K232821 · S.I.N. - Sistema DE Implante Nacional S.A. · DZE · May 31, 2024 · Dental
Device Facts
Record ID
K232821
Device Name
S.I.N. Dental Implant System
Applicant
S.I.N. - Sistema DE Implante Nacional S.A.
Product Code
DZE · Dental
Decision Date
May 31, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading. Unitite Slim, Unitite, Unitite Compact, Strong SW CM S.I.N Dental Implant System implants with lengths less than 7 mm are intended for delayed loading only. Epikut CM, Epikut Plus CM, Epikut S, Epikut S Plus S.I.N. Dental Implant System implants with lengths of 18, 20, 22, or 24 mm may be tilted up to 30°. When used in the mandible or maxilla with implants with lengths of 18, 20, 22, or 24 mm at an angulation of 30°, a minimum of four implants must be used and must be splinted. When placed in the maxilla with lengths of 18, 20, 22, or 24 mm at angulations between 0° and less than 30°, the S.I.N. Dental Implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants. Epikut CM, Epikut Plus CM, Strong SW CM Plus, Strong SW HE Plus, Strong SW HI Plus All digitally-designed custom abutments for use with Interface CAD-CAM abutments are to be sent to a S.I.N.-validated milling center for manufacture.
Device Story
Endosseous dental implants; titanium (ASTM F67) construction; HA-nanostructured surface treatment. Used by dentists/oral surgeons in clinical settings for tooth replacement. Implants placed in maxillary/mandibular bone; provide support for prosthetic restorations. Surface treatment enhances wettability and osteoblast differentiation (upregulation of RUNX2/ALP). Output is a stable foundation for dental prosthetics. Benefits include improved osseointegration and functional/esthetic rehabilitation. Digitally-designed custom abutments require S.I.N.-validated milling centers.
Clinical Evidence
No clinical data included. Bench testing and literature-based evidence provided. In vitro studies (Bezerra et al. 2017, da Silva et al. 2020) demonstrate HAnano surface increases wettability, promotes hydrophilic behavior, and modulates osteoblast signaling/differentiation compared to uncoated or acid-etched surfaces. Biocompatibility (ISO 10993), sterility (ISO 11137), and fatigue testing (ISO 14801) leveraged from predicate submissions.
Indicated for patients requiring single or multi-unit dental restorations in the maxillary or mandibular arch. Contraindicated for immediate loading in implants <7mm length. Specific requirements for tilted implants (up to 30°) include splinting and minimum implant counts (4 for 30° angulation; 2 for 0-30° angulation in maxilla).
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health and Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
S.I.N. - Sistema de Implante Nacional S.A. % Floyd Larson President PaxMed International, LLC 12264 EL Camino Real, Suite 400 San Diego, California 92130
May 31, 2024
Re: K232821
Trade/Device Name: S.I.N. Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: Mav 8, 2024 Received: May 8, 2024
Dear Floyd Larson:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K232821
#### Device Name
S.I.N. Dental Implant System
#### Indications for Use (Describe)
S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Unitite Slim, Unitite, Unitite Compact, Strong SW CM S.I.N Dental Implant System implants with lengths less than 7 mm are intended for delayed loading only.
#### Epikut CM, Epikut Plus CM, Epikut S, Epikut S Plus
S.I.N. Dental Implant System implants with lengths of 18, 20, 22, or 24 mm may be tilted up to 30°. When used in the mandible or maxilla with implants with lengths of 18, 20, 22, or 24 mm at an angulation of 30°, a minimum of four implants must be used and must be splinted. When placed in the maxilla with lengths of 18, 20, 22, or 24 mm at angulations between 0° and less than 30°, the S.I.N. Dental Implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants.
