The Solo+ Tympanostomy Tube Device is intended to deliver a tympanostomy tube through the tympanic membrane (TM) of the patient. It combines the separate functions of creating a myringotomy, and positioning and placing a ventilation tube across the TM.
Device Story
Single-use, sterile, all-in-one surgical instrument for tympanostomy tube placement. Device features a cartridge with a distal myringotomy knife and a handpiece with an activation button. Operator (physician) creates myringotomy using the knife, advances device until the tube flange contacts the tympanic membrane, and presses the activation button to retract the knife and deploy the tube. Used in office settings for pediatric patients. Provides middle ear ventilation. Eliminates need for separate instruments for myringotomy and tube placement; reduces need for operating room procedures.
Clinical Evidence
Multi-site clinical study in an office setting with 20 pediatric patients (40 ears, mean age 12.5 months). Primary endpoints: successful tube placement and procedure completion without additional instruments. Results: 100% (20/20) success rate for tube placement; 87.5% (35/40) of ears completed without additional instruments. No device or procedure-related serious adverse events reported. Data demonstrates safety and effectiveness comparable to the predicate Hummingbird TTS.
Technological Characteristics
Single-use, sterile, handheld mechanical instrument. Materials evaluated per ISO 10993-1 (biocompatibility). Delivery system categorized as surface device (contacting breached/compromised surfaces); tympanostomy tube categorized as implant device (contacting tissue/bone). Mechanical actuation via handpiece button to retract knife and deploy tube. No electronic or software components described.
Indications for Use
Indicated for pediatric patients 6-24 months old for the delivery of a tympanostomy (ventilation) tube through the tympanic membrane in office settings.
Regulatory Classification
Identification
A tympanostomy tube is a device that is intended to be implanted for ventilation or drainage of the middle ear. The device is inserted through the tympanic membrane to permit a free exchange of air between the outer ear and middle ear. A type of tympanostomy tube known as the malleous clip tube attaches to the malleous to provide middle ear ventilation. The device is made of materials such as polytetrafluoroethylene, polyethylene, silicon elastomer, or porous polyethylene.
Predicate Devices
Hummingbird Tympanostomy Tube System (HTTS), Preceptis Medical Inc. (K200952)
Reference Devices
Otological Ventilation Tube - Sheehy Design, Micromedics, Inc. (K830228)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" followed by the words "U.S. Food & Drug Administration".
May 20, 2024
AventaMed DAC Keith Jansen President and CEO Rubicon Centre, Rossa Avenue, Bishopstown, Cork T12 Y275 Ireland
Re: K232702
Trade/Device Name: Solo+ Tympanostomy Tube Device (TTD) Regulation Number: 21 CFR 874.3880 Regulation Name: Tympanostomy Tube Regulatory Class: Class II Product Code: ETD Dated: April 19, 2024 Received: April 19, 2024
Dear Keith Jansen:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Page 2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shuchen Peng -S
Shu-Chen Peng, Ph.D. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232702
Device Name Solo+ Tympanostomy Tube Device (TTD)
Indications for Use (Describe)
The Solo+ Tympanostomy Tube Device is intended to deliver a tympanostomy tube (also referred to as a ventilation tube) through the tympanic membrane of the patient and is indicated to be used in office settings for pediatric patients 6-24 months old.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-family: Arial, sans-serif;"> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span style="font-family: Arial, sans-serif;"> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for AventaMed, a medical company specializing in ENT surgery. The logo features the company name in a combination of blue and gray lettering, with the word "Aventa" in blue and "Med" in gray. Below the name is the tagline "Advancing ENT Surgery" in smaller, gray font. To the right of the name is a stylized blue image of an ear.
