K232473 · Zhuhai Yueja Medical Device Technology Co., Ltd. · FLL · Feb 9, 2024 · General Hospital
Device Facts
Record ID
K232473
Device Name
Infrared Thermometer: Model YJ600
Applicant
Zhuhai Yueja Medical Device Technology Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Feb 9, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
Infrared Thermometer is intended for the intermittent measurement of body temperature from the forehead on people of all ages. It can be used by consumers in the household environment and by health care providers.
Device Story
Device is a non-contact infrared thermometer; inputs infrared radiation emitted from human forehead; transforms infrared energy via thermopile sensor and internal circuitry into digital temperature reading; displays result on LCD. Used in household or clinical settings by adults; operator points device at forehead (1-3 cm distance). Provides visual (backlight color) and audible (buzzer) feedback based on temperature range; aids clinical decision-making by identifying fever states. Benefits patient through non-invasive, rapid temperature monitoring.
Clinical Evidence
Clinical study included 200 subjects across four age groups (0-3 months, 3 months-1 year, 1-5 years, >5 years). Primary endpoint was clinical repeatability per ISO 80601-2-56 and ASTM E1965-98. Results demonstrated repeatability within ±0.3°C for all age groups, meeting standard requirements.
Indicated for intermittent forehead body temperature measurement in all ages. Suitable for home and clinical use.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name and title on the right. The symbol on the left is a stylized representation of a human figure, while the text on the right reads "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue letters.
February 9, 2024
Zhuhai Yueja Medical Device Technology Co.,Ltd. Feng Yan Department manager Room 201, Building 2, No.2 Liushi Road Tangjiawan Town, High-tech Zone Zhuhai, Guangdong 519000 China
Re: K232473
Trade/Device Name: Infrared Thermometer: Model YJ600 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: January 11, 2024 Received: January 11, 2024
Dear Feng Yan:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Porsche Bennett
Porsche Bennett For David Wolloscheck, Ph.D. Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn,
{2}------------------------------------------------
General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232473
Device Name Infrared Thermometer: Model YJ600
Indications for Use (Describe)
Infrared Thermometer is intended for the intermittent measurement of body temperature from the forehead on people of all ages. It can be used by consumers in the household environment and by health care providers.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
| | |
| | Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K232473- 510(k) Summary
- Device Submitter: Zhuhai Yueja Medical Device Technology Co.,Ltd. Room 201, Building 2, No.2 Liushi Road, Tangjiawan Town, High-tech Zone, Zhuhai, 519000, China
- Contact Person: Feng Yan Department Manager Phone: +86-18826906029 E-mail: fengy@myzr.com.cn Date Prepared February 9, 2024
# l Subject Device
| Trade Name of Device: | Infrared Thermometer: |
|-----------------------|---------------------------------|
| | Model YJ600 |
| Regulation Number: | 21 CFR 880.2910 |
| Product Code: | FLL |
| Regulation Name: | Clinical Electronic Thermometer |
| Common name | Electronic Thermometer |
| Regulatory Class | II |
| Review Panel | General Hospital |
### II Predicate Device
| 510k Number | K200471 |
|-----------------------|---------------------------------|
| Trade Name of Device: | Infrared |
| | Thermometer, Model: |
| | RN-60A, RN-60B |
| Regulation Number: | 21CFR 880.2910 |
| Regulation Name: | Clinical electronic thermometer |
| Regulatory Class | II |
| Product Code: | FLL |
### III Device Descriptions
Infrared Thermometer is mainly composed of infrared probe components, main circuit board components, LCD display components and housing components. The Infrared Thermometer is intended for the intermittent measurement of body temperature from the forehead on people of all ages. Infants and children cannot operate the thermometer; it is recommended that adults take the measurement. The Infrared Thermometer can be used by consumers in the household environment and by healthcare providers. The operator
{5}------------------------------------------------
should be an adult with relevant experience. The operator can replace the battery, transmit data, and measure temperature.
# IV Indications for use
Infrared Thermometer is intended for the intermittent measurement of body temperature from the forehead on people of all ages. It can be used by consumers in the household environment and by health care providers.
