K211921 · S.I.N. - Sistema DE Implante Nacional S.A. · DZE · Oct 20, 2021 · Dental
Device Facts
Record ID
K211921
Device Name
S.I.N. Dental Implant System
Applicant
S.I.N. - Sistema DE Implante Nacional S.A.
Product Code
DZE · Dental
Decision Date
Oct 20, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Device Story
S.I.N. Dental Implant System consists of endosseous dental implants (Epikut CM, Epikut Plus CM, Epikut HE, Epikut Plus HE) designed for placement in maxillary or mandibular arches. Implants provide support for single or multi-unit restorations. Epikut CM/Plus CM feature Morse taper abutment interfaces; Epikut HE/Plus HE feature external hexagon interfaces. Surfaces are double acid-etched; Plus variants include hydroxyapatite (HA) coating. Used by dental clinicians in clinical settings. Implants are surgically placed; primary stability allows immediate loading. Output is a stable foundation for prosthetic restorations. Benefits include restoration of oral function and aesthetics.
Clinical Evidence
No clinical data were included in this submission. Substantial equivalence is supported by non-clinical bench testing, including fatigue testing (ISO 14801), biocompatibility (ISO 10993 series), sterilization validation (ISO 11137), shelf-life testing (ASTM F1929, F88), and surface characterization (SEM, XPS, TEM, XRD).
Technological Characteristics
Materials: Unalloyed titanium (ASTM F67). Surface: Double acid-etched; HA coating (HAnano). Interfaces: Morse taper (11.5°) and External Hex. Sterilization: Gamma irradiation (VDmax25, ISO 11137). Dimensions: Various diameters (3.5-5.0mm) and lengths (7-15mm).
Indications for Use
Indicated for patients requiring functional and esthetic rehabilitation of the edentulous mandible or maxilla via single-unit or multi-unit restorations. Suitable for one-stage surgical approach with immediate loading if primary stability is achieved.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
October 20, 2021
S.I.N. - Sistema de Implante Nacional S.A. % Kevin Thomas Vice President and Director of Regulatory Affairs PaxMed International, LLC 12264 El Camino Real, Suite 400 San Diego, California 92130
Re: K211921
Trade/Device Name: S.I.N. Dental Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: Class II Product Code: DZE Dated: September 17, 2021 Received: September 20, 2021
Dear Kevin Thomas:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Andrew Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic. Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K211921
Device Name
S.I.N. Dental Implant System
#### Indications for Use (Describe)
S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10pt;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-size: 10pt;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
|----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K211921 S.I.N. - Sistema de Implante Nacional S.A.
## S.I.N Dental Implant System
October 19, 2021
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name | S.I.N. - Sistema de Implante Nacional S.A. | | |
|---------------------------|--------------------------------------------|--------------------------------------------------|--|
| | Avenida Vereador Abel Ferreira, 1100 | | |
| | São Paulo, São Paulo | | |
| | 03340-000 Brazil | | |
| | Telephone | +55-11-21693000 ext 3236 | |
| Official Contact | | Denise Domiciano, Quality and Regulatory Manager | |
| Representative/Consultant | Kevin A. Thomas, PhD | | |
| | Floyd G. Larson, MS, MBA | | |
| | PaxMed International, LLC | | |
| | 12264 El Camino Real, Suite 400 | | |
| | San Diego, CA 92130 | | |
| | Telephone | +1 858-792-1235 | |
| | Fax | +1 858-792-1236 | |
| | Email | kthomas@paxmed.com | |
| | | flarson@paxmed.com | |
#### DEVICE NAME AND CLASSIFICATION
| Trade/Proprietary Name | S.I.N. Dental Implant System |
|------------------------|------------------------------------------------------------------------------------------------|
| Common Names | Endosseous dental implant |
| Regulation Number | 21 CFR 872.3640 |
| Regulation Name | Endosseous dental implant |
| Regulatory Class | Class II |
| Product Code | DZE |
| Classification Panel | Dental |
| Reviewing Office | Office of Health Technology 1 (Ophthalmic, Anesthesia, Respiratory,<br>ENT and Dental Devices) |
| Reviewing Division | Division of Health Technology 1 B (Dental Devices) |
#### PREDICATE DEVICE INFORMATION
Primary Predicate Device K200992, S.I.N. Dental Implant System, S.I.N. - Sistema de Implante Nacional S.A.
Additional Predicate Devices K193096, S.I.N. Dental Implant System, S.I.N. - Sistema de Implante Nacional S.A. K170398, S.I.N. Dental Implant System, S.I.N. - Sistema de Implante Nacional S.A.
