The LightSheer Duet System is indicated for use in surgical, aesthetic applications in the medical specialties of general and plastic surgery, and dermatology. The LightSheer Duet System is intended for use on all skin types (Fitzpatrick skin types I - VI), including tanned skin. The LightSheer Duet System with LightSheer ET 805nm Laser Handpiece is intended for: Treatment of benign vascular lesions, including angiomas, telangiectasia and other benign vascular lesions and leg veins Treatment of pseudofolliculitis barbae (PFB) Hair removal, permanent hair reduction* Treatment of benign pigmented lesions The LightSheer Duet System with LightSheer HS 805nm Laser Handpiece is intended for: Treatment of benign vascular lesions Treatment of benign pigmented lesions Hair removal, and permanent hair reduction* The LightSheer Desire System is indicated for use in surgical, aesthetic applications in the medical specialties of general and plastic surgery, and dermatology. The LightSheer Desire System is intended for use on all skin types (Fitzpatrick skin types I - VI), including tanned skin. The LightSheer Desire System with LightSheer ET/XC 805nm Laser Handpieces are intended for: Treatment of benign vascular lesions, including angiomas, telangiectasia and other benign vascular lesions and leg veins Treatment of pseudofolliculitis barbae (PFB) Hair removal, permanent hair reduction * Treatment of benign pigmented lesions The LightSheer Desire System with LightSheer HS 805nm Laser Handpiece is intended for: Treatment of benign vascular lesions Treatment of benign pigmented lesions Hair removal, and permanent hair reduction* The LightSheer Desire System with LightSheer ET/XC 1060mm and LightSheer HS 1060mm Handpieces are intended for: Treatment of benign vascular lesions, including angiomas, telangiectasia, port wine stains, leg veins and other benign vascular lesions Hair removal, permanent hair reduction* Treatment for Pseudofolliculitis Barbae (PFB) Treatment of benign pigmented lesions, including age spots, solar lentigines, cafe-au-lait spots, nevi of Ota/Ito, melasma, Becker's nevi and other benign pigmented lesions Treatment of wrinkles The LightSheer Desire Light System is indicated for use in surgical, aesthetic, and cosmetic applications in the medical specialties of general and plastic surgery, and dermatology. The Light System is intended for use on all skin types (Fitzpatrick skin types I - VI), including tanned skin. The LightSheer Desire Ligth System with LightSheer ET/XC 805nm Laser Handpieces are intended for: Treatment of benign vascular lesions, including angiomas, telangiectasia and other benign vascular lesions and leg veins Treatment of pseudofolliculitis barbae (PFB) Hair removal, permanent hair reduction* Treatment of benign pigmented lesions The LightSheer Desire Light System with LightSheer ET/XC 1060mm Handpiece is intended for: Treatment of benign vascular lesions, including angiomas, telangiectasia, port wine stains, leg veins and other benign vascular lesions. Hair removal, permanent hair reduction* Treatment for Pseudofolliculitis Barbae (PFB) Treatment of benign pigmented lesions, including age spots, solar lentigines, cafe-au-lait spots, nevi of Ota/Ito, melasma. Becker's nevi and other benign pigmented lesions Treatment of wrinkles LightSheer Infinity: The LightSheer Infinity System is indicated for use in surgical, aesthetic applications in the medical specialties of general and plastic surgery, and dermatology. The LightSheer Infinity System is intended for use on all skin types (Fitzpatrick skin types I - VI), including tanned skin. The LightSheer Infinity System with LightSheer ET/LR 805nm Laser Handpieces is intended for: Treatment of benign vascular lesions, including angiomas, telangiectasia and other benign vascular lesions and leg veins Treatment of pseudofolliculitis barbae (PFB) Hair removal, permanent hair reduction* Treatment of benign pigmented lesions The LightSheer Infinity System with LightSheer HS 805nm Laser Handpiece is intended for: Treatment of benign vascular lesions Treatment of benign pigmented lesions Hair removal, and permanent hair reduction* The LightSheer Infinity System with LightSheer ET/LR 1060mm and LightSheer HS 1060nm Handpieces is intended for: Treatment of benign vascular lesions, including angiomas, telangiectasia, port wine stains, leg veins and other benign vascular lesions. Hair removal, permanent hair reduction Treatment for Pseudofolliculitis Barbae (PFB) Treatment of benign pigmented lesions, including age spots, solar lentigines, cafe-au-lait spots, nevi of Ota/Ito, melasma, Becker's nevi and other benign pigmented lesions Treatment of wrinkles
Device Story
LightSheer family consists of pulsed diode array laser consoles (Duet, Infinity, Desire, Desire Light) used with interchangeable handpieces. Consoles provide GUI, power, vacuum, and cooling. Handpieces utilize either vacuum-assisted high-speed (HS) technology or contact cooling (ChillTip) technology. Systems deliver pulsed diode laser light at 805nm or 1060nm wavelengths. Operated by clinicians in surgical, aesthetic, and dermatology settings. Device provides laser energy to target tissue; output affects clinical decision-making by enabling treatment of vascular/pigmented lesions, hair reduction, and wrinkle reduction. Benefits include stable, long-term hair reduction and lesion clearance.