### Epikut CM, Epikut Plus CM, Strong SW CM Plus, Strong SW HE Plus, Strong SW HI Plus All digitally-designed custom abutments for use with Interface CAD-CAM abutments are to be sent to a S.I.N.-validated milling center for manufacture.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary K232821 S.I.N. - Sistema de Implante Nacional S.A. S.I.N Dental Implant System
May 22, 2024
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | S.I.N. - Sistema de Implante Nacional S.A.<br>Avenida Vereador Abel Ferreira, 2140<br>São Paulo, São Paulo<br>03340-000 Brazil<br>Telephone +55-11-21693000 ext 3236 | | | | | | | |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|
| Official Contact | Denise Domiciano, Quality and Regulatory Manager | | | | | | | |
| Representative/Consultant | Floyd G. Larson, MS, MBA<br>Kevin A. Thomas, PhD<br>PaxMed International, LLC<br>12264 El Camino Real, Suite 400<br>San Diego, CA 92130<br>Telephone +1 858-792-1235<br>Fax +1 858-792-1236<br>Email flarson@paxmed.com<br>kthomas@paxmed.com | | | | | | | |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | S.I.N. Dental Implant System |
|------------------------|------------------------------------------------------------------------------------------------|
| Common Names | Endosseous dental implant |
| Regulation Number | 21 CFR 872.3640 |
| Regulation Name | Endosseous dental implant |
| Regulatory Class | Class II |
| Product Code | DZE |
| Classification Panel | Dental |
| Reviewing Office | Office of Health Technology 1<br>(Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices) |
| Reviewing Division | Division of Health Technology 1 B (Dental Devices) |
#### PREDICATE DEVICE INFORMATION
Primary Predicate Device K170392, S.I.N. Dental Implant System, S.I.N. - Sistema de Implante Nacional S.A.
Reference Devices
| K193096, S.I.N. Dental Implant System, S.I.N. - Sistema de Implante Nacional S.A |
|----------------------------------------------------------------------------------|
| K200992, S.I.N. Dental Implant System, S.I.N. - Sistema de Implante Nacional S.A |
| K211921, S.I.N. Dental Implant System, S.I.N. - Sistema de Implante Nacional S.A |
| K221453, S.I.N. Dental Implant System, S.I.N. - Sistema de Implante Nacional S.A |
| K222231, S.I.N. Dental Implant System, S.I.N. - Sistema de Implante Nacional S.A |
{5}------------------------------------------------
#### INDICATIONS FOR USE STATEMENT
S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
#### Unitite Slim, Unitite, Unitite Compact, Strong SW CM
S.I.N Dental Implant System implants with lengths less than 7 mm are intended for delayed loading only.
#### Epikut CM, Epikut Plus CM, Epikut S, Epikut S Plus
S.I.N. Dental Implant System implants with lengths of 18, 20, 22, or 24 mm may be tilted up to 30°. When used in the mandible or maxilla with implants with lengths of 18, 20, 22, or 24 mm at an angulation of 30°, a minimum of four implants must be used and must be splinted. When placed in the maxilla with lengths of 18, 20, 22, or 24 mm at angulations between 0° and less than 30°, the S.I.N. Dental Implant System implants are only indicated for multiple unit restorations in splinted applications that utilize at least two implants.
Epikut CM, Epikut Plus CM, Strong SW CM Plus, Strong SW HE Plus, Strong SW HI Plus All digitally-designed custom abutments for use with Interface CAD-CAM abutments are to a S.I.N.-validated milling center for manufacture.
#### SUBJECT DEVICE DESCRIPTION
The purpose of this submission is to add labeling claims for a surface treatment designated HA™®, which is applied to implants of the S.I.N. Dental Implant System, as previously cleared in K170392. K193096. K200992, K211921, K221453, and K222231. The new labeling claims and the information supporting the claims are shown below in Performance Data.
The following information is a summary of the designs of the subject device implants, as they have been previously cleared with the HA"" surface treatment. Note that there are no changes to any aspect of the design, nor to the HA" surface treatment, as described in the respective 510(k) Premarket Notifications referenced above.
The previously cleared implants that are the subject of this submission includes eleven (11) product lines from six (6) 510(k)s. Implant body diameters, platform diameters and lengths for each product line are shown below in Table 10.1 Summary of Subject Device Designs and Sizes.
