KONG®-TL VBR System KONG®-TL VBR System devices are intended for use in the thoracolumbar spine (T1 to L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture). When used in the thoracolumbar spine, the KONG®-TL VBR System is intended to be used with FDA-cleared supplemental fixation appropriate for the implanted level, including icotec Pedicle Screw Systems. These implants may be used with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft. These implants are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracolumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graff used at the surgeon's discretion. When used with the VADER® Pedicle System, the KONG®-TL VBR System is intended to stabilize the thoracic and/or lumbar spine as an adjunct to fusion in patients with spinal infection (e.g., spondylodiscitis, osteomyelitis) and spinal instability due to infection, surgical debridement, or decompression. KONG®-C VBR System KONG®-C VBR System devices are intended for use in the cervical spine (from C2 to T1) in skeletally mature patients to replace a diseased or damaged vertebral body caused by tumor, fracture, or osteomyelitis, or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in cervical degenerative disorders. KONG®-C VBR System is intended to be used with supplemental fixation cleared by the FDA for use in the cervical spine. These implants may be used with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft. These implants are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion. When used with the icotec Anterior Cervical Plate System, the KONG®-C VBR System is intended to stabilize the cervical (from C2 to T1) spine as an adjunct to fusion in patients with spinal infection (e.g., spondylodiscitis, osteomyelitis) and spinal instability due to infection, surgical debridement, or decompression.
Device Story
Modular vertebral body replacement (VBR) system for anterior spinal stabilization; replaces diseased/damaged vertebral bodies; used in thoracic, lumbar, and cervical spine. System comprises expandable/lordotic body structures, end plates, and fixation screws. Implants feature partially hollow geometry for bone graft packing; surface treads/spikes enhance primary fixation. Manufactured from BlackArmor® (carbon fiber-reinforced PEEK) via composite flow molding; radiolucent; includes tantalum markers (ASTM F560) for visibility and cp-titanium coating (ASTM F1580) for bone apposition. Used by surgeons in clinical settings; requires supplemental fixation (e.g., pedicle screws, anterior cervical plates). Provides structural integrity to spinal column; restores stability in tumor/trauma/infection cases; reduces migration risk via large contact surfaces.
Clinical Evidence
Clinical data provided supports safety and effectiveness profile for VBR use; however, data did not support specific material/technological benefits for infection indications. Non-clinical testing performed per FDA MR safety guidance.
Technological Characteristics
Materials: BlackArmor® (Carbon/PEEK composite), tantalum markers (ASTM F560), cp-titanium coating (ASTM F1580). Modular design; expandable/lordotic body structures; various heights/widths/angles. Radiolucent. Sterile supply. No software/electronics.
Indications for Use
Indicated for patients with collapsed, damaged, or unstable vertebral bodies in the thoracolumbar (T1-L5) or cervical (C2-T1) spine due to tumor, trauma, or osteomyelitis; includes reconstruction following corpectomy for cervical degenerative disorders; indicated for stabilization as an adjunct to fusion in patients with spinal infection; intended for use in skeletally mature patients for cervical applications.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
Predicate Devices
KONG®-TL VBR System and KONG®-C VBR System (K200235)
{0}------------------------------------------------
April 4, 2024
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
icotec AG % Justin Eggleton VP, Musculoskeletal Regulatory Affairs MCRA, LLC 803 7th Street NW Washington, District of Columbia 20001
Re: K232790
Trade/Device Name: KONG®-TL VBR System and KONG® C VBR System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal Intervertebral Body Fixation Orthosis Regulatory Class: Class II Product Code: MQP, PLR Dated: March 5, 2024 Received: March 5, 2024
Dear Justin Eggleton:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Eileen
Digitally signed
by Eileen Cadel -
S
Cadel -S Date: 2024.04.04
13:33:57 -04'00'
for
Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K232790
Device Name KONG®-TL VBR System and KONG®-C VBR System
Indications for Use (Describe) KONG®-TL VBR System
KONG®-TL VBR System devices are intended for use in the thoracolumbar spine (T1 to L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture). When used in the thoracolumbar spine, the KONG®-TL VBR System is intended to be used with FDA-cleared supplemental fixation appropriate for the implanted level, including icotec Pedicle Screw Systems. These implants may be used with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft. These implants are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracolumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graff used at the surgeon's discretion.
When used with the VADER® Pedicle System, the KONG®-TL VBR System is intended to stabilize the thoracic and/or lumbar spine as an adjunct to fusion in patients with spinal infection (e.g., spondylodiscitis, osteomyelitis) and spinal instability due to infection, surgical debridement, or decompression.
