KONICAMINOLTA DI-X1 is a software device that receives digital x-ray images and data from various sources (i.e. R/F Units, digital radiographic devices or other imaging sources). Images and data can be stored, communicated, processed and displayed within the system and/or across computer networks at distributed locations. It is not intended for use in diagnostic review for mammography.
Device Story
Standalone software device installed on off-the-shelf servers/PCs; receives digital X-ray images (single/multi-frame) via DICOM 3.0 from imaging modalities. Processes images using various modes (REGIUS, FE, DM, BS, PL, Pk to Pk, PH, PH2, LM); performs measurements (distance, angle, area, CTR). Displays images and chronological measurement graphs for clinical review. Features RAID1 mirroring for data redundancy. Modifications include browser-based PC client for web reference, additional image processing modes, and chronological graph display for exam comparison. Used in clinical environments; operated by healthcare professionals to assist in diagnostic workflows. Benefits include efficient image management, remote access via web client, and longitudinal tracking of patient measurements.
Clinical Evidence
Bench testing only. No clinical studies were required to support substantial equivalence. Verification activities confirmed that all predetermined acceptance criteria were met.
Technological Characteristics
Standalone software for off-the-shelf servers/PCs. Connectivity: DICOM 3.0. Data storage: RAID1 mirroring. Image processing modes: REGIUS, FE-MODE, DM-MODE, BS-MODE, PL-MODE, Pk to Pk, PH-MODE, PH2-MODE, LM-MODE. Measurement tools: distance, angle, area, CTR. Deployment: Server-client architecture with web-based reference capability.
Indications for Use
Indicated for receiving, storing, communicating, processing, and displaying digital X-ray images and data from various diagnostic imaging sources (R/F units, digital radiographic devices). Not intended for diagnostic review of mammography.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
September 13, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Konica Minolta, Inc. % Jan Maniscalco Director of QA/RA Konica Minolta Healthcare Americas, Inc. 411 Newark-Pompton Turnpike WAYNE NJ 07470
Re: K212685
Trade/Device Name: KONICAMINOLTA DI-X1 Regulation Number: 21 CFR 892.2050 Regulation Name: Medical image management and processing system Regulatory Class: Class II Product Code: LLZ Dated: August 16, 2021 Received: August 24, 2021
Dear Jan Maniscalco:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
, for
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K212685
Device Name KONICAMINOLTA DI-X1
#### Indications for Use (Describe)
KONICAMINOLTA DI-X1 is a software device that receives digital x-ray images and data from various sources (i.e. RF Units, digital radiographic devices or other imaging sources). Images and data can be stored, communicated, processed and displayed within the system and/or across computer networks at distributed locations. It is not intended for use in diagnostic review for mammography.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <div> <span>☑</span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span>☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the Konica Minolta logo. The logo features a blue circle with several horizontal white lines across the center. Below the circle, the words "KONICA MINOLTA" are written in a bold, sans-serif font.
# 510(k) Summary K212685
| Company: | KONICA MINOLTA, INC.<br>1 Sakura-machi, Hino-shi, Tokyo, 191-8511 Japan | | |
|--------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Contact: | Tsutomu Fukui<br>Senior Manager of Regulatory & QMS Division<br>1 Sakura-machi, Hino-shi, Tokyo, 191-8511 Japan<br>Telephone: +81 42 589 8429<br>Email : tsutomu.fukui1@konicaminolta.com | | |
| Date Prepared: | September 08, 2021 | | |
| Device Name: | | | |
| Trade Name: | KONICAMINOLTA DI-X1 | | |
| Common Name: | Medical Image Management and Processing System | | |
| Regulation Number: | 21 CFR 892.2050 | | |
| Regulatory Class: | Class II | | |
| Product Code(s): | LLZ | | |
| Predicate Device: | K182431 - KONICAMINOLTA DI-X1<br>KONICA MINOLTA, INC.<br>Regulation Number: 21 CFR 892.2050<br>Regulatory Class: Class II<br>Product Codes: LLZ | | |
## Device Description
KONICAMINOLTA DI-X1 is a software device that performs image processing and display using X-ray digital images (single-frame images, multi-frame images) generated by various diagnostic imaging modality consoles. It is a standalone software device intended to install onto off-the -shelf Servers and PCs.
- Page 1 of 4 -
Giving Shape to Ideas
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Konica Minolta. The logo consists of a blue circle with several horizontal white lines across the middle. Below the circle is the text "KONICA MINOLTA" in a bold, sans-serif font.
KONICAMINOLTA DI-X1 receives X-ray digital images, including serial images, processes the received images, as well as displays and sends the resulting images to PACS and other devices. In addition, KONICAMINOLTA DI-X1 can display images through the browser connection with the client that displays and process images, and instruct transmission of images.
