K232236 · Quanta System Spa · GEX · Oct 20, 2023 · General, Plastic Surgery
Device Facts
Record ID
K232236
Device Name
Echo
Applicant
Quanta System Spa
Product Code
GEX · General, Plastic Surgery
Decision Date
Oct 20, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The device is intended to be used in surgical, aesthetic applications in the medical specialties of general and plastic surgery and dermatology, including: - Treatment of benign vascular lesions - Hair removal - Permanent* hair reduction * permanent hair reduction defined as reduced hair growth without maintenance when measured at 6, 9 and 12 months.
Device Story
Echo is a 160 W diode laser (808 nm) for professional use in general/plastic surgery and dermatology. Device delivers laser radiation via optical fibers connected to handpieces with fixed spot sizes. Features integrated skin cooling; handpieces include skin cooler housing. Activation via footswitch or finger-switch on handpiece. System is mobile electrical equipment with graphical user interface (GUI). Clinician operates device to treat benign vascular lesions or perform hair removal/reduction. Output is laser energy delivered to patient skin; cooling system manages thermal effects. Clinical benefit includes effective hair reduction and vascular lesion treatment.
Clinical Evidence
Bench testing only. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 62366-1, IEC 62304, IEC 60601-2-22, IEC 60825-1, and ISO 10993-1 standards.
Technological Characteristics
160 W diode laser, 808 nm wavelength. Contact skin cooling system. Mobile form factor. Power supply 100-240V, 50-60 Hz. Pulse width 5-100 ms; fluence 5-90 J/cm2; spot sizes 12, 18, 24 mm. GUI-based user interaction. Software developed per IEC 62304.
Indications for Use
Indicated for surgical, aesthetic, and cosmetic applications in general surgery, plastic surgery, and dermatology. Includes treatment of benign vascular lesions, hair removal, and permanent hair reduction.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the full name of the agency on the right. The symbol on the left is a stylized representation of a human figure, while the FDA acronym is in blue, followed by the full name of the agency, "U.S. Food & Drug Administration," in a smaller font size.
October 20, 2023
Quanta System Spa Dario Bandiera Regulatory Affairs Manager Via Acquedotto, 109 Samarate, VA 21017 Italy
Re: K232236
Trade/Device Name: Echo Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: July 26, 2023 Received: July 27, 2023
Dear Dario Bandiera:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K232236
Device Name Echo
Indications for Use (Describe)
The device is intended to be used in surgical, aesthetic applications in the medical specialties of general and plastic surgery and dermatology, including:
- Treatment of benign vascular lesions
- Hair removal
- Permanent* hair reduction
* permanent hair reduction defined as reduced hair growth without maintenance when measured at 6, 9 and 12 months.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
| 510(K)# | 232236 | |
|---------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|--|
| Applicant /<br>Manufacturer<br>and address: | Quanta System S.p.A., Via Acquedotto 109, 21017<br>Samarate (VA), Italy | |
| Contact person: | Dario Bandiera<br>RA Manager<br>Quanta System S.p.A.<br>Email: dario.bandiera@quantasystem.com<br>Phone: +39-0331-376797 | |
| Date Prepared: | 10/02/2023 | |
| Trade/device name: | Echo | |
| Classification: | Class II | |
| Classification Name: | Laser surgical instrument for use in general and<br>plastic surgery and in dermatology. | |
| Regulation Number: | 21 CFR 878.4810 | |
| Product Code: | GEX | |
| Predicate devices: | MeDioStar (K192483) and LightSheer Desire<br>(K170179) | |
{5}------------------------------------------------
#### Device description 1
The laser device Echo is a 160 W diode laser emitting at 808 nm laser wavelength. The device is a therapeutic and aesthetic medical laser system for professional use only.
For the release of laser radiation to the patient, the device uses, as delivery system, a fiber with a handpiece plugged on its end.
Echo is a transportable mobile electrical equipment with a graphical user interface (GUI) for user-device interaction.
Laser radiation is delivered through optical fibers connected to handpieces having fix spot dimension.
The device is equipped with an integrated skin cooler. In this case, a specific housing called Skin Cooler handpiece is provided to provide skin cooling and housing the laser handpiece at the same time.
Laser emission can be activated by the footswitch or by a finger-switch placed on Skin Cooler handpieces.