K170392: Unitite Slim, Unitite, Unitite Compact K193096: Strong SW CM Plus, Strong SW HE Plus, Strong SW HI Plus K200992: Strong SW Plus K211921: Epikut Plus CM, Epikut Plus HE K221453: Epikut Plus CM (additional lengths) K22231: Epikut S Plus
{6}------------------------------------------------
| 510k | Implant Line | Image<br>Not to<br>Scale | Body Ø,<br>mm | Platform<br>Ø,<br>mm | Length, mm | | | | | | |
|---------|-------------------|--------------------------|---------------|----------------------|------------|-----|------|------|----|----|--|
| | Unitite Slim | Image: Unitite Slim | 2.9 | 2.9 | | 10 | 11.5 | 13 | | | |
| K170392 | Unitite | Image: Unitite | 3.5 | 3.5 | 8.5 | 10 | 11.5 | 13 | 15 | | |
| | | Image: Unitite | 4.3 | 4.3 | 8.5 | 10 | 11.5 | 13 | 15 | | |
| | | Image: Unitite | 5.0 | 5.0 | 8.5 | 10 | 11.5 | 13 | 15 | | |
| | Unitite Compact | Image: Unitite Compact | 4.0 | 4.0 | 5 | 6 | 7 | | | | |
| | | Image: Unitite Compact | 5.0 | 5.0 | 5 | 6 | 7 | | | | |
| | | Image: Unitite Compact | 6.0 | 6.0 | 5 | 6 | 7 | | | | |
| K193096 | Strong SW CM Plus | Image: Strong SW CM Plus | 3.5 | 3.5 | 8.5 | 10 | 11.5 | 13 | 15 | | |
| | | Image: Strong SW CM Plus | 3.8 | 3.8 | 8.5 | 10 | 11.5 | 13 | 15 | | |
| | | Image: Strong SW CM Plus | 4.5 | 4.5 | 8.5 | 10 | 11.5 | 13 | 15 | | |
| | | Image: Strong SW CM Plus | 5.0 | 5.0 | 8.5 | 10 | 11.5 | 13 | 15 | | |
| | Strong SW HE Plus | Image: Strong SW HE Plus | 3.5 | 3.65 | 7 | 8.5 | 10 | 11.5 | 13 | 15 | |
| | | Image: Strong SW HE Plus | 3.75 | 4.1 | 7 | 8.5 | 10 | 11.5 | 13 | 15 | |
| | | Image: Strong SW HE Plus | 4.0 | 4.1 | 7 | 8.5 | 10 | 11.5 | 13 | 15 | |
| | | Image: Strong SW HE Plus | 4.5 | 4.5 | | 8.5 | 10 | 11.5 | 13 | 15 | |
| | | Image: Strong SW HE Plus | 5.0 | 5.0 | 7 | 8.5 | 10 | 11.5 | 13 | 15 | |
| | Strong SW HI Plus | Image: Strong SW HI Plus | 3.8 | 3.8 | | 8.5 | 10 | 11.5 | 13 | 15 | |
| | | Image: Strong SW HI Plus | 4.5 | 4.5 | | 8.5 | 10 | 11.5 | 13 | 15 | |
| | | Image: Strong SW HI Plus | 5.0 | 5.0 | | 8.5 | 10 | 11.5 | 13 | 15 | |
| K200992 | Strong SW Plus | Image: Strong SW Plus | 3.5 | 3.5 | | 8.5 | 10 | 11.5 | 13 | 15 | |
| | | Image: Strong SW Plus | 3.8 | 3.8 | | 8.5 | 10 | 11.5 | 13 | 15 | |
| | | Image: Strong SW Plus | 4.5 | 4.5 | | 8.5 | 10 | 11.5 | 13 | 15 | |
| | | Image: Strong SW Plus | 5.0 | 5.0 | | 8.5 | 10 | 11.5 | 13 | 15 | |
| K211921 | Epikut Plus CM | Image: Epikut Plus CM | 3.5 | 3.5 | | 8.5 | 10 | 11.5 | 13 | 15 | |
| | | | 3.8 | 3.8 | | 8.5 | 10 | 11.5 | 13 | 15 | |
| | | | 4.5 | 4.5 | | 8.5 | 10 | 11.5 | 13 | 15 | |
| | | Image: Epikut Plus CM | 5.0 | 5.0 | | 8.5 | 10 | 11.5 | 13 | 15 | |
| | Epikut Plus HE | Image: Epikut Plus HE | 3.5 | 3.65 | 7 | 8.5 | 10 | 11.5 | 13 | 15 | |
| | | Image: Epikut Plus HE | 4.5 | 4.5 | | 8.5 | 10 | 11.5 | 13 | 15 | |
| | | Image: Epikut Plus HE | 5.0 | 5.0 | 7 | 8.5 | 10 | 11.5 | 13 | 15 | |
## Table 10.1 – Summary of Subject Device Designs and Sizes
{7}------------------------------------------------
| 510k | Implant Line | Image<br>Not to<br>Scale | Body Ø, mm | Platform Ø, mm | Len<br>gth, mm | | | | | | | | | | |
