K200471 · Ningbo Ranor Medical Science & Technology Co., Ltd. · FLL · Jun 19, 2020 · General Hospital
Device Facts
Record ID
K200471
Device Name
Infrared Thermometer Model Number RN-60A,RN-60B
Applicant
Ningbo Ranor Medical Science & Technology Co., Ltd.
Product Code
FLL · General Hospital
Decision Date
Jun 19, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Infrared thermometer is a non-contact infrared thermometer intended for the intermittent measurement of human body temperature from forehead for people of one month old and above. The device is reusable for home use and clinical use.
Device Story
Hand-held, battery-powered non-contact infrared thermometer; measures human body temperature via forehead infrared energy emission. Uses CMOS-compatible thermopile sensor to detect infrared radiation; converts pressure difference from sensor into digital temperature reading. Displays results on LCD screen. Used in home or clinical settings by patients or clinicians. Provides intermittent temperature monitoring to assist in clinical decision-making regarding patient health status.
Clinical Evidence
Clinical accuracy validation conducted per ASTM E1965-98 (2016). Study design: randomized, simple blind, homologous control, pairing design. Sample size: 178 subjects across three age groups (1-3 months, 3 months-1 year, 1-5 years, and >5 years). Results demonstrated compliance with ASTM E1965-98 requirements.
Technological Characteristics
Materials: ABS plastic. Sensing: CMOS-compatible thermopile sensor. Energy: 2x AAA batteries. Form factor: Hand-held. Connectivity: Standalone. Sterilization: Not specified. Software: Embedded firmware. Standards: ISO 80601-2-56, IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, ASTM E1965-98, ISO 10993-5, ISO 10993-10.
Indications for Use
Indicated for intermittent forehead temperature measurement in patients one month of age and older. Suitable for home and clinical use.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
Predicate Devices
Infrared Thermometer, Model DT-8836T, DT-8836P (K191251)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. Food & Drug Administration" in blue.
June 19, 2020
Ningbo Ranor Medical Science & Technology Co., Ltd. % Mr. Boyle Wang Correspondent Shanghai Truthful Information Technology Co., Ltd. RM.608, No.738, Shangcheng Rd., Pudong Shanghai, Shanghai 200120 China
Re: K200471
Trade/Device Name: Infrared Thermometer, Model: RN-60A, RN-60B Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical Electronic Thermometer Regulatory Class: Class II Product Code: FLL Dated: May 18, 2020 Received: May 21, 2020
Dear Mr. Boyle Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmr/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Tina Kiang, Ph.D. Director DHT3C: Division of Drug Delivery and General Hospital Devices. and Human Factors OHT3: Office of Gastrorenal. ObGyn. General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K200471
Device Name
Infrared Thermometer, Model: RN-60A, RN-60B
Indications for Use (Describe)
The Infrared thermometer is a non-contact infrared thermometer intent measurement of human body temperature from forehead for people of one month old and above. The device is reusable for home use and clinical use.
| Type of Use (Select <i>one</i> or <i>both</i> , as applicable) |
|----------------------------------------------------------------|
|----------------------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Tab #06 510(k) Summary K200471
This summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR 807.92.
#### 1.0 submitter's Information
Name: Ningbo Ranor Medical Science & Technology Co., Ltd. Address: No. 127 Fenghui Road, Wangchun Industrial Park, Haishu District, Ningbo, China Tel: 86-574-89258788 Fax: 86- 574-88219485 Contact: Emma Hu Date of Preparation: May 28, 2020
## Designated Submission Correspondent
Mr. Boyle Wang Shanghai Truthful Information Technology Co., Ltd. Room 608, No. 738 Shangcheng Rd., Pudong Shanghai, 200120 China Tel: +86-21-50313932 Email: Info@truthful.com.cn
## 2.0 Device Information
Trade name: Infrared Thermometer Common name: Infrared Thermometer Classification name: Clinical electronic thermometer Model(s): RN-60A,RN-60B
#### 3.0 Classification
Production code: FLL Regulation number: 21CFR 880.2910 Class II Classification: Panel: General Hospital
#### 4.0 Predicate Device Information
Manufacturer: Shenzhen Calibeur Industries Co., Ltd. Device: Infrared Thermometer, Model DT-8836T, DT-8836P 510(k) number: K191251
{4}------------------------------------------------
## 5.0 Device Description
The Infrared Thermometer, Models RN-60A and RN-60B are hand-held device powered by 2*AAA batteries and designed to measure human body temperature without contacting patient' forehead.