#### Comparison with the predicate 2
In Table 1, the main specifications of the subject device are summarized and compared to the predicates:
| Specification | Subject device | Predicate 1 | Predicate 2 | Comments |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|
| Device Name | Echo | MeDioStar | Lightsheer Desire | |
| K number | | K192483 | K170179 | -- |
| Manufacturer | Quanta System S.p.A. | Asclepion GmbH | Lumenis | -- |
| Product Code | GEX | GEX | GEX | -- |
| Laser Sources | Diode laser | Diode laser | Diode laser | Same |
| Laser<br>Wavelengths | 808 nm +/- 20 nm | 755-950 nm | 805-1060 nm | Whitin the range<br>of the predicate<br>devices |
| Indications for<br>use | Intended for surgical,<br>aesthetic and cosmetic<br>applications in the<br>medical specialties of<br>general and plastic<br>surgery and<br>dermatology.<br>Intended for the<br>treatment of benign<br>vascular lesions.<br>Intended for hair<br>removal and permanent<br>hair reduction. | Intended for surgical, aesthetic<br>and cosmetic applications in the<br>medical<br>specialties of general and<br>plastic surgery and<br>dermatology.<br>Intended for the treatment of<br>benign vascular lesions.<br>Intended for hair removal,<br>permanent hair reduction and<br>the treatment of benign<br>pigmented lesions. | indicated for use in<br>surgical, aesthetic, and<br>cosmetic applications<br>in the medical<br>specialties of general<br>and plastic surgery,<br>and dermatology. The<br>LightSheer<br>Desire<br>System is intended for<br>use on all skin types<br>(Fitzpatrick skin types I<br>— VI), including tanned<br>skin.<br>The LightSheer Desire<br>System<br>with<br>LightSheer ET/XC<br>805nm Laser<br>Handpieces are<br>intended for:<br>• Treatment of benign<br>vascular lesions,<br>including angiomas,<br>hemangiomas,<br>telangiectasia and<br>other benign vascular<br>lesions and leg veins | Same |
Table 1: Main specifications comparison table.
{6}------------------------------------------------
| Specification | Subject device | Predicate 1 | Predicate 2 | Comments |
|--------------------------|----------------------------------------------------------------------------------|------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------|
| Device Name | Echo | MeDioStar | Lightsheer Desire | |
| | | | Treatment of<br>pseudofolliculitis<br>barbae (PFB)<br>Hair removal,<br>permanent hair<br>reduction* Treatment of benign<br>pigmented lesions | |
| Pulse width<br>(max) | 5 - 100 ms | 3 - 400 ms | Up to 400 ms | Whitin the range<br>of the predicate<br>devices |
| Fluence (max) | 5-90 J/cm2 | 210 J/cm²<br>Max. Fluence Hair reduction<br>60 J/cm² | 10-100 J/cm2 | Same as the<br>predicate device 2 |
| Spot Size | 12, 18, 24 mm (round<br>diameter) respective<br>area (cm²) 1,13 –<br>2,54 - 4,52 | square 10x10, 15x10, 30x10,<br>31.5x31.5 respective area<br>(cm²) 1 - 1,5 -3 - 9.9 | Square 9x9, 27x9,<br>22x35<br>respective area<br>(cm²) 0.81 - 2.43 -<br>7.7 | Areas within the<br>range of the<br>predicate devices |
| Repetition Rate<br>(max) | Up to 10 Hz | Up to 20 Hz | Up to 3 Hz | Whitin the range<br>of the predicate<br>device 1 |
| Skin Cooling | Contact skin cooling | contact skin cooling system<br>(metal probe cooled by a peltier<br>element) | Chilltip contact cooling | Same as Predicate<br>devices (contact<br>mode) |
| Power Supply | Single phase,<br>100-240V<br>50-60 Hz | Single phase,<br>100-240V<br>50-60 Hz | Single phase,<br>100-240V<br>50-60 Hz | Same as predicate<br>devices |
#### Indication for use 3
The device is Intended for surgical, aesthetic and cosmetic applications in the medical specialties of general and plastic surgery and dermatology.
Intended for the treatment of benign vascular lesions.
Intended for hair removal and permanent hair reduction*.
* Permanent hair reduction defined as reduced hair growth with or without maintenance when measured at 6, 9 and 12 months.
#### Non-clinical tests 4
The present device was subject to non-clinical testing according to the following standards (Table 2):
{7}------------------------------------------------
Table 2: Applied standards.
| Standard | |
|--------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| IEC 60601-1: 2005/AC 1:2006/A1:2012/AC1:2014/AMD2:2020 | Medical electrical equipment -<br>Part 1: General requirements for basic safety and essential performance |
| IEC 60601-1-2:2014/A1:2020 | Medical electrical equipment - Part 1-2: General requirements<br>for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances |
- Requirements and tests
IEC 60601-1-6: 2010/AMD1: 2013 Collateral standard: Usability
IEC 62366-1: 2015/COR1: 2016 Part 1: Application of usability engineering to medical devices IEC 62304: 2006/AMD1: 2015 Medical device software (SW)
IEC 60601-2-22: 2019 Medical electrical equipment - Part 2-22: Particular requirements for basic safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
IEC 60825-1: 2014 Safety of laser products – Part 1: Equipment classification and requirements ISO 10993-1: 2018 Biological evaluation of medical devices
The results of the non-clinical performance standards testing support that the subject device can be used safely and effectively.
#### Substantial equivalence discussion ട
Echo has the same intended use of the predicate devices and comparable technical specifications.
#### ട Conclusions
Non-clinical tests conducted support that the device can be used safely and effectively for the proposed indications for use. The differences in technological characteristics between the subject and predicate devices do not raise new questions regarding safety and effectiveness for the proposed indications for use. Thus, the subject device is considered to be substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.