|---------|----------------|--------------------------|------------|----------------|----------------|-----|----|------|----|----|----|----|----|----|----|
| K221453 | Epikut Plus CM | Image: Epikut Plus CM | 3.8 | 3.8 | | | | | | | | 18 | 20 | 22 | 24 |
| | | | 4.0 | 4.0 | | | | | | | | 18 | 20 | 22 | 24 |
| | | | 4.5 | 4.5 | | | | | | | | 18 | 20 | 22 | 24 |
| K222231 | Epikut S Plus | Image: Epikut S Plus | 3.5 | 3.5 | | 8.5 | 10 | 11.5 | 13 | 15 | | | | | |
| | | | 3.8 | 3.8 | | 8.5 | 10 | 11.5 | 13 | 15 | 18 | 20 | 22 | 24 | |
| | | | 4.0 | 4.0 | | 8.5 | 10 | 11.5 | 13 | 15 | 18 | 20 | 22 | 24 | |
| | | | 4.5 | 4.5 | | 8.5 | 10 | 11.5 | 13 | 15 | 18 | 20 | 22 | 24 | |
| | | | 5.0 | 5.0 | | 8.5 | 10 | 11.5 | 13 | 15 | | | | | |
All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67. The HA1000 surface treatment is identical to that cleared in K211921.
#### PERFORMANCE DATA
Non-clinical data submitted, referenced, or relied upon to demonstrate substantial equivalence included: in vitro testing and animal testing referenced from published literature to support the labeling claims for HAmano included in the submission, as shown below:
| Labeling Claim | Reference** | Discussion |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| The HAnano surface increases the surface wettability.* | Bezerra, et al.,<br>2017 | Page 6: Water contact angle was<br>significantly less on HAnano treated<br>titanium surface than on dual acid-<br>etched titanium surface |
| The HAnano surface is hydrophilic.* | Bezerra, et al.,<br>2017 | Page 6: Water contact angle was<br>significantly less on HAnano treated<br>titanium surface than on dual acid-<br>etched titanium surface |
| In vitro studies using pre-osteoblast cells, evaluating the HAnano<br>coating by assessment of cell signaling dynamics, have demonstrated a<br>greater transition from adhesion/proliferative-related signaling to<br>differentiation-related signaling compared to uncoated titanium, thus<br>culminating in changes in osteogenic gene marker expression, which is<br>important in bone regeneration processes.* | Bezerra, et al.,<br>2017 | Page 10, correcting name of surface<br>(HAnano vs. Nano Hydroxyapatite-<br>Blasted) |
| Cells grown on HAnano surfaces demonstrated a spreading morphology<br>different than that of cells grown on machined surfaces.* | Bezerra, et al.,<br>2017 | Page 4, Figure 2 caption, correcting<br>name of surface |
| The HAnano Surface promotes osteoblast differentiation by significantly<br>upregulating RUNX2 and ALP activity when pre-osteoblasts were<br>grown directly on titanium discs.* | Bezerra, et al.,<br>2017 | Page 10 |
| Osteoblasts were observed to have a morphology that spread across the<br>HAnano surface differently than over a hydrophilic dual acid-etched<br>surface (SLActive) in in vitro experiments.* | da Silva, et al.,<br>2020 | Page 7. "Firstly, our data shows there<br>is a differential behavior of<br>osteoblasts adhering on both surfaces;<br>osteoblast spreads better over<br>HAnano®, while osteoblast on<br>SLActive presented a fusiform<br>morphology." |
*Results of in vitro testing are not necessarily predictive of human clinical performance.
