K232456 · Mediott, Inc. · LLZ · Aug 11, 2025 · Radiology
Device Facts
Record ID
K232456
Device Name
RW-1
Applicant
Mediott, Inc.
Product Code
LLZ · Radiology
Decision Date
Aug 11, 2025
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
This software is a medical device intended for the evaluation of DICOM images. It receives, stores, processes, and displays sequential DICOM images primarily obtained through low-dose chest fluoroscopy (e.g., RF and AX modalities). This software is not intended to be used for primary diagnosis. Reference images such as scintigraphy or CT scans may be displayed for supplementary purposes.
Device Story
Software-only medical imaging system for COTS PCs; receives, stores, processes, and displays sequential DICOM images from chest fluoroscopy (RF, AX). Operates as standalone application; integrates visualization and processing. Healthcare providers use device to view images and perform area measurements; supports time-series comparison via graph display. Facilitates supplementary review of pulmonary images; not for primary diagnosis. Benefits include streamlined image management and evaluation workflow.
Clinical Evidence
No clinical data. Performance supported by system-level verification and validation (V&V) activities confirming functional correctness, repeatability, and robustness of deterministic software algorithms.
Indicated for the evaluation of pulmonary images in patients undergoing low-dose chest fluoroscopy (RF and AX modalities). Not intended for primary diagnosis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}
FDA U.S. FOOD & DRUG ADMINISTRATION
August 11, 2025
Mediott Inc.
Akio Oki
Senior Consultant
18-2 Shitamachi, Akagi Kagura Sacas
SHINJUKU-KU, TOKYO 162-0803
JAPAN
Re: K232456
Trade/Device Name: RW-1
Regulation Number: 21 CFR 892.2050
Regulation Name: Medical Image Management And Processing System
Regulatory Class: Class II
Product Code: LLZ
Dated: July 4, 2025
Received: July 7, 2025
Dear Akio Oki:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
U.S. Food & Drug Administration
10903 New Hampshire Avenue
Silver Spring, MD 20993
www.fda.gov
{1}
K232456 - Akio Oki
Page 2
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
All medical devices, including Class I and unclassified devices and combination product device constituent parts are required to be in compliance with the final Unique Device Identification System rule ("UDI Rule"). The UDI Rule requires, among other things, that a device bear a unique device identifier (UDI) on its label and package (21 CFR 801.20(a)) unless an exception or alternative applies (21 CFR 801.20(b)) and that the dates on the device label be formatted in accordance with 21 CFR 801.18. The UDI Rule (21 CFR 830.300(a) and 830.320(b)) also requires that certain information be submitted to the Global Unique Device Identification Database (GUDID) (21 CFR Part 830 Subpart E). For additional information on these requirements, please see the UDI System webpage at https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/unique-device-identification-system-udi-system.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reporting-mdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See
{2}
K232456 - Akio Oki
Page 3
the DICE website (https://www.fda.gov/medical-devices/device-advice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,

Jessica Lamb
Assistant Director, Imaging Software Team
DHT8B: Division of Radiologic Imaging Devices and Electronic Products
OHT8: Office of Radiological Health
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{3}
# 510(k) Summary
510(k) Number: K232456
Date of Preparation: August 7th, 2025
## Applicant
Company name: Mediott Inc.
Address: Kagura Sacas, 18-2, Akagi-shitamachi, Shunjuku-ku, Tokyo, 162-0803, Japan
## Contact
Contact Person: Akio OKI
Title: Senior Consultant
Phone: +81 70 7793 8055
Fax: None
e-mail: akio.oki@radwisp.com
## Name of Device(s)
Product Name: RW-1
Regulation Name(s): System, Image Processing, Radiological
Classification Name: Medical image management and processing system
Regulation Number: 892.2050
Regulatory Class: II
Product Code: LLZ
Panel: Radiology
## Predicate Device
- KONICAMINOLTA DI-X1 (K212685) by Konica Minolta, Inc.
## Device Description
The subject device is a software-only medical imaging system intended for installation on commercial off-the-shelf personal computers. It receives, stores, processes, and displays sequential DICOM images, primarily obtained from chest fluoroscopy (e.g., RF, AX modalities). The software is compatible with external systems such as hospital PACS via DICOM-compliant communication protocols.