The Infrared Thermometers RN-60A and RN-60B measure the temperature by using the principle of receiving infrared. An object with a temperature higher than absolute zero will radiate a certain amount of infrared energy. According to the infrared energy and wavelength, the surface temperature of the object can be determined. The temperature of human forehead is relatively constant, so the temperature of human forehead can be measured according to this principle. The thermometer adopts a new CMOS compatible thermopile sensor with good sensitivity. The pressure difference generated by the sensor after receiving the infrared signal is amplified and analyzed then display on the LCD screen in digital form.
The measurement distance of the subject device is 3~5 cm from the forehead.
The only differences among the RN-60A and RN-60B thermometers are size, weight, shape of enclosure.
#### 6.0 Indication for Use Statement
The Infrared thermometer is a non-contact infrared thermometer intended for the intermittent measurement of human body temperature from forehead for people of one month old and above. The device is reusable for home use and clinical use.
| Item | Subject Device<br>K200471 | Predicate Device<br>K191251 | Remark |
|-----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|
| Type of<br>Thermometer | Infrared Thermometer<br>RN-60A,RN-60B | Infrared thermometer<br>DT-8836T,DT-8836P | -- |
| Product Code | FLL | FLL | Same |
| Regulation No. | 21 CFR 880.2910 | 21 CFR 880.2910 | Same |
| Class | II | II | Same |
| Intended Use&<br>Indications for<br>use | The Infrared thermometer is<br>a non-contact infrared<br>thermometer intended for<br>the<br>intermittent<br>measurement of human<br>body temperature from | The Infrared<br>thermometer is a non-<br>contact infrared<br>thermometer intended<br>for the intermittent<br>measurement of human | Different<br>1 |
| | forehead for people of one<br>month old and above.<br>The device is reusable for<br>home use and clinical use. | body temperature from<br>forehead for people of<br>all ages. The device is<br>reusable for home use<br>and clinical use. | |
| Prescription/over-the-counter use | over-the-counter use | over-the-counter use | Same |
| Measurement technology | Infrared radiation detection<br>that converts a user's<br>forehead temperature using<br>the infrared energy emitted<br>in the area around the user's<br>forehead to a reference site<br>equivalent temperature | Infrared radiation<br>detection that converts a<br>user's forehead<br>temperature using the<br>infrared energy emitted in<br>the area around the<br>user's forehead to a<br>reference site equivalent<br>temperature | Same |
| Measurement place | Forehead | Forehead | Same |
| Measurement Range | Forehead mode:<br>$32.0°C ~42.9°C$<br>(89.6 to 109.2 ° F) | Forehead mode:<br>$32.0°C ~42.5°C$<br>(89.6 to 108.5 ° F) | |
| Accuracy | Forehead mode:<br>$±0.2°C (0.4°F)$ within<br>(96.8°F ~ 102.2°F),<br>$±0.3°C(0.5°F)$<br>when $<36.0°C$ (96.8°F)<br>and $>39.0°C$ (102.2°F) | Forehead mode:<br>$±0.2°C (0.4°F)$ within<br>$35.0°C ~ 42.0°C$<br>(95.0°F ~ 107.6°F),<br>$±0.3°C(0.5°F)$ when<br>$<35.0°C$ (95.0 °F) and<br>$>42.0°C$ (107.6°F) | Different<br>2 |
| Display resolution | $0.1°C(0.1°F)$ | $0.1°C(0.1°F)$ | Same |
| C/F switchable | Yes | Yes | Same |
| Measurement distance | 3~5 cm | <3cm | Different<br>3 |
| Response time | 1s | 1s | Same |
| Sensor type | Thermopile | Thermopile | Same |
| Memory | 60 sets | 60 sets | Same |
| Buzzer | Yes | Yes | Same |
| Auto power-off while no operation | Yes | Yes | Same |
| Power supply | 2*AAA battery | 2 * 1.5V AAA | Same |
| Display screen | LCD | LCD | Same |
| Operation<br>Environment | $5.0°C~40.0°C$<br>$(41°F~104°F)$<br>$15%≤RH≤90%$<br>$70.0kPa-106.0kPa$ | $10~40°C (50°F ~104 °F)$<br>RH 15~95% | Different<br>4 |
| Storage<br>Environment | $-25.0℃~70.0°F$<br>$(-13°F~158°F)$<br>RH≤95%<br>$50.0kPa-106.0kPa$ | $-25~+55°$<br>$(-13~+131°F)$<br>RH:15~95% | |
| Dimension | RN-60A:46*70*182mm<br>RN-60B:42*64*185mm | 153.8*62.4*62.4 mm | -- |
| Weight | RN-60A:82g<br>RN-60B:80g | 96g | -- |