{8}------------------------------------------------
** References
Bezerra F, Ferreira MR, Fontes GN, da Costa Fernandes CJ, Andia DC, Cruz NC, da Silva RA, Zambuzzi WF. Nano hydroxyapatite-blasted titanium surface affects pre-osteoblast morphology by modulating critical intracellular pathways. Biotechnol Bioeng. 2017 Aug;114(8):1888-1898.
da Silva RA, da Silva Feltran G, Ferreira MR, Wood PF, Bezerra F, Zambuzzi WF. The Impact of Bioactive Surfaces in the Early Stages of Osseointegration: An In Vitro Comparative Study Evaluating the HAnano® and SLActive® Super Hydrophilic Surfaces. Biomed Res Int. 2020 Sep 13.
Other non-clinical data submitted, referenced, or relied upon to demonstrate substantial equivalence included:
- · gamma irradiation sterilization for all subject devices (to a sterility assurance level of 10 % by selecting and substantiating a 25 kGy dose using method VDmax25, according to ISO 11137-1 and ISO 11137-2 (referenced from K211921);
- bacterial endotoxin testing (referenced from K211921) including Limulus amebocyte lysate (LAL) test according to ANSI/AAMI ST72 on samples of water used in manufacturing on a weekly basis and on samples from sterilized product on a quarterly basis to demonstrate all sterile product meets a limit of < 20 EU/device;
- · shelf life testing (referenced from K211921) including testing of samples after 4 years of real time aging according to ASTM F1929 and F88/F88M (packaging sterile barrier) and sterility testing of product:
- · biological evaluation was performed according to ISO 10993-1 and test results leveraged from K170392 for ISO 10993-3 (genotoxicity), ISO 10993-5 (cytotoxicity), ISO 10993-6 (implantation), ISO 10993-10 (sensitization, irritation), and ISO 10993-11 (systemic toxicity) to support biocompatibility of the subject device;
- · information for the HA "and hydroxyapatite coating was leveraged from K 170392;
- · non-clinical analysis and testing to evaluate the metallic subject devices in the MR environment according to ASTM F2052 (magnetically induced displacement force), ASTM F2213 (magnetically induced torque), ASTM F2182 (RF induced heating), and ASTM F2119 (image artifact), and the FDA guidance document Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment (issued May 2021).
- · Fatigue testing according to ISO 14801 referenced from K170392, K193096, K200992, K211921, K221453, and K222231 is applicable to the subject device.
No clinical data were included in this submission.
#### EQUIVALENCE TO MARKETED DEVICES
The primary predicate device K170392 is in support of substantial equivalence for the implant designs, materials, HA1200 surface, manufacturing, sterilization, shelf-life and biocompatibility. The reference devices are in support of the specific designs of each of the HA™® coated implants.
The Indications for Use Statement (IFUS) for the subject device includes language concerning placement in the maxillary or mandibular arches and regarding immediate loading that is identical to the language in K170392. K193096. K200992. K211921. K221453 and K222231. The IFUS for the subject device also includes identical language to that included in K221453 and K222231 regarding longer implants and identical language to that included in K170392 regarding short implants.
The IFUS for the subject device includes language that implants may be tilted up to 30°, and language that requires, for implants with lengths of 18, 20, 22 or 24 mm and angulation of 30°, that a minimum of four (4) implants must be used and must be splinted. It also includes language that, when placed in the
{9}------------------------------------------------
maxilla with lengths of 18, 20, 22, or 24 mm at angulations between 0° and less than 30°, implants are indicated only for multiple unit restorations in splinted applications that utilize at least two implants. This language is identical to the language in the IFUS of the reference devices K221453 and K222231.
The basis for the belief of S.I.N. - Sistema de Implante Nacional S.A. that the subject device is substantially equivalent to the predicate devices is summarized in the following Table of Substantial Equivalence.
#### CONCLUSION
The subject device, the primary predicate device, and the reference devices have the same intended use, have similar technological characteristics, and are made of identical materials. The subject device, the primary predicate, and the reference devices encompass the same range of physical dimensions, are packaged in similar materials, and are sterilized using similar methods.