KONG®-C VBR System
KONG®-C VBR System devices are intended for use in the cervical spine (from C2 to T1) in skeletally mature patients to replace a diseased or damaged vertebral body caused by tumor, fracture, or osteomyelitis, or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in cervical degenerative disorders. KONG®-C VBR System is intended to be used with supplemental fixation cleared by the FDA for use in the cervical spine.
These implants may be used with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft. These implants are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion.
When used with the icotec Anterior Cervical Plate System, the KONG®-C VBR System is intended to stabilize the cervical (from C2 to T1) spine as an adjunct to fusion in patients with spinal infection (e.g., spondylodiscitis, osteomyelitis) and spinal instability due to infection, surgical debridement, or decompression.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| <div> <span> <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | |
| <div> <span> □ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | |
| CONTINUE ON A SEPARATE PAGE IF NEEDED. | |
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
| Device Trade Name: | KONG®-C VBR System and KONG®-TL VBR System |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | icotec ag<br>Industriestrasse 12<br>9450 Altstätten<br>Switzerland<br>www.icotec-medical.com<br>Phone: +41 71 757.0000 |
| Contact: | Ms. Marina Hess<br>CQO/Management Representative<br>icotec ag |
| Prepared by: | Mr. Justin Eggleton<br>Vice President, Head of Musculoskeletal Regulatory Affairs<br>MCRA, LLC<br>803 7th Street NW<br>Washington, DC 20001<br>jeggleton@mcra.com |
| Date Prepared: | April 1, 2024 |
| Classifications: | 21 CFR §888.3060, Spinal intervertebral body fixation orthosis |
| Class: | II |
| Product Codes: | MQP, PLR |
| Primary Predicate: | KONG®-TL VBR System and KONG®-C VBR System (K200235) |
| Reference Device: | icotec VADER® Pedicle System (K232628) |
{5}------------------------------------------------
#### Indications For Use:
## KONG®-TL VBR System
KONG®-TL VBR System devices are intended for use in the thoracolumbar spine (T1 to L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e., fracture). When used in the thoracolumbar spine, the KONG®-TL VBR System is intended to be used with FDA-cleared supplemental fixation appropriate for the implanted level, including icotec Pedicle Screw Systems. These implants may be used with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft. These implants are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracolumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion.
When used with the VADER® Pedicle System, the KONG®-TL VBR System is intended to stabilize the thoracic and/or lumbar spine as an adjunct to fusion in patients with spinal infection (e.g., spondylodiscitis, osteomyelitis) and spinal instability due to infection, surgical debridement, or decompression.
## KONG®-C VBR System
KONG®-C VBR System devices are intended for use in the cervical spine (from C2 to T1) in skeletally mature patients to replace a diseased or damaged vertebral body caused by tumor. fracture, or osteomyelitis, or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in cervical degenerative disorders. KONG®-C VBR System is intended to be used with supplemental fixation cleared by the FDA for use in the cervical spine.
These implants may be used with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft. These implants are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion.
When used with the icotec Anterior Cervical Plate System, the KONG®-C VBR System is intended to stabilize the cervical (from C2 to T1) spine as an adjunct to fusion in patients with spinal infection (e.g., spondylodiscitis, osteomyelitis) and spinal instability due to infection, surgical debridement, or decompression.
{6}------------------------------------------------
## Device Description: KONG®-TL VBR System
The KONG®-TL VBR System is a vertebral body replacement system for anterior stabilization of the thoracic and lumbar spine. This device serves as a replacement for vertebral bodies in tumorous and traumatic diseases which lead to instabilities in the anterior support or compression of neural structures or diseases that make an infection repair necessary. The KONG®-TL VBR System is intended for use with an additional dorsal or anterolateral fixation system (e.g., icotec Pedicle System). The KONG®-TL VBR System is a modular design featuring an expandable body, extensions, and end plates with various heights, widths, and angles. The system consists of expandable body structures that range in height from 19 to 96 mm, have a diameter of 19 mm and surface treads/spikes on its cranial and caudal end plate surfaces that serve to guide and anchor the implant. The system also consists of rectangular and round end plate configurations. The end plate round configuration has a geometry that ranges from 0°, 4°, 8° and 12° and size that ranges from 26 to 43 mm. The end plate rectangular configuration has geometry that ranges from 0°, 4°, 8° and 12° and size that ranges from 33 to 52 mm. The end plates are locked with a screw onto the main body. The tread surfaces and spikes help to improve primary fixation stability of the end plates. The large contact surface of the end plates reduces the risk of migration. The screw for fixation of the end plates to the body has a length of 6 mm and a thread size of M8.