# 510(k) Summary
# General Information
| Date Prepared: | 20th May 2024 |
|-----------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Classification: | Class II, 21 CFR 874.3880 Tympanostomy Tube |
| Product Code: | ETD |
| Trade Name: | Solo+ Tympanostomy Tube Device (TTD) |
| Common Name: | Tympanostomy Tube |
| Model Numbers: | Solo+ Tympanostomy Tube Handpiece (Catalogue#: 12115-100-000)<br>Solo+ Tympanostomy Tube Cartridge (Catalogue#: 12115-200-000) |
| Owners Name | AventaMed DAC<br>Rubicon Centre<br>Rossa Avenue<br>Bishopstown Cork<br>T12 Y275<br>Ireland |
| Contact Person: | Keith Jansen, President and CEO |
| AventaMed DAC: | Rubicon Centre<br>Rossa Avenue<br>Bishopstown<br>Cork<br>T12 Y275<br>Ireland<br>Tel: +353 21 492 8980<br>Email: keith.jansen@aventamed.com |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for AventaMed, a company that specializes in advancing ENT surgery. The logo features the company name in a combination of blue and gray lettering, with the word "Aventa" in blue and "Med" in gray. Below the name is the tagline "Advancing ENT Surgery" in smaller, gray font. To the right of the name is a stylized blue image of an ear.
# Intended Use
The Solo+ Tympanostomy Tube Device is intended to deliver a tympanostomy tube through the tympanic membrane (TM) of the patient. It combines the separate functions of creating a myringotomy, and positioning and placing a ventilation tube across the TM.
# Indications for Use
The Solo+ Tympanostomy Tube Device is intended to deliver a tympanostomy tube (also referred to as a ventilation tube) through the tympanic membrane of the patient and is indicated to be used in office settings for pediatric patients 6-24 months old.
# Predicate Device
- Hummingbird Tympanostomy Tube System (HTTS), Preceptis Medical Inc. (K200952) .
# Reference Device
- . Otological Ventilation Tube - Sheehy Design, Micromedics, Inc. (K830228)
# Device Description
The Solo+ Tympanostomy Tube Device TTD (Solo+ TTD) is a single use, sterile, "all-in-one" surgical instrument that rapidly places a tympanostomy tube across the tympanic membrane of a patient. It combines the traditionally separate functions of creating a myringotomy, and positioning and placing a tympanostomy tube across the tympanic membrane. Placement of the tube provides ventilation to the middle ear space through the tympanic membrane.
To use the device, the user creates a myringotomy with the device's myringotomy knife, which is located at its distal tip of the Cartridge. The user advances the device until the tympanostomy tube outer flange reaches the tympanic membrane. The user then actuates the device by pressing the activation button on the Handpiece. This retracts the myringotomy knife construct and deploys the tube across the tympanic membrane.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for AventaMed, a company that specializes in advancing ENT surgery. The logo features the company name in a combination of blue and gray lettering, with the word "Aventa" in blue and "Med" in gray. Below the name is the tagline "Advancing ENT Surgery" in smaller, gray font. To the right of the name is a stylized blue image of an ear.
# Materials
The Solo+ TTD is comprised of materials that are commonly used in medical device applications. The biological safety tests were performed in accordance with ISO 10993-1 (Biological evaluation of medical devices -- Part 1: Evaluation and testing). For testing the device has been categorized in two parts, the Solo+ Delivery System which has a category of a surface device contacting breached or compromised surfaces and the Solo+ Tympanostomy Tube which has a category of Implant Device contacting tissue/bone.
The tests performed to demonstrate the biocompatibility of the device were:
- Physical and Chemical Characterization
- . Cytotoxicity
- Sensitization ●
- Irritation / Intracutaneous Reactivity ●
- Acute Systemic Toxicity
- Subchronic Toxicity
- Subacute Toxicity ●
- Implantation Short-Term and Long Term ●
- Material- Mediated Pyrogenicity ●
## Comparison of Technological Characteristics of Solo+ TTD compared to the Predicate Device (Preceptis, Hummingbird TTS) and Reference Device (Micromedics, Otological Ventilation Tube - Sheehy Design)
The Solo+ TTD and the equivalent commercialized predicate device Preceptis Hummingbird TTS were evaluated for substantial equivalence. No significant difference in clinical, and biological parameters was identified between the devices have the same intended patient and user populations and intended use; to deliver a tympanostomy tube through the tympanic membrane of the patient. Both devices combine the separate functions of creating a myringotomy, and positioning and placing a ventilation tube across the tympanic membrane. Technological equivalence is demonstrated via comparative test data and assessment of the
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for AventaMed, a company specializing in advancing ENT surgery. The logo features the company name in a combination of blue and gray lettering, with the word "Aventa" in blue and "Med" in gray. Below the name is the tagline "Advancing ENT Surgery" in smaller, gray font. To the right of the name is a stylized blue drawing of an ear.
primary technological characteristics. Biological equivalence is demonstrated via reference to the material used in each device and the biocompatibility testing which demonstrates that the devices are biologically safe for their intended use.