The device operates as a standalone application, with all processing and visualization functionalities integrated into a single software package.
## Indications for Use
{4}
This software is a medical device intended for the evaluation of DICOM images. It receives, stores, processes, and displays sequential DICOM images primarily obtained through low-dose chest fluoroscopy (e.g., RF and AX modalities).
This software is not intended to be used for primary diagnosis. Reference images such as scintigraphy or CT scans may be displayed for supplementary purposes.
# Technological Characteristics
The subject device and the predicate device are both software-based medical imaging systems that receive, process, display, and transmit X-ray digital images.
While there are differences in the number of display functions, measurement tools, and system configuration, the subject device retains the core capabilities necessary for its intended use.
These differences represent either reductions in functionality or architectural simplifications. They do not raise new questions of safety or effectiveness.
| | Subject Device | Predicate Device |
| --- | --- | --- |
| Device Name | RW-1 | KONICAMINOLTA DI-X1 |
| 510(k) number | This submission | K212685 |
| Indications for Use | This software is a medical device intended for the evaluation of pulmonary images. It receives, stores, processes, and displays sequential DICOM images primarily obtained through low-dose chest fluoroscopy (e.g., RF and AX modalities).This software is not intended to be used for primary diagnosis. Reference images such as pulmonary scintigraphy or CT scans may be displayed for supplementary purposes. | KONICAMINOLTA DI-X1 is a software device that receives digital x-ray images and data from various sources (i.e. R/F Units, digital radiographic devices or other imaging sources). Images and data can be stored, communicated, processed and displayed within the system and/or across computer networks at distributed locations. It is not intended for use in diagnostic review for mammography |
| Input Image | DICOM 3.0DICOM Modality (RF, DX, CR, AX) | DICOM 3.0DICOM Modality (RF, DX, CR) |
| I/O Data | DICOM Storage | DICOM Storage |
| Display Functions | Adjustment of density and gradation, Rotation, Scaling, PanningScreen display (listing, viewer)Image display (Cine, Comparison)Graph display (Time-series comparison) | Adjustment of density and gradation, Rotation and reversal, Scaling, PanningScreen display (listing, viewer)Image display (Cine, Comparison, Annotation, Overlay) |
{5}
| | | • Graph display (Time-series comparison) |
| --- | --- | --- |
| Measurement Functions | • Area | • Distance
• Angle
• Area
• CTR (Only the image of the front of the chest) |
| Client | • RW-1 client (Standalone application) | • DI-X1 client
• DI-X1 Server client
• PC client (WEB reference) |
# Performance (Non-Clinical) Testing
Non-clinical performance testing was conducted as part of the comprehensive system-level verification and validation (V&V) activities for the subject device. These tests were designed to confirm that the implemented software algorithms operate reliably and consistently under representative conditions. The primary focus was on ensuring functional correctness, repeatability, and robustness of the device functions, consistent with industry standards for software-based medical devices. Statistical exhaustiveness was not required due to the deterministic nature of the implemented algorithms.
No separate standalone bench tests were performed beyond these system-level V&V activities, as the system-level testing was considered sufficient to evaluate all performance-critical features under anticipated use conditions. This approach aligns with standard practices for medical software, where system-level integration tests typically serve as the main source of performance evidence.
Collectively, these results support the conclusion that the subject device performs as intended and is substantially equivalent to the predicate device with respect to safety and effectiveness.
# Conclusion
Based on the comparison of technological characteristics and intended use, as well as non-clinical performance testing, it is concluded that the subject device is substantially equivalent to the predicate device. The observed differences do not raise new questions of safety or effectiveness and reflect reductions in scope or architectural simplification.
{6}
# 510(k) Summary
510(k) Number: K232456
Date of Preparation: August 7th, 2025
## Applicant
Company name: Mediott Inc.
Address: Kagura Sacas, 18-2, Akagi-shitamachi, Shunjuku-ku, Tokyo, 162-0803, Japan
## Contact
Contact Person: Akio OKI
Title: Senior Consultant
Phone: +81 70 7793 8055
Fax: None
e-mail: akio.oki@radwisp.com
## Name of Device(s)
Product Name: RW-1
Regulation Name(s): System, Image Processing, Radiological
Classification Name: Medical image management and processing system
Regulation Number: 892.2050
Regulatory Class: II
Product Code: LLZ
Panel: Radiology
## Predicate Device
- KONICAMINOLTA DI-X1 (K212685) by Konica Minolta, Inc.
