K132048 · Molnlycke Health Care Us, LLC · DWJ · Nov 6, 2013 · Cardiovascular
Device Facts
Record ID
K132048
Device Name
BARRIER EASYWARM ACTIVE SELF-WARMING BLANKET
Applicant
Molnlycke Health Care Us, LLC
Product Code
DWJ · Cardiovascular
Decision Date
Nov 6, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.5900
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The BARRIER® EasyWarm® Active Self-Warming Blanket is intended to help prevent hypothermia by providing warmth to the patient during the perioperative period.
Device Story
Disposable 152 x 92 cm blanket containing 12 sealed pouches with warming pads; pads contain activated coal, clay, iron, water, salt, and sodium polyacrylate. Exothermic chemical reaction (iron oxidation) initiates upon exposure to air after removal from vacuum packaging. Provides continuous thermal coverage for up to 10 hours; reaches 40°C in approximately 30 minutes. Used in perioperative settings; operated by clinical staff. Blanket features cut-outs for positioning flexibility. No external power source or hosing required. Reduces cutaneous heat loss to maintain patient core body temperature; prevents perioperative hypothermia.
Clinical Evidence
Prospective multi-center randomized controlled clinical study. Subjects in the treatment arm showed significantly higher perioperative core body temperatures and lower incidence of hypothermia (pre-, intra-, and post-operatively) compared to control. No serious adverse events reported. Independent evaluation confirmed average skin temperature range of 35-36°C and maximum of 40°C, well below the 43°C safety threshold for skin surface warming.
Technological Characteristics
Disposable 152 x 92 cm blanket. Heating principle: exothermic chemical reaction (iron oxidation). Components: activated coal, clay, iron, water, salt, sodium polyacrylate. No external power, hosing, or electrical components. Biocompatibility per ISO 10993-1 (skin irritation, sensitization, cytotoxicity).
Indications for Use
Indicated for patients undergoing surgical procedures (pre-, intra-, and post-operative) to prevent hypothermia and maintain normothermia by reducing cutaneous heat loss.
Regulatory Classification
Identification
A thermal regulating system is an external system consisting of a device that is placed in contact with the patient and a temperature controller for the device. The system is used to regulate patient temperature.
{0}------------------------------------------------
Traditional 510(k) Premarket Notification
:
K132048
## 510(k) SUMMARY
# BARRIER® EasyWarm® Active Self-Warming Blanket
This 510(k) summary information is being submitted in accordance with the requirements of 21 CFR 807.92(c).
| Date Prepared: | October 15, 2013 | NOV 06 2013 |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Applicant: | Mölnlycke Health Care US, LLC<br>5550 Peachtree Parkway, Suite 500<br>Norcross, GA 30092<br>Registration number: 3004763499<br>Owner/Operator Number: 8030877 | |
| Official Correspondent: | Curtis Truesdale<br>Director, Regulatory Affairs for the Americas<br>Tel: 678-250-7928<br>Fax: 678-245-7746<br>e-mail: curtis.truesdale@molnlycke.com | |
| Trade/Proprietary Name: | BARRIER® Easy Warm® Active Self-Warming Blanket | |
| Common Name: | Warming Blanket | |
| Regulation Name: | Thermal Regulating System | |
| Device Class: | Class II | |
| Regulation Number: | 21 CFR 870.5900 | |
| Product Code: | DWJ | |
| Predicate Device Name(s): | HotDog Patient Warming System (K112488) | |
Description of Device:
The BARRIER® EasyWarm® Active Self-Warming Blanket is a disposable blanket 152 x 92 cm in size. The blanket contains warming pads (or heating elements) that are contained within sealed pouches that are advantageously positioned and sewn into the blanket's fabric for appropriate distribution of heat. The device is supplied in a vacuum sealed packaging. Once the blanket is removed from its' packaging, the blanket produces heat via an exothermic chemical reaction that takes place within the warming pads upon exposure to air.
Page 1 of 4
{1}------------------------------------------------
The BARRIER® EasyWarm® Active Self-Warming Blanket is intended to be used in the pre-, intra- and post-operative phases of a patient's surgical experience to maintain normothermia. The blanket provides continuous thermal coverage, affording the patient active warming up to 10 hours. The blanket is removed from packaging at least 30 minutes prior to use to initiate the warming of blanket. The blanket reaches 40 ℃ in approximately 30 minutes. Once the blanket is placed on the patient, it helps prevent the patient from cooling during the perioperative period. The blanket is designed to help patients maintain body temperature by decreasing cutaneous heat loss. The maintenance of normothermia is accomplished through safe administration of skin surface warming facilitated by the active self-warming blanket.
#### Intended Use/Indication for Use:
The BARRIER® Easy Warm® Active Self-Warming Blanket is intended to help prevent hypothermia by providing warmth to the patient during the perioperative period.