The KONG®-TL VBR System is made of BlackArmor® material, a carbon fiber reinforced polyetheretherketone (Carbon/PEEK) composite. Due to the excellent biocompatibility and material properties. Carbon/PEEK has been in clinical use in spine surgery for more than 20 years world-wide. The BlackArmor® material is a combination of continuous, high strength carbon fiberreinforced PEEK and icotec's composite flow molding (CFM) process. The result is a non-metallic implant component with an interwoven 3D fiber architecture that provides strength and endurance. BlackArmor® is radiolucent in all diagnostic imaging modes (MRI, CT and X-ray) and will therefore not create imaging artifacts. Embedded tantalum markers (per ASTM F560) ensure the required radiologic visibility of the implant during surgery and follow-up. The KONG®-TL VBR System incorporates a rough cp-titanium coating (per ASTM F1580), designed to improve direct bone apposition.
The partially hollow geometry of the implants (endplates and extension) allows them to be packed with bone graft. The implants are supplied sterile and are available in a variety of heights, footprints and anatomical shapes to accommodate patient anatomy.
## KONG®-C VBR System
The KONG®-C VBR System is a vertebral body replacement system for anterior stabilization of the cervical spine. This device serves as a replacement for vertebral bodies in tumorous and traumatic diseases which lead to instabilities in the anterior support or compression of neural structures or diseases that make an infection repair necessary. The KONG®-C VBR System is a modular design with various heights, widths, and angles. The system consists of a lordotic body structure that ranges in height from 5 - 55mm with a width and depth of 13mm, and surface treads on its cranial and caudal end plate surfaces that serve to guide and anchor the implant. The overall height of the VBR device including body and end plates ranges from 16 to 66 mm. The end plates
{7}------------------------------------------------
have a geometry that range from 0 to 6° and are manufactured in the following sizes: 14 x 13mm, 16 x 13mm. and 18 x 14mm. The tread surfaces help to improve primary fixation stability of the end plates. The large contact surface of the end plates reduces the risk of migration. The screw to secure the end plates to the body has a length of 7 mm and a thread size of M6.
The KONG®-C VBR System is composed of BlackArmor® material, a carbon fiber reinforced polyetheretherketone (Carbon/PEEK) composite. Due to the excellent biocompatibility and material properties, Carbon/PEEK has been in clinical use in spine surgery for more than 20 years world-wide. The BlackArmor® material is a combination of continuous, high strength carbon fiberreinforced PEEK and icotec's composite flow molding (CFM) process. The result is a non-metallic implant component with an interwoven 3D fiber architecture that provides strength and endurance. BlackArmor® is radiolucent in all diagnostic imaging modes (MRI, CT and X-ray) and will therefore not create imaging artifacts. Embedded tantalum markers (per ASTM F560) ensure the required radiologic visibility of the implant during surgery and follow-up. The KONG®-C VBR System incorporates a rough cp-titanium coating (per ASTM F1580), designed to improve direct bone apposition.
The implants are supplied sterile and are available in a variety of heights, footprints and anatomical shapes to accommodate patient anatomy.
#### Predicate Device:
icotec ag submits the following information in this Premarket Notification to demonstrate that, for the purposes of FDA's regulation of medical devices. KONG®-TL VBR System and KONG®-C VBR System are substantially equivalent in indications, design principles, and performance to the following predicate devices, which have been determined by FDA to be substantially equivalent to pre-amendment devices:
Primary Predicate: KONG®-TL VBR System and KONG®-C VBR System (K200235) Reference Device: icotec VADER Pedicle System (K222789)
## Performance Testing Summary:
The clinical data provided supports a substantially equivalent safety and effectiveness profile for the use of vertebral body replacement devices for the indications for use. However, the provided clinical data did not support that the specific material or unique technological characteristics of the subject device components provides additional benefit relative to other vertebral body replacement devices for the indications for use related to infection.
Additionally, non-clinical performance testing was performed in accordance with the FDA Guidance Document "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment" to support labeling the subject device as MR Conditional.
## Substantial Equivalence:
The subject device was demonstrated to be substantially equivalent to the predicate device (K200235) with respect to indications for use, intended use, design, and performance.
Non-clinical testing data and clinical data demonstrate the performance of the subject device is substantially equivalent to that of the predicate device, and support the performance of the subject device in its expanded indications for use.
{8}------------------------------------------------
The updated manufacturing does not raise new issues or concerns of safety or efficacy.
### Conclusion:
The subject device and the predicate devices have intended use, have similar technological characteristics, and are made of similar materials. The subject and predicate devices are packaged in similar materials and are sterilized using similar methods. The data included in this submission demonstrate substantial equivalence to the predicate devices listed above. The KONG®-TL VBR System and KONG®-C VBR System are as safe, as effective, and performs as well as, or better, than the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.