K232508 · Molnlycke Health Care Us, LLC · DWJ · Sep 15, 2023 · Cardiovascular
Device Facts
Record ID
K232508
Device Name
Barrier EasyWarm Active Self-Warming Blanket
Applicant
Molnlycke Health Care Us, LLC
Product Code
DWJ · Cardiovascular
Decision Date
Sep 15, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.5900
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Barrier EasyWarm Active Self-Warming Blanket is intended to help prevent hypothermia by providing warmth to the patient during the perioperative period.
Device Story
Barrier EasyWarm Active Self-Warming Blanket is a non-sterile, single-use thermal regulating device; prevents hypothermia during perioperative period. Device consists of polypropylene blanket containing sealed warming pads; pads contain iron, active coal, water, salt, and clay with sodium polyacrylate cover. Operation relies on exothermic chemical reaction (oxidation of iron) triggered by exposure to air upon removal from vacuum-sealed packaging. Available in one-piece or two-piece configurations; two-piece design uses polyamide hook-and-loop closures for flexible positioning. Used in healthcare facilities; operated by clinical staff. Provides passive/active warming to patient skin; average temperature 34-39°C; maximum 41°C. Benefits patient by maintaining normothermia during surgery.
Clinical Evidence
Bench testing performed for rise time, maintenance time, and temperature specifications. Clinical thermal performance testing conducted on healthy human volunteers using IR camera measurements to confirm safety and effectiveness of the warming effect.
Indicated for patients during the perioperative period to help prevent hypothermia.
Regulatory Classification
Identification
A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.
Predicate Devices
Barrier EasyWarm Active Self-Warming Blanket (K132048)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 15, 2023
Molnlycke Health Care US, LLC Megan Bevill Regulatory Affairs Director, Americas and Antiseptics 5445 Triangle Parkway, Suite 400 Peachtree Corners, Georgia 30092
Re: K232508
Trade/Device Name: Barrier EasyWarm Active Self-Warming Blanket Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal Regulating System Regulatory Class: Class II Product Code: DWJ Dated: August 16, 2023 Received: August 18, 2023
Dear Megan Bevill:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Eric E. Richardson - S
for Nicole Gillette Assistant Director DHT2B: Division of Circulatory Support, Structural and Vascular Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2023 See PRA Statement below.
Submission Number (if known)
K232508
Device Name
Barrier EasyWarm Active Self-Warming Blanket
Indications for Use (Describe)
The Barrier EasyWarm Active Self-Warming Blanket is intended to help prevent hypothermia by providing warmth to the patient during the perioperative period.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) SUMMARY
This 510(k) summary information is being submitted in accordance with the requirements of 21 CFR 807.92(c).
| Date Prepared: | September 8, 2023 |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Applicant: | Mölnlycke Health Care US, LLC<br>5445 Triangle Parkway, Suite 400<br>Peachtree Corners, GA 30092<br>Registration number: 3004763499<br>Owner/Operator Number: 8030877 |
| Official Correspondent: | Megan Bevill<br>Regulatory Affairs Director, Americas and Antiseptics<br>Tel: 770-547-9196<br>email: megan.bevill@molnlycke.com |
| Trade/Proprietary Names: | Barrier EasyWarm Active Self-Warming Blanket |
| Regulation Name: | Thermal regulating system |
| Device Class: | II |
| Regulation Number: | 21 CFR 870.5900 |
| Product Code: | DWJ |
| Predicate Device Name(s): | Barrier EasyWarm Active Self-Warming Blanket (K132048) |
## Reason for 510(k) Submission:
This premarket notification is being submitted to obtain clearance for minor design and labeling modifications to the Barrier EasyWarm Active Self-Warming Blanket, originally cleared for market under premarket notification K132048. Minor design and labeling modifications include the following:
- Introduction of a two-piece design to the assortment ●
- . Changes to the warmers, resulting in a shift to a higher average temperature performance
- Upgrading of existing labeling statements to Contraindications
- Other minor labeling changes (e.g. administrative, formatting, and placement/presentation of ● information changes)
#### Description of Device:
The Barrier EasyWarm Active Self-Warming Blanket is a non-sterile, single use blanket that is intended to prevent hypothermia by providing warmth to the patient during the period. The blanket contains warming pads (or heating elements) that are contained within sealed pouches that are advantageously positioned and sewn into the blanket's fabric for appropriate distribution of heat. The warming pads contain iron, active coal, water, salt, clay with a sodium polyacrylate cover. The device is supplied in a vacuum sealed packaging. Once the blanket is removed from its packaging, the blanket produces heat via an exothermic chemical reaction that takes place within the warming pads upon exposure to air. The chemical reaction results from the oxidation of iron.