## Device Description
The subject device is a software-only medical imaging system intended for installation on commercial off-the-shelf personal computers. It receives, stores, processes, and displays sequential DICOM images, primarily obtained from chest fluoroscopy (e.g., RF, AX modalities). The software is compatible with external systems such as hospital PACS via DICOM-compliant communication protocols.
The device operates as a standalone application, with all processing and visualization functionalities integrated into a single software package.
## Indications for Use
{7}
This software is a medical device intended for the evaluation of DICOM images. It receives, stores, processes, and displays sequential DICOM images primarily obtained through low-dose chest fluoroscopy (e.g., RF and AX modalities).
This software is not intended to be used for primary diagnosis. Reference images such as scintigraphy or CT scans may be displayed for supplementary purposes.
# Technological Characteristics
The subject device and the predicate device are both software-based medical imaging systems that receive, process, display, and transmit X-ray digital images.
While there are differences in the number of display functions, measurement tools, and system configuration, the subject device retains the core capabilities necessary for its intended use.
These differences represent either reductions in functionality or architectural simplifications. They do not raise new questions of safety or effectiveness.
| | Subject Device | Predicate Device |
| --- | --- | --- |
| Device Name | RW-1 | KONICAMINOLTA DI-X1 |
| 510(k) number | This submission | K212685 |
| Indications for Use | This software is a medical device intended for the evaluation of pulmonary images. It receives, stores, processes, and displays sequential DICOM images primarily obtained through low-dose chest fluoroscopy (e.g., RF and AX modalities).This software is not intended to be used for primary diagnosis. Reference images such as pulmonary scintigraphy or CT scans may be displayed for supplementary purposes. | KONICAMINOLTA DI-X1 is a software device that receives digital x-ray images and data from various sources (i.e. R/F Units, digital radiographic devices or other imaging sources). Images and data can be stored, communicated, processed and displayed within the system and/or across computer networks at distributed locations. It is not intended for use in diagnostic review for mammography |
| Input Image | DICOM 3.0DICOM Modality (RF, DX, CR, AX) | DICOM 3.0DICOM Modality (RF, DX, CR) |
| I/O Data | DICOM Storage | DICOM Storage |
| Display Functions | Adjustment of density and gradation, Rotation, Scaling, PanningScreen display (listing, viewer)Image display (Cine, Comparison)Graph display (Time-series comparison) | Adjustment of density and gradation, Rotation and reversal, Scaling, PanningScreen display (listing, viewer)Image display (Cine, Comparison, Annotation, Overlay) |
{8}
| | | • Graph display (Time-series comparison) |
| --- | --- | --- |
| Measurement Functions | • Area | • Distance
• Angle
• Area
• CTR (Only the image of the front of the chest) |
| Client | • RW-1 client (Standalone application) | • DI-X1 client
• DI-X1 Server client
• PC client (WEB reference) |
# Performance (Non-Clinical) Testing
Non-clinical performance testing was conducted as part of the comprehensive system-level verification and validation (V&V) activities for the subject device. These tests were designed to confirm that the implemented software algorithms operate reliably and consistently under representative conditions. The primary focus was on ensuring functional correctness, repeatability, and robustness of the device functions, consistent with industry standards for software-based medical devices. Statistical exhaustiveness was not required due to the deterministic nature of the implemented algorithms.
No separate standalone bench tests were performed beyond these system-level V&V activities, as the system-level testing was considered sufficient to evaluate all performance-critical features under anticipated use conditions. This approach aligns with standard practices for medical software, where system-level integration tests typically serve as the main source of performance evidence.
Collectively, these results support the conclusion that the subject device performs as intended and is substantially equivalent to the predicate device with respect to safety and effectiveness.
# Conclusion
Based on the comparison of technological characteristics and intended use, as well as non-clinical performance testing, it is concluded that the subject device is substantially equivalent to the predicate device. The observed differences do not raise new questions of safety or effectiveness and reflect reductions in scope or architectural simplification.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.