Where there is a difference in geometry of the tympanostomy tube to the Preceptis Hummingbird TTS tympanostomy tube, equivalence is demonstrated to the reference device; the Micromedics Otological Ventilation Tube - Sheehy Design via review of the appropriate technological characteristics.
Via the comparative analysis and the test data generated and reviewed, the Solo+ TTD Device has been shown to be substantially equivalent to the commercialized Hummingbird TTS.
## Non-Clinical Information
The determination of substantial equivalence is also based on an assessment of non-clinical engineering tests, as listed in Table 1 and Table 2.
# Table 1: List of Design Verification Tests
| Design Verification Tests | |
|--------------------------------------|----------------------------------|
| Device Visual and Dimensional Checks | Packaging testing |
| Device Puncture Measurement | Biocompatibility Testing |
| Simulated Use Testing | MRI safety testing |
| Cartridge Functional Tests | Sterilization Validation |
| Handpiece Functional Tests | Residuals testing |
| Aging Testing | Bioburden & Endotoxin Validation |
| Transportation Testing | |
# Table 2: List of Design Validation Tests
| Design Validation Tests | |
|------------------------------|--------------------|
| IFU/Label comprehension | Device placement |
| Package suitability | Device ease of use |
| Device functionality testing | Device Disposal |
| Visualization assessment | |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for AventaMed. The logo features the name "AventaMed" in a combination of blue and gray colors, with the "Aventa" portion in blue and the "Med" portion in gray. Below the name, there is a tagline that reads "Advancing ENT Surgery." To the right of the name, there is a stylized blue image of an ear.
The test results demonstrate that the Solo+ TTD Device meets the requirements in the applicable standards and specifications and is substantially equivalent to a legally marketed predicate device.
# Clinical Information
A clinical study using the Solo+ TTD has demonstrated substantial equivalence to the predicate device, Preceptis Hummingbird TTS intra-operatively and through 1st post-operative follow-up in an in-office setting.
# In-office setting: 6-24 months old.
In a multi-site study, a total of 20 patients (40 ears), underwent tympanostomy procedures using the Solo+ TTD across 2 sites in an in-office setting with topical anesthesia and protective stabilization. The mean age of the patients is 12.5 months (range 6 to 19 months) and a median age of 12.5 months.
Results:
- 20/20 (100%) children received tympanostomy tubes as planned (success) .
- There were no device or procedure related serious adverse events. .
- 87.5% (35/40) of ears were completed using only the Solo+ TTD without the need for . additional instruments to aid in the placement.
The clinical data described above demonstrates that in an in-office setting, the Solo+ TTD is as safe and as effective as the Hummingbird TTS in pediatric patients 6-24 months old. A table comparing study endpoint results with the comparative study for the Hummingbird TTS (K200952) is shown in Table 3.
| Table 3: Study endpoints compared between Solo+ TTD and Hummingbird TTS | | |
|-------------------------------------------------------------------------|--|--|
|-------------------------------------------------------------------------|--|--|
| Clinical Outcome | AventaMed DAC<br>Solo+ TTD | Preceptis Hummingbird TTS<br>(K200952) |
|---------------------------------------------------------------------------------------------------------------|----------------------------|----------------------------------------|
| Successful Placement of the Device<br>without need for an operating room<br>procedure | 100 % (20/20 patients) | 98.9% |
| Delivery Success (placed without<br>the need for additional instruments<br>to aid in placement of the device) | 87.5% (35/40 ears) | 96.9% |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the logo for AventaMed, a medical company. The logo features the company name in a combination of gray and teal colors, with the word "Aventa" in gray and "Med" in teal. Below the name is the tagline "Advancing ENT Surgery" in a smaller font. To the right of the company name is a stylized image of an ear in teal.
# Conclusion
These clinical results, together with the comparison of technological and performance characteristics, demonstrate that there are no differences between the two devices that raise different questions of safety and effectiveness relevant to use of the Solo+ TTD in an office setting. The Solo+ TTD has been demonstrated to be substantially equivalent to the Hummingbird TTS.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.