{4}------------------------------------------------
### INDICATIONS FOR USE STATEMENT
S.I.N. Dental Implant System is intended for placement in the maxillary or mandibular arch to provide support for single-unit or multi-unit restorations. When a one-stage surgical approach is applied, the S.I.N. Dental Implant System is intended for immediate loading when good primary stability is achieved and with appropriate occlusal loading.
#### SUBJECT DEVICE DESCRIPTION
The purpose of this submission is to add components to the S.I.N. Dental Implant System components cleared in K200992, K193096, and K170398.
This submission includes dental implants Epikut CM with a Morse taper (CM) abutment interface and a double acid-etched endosseous surface, and Epikut Plus CM implants with an endosseous surface produced by double acid-etching followed by application of a hydroxyapatite coating (HA™Y). These implant endosseous surfaces are identical to those cleared in K200992.
This submission also includes dental implants Epikut HE with an external hexagon (HE) abutment interface and a double acid-etched endosseous surface, and Epikut Plus HE implants with an endosseous surface produced by double acid-etching followed by application of a hydroxyapatite coating (HA1000). These implant endosseous surfaces also are identical to those cleared in K200992.
| Implant Lines | Body Ø, mm | Platform Ø, mm | | Lengths, mm | | | | |
|----------------|------------|----------------|-----|-------------|----|------|----|----|
| Epikut CM | 3.5 | 3.5 | | 8.5 | 10 | 11.5 | 13 | 15 |
| Epikut Plus CM | 3.8 | 3.8 | | 8.5 | 10 | 11.5 | 13 | 15 |
| | 4.5 | 4.5 | | 8.5 | 10 | 11.5 | 13 | 15 |
| | 5.0 | 5.0 | | 8.5 | 10 | 11.5 | 13 | 15 |
| Epikut HE | 3.5 | 3.65 | 7.0 | 8.5 | 10 | 11.5 | 13 | 15 |
| Epikut Plus HE | 4.5 | 4.5 | | 8.5 | 10 | 11.5 | 13 | 15 |
| | 5.0 | 5.0 | 7.0 | 8.5 | 10 | 11.5 | 13 | 15 |
The subject device dental implants are summarized in the following table.
All subject device dental implants are manufactured from unalloyed titanium conforming to ASTM F67. The acid etching procedure is applied to all subject device dental implants. The acid etching process in this submission is identical to the process used to manufacture the dental implants cleared in K200992. K193096, and K170398. The Epikut Plus CM and Epikut Plus HE implants have the identical HA™ surface treatment cleared in K200992 and K193096.
The subject device dental implants are compatible with abutments and prosthetic components cleared previously in K200992. K193096. K170398. and K051859. Epikut HE and Epikut Plus HE dental implants are not for use with angled abutments.
#### PERFORMANCE DATA
Non-clinical data submitted, referenced, or relied upon to demonstrate substantial equivalence included: gamma irradiation sterilization for all subject devices (to a sterility assurance level of 10th by selecting and substantiating a 25 kGy dose using method VDmax25, according to ISO 11137-1 and ISO 11137-2, referenced from K200992);
{5}------------------------------------------------
bacterial endotoxin testing (referenced from K200992) including Limulus amebocyte lysate (LAL) test according to ANSI/AAMI ST72 on samples of water used in manufacturing on a weekly basis and on samples from sterilized product on a quarterly basis to demonstrate all sterile product meets a limit of < 20 EU/device;
shelf life testing (referenced from K200992) including testing of samples after 4 years of real time aging according to ASTM F1929 and F88/F88M (packaging sterile barrier) and sterility testing of product;
biocompatibility testing for the HAnao hydroxyapatite coating leveraged from K200992 and provided in K170392 included ISO 10993-3 (genotoxicity), ISO 10993-5 (cytotoxicity), ISO 10993-6 (implantation), ISO 10993-10 (sensitization, irritation), and ISO 10993-11 (systemic toxicity);
and characterization of the HAnao hydroxyapatite coating leveraged from K200992 and provided in K170392 included scanning electron microscopy (SEM), x-ray photoelectron spectroscopy (XPS), transmission electron microscopy (TEM), x-ray diffraction (XRD), and testing of the adherence of the coating.
No clinical data were included in this submission.
#### EQUIVALENCE TO MARKETED DEVICES
The subject device is substantially equivalent in indications and design principles to the predicate device and the additional predicate devices listed above. Provided at the end of this summary is a table comparing the Indications for Use Statements and the technological characteristics of the subject device, the primary predicate device, and the additional predicate devices.
The Indications for Use Statement (IFUS) for the subject device is identical to that of the primary predicate K200992, and is very similar to the IFUS of the additional predicates K193096 and K170398. The minor difference between the IFUS for the subject device and the additional predicate K193096 is the language concerning CAD-CAM abutments in K193096. The minor difference between the IFUS for the subject device and the additional predicate K170398 is the language concerning Revolution Company implants in K170398. These minor differences do not impact safety or effectiveness because all IFUS express equivalent intended use to facilitate functional and esthetic rehabilitation of the edentulous mandible or maxilla.
