-1:2019 and ISO 11607-2:2019 is carried out.
- 3) Transport
Subject devices are processed by compression, first vibration, shock and second vibration, and then check. The result indicates that packaging is not damaged.
- 4) Shelf Life
Accelerated aging was used to simulate the storage of 5 years, then the physical performance tests, chemical performance tests as well as the package integrity tests were performed on the accelerated aged samples. The test results demonstrate that the aged samples complied with the pre-determined acceptance criteria.
- 5) Comparative Claims/Performance Comparison with predicate devices
The test was conducted to the predicate device and subject device to compare their performance, including scale mark identification, puncture force, biopsy sample testing, stiffness, resistance to breakage, resistance to corrosion, joint strength, total heavy metal content.
{11}------------------------------------------------
# 6) EO/ECH Resudial
Examination of the EO and ECH residuals have been conducted in accordance with ISO 10993-7:2008+Amd.1:2019 to evaluate whether the sterilized proposed device comply with the above selected allowable limits, and the results meet the requirements.
# 9. Preclinical (Animal) Studies
Bench testing is sufficient to demonstrate performance of the device. No preclinical testing of the subject device is necessary.
# 10. Clinical Test
No. Use of biopsy needle is proven technology and is well accepted by the medical community. Performance Test is sufficient to demonstrate safety and effectiveness of the subject devices with the predicate devices.
# 11. Conclusion
The differences between both devices are insignificant in terms of safety and effectiveness. The conclusions drawn from information above demonstrate that the proposed devices are as safe, as effective, and performs as well as the legally marketed predicate devices and raises no new risks of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.