The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
{10}------------------------------------------------
# Table of Substantial Equivalence
| | Subject Device | Primary Predicate Device | Reference Devices | | | | | |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| | S.I.N. Dental Implant System<br>S.I.N. - Sistema de Implante<br>Nacional S.A. | K170392<br>S.I.N. Dental Implant System<br>S.I.N. - Sistema de Implante<br>Nacional S.A. | K193096<br>S.I.N. Dental Implant System<br>S.I.N. - Sistema de Implante<br>Nacional S.A. | K200992<br>S.I.N. Dental Implant System<br>S.I.N. - Sistema de Implante<br>Nacional S.A. | K211921<br>S.I.N. Dental Implant System<br>S.I.N. - Sistema de Implante<br>Nacional S.A. | K221453<br>S.I.N. Dental Implant System<br>S.I.N. - Sistema de Implante<br>Nacional S.A. | K222231<br>S.I.N. Dental Implant System<br>S.I.N. - Sistema de Implante<br>Nacional S.A. | |
| Indications for Use<br>Statement | S.I.N. Dental Implant System is intended<br>for placement in the maxillary or<br>mandibular arch to provide support for<br>single-unit or multi-unit restorations.<br>When a one-stage surgical approach is<br>applied, the S.I.N. Dental Implant<br>System is intended for immediate<br>loading when good primary stability is<br>achieved and with appropriate occlusal<br>loading.<br>Unitite Slim, Unitite, Unitite Compact,<br>Strong SW CM<br>S.I.N Dental Implant System implants<br>with lengths less than 7 mm are intended<br>for delayed loading only.<br>Epikut CM, Epikut Plus CM, Epikut S,<br>Epikut S Plus<br>S.I.N. Dental Implant System implants<br>with lengths of 18, 20, 22, or 24 mm may<br>be tilted up to 30°. When used in the<br>mandible or maxilla with implants with<br>lengths of 18, 20, 22, or 24 mm at an<br>angulation of 30°, a minimum of four<br>implants must be used and must be<br>splinted. When placed in the maxilla<br>with lengths of 18, 20, 22, or 24 mm at<br>angulations between 0° and less than 30°, the S.I.N. Dental Implant System<br>implants are only indicated for multiple<br>unit restorations in splinted applications<br>that utilize at least two implants.<br>Epikut CM, Epikut Plus CM, Strong SW<br>CM Plus, Strong SW HE Plus, Strong<br>SW HI Plus<br>All digitally-designed custom abutments<br>for use with Interface CAD-CAM<br>abutments are to be sent to a S.I.N.-<br>validated milling center for manufacture. | S.I.N. Dental Implant System is intended<br>for placement in the maxillary or<br>mandibular arch to provide support for<br>single-unit or multi-unit restorations.<br>When a one-stage surgical approach is<br>applied, the S.I.N. Dental Implant<br>System is intended for immediate<br>loading when good primary stability is<br>achieved and with appropriate occlusal<br>loading. Implants with lengths less than 7<br>mm are intended for delayed loading<br>only. | S.I.N. Dental Implant System is intended<br>for placement in the maxillary or<br>mandibular arch to provide support for<br>single-unit or multi-unit restorations.<br>When a one-stage surgical approach is<br>applied, the S.I.N. Dental Implant<br>System is intended for immediate<br>loading when good primary stability is<br>achieved and with appropriate occlusal<br>loading.<br>All digitally-designed custom abutments<br>for use with Interface CAD-CAM<br>abutments are to be sent to a<br>S.I.N.-validated milling center for<br>manufacture. | S.I.N. Dental Implant System is intended<br>for placement in the maxillary or<br>mandibular arch to provide support for<br>single-unit or multi-unit restorations.<br>When a one-stage surgical approach is<br>applied, the S.I.N. Dental Implant<br>System is intended for immediate<br>loading when good primary stability is<br>achieved and with appropriate occlusal<br>loading. | S.I.N. Dental Implant System is intended<br>for placement in the maxillary or<br>mandibular arch to provide support for<br>single-unit or multi-unit restorations.