K200952 · Preceptis Medical, Inc. · ETD · Jun 5, 2020 · Ear, Nose, Throat
Device Facts
Record ID
K200952
Device Name
Hummingbird Tympanostomy Tube System
Applicant
Preceptis Medical, Inc.
Product Code
ETD · Ear, Nose, Throat
Decision Date
Jun 5, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.3880
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Hummingbird® Tympanostomy Tube System (HTTS) is intended to deliver a tympanostomy tube (also referred to as a ventilation tube) through the tympanic of the patient and is indicated to be used in office settings for children 6-24 months old.
Device Story
Hummingbird Tympanostomy Tube System (HTTS) is a single-use, sterile manual surgical instrument for myringotomy and ventilation tube placement. Device features a preloaded ventilation tube within a sharpened sheath. Surgeon manually advances sheath to create myringotomy and position tube under direct visualization; manual actuator on handle retracts sheath to release tube. Used in ENT office settings by physicians; requires patient immobilization (papoose/swaddling) and local anesthesia. Eliminates need for general anesthesia in pediatric patients. Benefits include reduced procedural risk and avoidance of general anesthesia for children 6-24 months old.
Clinical Evidence
Multi-site clinical study in 180 children (360 ears, mean age 13 months) undergoing office-based tympanostomy. 98.9% (178/180) success rate for tube delivery. Procedural adverse event rate 0.56% (2/360 ears). 84.5% of ears completed in one surgical pass; 97.7% in two or fewer. 98.3% of children judged calm post-procedure. Independent video review by clinicians confirmed acceptable tolerance in all 18 cases reviewed. Parent surveys indicated 93.1% support for the alternative to general anesthesia.
Technological Characteristics
Single-use, sterile manual surgical instrument. Combines myringotomy creation and ventilation tube placement. Features a sharpened sheath and manual actuator for tube release. Identical design and materials to predicate K142282, excluding the suction feature. Shelf life: 27 months.
Indications for Use
Indicated for delivery of tympanostomy tubes through the tympanic membrane in children 6-24 months old in office settings.
Regulatory Classification
Identification
A tympanostomy tube is a device that is intended to be implanted for ventilation or drainage of the middle ear. The device is inserted through the tympanic membrane to permit a free exchange of air between the outer ear and middle ear. A type of tympanostomy tube known as the malleous clip tube attaches to the malleous to provide middle ear ventilation. The device is made of materials such as polytetrafluoroethylene, polyethylene, silicon elastomer, or porous polyethylene.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters 'FDA' in white. Next to the blue square, the words 'U.S. FOOD & DRUG' are written in blue, with 'ADMINISTRATION' written in a smaller font size below it.
June 5, 2020
Preceptis Medical, Inc Steve Anderson Chief Executive Officer 10900 89th Ave N Suite 4 Maple Grove, Minnesota 55369
Re: K200952
Trade/Device Name: Hummingbird Tympanostomy Tube System Regulation Number: 21 CFR 874.3880 Regulation Name: Tympanostomy Tube Regulatory Class: Class II Product Code: ETD Dated: April 4, 2020 Received: April 9, 2020
### Dear Steve Anderson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shu-Chen Peng
for Michael J. Ryan Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200952
Device Name
Hummingbird® Tympanostomy Tube System
Indications for Use (Describe)
The Hummingbird® Tympanostomy Tube System (HTTS) is intended to deliver a tympanostomy tube (also referred to as a ventilation tube) through the tympanic of the patient and is indicated to be used in office settings for children 6-24 months old.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Submitter<br>Information: | Preceptis Medical, Inc.<br>10900 89th Avenue North, Suite 4<br>Maple Grove, MN 55369<br>763.568.7819 |
|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Steve Anderson, CEO |
| Date<br>Prepared: | 5 June 2020 |
| Trade Name | Hummingbird® Tympanostomy Tube System (TTS) |
| Product Code | Ear, Nose and Throat Devices: ETD (21 CFR Part 874.3880) |
| Classification | Class II |
| Common<br>Name | Tympanostomy Tube Inserter with pre-loaded ventilation tube |
| Predicate<br>Devices | Preceptis Tympanostomy Tube System, K142282 |
| Device<br>Description | The Hummingbird® Tympanostomy Tube System (HTTS), which<br>includes a preloaded ventilation tube, is a single-use, sterile manual<br>surgical instrument. The HTTS is used to create a myringotomy in<br>the tympanic membrane and place a ventilation tube. The surgeon<br>manually advances the sharpened sheath to create a myringotomy<br>and simultaneously positions the ventilation tube within the<br>myringotomy, always under direct visualization. The surgeon then<br>manually retracts the sharpened sheath away from the myringotomy<br>using the manual actuator located on the handle. The retraction of<br>the sheath releases the tube within the myringotomy. |