The subject device Epikut CM and Epikut Plus CM implants are similar in design and sizes to the Strong SW and Strong SW Plus implants, respectively, cleared in K200992. The subject device Epikut CM and Evikut Plus CM implants and the Strong SW Plus implants have the same 11.5° internal Morse taper abutment connection, and are provided in the same sizes of body/platform diameters and overall lengths. The subject device Epikut CM implants and the Strong SW implants have the same acidetched endosseous surface treatment. The subject device Epikut Plus CM implants and the Strong SW Plus implants have the same acid-etched and HA™®º endosseous surface treatment.
The subject device Epikut HE implants are similar in design and sizes to the Strong SW HE implants cleared in K170398. The subject device Epikut HE implants and the Strong SW HE implants have the external hexagon abutment interface and are provided in the same sizes of body/platform diameters and overall lengths. The subject device Epikut HE implants and the Strong SW HE implants have the same acid-etched endosseous surface treatment.
The subject device Epikut Plus HE implants are similar in design and sizes to the Strong SW HE Plus
{6}------------------------------------------------
implants cleared in K193096. The subject device Epikut Plus HE implants and the Strong SW HE Plus implants have the external hexagon abutment interface and are provided in the same sizes of body/platform diameters and overall lengths. The subject device Epikut Plus HE implants and the Strong SW HE Plus implants have the same acid-etched and HA™® endosseous surface treatment.
All subject device dental implants are manufactured from the same unalloyed titanium and have the same acid-etched surface treatment as the dental implants cleared in K200992, K193096, and K170398. In addition to the acid-etch, the subject device Epikut Plus CM and Epikut Plus HE implants also have the identical HAmano surface cleared in K200992 and K193096.
Minor differences in the exact dimensions of the subject device components as compared to those of the primary predicate device and additional predicate devices do not impact safety, effectiveness, or substantial equivalence. The cross sections of the subject device implants (Epikut CM and Epikut Plus CM) were compared to the cross sections of the worst-case implants cleared in K200992. The critical dimension evaluated was the minimum wall thickness from the top of the implant down to the potting level for ISO 14801 testing. This analysis demonstrated that the subject device CM implants do not introduce an additional or new worst-case construct that requires mitigation by additional mechanical testing. Therefore, the fatigue testing according to ISO 14801 provided in K200992 is applicable to the subject device CM implants.
All subject device implants components are provided sterile by gamma irradiation, the same sterilization method used in K200992, K193096, and K170398.
#### CONCLUSION
The subject device, the primary predicate device, and the additional predicate devices have the same intended use, have similar technological characteristics, and are made of identical or similar materials. The subject device, the primary predicate, and the additional predicate devices encompass the same range of physical dimensions, are packaged in similar materials, and are sterilized using similar methods.
The data included in this submission demonstrate substantial equivalence to the predicate devices listed above.
{7}------------------------------------------------
# Table of Substantial Equivalence
| | Subject Device | Primary Predicate Device | Additional Predicate | | Additional Predicate |
|---------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | S.I.N. Dental Implant System<br>S.I.N. - Sistema de Implante Nacional S.A. | K200992<br>S.I.N. Dental Implant System<br>S.I.N. - Sistema de Implante Nacional S.A. | K193096<br>S.I.N. Dental Implant System<br>S.I.N. - Sistema de Implante Nacional S.A. | | K170398<br>S.I.N. Dental Implant System<br>S.I.N. - Sistema de Implante Nacional S.A. |