Clinical Evidence
Bench testing only. No clinical data presented. Performance verified through risk analysis, electrical/laser safety testing (IEC 60601 series), and software verification and validation.
Technological Characteristics
Pulsed diode array laser systems. Wavelengths: 805nm (790-950nm) or 1060nm (1040-1080nm). Handpieces: HS (vacuum-assisted) or ET/XC/LR (ChillTip contact cooling). Consoles include GUI, vacuum pump, cooling system, and power electronics. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 60601-2-22.
Indications for Use
Indicated for patients of all skin types (Fitzpatrick I-VI), including tanned skin, requiring treatment of benign vascular lesions (angiomas, telangiectasia, port wine stains, leg veins), benign pigmented lesions (age spots, solar lentigines, cafe-au-lait spots, nevi of Ota/Ito, melasma, Becker's nevi), pseudofolliculitis barbae (PFB), wrinkles, and permanent hair reduction.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Lumenis LightSheer Duet Light Laser System (K053628)
Lumenis LightSheer Desire Light Laser System (K151947)
LightSheer HS and ET 1060nm HPs for Duet (K133319)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, stacked on top of each other. The profiles are depicted in a simple, line-art style.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 18, 2017
Lumenis Ltd. Amaya Levie Head Of RA Surgical Bu 6 Hakidma Street Po Box 240 Yokneam, 2069204 IL
Re: K170179
Trade/Device Name: Lightsheer Desire: Lightsheer Desire Light: Lightsheer Duet; Lightsheer Infinity Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: GEX Dated: August 17, 2017 Received: August 21, 2017
Dear Amaya Levie:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Actinclude requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply
{1}------------------------------------------------
with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education (DICE) at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely.
# Jennifer R. Stevenson -S3
For Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K170179
#### Device Name
The LightSheer Family of Pulsed Diode Array Laser Systems (LightSheer Desire, LightSheer Desire Light and LightSheer Infinity)
#### Indications for Use (Describe)
LightSheer Duet:
The LightSheer Duet System is indicated for use in surgical, aesthetic applications in the medical specialties of general and plastic surgery, and dermatology. The LightSheer Duet System is intended for use on all skin types (Fitzpatrick skin types I - VI), including tanned skin.
The LightSheer Duet System with LightSheer ET 805nm Laser Handpiece is intended for:
· Treatment of benign vascular lesions, including angiomas, telangiectasia and other benign vascular lesions and leg veins
- · Treatment of pseudofolliculitis barbae (PFB)
- · Hair removal, permanent hair reduction*
- · Treatment of benign pigmented lesions
The LightSheer Duet System with LightSheer HS 805nm Laser Handpiece is intended for:
- · Treatment of benign vascular lesions
- · Treatment of benign pigmented lesions
- · Hair removal, and permanent hair reduction*
#### LightSheer Desire:
The LightSheer Desire System is indicated for use in surgical, aesthetic applications in the medical specialties of general and plastic surgery, and dermatology. The LightSheer Desire System is intended for use on all skin types (Fitzpatrick skin types I - VI), including tanned skin.