| Indications<br>For Use | The Hummingbird® Tympanostomy Tube System (HTTS) is<br>intended to deliver a tympanostomy tube (also referred to as a<br>ventilation tube) through the tympanic membrane of the patient<br>and is indicated to be used in office settings for children 6-24 months<br>old. |
| Technical<br>Characteristics | The HTTS is intended to deliver a tympanostomy tube (also<br>referred to as a ventilation tube) through the tympanic membrane<br>of the patient. It combines the separate functions of creating a<br>myringotomy, positioning and placing a ventilation tube across the<br>tympanic membrane. |
| Comparison to<br>the predicate | The HTTS is identical to the predicate (except the removal of the<br>suction feature), including the Intended Use. All other technological<br>characteristics (e.g., design, material, chemical composition,<br>functionality) are identical and non-clinical performance data tests<br>(except shelf life testing) were not repeated from K142282.<br><br>Shelf life of the device is increased to 27 months (supported by non-<br>clinical testing).<br><br>Indications for use are different from the predicate device as the subject<br>device indications are expanded to include in-office use in children 6-<br>24 months old. |
| Performance<br>Data | In order to evaluate the pediatric-specific performance of the HTTS in<br>an office setting, a multi-site clinical study in children 6-24 months old<br>was performed. A total of 180 children (360 ears) underwent<br>tympanostomy procedures in an ENT office using the HTTS. The<br>mean age of the patients was 13 months (range of 6-24 months).<br>Immobilization (papoose and/or swaddling with a nurse or MA holding<br>the child's head) was used on all patients, and Phenol was used on 173<br>out 180 patients. Results:<br>178/180 (98.9%) children received ventilation tubes in the<br>office as planned. The rate of procedural adverse events was 2/360 ears<br>(0.56%), and there were no serious or unanticipated<br>adverse events reported. The median bi-lateral procedure time was 5:00 (range of<br>2:00-15:32). The recovery of the child was evaluated by the ENT and<br>staff, and 177/180 (98.3%) children were judged as calm<br>and/or no inappropriate crying before leaving the clinic. In 7 ears, tube delivery was completed using additional<br>instruments to the HTTS. 84.5% of ears were completed in one surgical pass; 97.7%<br>were completed in two passes or less; and 2.3% required<br>more than 2 passes. |
{4}------------------------------------------------
{5}------------------------------------------------
- In 130 parent surveys collected at follow-up, 93.1% of parents strongly agree or agree that it was important to have an alternative to general anesthesia and that they would recommend the HTTS office procedure to other parents.
- A committee of clinicians with ENT and/or pediatric . specialty independently reviewed 18 procedure videos to evaluate how the child tolerated the procedure. At each stage of the procedure, the clinician reviewers rated the response of the child in the following categories: no response, mild, moderate or significant. In all 18 cases, each of the reviewers deemed that the child tolerated the procedure acceptably. A bar graph tabulation of the committee video review is as follows:
Image /page/5/Figure/3 description: The image is a bar graph titled "Patient Response Observed by Procedure Phase". The graph shows the percentage of patients reviewed for different procedure phases, including patient entering the room, papoose/swaddling, ear wax removal, topical anesthetic application and tube insertion, and 3 minutes post procedure. The patient responses are categorized as no response, mild, moderate, and significant.
## Precautions for office use
When using the HTTS for tympanostomy tube placement in children in an otolaryngology office setting:
- Assess suitability of the HTTS procedure using shared ● decision making between the parents and the physician.
- . Local anesthesia should be used on the tympanic membrane to increase the child's comfort.
- . Immobilization with a papoose and/or swaddling should be used to mitigate the child's body movement, and the head should be gently restrained by a nurse or medical assistant.
{6}------------------------------------------------
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
| Clinical Metric | Pediatric Office Study | Predicate Conscious Sedation Study (K142282) |
|---------------------------------------------|------------------------|----------------------------------------------|
| Median age of child | 12 months | 21 months |
| Successful rate in office or under sedation | 98.9% | 88.3% |
| Efficacy endpoint (HTTS delivery of tube) | 96.9% | 92% |
| Safety endpoint (procedural AE rate) | 0.56% | 0.0% |
| % ears completed in one surgical pass | 84.5% | 82% |
| Early extrusion rate | 3.0% | 3.2% |
| Plugging rate | 11.3% | 12.3% |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.