#### Intended Use/Indication for Use:
The Barrier EasyWarm Active Self-Warming Blanket is intended to help prevent hypothermia by providing warmth to the patient during the perioperative period.
{4}------------------------------------------------
## Technological Characteristics:
| Feature | Barrier EasyWarm Active Self-<br>Warming Blanket | Barrier EasyWarm Active Self-<br>Warming Blanket | Substantial Equivalence Comments |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)<br>clearance | Subject of submission | K132048 | NA |
| Rationale for<br>inclusion | Subject of submission | Predicate device | NA |
| Manufacturer | Mölnlycke Health Care | Mölnlycke Health Care | NA |
| Regulation | 21 CFR 870.5900 | 21 CFR 870.5900 | Same regulation as predicate |
| Class name | Thermal regulating system | Thermal regulating system | Same classification as predicate |
| Class | II | II | Same classification as predicate |
| Product code | DWJ | DWJ | Same product code as predicate |
| Indication for<br>use/Intended<br>use | The Barrier EasyWarm Active Self-<br>Warming Blanket is intended to help<br>prevent hypothermia by providing<br>warmth to the patient during the<br>perioperative period. | The Barrier EasyWarm Active Self-<br>Warming Blanket is intended to help<br>prevent hypothermia by providing<br>warmth to the patient during the<br>perioperative period. | Same indications for use/intended use<br>as predicate |
| Use<br>environment | Healthcare facilities | Healthcare facilities | Same use environment as predicate |
| Presentation | One piece blanket (92 x 152 cm) with 12<br>warmers<br>Two piece blanket (110/150 x 200 cm,<br>assembled) with 12 warmers | One piece blanket (92 x 152 cm) with 12<br>warmers | The introduction of the new two-piece<br>blanket style does not impact safety or<br>effectiveness of the Barrier EasyWarm<br>Self-Warming Blanket. The overall<br>design, materials of construction, and<br>warmer technology are the same. The<br>two piece design allows the blanket to<br>be separated for upper and lower body<br>coverage in different surgical positions.<br>The two pieces can be connected via a<br>hook and loop closure. |
| Sterility | Non-sterile | Non-sterile | Same |
| Shelf Life | 2 years | 3 years | Based on real-time stability data<br>available for the modified design |
| Materials of<br>construction | Polypropylene blanket (blue for one-<br>piece blanket, blue topside and white<br>underside for two-piece blanket)<br>Polyester sewing thread<br>Polypropylene hem | Polypropylene blanket (blue)<br>Polyester sewing thread<br>Polypropylene hem | The materials of construction are the<br>same for the subject and predicate<br>devices, with the following exceptions: |
| | Polyamide hook and loop closure (two<br>piece design only)<br>Warmer containing activated charcoal,<br>clay, iron, water, salt, and sodium<br>polyacrylate | Warmer containing activated charcoal,<br>clay, iron, water, salt, and sodium<br>polyacrylate | • Use of white polypropylene on<br>one side of two-piece blanket;<br>the white polypropylene is<br>identical to the blue<br>polypropylene cleared under<br>K132048 but with a white<br>pigment instead of blue<br>• Addition of polyamide hook and<br>loop closure on the two-piece<br>blanket<br>Replacement of the color pigment in a<br>portion of the blanket and addition of a<br>hook and loop closure do not impact<br>safety or effectiveness of the device. |
| Performance<br>specification | Ave. skin temperature reached: 34-39°C<br>Max skin temperature reached: 41°C<br>Useful life: 10 hours | Ave. skin temperature reached: 35-36°C<br>Max skin temperature reached: 40°C<br>Useful life: 10 hours | The average temperature performance<br>specification has been shifted higher<br>while maintaining the same specification<br>for maximum skin temperature reached.<br>As the maximum skin temperature<br>reached is still below the original limit of<br>42°C, there are no new issues of safety<br>or effectiveness. |
{5}------------------------------------------------
{6}------------------------------------------------
#### Performance Testing:
The subject devices were subjected to warmer/blanket performance testing (rise time, maintaining time, average temperature during 10 hours use, maximum temperature) to ensure performance according to specification.
In addition, the subject devices were subjected to thermal performance testing (skin temperature measurements via IR camera) in healthy human volunteers to demonstrate that the warming effect is safe and clinically effective.
#### Conclusion:
Substantial equivalence has been demonstrated through a comparison of intended use, design, and technological characteristics as well as performance testing. The subject EasyWarm Active Self-Warming Blanket is at least as safe and effective, and performs at least as well as the predicate device, Easy Warm Active Self-Warming Blanket cleared under K132048.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.