K200235 · Icotec AG · MQP · May 29, 2020 · Orthopedic
Device Facts
Record ID
K200235
Device Name
KONG-TL VBR System, KONG-C VBR System
Applicant
Icotec AG
Product Code
MQP · Orthopedic
Decision Date
May 29, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3060
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
KONG®-C VBR System devices are intended for use in the cervical spine (from C2 to T1) in skeletally mature patients to replace a diseased or damaged vertebral body caused by tumor, fracture, or osteomyelitis, or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in cervical degenerative disorders. KONG®-C VBR System is intended to be used with supplemental fixation cleared by the FDA for use in the cervical spine. When used at more than two levels, supplemental fixation should include posterior fixation which is cleared by the FDA. These implants may be used with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft. These implants are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion. KONG®-TL VBR System devices are intended for use in the thoracolumbar spine (T1 to L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e. fracture). When used in the thoracolumbar spine, the KONG®-TL VBR System is intended to be used with FDA-cleared supplemental fixation appropriate for the implanted level, including icotec Pedicle Screw Systems. These implants may be used with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft. These implants are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracolumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion.
Device Story
Vertebral body replacement (VBR) system; provides spinal column stability; replaces diseased/damaged/collapsed vertebral bodies. Implants consist of carbon fiber-reinforced PEEK (BlackArmor®) with rough titanium coating (Ti-iT®). Available in various heights, footprints, and lordosis angles. Used in cervical or thoracolumbar spine; requires supplemental FDA-cleared fixation. Surgeon-operated in clinical/surgical setting. Restores spinal integrity; supports bone graft (autograft/allogenic). Benefits patients by stabilizing spine post-corpectomy or trauma.
Clinical Evidence
Bench testing only. No clinical data provided. Performance validated via static and dynamic compression, static and dynamic torsion per ASTM F2077, subsidence per ASTM F2267, and expulsion testing of worst-case constructs.
Indicated for skeletally mature patients requiring vertebral body replacement in the cervical (C2-T1) or thoracolumbar (T1-L5) spine due to tumor, fracture, osteomyelitis, or collapse. Used for reconstruction following corpectomy for decompression of spinal cord/neural tissues in degenerative disorders. Contraindications include patients where fusion is not intended or possible, though limited-time use is permitted for advanced-stage tumor patients with short life expectancy.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 29, 2020
icotec ag % Mr. Samuel Pollard Associate Director, Regulatory Affairs MCRA, LLC 1050 K Street NW, Suite 1000 Washington, District of Columbia 20001
Re: K200235
Trade/Device Name: KONG-TL® VBR System, KONG-C® VBR System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: MOP, PLR Dated: May 7, 2020 Received: May 7, 2020
Dear Mr. Pollard:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brent Showalter, Ph.D. Acting Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200235
Device Name
#### KONG®-TL VBR System and KONG®-C VBR System
Indications for Use (Describe)
Cervical
KONG®-C VBR System devices are intended for use in the cervical spine (from C2 to T1) in skeletally mature patients to replace a diseased or damaged vertebral body caused by tumor, fracture, or osteomyelitis, or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in cervical degenerative disorders. KONG®-C VBR System is intended to be used with supplemental fixation cleared by the FDA for use in the cervical spine.
These implants may be used with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft. These implants are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion.
#### Thoracolumbar
KONG®-TL VBR System devices are intended for use in the thoracolumbar spine (T1 to L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e. fracture). When used in the thoracolumbar spine, the KONG®-TL VBR System is intended to be used with FDA-cleared supplemental fixation appropriate for the implanted level, including icotec Pedicle Screw Systems.
These implants may be used with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft. These implants are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracolumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <span> <b> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </b> </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
|---------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K200235 - 510(k) Summary
| Device Trade Name: | KONG®-TL VBR System and<br>KONG®-C VBR System |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer: | icotec ag<br>Industriestrasse 12<br>9450 Altstätten<br>Switzerland |
| Contact: | Ms. Marina Hess<br>CQO/Management Representative<br>icotec ag<br>Industriestrasse 12<br>9450 Altstätten, Switzerland<br>Phone: +41 71 757 0024<br>Fax: +41 71 757 0001<br>marina.hess@icotec.ch |
| Prepared by: | Mr. Samuel Pollard<br>Associate Director, Regulatory Affairs<br>MCRA, LLC<br>1050 K Street NW, Suite 1000<br>Washington, DC 20005<br>Phone: 202.552.5819<br>Fax: 202.552.5798<br>spollard@mcra.com |
| Date Prepared: | January 30, 2020 |
| Classification:<br>Class: | 21 CFR §888.3060, Spinal intervertebral body fixation orthosis<br>II |
| Product Code: | MQP, PLR |
| Primary Predicates: | Omnia Medical VBR (K191778), |
| Additional Predicates: | NuVasive Monolith Cervical Corpectomy System (K180550), Ulrich<br>Small VBR (K192117), Solidity™ Vertebral Body Replacement<br>(K181921), CAPRI Corpectomy Cage System (K180665), Obelisc<br>Vertebral Body Replacement Device (K060416), Synthes CRL<br>System (K103320) |
| Reference Devices: | icotec VADER®one Pedicle System MIS and LightMore® Pedicle<br>System 6.0 (K190545), icotec Interbody Cage System (K172480) |
{5}------------------------------------------------
# Indications for Use:
# Cervical
KONG®-C VBR System devices are intended for use in the cervical spine (from C2 to T1) in skeletally mature patients to replace a diseased or damaged vertebral body caused by tumor, fracture, or osteomyelitis, or for reconstruction following corpectomy performed to achieve decompression of the spinal cord and neural tissues in cervical degenerative disorders. KONG®-C VBR System is intended to be used with supplemental fixation cleared by the FDA for use in the cervical spine. When used at more than two levels, supplemental fixation should include posterior fixation which is cleared by the FDA.