The personal computer used in KONICAMINOLTA DI-X1 stores the same data in two hard disks in real time using RAID1 mirroring function. Thus, even if one hard disk is defective, operations can be continued with the other hard disk which has the same data.
The modifications are made on software to the identified predicate device to add the PC client to connect to the server using the browser on a personal computer to display images for a WEB reference. In addition, additional imaging processing MODES are implemented into the subject device. The subject device also modifies the graphical display to compare the past exam. graphs based on the measurement values in a chronological order.
#### Indications for Use
KONICAMINOLTA DI-X1 is a software device that receives digital x-ray images and data from various sources (i.e. R/F Units, digital radiographic devices or other imaging sources). Images and data can be stored, communicated, processed and displayed within the system and/or across computer networks at distributed locations. It is not intended for use in diagnostic review for mammography.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Konica Minolta. The logo features a blue circle with several horizontal white lines running across the middle. Below the circle, the words "KONICA MINOLTA" are written in a bold, sans-serif font.
## Comparison Table
The comparison to the predicate decides was summarized in the Table below.
| | Subject Device | Predicate Device |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | KONICAMINOLTA DI-X1 | KONICAMINOLTA DI-X1 |
| 510(K) Number | This Submission | K182431 |
| Indications for<br>Use | KONICAMINOLTA DI-X1 is a software<br>device that receives digital x-ray images<br>and data from various sources (i.e. R/F<br>Units, digital radiographic devices or<br>other imaging sources). Images and data<br>can be stored, communicated, processed<br>and displayed within the system and/or<br>across computer networks at distributed<br>locations. It is not intended for use in<br>diagnostic review for mammography. | KONICAMINOLTA DI-X1 is a software<br>device that receives digital x-ray images<br>and data from various sources (i.e. R/F<br>Units, digital radiographic devices or other<br>imaging sources). Images and data can be<br>stored, communicated, processed and<br>displayed within the system and/or across<br>computer networks at distributed<br>locations. It is not intended for use in<br>diagnostic review for mammography. |
| Input image | ● DICOM 3.0<br>● DICOM Modality (RF, DX, CR) | ● DICOM 3.0<br>● DICOM Modality (RF, DX, CR) |
| I/O Data | ● DICOM Storage | ● DICOM Storage |
| Image<br>Processing | ● REGIUS<br>● FE-MODE<br>● DM-MODE<br>● BS-MODE<br>● PL-MODE<br>● Pk to Pk<br>● PH-MODE<br>● PH2-MODE<br>● LM-MODE | ● REGIUS<br>● FE-MODE<br>● DM-MODE<br>● BS-MODE<br>● PL-MODE<br>● Pk to Pk |
| Display<br>Functions | ● Adjustment of density and<br>gradation, Rotation and reversal,<br>Scaling, Panning<br>● Screen display (listing, viewer)<br>● Image display (Cine, Comparison,<br>Annotation, Overlay)<br>● Graph display (Time-series<br>comparison) | ● Adjustment of density and gradation,<br>Rotation and reversal, Scaling,<br>Panning<br>● Screen display (listing, viewer)<br>● Image display (Cine, Comparison,<br>Annotation, Overlay)<br>● Graph display |
| Measurement | ● Distance<br>● Angle<br>● Area<br>● CTR (Only the image of the front of<br>the chest) | ● Distance<br>● Angle<br>● Area<br>● CTR |
| Client | ● DI-X1 client<br>● DI-X1 Server client<br>● PC client (WEB reference) | ● DI-X1 client<br>● DI-X1 Server client |
— Page 3 of 4 —
Giving Shape to Ideas
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the Konica Minolta logo. The logo features a blue globe-like shape at the top with several horizontal white lines running across it. Below the globe is the text "KONICA MINOLTA" in a simple, sans-serif font.
#### Performance Data
All the verification activities required by the specification and the risk analysis for the KONICAMINOLTA DI-X1 were performed and the results demonstrated that the predetermined acceptance criteria were met. No clinical studies were required to support the substantial equivalence.
## Conclusion
KONICAMINOLTA DI-X1 has the same intended use and indications for use, technological characteristics, and principal operations. The technological differences raised no new issues of safety or effectiveness as compared to its predicate device (K182431). Performance tests demonstrate that the KONICAMINOLTA DI-X1 performs according to specifications and functions as intended. All the information to demonstrate assurance of our evaluation is attached to relevant sections of this submission.
Therefore, as for our conclusion, the KONICAMINOLTA DI-X1 is substantially equivalent to the predicate device and presents no new questions of safety or effectiveness.
Page 4 of 4
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.