### Technological Characteristics:
The BARRIER® EasyWarm® Active Self-Warming Blanket and HotDog Patient Warming System (K112488) device have similar principles of operation - to prevent or maintain normothermia.
The differences in these two products are the mechanism used to provide the heat through the blanket. The BARRIER® EasyWarm® Active Self-Warming Blanket is a disposable 152 x 92 cm blanket that produces heat via an exothermic chemical reaction initiated by exposure to air, resulting from the oxidation of iron. The chemicals are securely contained in 12 individually sealed pockets, strategically placed to provide heat to the subject for optimal thermal warming. The components of the warmers are -activated coal, clay, iron, water, salt and sodium polyacrylate.
BARRIER® EasyWarm® has no external attachments, requires no power source, is quiet, and carries no risk of "hosing" or blowing pathogens. It is available in one design with three cut-outs positioned to provide the ability to position the blanket longwise on the patient. Turning the blanket 90 degrees will allow exposure of abdomen and legs if the surgical procedure warrants this.
There is asymmetry of the heating pads to allow the blanket to provide access to one extremity while maintaining coverage of the rest of the patient. The blanket should be placed with the heating pads away from the patient.
The Active Self-Warming Blanket does not contain any novel features. Non-warming blankets of the same material have been used for more than three years and this material has been used by Mölnlycke Health Care for other applications such as surgical drapes and surgical gowns.
{2}------------------------------------------------
## Evaluation of Device Performance
#### Bench testing:
A series of tests and evaluations were conducted in support of the device design verification activities were completed to characterize the device with respect to the temperature achieved (maximum and minimums), temperature rise time, duration of temperatures are maintained.
## Biocompatibility Evaluation:
A biocompatibility assessment was completed for the device per ISO 10993-1 Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing, which included the following studies:
- Skin Irritation;
- Closed Patch Sensitization
- Cytotoxity .
## Stability (Shelf-life) Testing:
Stability testing was conducted to demonstrate that the integrity of the device is maintained for the labeled shelf-life period.
#### Clinical Testing:
The results of from a prospective multi-center randomized controlled clinical study. conducted for the BARRIER® EasyWarm® Active Self-Warming Blanket, demonstrated that skin temperatures reached with the use of the device were safety and clinically effective at maintaining core body temperatures. The following is a summary of the primary endpoint results:
- Subjects in the BARRIER® EasyWarm® treatment arm had significantly higher . perioperative core body temperature, relative to subjects in the control treatment arm;
- Subjects in the BARRIER® EasyWarm® treatment arm had a significantly lower . incidence of hypothermia preoperatively, intraoperatively, and postoperatively compared with the subjects in the control treatment arm: and
- . There were no serious adverse events in this clinical investigation.
In addition to the aforementioned clinical study, an independent non-interventional evaluation was performed to determine the average skin temperature range and maximum skin temperatures reached with the warming blanket. The average skin temperature range and the maximum skin temperature achieved with the use of the device are 35-36° C and 40° C respectively. The skin temperatures achieved with the BARRIER® EasyWarm® Active Self-Warming Blanket have been shown to be effective at maintaining core body temperature, with very minimal risks of causing thermal injury to the skin. Skin
{3}------------------------------------------------
temperatures not exceeding 43° C are generally considered safe for skin surface warming. The reduction in cutaneous heat loss via active skin warming techniques is the scientific principle behind the maintenance and/or management of core body temperature; most of metabolic heat is lost the through skin.
## Conclusion:
Based on the performance.testing, it can be.concluded that the BARRIER® Easy Warm® Active Self-Warming Blanket is substantially equivalent to the HotDog Patient Warming System (KI12488) predicate.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes coiled around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 14, 2014
Mölnlycke Health Care US, LLC Curtis Truesdale, Director, Regulatory Affairs for the Americas 5550 Peachtree Parkway Suite 500 Norcross, GA 30092
Re: K132048 Trade/Device Name: BARRIER® EasyWarm® Active Self-Warming Blanket Regulation Number: 21 CFR 870.5900 Regulation Name: Thermal Regulating System Regulatory Class: II Product Code: DWJ Dated: October 15, 2013 Received: October 17, 2013
Dear Mr. Truesdale:
This letter corrects our substantially equivalent letter of November 06, 2013.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{5}------------------------------------------------
Page 2 - Mr. Curtis Truesdale
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M.A. Willhemen
for Bram Zuckerman, MD Director, Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## INDICATONS FOR USE
510(k) Number (if known): K132048
Device Name: BARRIER® EasyWarm® Active Self-Warming Blanket
The BARRIER® EasyWarm® Active Self-Warming Blanket is intended to help prevent hypothermia by providing warmth to the patient during the perioperative period.
Prescription Use X (21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
M. A. Wilhelmen
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.