These implants may be used with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft. These implants are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the cervical spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion.
## Thoracolumbar
KONG®-TL VBR System devices are intended for use in the thoracolumbar spine (T1 to L5) to replace a collapsed, damaged, or unstable vertebral body due to tumor or trauma (i.e. fracture). When used in the thoracolumbar spine, the KONG®-TL VBR System is intended to be used with FDA-cleared supplemental fixation appropriate for the implanted level, including icotec Pedicle Screw Systems.
These implants may be used with autograft or allogenic bone graft comprising cancellous and/or corticocancellous bone graft. These implants are intended to restore the integrity of the spinal column even in the absence of fusion for a limited time period in patients with advanced stage tumors involving the thoracolumbar spine in whom life expectancy is of insufficient duration to permit achievement of fusion, with bone graft used at the surgeon's discretion.
## Device Description:
The KONG®-TL VBR System and the KONG®-C VBR System are vertebral body replacement systems designed to improve the stability of the cervical and thoracolumbar spine. The devices are manufactured from high strength carbon fiber reinforced polyetheretherketone (Carbon/ PEEK. BlackArmor®) and incorporate a rough titanium coating (Ti-TT®). The devices are intended to be used with supplemental spinal fixation.
Each implant is provided sterile and is available in an assortment of heights, footprints, and lordosis angles to accommodate patient anatomy.
## Primary Predicate Devices:
The KONG®-TL VBR System and the KONG®-C VBR System is substantially equivalent to the Omnia Medical VBR (K191778).
# Additional Predicate Devices:
{6}------------------------------------------------
- NuVasive Monolith Cervical Corpectomy System (K180550) ●
- Ulrich Small VBR (K192117)
- SolidityTM Vertebral Body Replacement (K181921)
- CAPRI Corpectomy Cage System (K180665)
- Obelisc Vertebral Body Replacement Device (K060416)
- Synthes XRL System (K103320)
## Reference Devices:
The following devices are used as reference devices throughout this submission:
- icotec VADER®one Pedicle System MIS and LightMore® Pedicle System 6.0 0 (K190545).
- icotec Interbody Cage System (K172480)
## Performance Testing Summary:
The non-clinical tests performed by the company include static and dynamic compression, static and dynamic torsion per ASTM F2077, subsidence per ASTM F2267, and expulsion testing of the worst-case construct. The results of the performed tests demonstrate that the KONG®-C VBR System and the KONG®-TL VBR System are substantially equivalent to legally marketed predicate devices.
#### Technological Characteristics:
the KONG®-C VBR System and the KONG®-TL VBR System possesses the same technological characteristics as one or more of the predicate devices. These characteristics include, the basic design (expanding corpectomy spacer), the carbon fiber reinforced polyetheretherketone (Carbon/PEEK, BlackArmor®) and a rough titanium coating (Ti-iT®), the endplate sizes (dimensions are comparable to those offered by predicate systems), the spiked design (the spiked feature is seen among multiple VBR system endplates) and the fundamental scientific technology.
#### Substantial Equivalence:
The subject devices were demonstrated to be substantially equivalent to predicates cited in the above with respect to indications, design, materials, function, manufacturing, and performance. The nonclinical tests performed by the include static and dynamic compression, static and dynamic torsion per ASTM F2077, subsidence per ASTM F2267, and expulsion testing of the worst-case construct. The results of the performed tests demonstrate that the KONG®-C VBR System and the KONG®-TL VBR System is substantially equivalent to legally marketed predicate devices.
#### Conclusion:
The purpose of the traditional 510(k) is to receive regulatory clearance to introduce the KONG® TL VBR System and the KONG®-C VBR System to interstate commerce. Substantial equivalence has been demonstrated to the cited predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.