# V Technological Characteristics Comparison
VI-1: Comparison of Subject and Predicate Devices
| Device<br>Characteristic | Subject Device | Predicate Device<br>(K200471) | Discussion |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------|
| Type of<br>Thermometer | Infrared Thermometer:<br>Model YJ600 | Infrared<br>Thermometer,<br>Model:<br>RN-60A, RN-60B | - |
| Product Code | FLL | FLL | Same |
| Regulation No. | 21 CFR 880.2910 | 21 CFR 880.2910 | Same |
| Class | II | II | Same |
| Indications for Use | Infrared Thermometer is<br>intended for the intermittent<br>measurement of body<br>temperature from<br>forehead on people of all<br>ages. It can be used by<br>Consumers in<br>household environment and<br>by health care providers. | The Infrared<br>thermometer is a<br>non-contact infrared<br>thermometer<br>intended for<br>intermittent<br>the measurement of<br>all human body<br>temperature from<br>the forehead for<br>people of one month<br>old and above. The<br>device is reusable for<br>home use and<br>clinical use. | Different 1 |
| Patient population | All ages | One month old<br>and above | Different 2 |
| Operation Principle | According to the<br>characteristics of natural<br>objects, all objects with<br>temperatures higher | The Infrared<br>Thermometers<br>RN-60A and RN-60B<br>measure the | Different 3 |
| | | | |
| | zero (-273.15 °C ) are constantly emitting infrared capabilities to surrounding space. Its radiation characteristics, radiation energy and wavelength distribution are closely related to the surface temperature of the object. The human body, like other organisms, radiates infrared energy around itself, and its wavelength is generally 9-14µm, which is in the infrared band of 0.76-100 μ m. Because the light in this wavelength range is not absorbed by air, the infrared radiation from human body has nothing to do with the environmental impact, but only with the energy released by human body. | temperature by using the principle of receiving infrared. An object with a temperature higher than absolute zero and will radiate a certain amount of infrared energy. According to the infrared energy and wavelength, the surface temperature of the object can be determined. The temperature of human forehead is relatively constant, so the temperature of human forehead can be measured according to this principle. The thermometer adopts a new CMOS thermopile sensor with good sensitivity. The pressure difference generated by the sensor after receiving the infrared signal is amplified and analyzed then display on the LCD screen in digital form. | |
| Prescription/over-the-counter use | over-the-counter use | over-the-counter use | Same |
| Measurement technology | Infrared radiation detection that converts a user's forehead temperature using | Infrared radiation detection that converts a user's | Same |
| | the infrared energy emitted<br>in the area around the user's<br>forehead to a reference site<br>equivalent temperature. | forehead<br>temperature<br>using<br>the infrared energy<br>emitted in the area<br>around the user's<br>forehead to a<br>reference site<br>equivalent<br>temperature. | |
| Measurement site | Forehead | Forehead | Same |
| Measurement Range | $22.0 °C~42.9 °C$ | Forehead mode:<br>$32.0°C ~42.9°C$<br>(89.6 to 109.2 ° F) | Different 4 |
| Accuracy | $±0.3°C$ | Forehead mode:<br>$±0.2°C (0.4°F)$ within<br>(96.8°F ~<br>102.2°F), $±0.3°C(0.5°$<br>F) when $<36.0°C$<br>(96.8°F) and ><br>$39.0°C$ (102.2°F) | Different 5 |
| Display resolution | $0.1°C$ | $0.1°C(0.1°F)$ | Same |
| C/F switchable | Yes | Yes | Same |
| Measurement distance | 1-3 cm | 3~5 cm | Different 6 |
| Measurement time | 2s | 1s | Different 7 |
| Sensor type | high precision infrared<br>sensor | Thermopile | Different 8 |
| Memory | 32 sets | 60 sets | Different 9 |
| Buzzer | Yes | Yes | Same |
| Auto power-off while no<br>operation | Yes | Yes | Same |
| Power supply | 2 AAA size batteries | 2*AAA battery | Same |
| Display screen | LCD | LCD | Same |
| Operation Environment | +15°C~+40C RH<95%<br>non-condensing,70kPa~106k<br>Pa | $5.0°C~40.0°C$<br>(41°F~104°F)<br>15%≤RH≤90%<br>70.0kPa-106.0kPa | Different<br>10 |
| Storage Environment | -20°C~+55°C<br>humidity<93%,<br>70kPa~106kPa | -25.0℃~70.0°F<br>(-13°F~158°F)<br>RH≤95%<br>50.0kPa-106.0kPa | Different<br>11 |
| Conformance<br>standard | ISO80601-2-56<br>(performance),<br>IEC60601-1(Safety),<br>IEC60601-1-2(EMC)<br>IEC 60601-1-11(Home use)<br>ASTM E1965-98<br>IEC60601-1-6(Usability) | ISO80601-2-56<br>(performance),<br>IEC60601-1(Safety),<br>IEC60601-1-2(EMC)<br>IEC<br>60601-1-11(Home<br>use)<br>ASTM E1965-98 | Same |
| Patient contact<br>materials | ABS | ABS | Same |
| Biocompatibility | ISO 10993-5<br>ISO 10993-10 | ISO 10993-5<br>ISO 10993-10 | Same |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
# Different 1
There are minor differences in the wording of the indications for use. In addition, the subject device is indicated for people of all ages, while the predicate device is indicated for people one month old and above. Clinical accuracy testing was conducted to ISO80601-2-56. Therefore, the difference does not raise different questions of safety and effectiveness.