| Indications for Use Statement | S.I.N. Dental Implant System is intended for placement in the<br>maxillary or mandibular arch to provide support for single-unit or<br>multi-unit restorations. When a one-stage surgical approach is applied,<br>the S.I.N. Dental Implant System is intended for immediate loading<br>when good primary stability is achieved and with appropriate occlusal<br>loading. | S.I.N. Dental Implant System is intended for placement in the<br>maxillary or mandibular arch to provide support for single-unit or<br>multi-unit restorations. When a one-stage surgical approach is applied,<br>the S.I.N. Dental Implant System is intended for immediate loading<br>when good primary stability is achieved and with appropriate occlusal<br>loading. | S.I.N. Dental Implant System is intended for placement in the<br>maxillary or mandibular arch to provide support for single-unit or<br>multi-unit restorations. When a one-stage surgical approach is applied,<br>the S.I.N. Dental Implant System is intended for immediate loading<br>when good primary stability is achieved and with appropriate occlusal<br>loading.<br>All digitally-designed custom abutments for use with Interface<br>CAM abutments are to be sent to a S.I.N.-validated milling<br>manufacture. | | S.I.N. Dental Implant System is intended for placement in the<br>maxillary or mandibular arch to provide support for single-unit or<br>multi-unit restorations. When a one-stage surgical approach is applied,<br>the S.I.N. Dental Implant System is intended for immediate loading<br>when good primary stability is achieved and with appropriate occlusal<br>loading. Revolution Compact with a 6 mm length is intended for<br>delayed loading only. |
| Reason for Predicate Device | Not applicable | Implant design; HAnano implant surface | Implant design; HAnano implant surface | | Implant design |
| Product Codes | DZE | DZE, NHA | DZE, NHA | | DZE, NHA |
| Intended Use | Functional and esthetic rehabilitation of the edentulous<br>mandible or maxilla | Functional and esthetic rehabilitation of the edentulous<br>mandible or maxilla | Functional and esthetic rehabilitation of the edentulous<br>mandible or maxilla | | Functional and esthetic rehabilitation of the edentulous<br>mandible or maxilla |
| Implant Designs | | | | | |
| Prosthetic Interface Connections | Morse taper (CM, 11.5°) and External Hex (HE) | Morse taper (CM, 11.5°) | Morse taper (CM, 16°) and External Hex (HE) | | External hex (HE), and Internal Hex (HI) |
| Body/Platform Diameters, mm<br>Lengths, mm<br>Interface | Epikut CM<br>3.5/3.5; 3.8/3.8; 4.5/4.5; 5.0/5.0<br>8.5, 10, 11.5, 13, 15<br>Morse taper interface (CM, 11.5°) | Strong SW<br>3.5/3.5; 3.8/3.8; 4.5/4.5; 5.0/5.0<br>8.5, 10, 11.5, 13, 15<br>Morse taper interface (CM, 11.5°) | | | |
| Body/Platform Diameters, mm<br>Lengths, mm<br>Interface | Epikut Plus CM<br>3.5/3.5; 3.8/3.8; 4.5/4.5; 5.0/5.0<br>8.5, 10, 11.5, 13, 15<br>Morse taper interface (CM, 11.5°) | Strong SW Plus<br>3.5/3.5; 3.8/3.8; 4.5/4.5; 5.0/5.0<br>8.5, 10, 11.5, 13, 15<br>Morse taper interface (CM, 11.5°) | | | |
| Body/Platform Diameters, mm<br>Lengths, mm<br>Interface | Epikut HE<br>3.5/3.65; 4.5/4.5; 5.0/5.0<br>7*, 8.5, 10, 11.5, 13, 15<br>* no 7 mm length for 4.5 mm body<br>External hex interface (HE) | | | | Strong SW HE<br>3.5/3.65; 3.75/4.1; 4.0/4.1; 4.5/4.5; 5.0/5.0 |
| Body/Platform Diameters, mm<br>Lengths, mm<br>Interface | Epikut Plus HE<br>3.5/3.65; 4.5/4.4; 5.0/5.0<br>7*, 8.5, 10, 11.5, 13, 15<br>* no 7 mm length for 4.5 mm body<br>External hex interface (HE) | | Strong SW HE Plus<br>3.5/3.65; 3.75/4.1; 4.0/4.1; 4.5/4.5; 5.0/5.0<br>7*, 8.5, 10, 11.5, 13, 15<br>* no 7 mm length for 4.5 mm body<br>External hex interface (HE) | | 7*, 8.5, 10, 11.5, 13, 15 |
| Implant Material | All implants: unalloyed titanium, ASTM F67 | All implants: unalloyed titanium, ASTM F67 | All implants: unalloyed titanium, ASTM F67 | | * no 7 mm length for 4.5 mm body<br>External hex interface (HE) |
| Implant Endosseous Surface | All implants: acid-etched;<br>HAnano applied to the Epikut Plus CM and Epikut Plus HE<br>implants | All implants: acid-etched;<br>HAnano applied to the Strong SW implants | All implants: acid-etched;<br>HAnano applied to the Strong SW implants | | |
| Implant Placement | CM designs: 1.5 mm intra-bony (sub-crestal)<br>HE designs: at bone level (at the crest) | CM designs: 1.5 mm intra-bony (sub-crestal) | HE designs: at bone level (at the crest) | | All implants: unalloyed titanium, ASTM F67 |
| How Provided | | | | | All implants: acid-etched |
| Implants | Sterile by gamma irradiation | Sterile by gamma irradiation | Sterile by gamma irradiation | | HE designs: at bone level (at the crest) |
| Usage | Single patient, single use | Single patient, single use | Single patient, single use | | Sterile by gamma irradiation |
| | Single patient, single use | | | | |
| | | | | | |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.