<br>When a one-stage surgical approach is<br>applied, the S.I.N. Dental Implant<br>System is intended for immediate<br>loading when good primary stability is<br>achieved and with appropriate occlusal<br>loading. | S.I.N. Dental Implant System is intended<br>for placement in the maxillary or<br>mandibular arch to provide support for<br>single-unit or multi-unit restorations.<br>When a one-stage surgical approach is<br>applied, the S.I.N. Dental Implant<br>System is intended for immediate<br>loading when good primary stability is<br>achieved and with appropriate occlusal<br>loading.<br>S.I.N. Dental Implant System implants<br>with lengths of 18, 20, 22, or 24 mm may<br>be tilted up to 30°. When used in the<br>mandible or maxilla with implants with<br>lengths of 18, 20, 22, or 24 mm at an<br>angulation of 30°, a minimum of four<br>implants must be used and must be<br>splinted. When placed in the maxilla<br>with lengths of 18, 20, 22, or 24 mm at<br>angulations between 0° and less than 30°, the S.I.N. Dental Implant System<br>implants are only indicated for multiple<br>unit restorations in splinted applications<br>that utilize at least two implants.<br>All digitally-designed custom abutments<br>for use with Interface CAD-CAM<br>abutments are to be sent to a S.I.N.-<br>validated milling center for manufacture. | S.I.N. Dental Implant System is intended<br>for placement in the maxillary or<br>mandibular arch to provide support for<br>single-unit or multi-unit restorations.<br>When a one-stage surgical approach is<br>applied, the S.I.N. Dental Implant<br>System is intended for immediate<br>loading when good primary stability is<br>achieved and with appropriate occlusal<br>loading.<br>S.I.N. Dental Implant System implants<br>with lengths of 18, 20, 22, or 24 mm may<br>be tilted up to 30°. When used in the<br>mandible or maxilla with implants with<br>lengths of 18, 20, 22, or 24 mm at an<br>angulation of 30°, a minimum of four<br>implants must be used and must be<br>splinted. When placed in the maxilla<br>with lengths of 18, 20, 22, or 24 mm at<br>angulations between 0° and less than 30°, the S.I.N. Dental Implant System<br>implants are only indicated for multiple<br>unit restorations in splinted applications<br>that utilize at least two implants. | |
| Reason for Predicate Device | Not applicable | Implant design, materials;<br>manufacturing, sterilization | Implant design, materials;<br>manufacturing, sterilization | Implant design, materials;<br>manufacturing, sterilization | Implant design, materials;<br>manufacturing, sterilization | Implant design, materials;<br>manufacturing, sterilization | Implant design, materials;<br>manufacturing, sterilization | |
| Product Code(s) | DZE | DZE, NHA | DZE, NHA | DZE, NHA | DZE | DZE, NHA | DZE, NHA | |
| Intended Use | Functional and esthetic<br>rehabilitation of the edentulous<br>mandible or maxilla | Functional and esthetic<br>rehabilitation of the edentulous<br>mandible or maxilla | Functional and esthetic<br>rehabilitation of the edentulous<br>mandible or maxilla | Functional and esthetic<br>rehabilitation of the edentulous<br>mandible or maxilla | Functional and esthetic<br>rehabilitation of the edentulous<br>mandible or maxilla | Functional and esthetic<br>rehabilitation of the edentulous<br>mandible or maxilla | Functional and esthetic<br>rehabilitation of the edentulous<br>mandible or maxilla | |
| | Subject Device | Primary Predicate Device | Reference Devices | | | | | |
| | S.I.N. Dental Implant System<br>S.I.N. - Sistema de Implante<br>Nacional S.A. | K170392<br>S.I.N. Dental Implant System<br>S.I.N. - Sistema de Implante<br>Nacional S.A. | K193096<br>S.I.N. Dental Implant System<br>S.I.N. - Sistema de Implante<br>Nacional S.A. | K200992<br>S.I.N. Dental Implant System<br>S.I.N. - Sistema de Implante<br>Nacional S.A. | K211921<br>S.I.N. Dental Implant System<br>S.I.N. - Sistema de Implante<br>Nacional S.A. | K221453<br>S.I.N. Dental Implant System<br>S.I.N. - Sistema de Implante<br>Nacional S.A. | K222231<br>S.I.N. Dental Implant System<br>S.I.N. - Sistema de Implante<br>Nacional S.A. | |
| Designs |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.