### Different 2
The subject device is indicated for people of all ages, while the predicate device is indicated for people one month old and above. Clinical accuracy testing was conducted to ISO80601-2-56. Therefore, the difference does not raise different questions of safety and effectiveness.
### Different 3
The description of operation principle is different. However, its working principle is the same, both use the principle of receiving infrared rays to measure the temperature of the human body. Therefore, the difference does not raise different questions of safety and effectiveness.
### Different 4
The minimum measurement temperature of the subject device is lower than that of the predicate device. Performance testing was conducted to ISO 80601-2-56. Therefore, the difference does not raise different questions of safety and effectiveness.
### Different 5
The accuracy is different, however compliant with ISO 80601-2-56. Therefore, the difference does not raise different questions of safety and effectiveness.
{9}------------------------------------------------
#### Different 6
The measurement distance is different; however, performance testing was conducted to demonstrate the difference does not raise different questions of safety and effectiveness.
#### Different 7
The measurement time is different; however, performance testing was conducted to demonstrate the difference does not raise different questions of safety and effectiveness.
#### Different 8
The sensor type is different; however, performance testing was conducted to demonstrate the difference does not raise different questions of safety and effectiveness.
#### Different 9
The memory is different; however, the difference does not raise different questions of safety and effectiveness.
#### Different 10
The operation environment is different, however compliant with ISO 80601-2-56. Therefore, the difference does not raise different questions of safety and effectiveness.
#### Different 11
The storage environment is different, however compliant with ISO 80601-2-56. Therefore, the difference does not raise different questions of safety and effectiveness.
### VI Summary of Non-clinical Testing (Bench)
The non-clinical testing for Infrared Thermometer: Model YJ600 was performed to demonstrate verification testing in conformance with the acceptance criteria of test methods and recognized consensus standards shown below.
| Inspecting item | | Inspection content and acceptance standard | Result | |
|----------------------------------------|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| 1. Appearance<br>and structure<br>test | 1.1 | The shape of the thermometer should<br>be correct, the surface should be<br>smooth and clean, and there should<br>be no obvious scratches, sharp<br>edges, burrs and deformation. | Pass | |
| | 1.2 | The text and symbols on the<br>thermometer control panel and<br>function keys should be clear,<br>accurate and firm. | Pass | |
| | 1.3 | The control devices of the<br>thermometer should be flexible and<br>reliable, and the fasteners should not | Pass | |
| | | | he loose. | |
| | 1.4 | | The top of the thermometer probe should be smooth and the edges should be free of burrs. | Pass |
| | 1.5 | | The function keys of the thermometer should be clearly marked, indicating that the function keys of the thermometer should be clearly marked and indicated. | Pass |
| 2. Temperature range test | display | | In body temperature mode, the temperature display range of the thermometer $22.0°C~42.9°C$ | Pass |
| 3. Maximum allowable error test | | | The maximum allowable error of the thermometer is $\pm 0.3° C$ within the temperature display range of $22.0° C$ to $42.0°C$ . | Pass |
| 4. Anti-drop test | | | The thermometer shall meet the requirements of 6.3.1 after falling freely from a height of 1m in vertical distance to a hard surface with three different initial postures during normal use. | Pass |
| 5. Indicative unit test | 5.1 Resolution | | The resolution of the indication unit of the thermometer is $0.1°C$ ( $0.1° F$ ). | Pass |
| | 5.2 Display | | The height of the reading value on the thermometer display is >4mm. | Pass |
| | 5.3 Prompt function | | In the body temperature mode, when the measured value of the thermometer is lower than $22.0°C$ , there should be a prompt sound "Di di" twice and "LO" should be displayed; when the measured value is higher than $42.9°C$ , a prompt sound " Beep" twice and display "HI" | Pass |
| | 5.4<br>Low | The<br>battery<br>of the<br>voltage | Pass | |
| | voltage | thermometer is lower than 2.5±0.2V. | | |
| | prompt | " icon flashes on<br>When the " | | |
| | function | | | |
| | | the screen, it means that the battery | | |
| | | is low and needs to be replaced. | | |
| | 5.5 Mode | a) The thermometer should have a | Pass | |
| | | temperature measurement mode. | | |
| | | The<br>calibration<br>mode<br>b)<br>for | | |
| | | calibration purposes<br>shall<br>be | | |
| | | obtained by a conversion technique | | |
| | | directly sets the infrared<br>that | | |
| | | thermometer to this mode. | | |
| 6 Function and | 6.1<br>Unit | The thermometer has a temperature | Pass | |
| performance | switching | unit switching function, which can | | |