| Conformance<br>standard | ISO80601-2-56<br>(performance),<br>IEC60601-1(Safety),<br>IEC60601-1-2(EMC)<br>IEC 60601-1-11(Home use)<br>ASTM E1965-98 | ISO80601-2-56<br>(performance),<br>IEC60601-1(Safety),<br>IEC60601-1-2(EMC)<br>IEC 60601-1-11(Home<br>use)<br>ASTM E1965-98 | Same |
| Patient contact<br>materials | ABS | ABS | Same |
| Biocompatibility | ISO 10993-5<br>ISO 10993-10 | ISO 10993-5<br>ISO 10993-10 | Same |
## 7.0 Comparison to the Predicate Device
{5}------------------------------------------------
{6}------------------------------------------------
## Analysis:
From the comparison table, the subject devices and predicate device have the same measurement place, display resolution, display screen, auto power-off while no operation and conformance standard. There are slightly differences between the devices and predicate device as follows:
Different 1:The restriction in use for people of one month old and above, that is the subset patient population of the predicate device, thus no new safety and
effectiveness concerns raised due to the difference.
{7}------------------------------------------------
Different 2: Both devices have different measurement range, but they have the
same accuracy and the measurement range of subject devices meet the
requirements of ASTME1965-98.The different does not raise new performance questions.
Different 3: Measurement distance of the subject devices is 3-5cm, the predicate device's is ≤3cm. But the performance test result of subject device shows the accuracy meets the requirements within the distance range. The different does not raise new performance questions.
Different 4: Both devices have slightly different Operation & Storage Environment, but the subject devices meet the requirements of IEC60601-1 and ISO80601-2-56. The different does not raise new performance questions.
## 8.0 Non-Clinical Test Conclusion
#### Non-clinical data:
Non-clinical tests were conducted to verify that the subject devices met all design specifications. The test results demonstrated that the proposed device complies with the following standards:
● IEC 60601-1:2012 Medical Electrical Equipment - Part 1: General Requirements for Basic Safety and essential performance
●IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General
requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
- ISO 80601-2-56:2017+A1:2018 Medical electrical equipment Particular
requirements for the basic safety and essential performance of clinical thermometers for body temperature measurement.
●IEC 60601-1-11:2015 Medical electrical equipment -- Part 1-11: General
requirements for basic safety and essential performance -- Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
{8}------------------------------------------------
- · ISO 10993-5:2009 Biological evaluation of medical devices -- Part 5: Tests
for in vitro cytotoxicity
- ISO10993-10:2010 Biological evaluation of medical devices -- Part 10: Tests
for irritation and skin sensitization
#### Software verification and validation testing:
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", dated May 11, 2005.
#### 9.0 Clinical Accuracy Validation Test Conclusion
Clinical tests were conducted per ASTM E1965-98(Reapproved 2016) Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature. This clinical study is a randomization, simple blind homologous control, pairing design of clinical investigation, consists of a minimum of 178 subjects which were divided into three group age
ranges- A Infants group (Group A1 - 1 month up to 3 months; Group A2 - 3 months up to 1 year), B Children group (greater than 1 to 5 years old) and C Over 5 years old (Above 5 years old). Each group at least has 30 subjects.
Based on the result, it is demonstrated the clinical performance of the subject device complied with the requirement of ASTM E1965-98 (2016).
#### 10.0 Conclusion
Performance testing contained in this submission demonstrates the minor differences in technological characteristics between the subject device and the predicate do not raise different questions of safety and effectiveness. And based on the performance testing and comformance with acceptable voluntary standards, we believe the Infrared Thermometer RN-60A and RN-60B are substantially equivalent to its predicate device in K191251.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.