| test | function | switch between "℃" and "F". | | |
| | 6.2 Sound | a) In body temperature mode, when | Pass | |
Table VII-1: Performance testing was conducted on the subject device
{10}------------------------------------------------
{11}------------------------------------------------
| Prompt<br>Function | the measurement result is in the<br>range of 22° C to 37.4°C, a green<br>backlight will be displayed and a short<br>sound will arise.<br>b) In body temperature mode, when<br>the measurement result is in the<br>range of 37.5° C to 38.4° C, an orange<br>backlight will be displayed and 1 long<br>sound + 3 short sounds will arise.<br>c) In body temperature mode, when<br>the measurement result is in the<br>range of 38.5° C to 42.9° C, a red<br>backlight will be displayed and 1 long<br>sound + 5 short sounds will arise. | |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| 6.3<br>Measuring<br>time | The thermometer should complete<br>the temperature measurement within<br>2s. | Pass |
{12}------------------------------------------------
| | 6.4<br>Memory<br>query | 32 groups of local data can be queried. | Pass |
|-------------------------------------|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|------|
| 7. Cleaning and disinfection test | | After the thermometer is cleaned/disinfected according to the instructions in the instruction manual, it shall meet the requirements of 6.3.1. | Pass |
| 8. Self-inspection function test | | All LCDs on the screen should be fully displayed every time the thermometer is turned on. | Pass |
| 9. Automatic shutdown function test | | In standby mode, when there is no button action, it will automatically shut down within 20 seconds. | Pass |
| 10. Instruction Manual Inspection | | The instructions for use of the thermometer shall at least include the following: | Pass |
| 11.<br>Electrical<br>safety inspection | 11.1 | a) Temperature display range,<br>temperature unit, maximum allowable<br>error, normal working and storage<br>conditions.<br>b) Conversion method between<br>calibration mode and body<br>temperature mode;<br>c) The clinical accuracy or clinical<br>bias of the subject population, body<br>part and thermometer. | Case Leakage Current and Patient<br>Leakage Current: | Pass | | | | | | |
|----------------------------------------|------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|------|--|--|--|--|--|--|
| | | | Test item Normal<br>status Case leakage<br>current <100uA Patient d.c <10uA | | | | | | | |
{13}------------------------------------------------
| | | leakage<br>(BF) | a.c | <100uA | |
|--------------------------|------|--------------------------------------------------------------------------------------------------------------------|-----|--------|------|
| | 11.2 | Dielectric Strength:<br>Dielectric strength, apply 500Vd.c.,<br>the device under test should not be<br>broken down | | | Pass |
| 12. Packaging inspection | | The actual product is consistent with<br>the packing list, and the content on<br>the nameplate is correct. | | | Pass |
# Others test:
| ID# | Test | Method | Acceptance Criteria | Conclusion |
|-----|-----------------------------------------------------------|-------------------|--------------------------------------|------------|
| 1 | Performance test | ASTM E 1965-98 | Meet the requirement of the standard | Pass |
| | | ISO 80601-2-56 | Meet the requirement of the standard | Pass |
| 2 | Electromagnetic<br>Compatibility and Electrical<br>Safety | IEC 60601-1 | Meet the requirement of the standard | Pass |
| | | IEC 60601-1-2 | Meet the requirement of the standard | Pass |
| | | IEC 60601-1-11 | Meet the requirement of the standard | Pass |
| | | IEC/TR 60601-4-2 | Meet the requirement of the standard | Pass |
| 3 | Usability Test | IEC 60601-1-6 | Meet the requirement of the standard | Pass |
| 4 | Biocompatibility Testing (skin contact < 24h) | | | |
| 4.1 | In Vitro Cytotoxicity | ISO 10993-5:2009 | Non-cytotoxic | Pass |
| 4.2 | Skin Sensitization Test | ISO 10993-10:2010 | Non-sensitizer | Pass |
| 4.3 | Skin Irritation Test | ISO 10993-10:2010 | Non-Irritation | Pass |
{14}------------------------------------------------
# VII Clinical Test Conclusion
YJ600 was tested to ISO 80601-2-56:2017/Amd 1:2018 Medical electrical equipment --Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement — Amendment 1 and ASTM E1965-98(Reapproved 2016) Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature.
This study included 200 subjects (the age range include 0 up to 3 months to one year, older than one year to five years, older than five years, detail groups as below)
| Age group | Age Range | Number | Gender distribution | | |
|-----------|-----------------------------------|--------|---------------------|--------|------------|
| | | | Gender | Number | Percentage |
| A | 0 up to 3 months | 50 | Female | 22 | 44% |
| | | | Male | 28 | 56% |
| B | 3 months to one year | 50 | Female | 23 | 46% |
| | | | Male | 27 | 54% |
| C | older than one year to five years | 50 | Female | 28 | 56% |
| | | | Male | 22 | 44% |
| D | older than five years | 50 |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.