The LightSheer Desire System with LightSheer ET/XC 805nm Laser Handpieces are intended for:
· Treatment of benign vascular lesions, including angiomas, telangiectasia and other benign vascular lesions and leg veins
- · Treatment of pseudofolliculitis barbae (PFB)
- · Hair removal, permanent hair reduction *
- Treatment of benign pigmented lesions
The LightSheer Desire System with LightSheer HS 805nm Laser Handpiece is intended for:
- · Treatment of benign vascular lesions
- · Treatment of benign pigmented lesions
- · Hair removal, and permanent hair reduction*
The LightSheer Desire System with LightSheer ET/XC 1060mm and LightSheer HS 1060mm Handpieces are intended for: · Treatment of benign vascular lesions, including angiomas, telangiectasia, port wine stains, leg veins and other benign vascular lesions
· Hair removal, permanent hair reduction*
- · Treatment for Pseudofolliculitis Barbae (PFB)
- · Treatment of benign pigmented lesions, including age spots, solar lentigines, cafe-au-lait spots, nevi of Ota/Ito,
- melasma, Becker's nevi and other benign pigmented lesions
· Treatment of wrinkles
{3}------------------------------------------------
#### LightSheer Desire Light:
The LightSheer Desire Light System is indicated for use in surgical, aesthetic, and cosmetic applications in the medical specialties of general and plastic surgery, and dermatology. The Light System is intended for use on all skin types (Fitzpatrick skin types I - VI), including tanned skin.
The LightSheer Desire Ligth System with LightSheer ET/XC 805nm Laser Handpieces are intended for:
· Treatment of benign vascular lesions, including angiomas, telangiectasia and other benign vascular lesions and leg veins
· Treatment of pseudofolliculitis barbae (PFB)
· Hair removal, permanent hair reduction*
· Treatment of benign pigmented lesions
The LightSheer Desire Light System with LightSheer ET/XC 1060mm Handpiece is intended for:
· Treatment of benign vascular lesions, including angiomas, telangiectasia, port wine stains, leg veins and other benign vascular lesions.
· Hair removal, permanent hair reduction*
· Treatment for Pseudofolliculitis Barbae (PFB)
· Treatment of benign pigmented lesions, including age spots, solar lentigines, cafe-au-lait spots, nevi of Ota/Ito,
melasma. Becker's nevi and other benign pigmented lesions
· Treatment of wrinkles
LightSheer Infinity:
The LightSheer Infinity System is indicated for use in surgical, aesthetic applications in the medical specialties of general and plastic surgery, and dermatology. The LightSheer Infinity System is intended for use on all skin types (Fitzpatrick skin types I - VI), including tanned skin.
The LightSheer Infinity System with LightSheer ET/LR 805nm Laser Handpieces is intended for:
· Treatment of benign vascular lesions, including angiomas, telangiectasia and other benign vascular lesions and leg veins
· Treatment of pseudofolliculitis barbae (PFB)
· Hair removal, permanent hair reduction*
· Treatment of benign pigmented lesions
The LightSheer Infinity System with LightSheer HS 805nm Laser Handpiece is intended for:
- · Treatment of benign vascular lesions
- · Treatment of benign pigmented lesions
· Hair removal, and permanent hair reduction*
The LightSheer Infinity System with LightSheer ET/LR 1060mm and LightSheer HS 1060nm Handpieces is intended for: · Treatment of benign vascular lesions, including angiomas, telangiectasia, port wine stains, leg veins and other benign vascular lesions.
· Hair removal, permanent hair reduction
· Treatment for Pseudofolliculitis Barbae (PFB)
· Treatment of benign pigmented lesions, including age spots, solar lentigines, cafe-au-lait spots, nevi of Ota/Ito,
melasma, Becker's nevi and other benign pigmented lesions
• Treatment of wrinkles
*Permanent hair reduction is defined as the long-term. stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.
Type of Use (Select one or both, as applicable)
ال Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
{4}------------------------------------------------
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{5}------------------------------------------------
# 510(k) SUMMARY ULumenis®
Lumenis LightSheer Family of Diode Pulsed Array Laser Systems
| Applicant Name: | Lumenis Ltd. |
|----------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | 6 Hakidma Street PO Box 240 |
| | Yokneam Industrial Park, |
| | Yokneam 2069204, Israel |
| | Tel: +972-4-9599000 |
| | Fax: +972-4-9599046 |
| Contact Person: | Amaya De Levie |
| | Lumenis Ltd. |
| | 6 Hakidma Street PO Box 240 |
| | Yokneam Industrial Park, |
| | Yokneam 2069204, Israel |
| | Tel: +972-4-9599119 |
| | amaya.de-levie@lumenis.com |
| Date Prepared: | September 14, 2017 |
| Trade Name: | Lumenis LightSheer Family of Diode Pulsed Array Laser Systems |
| Classification Name: | Powered laser surgical instrument |
| Product Code: | GEX |
| Device Class: | Class II |
| Regulation Number: | 21 CFR 878.4810 |
| Panel: | General & Plastic Surgery |
| Predicate Device: | K053628 Lumenis LightSheer Duet Light Laser System<br>K151947 Lumenis LightSheer Desire Light Laser System<br>K133319 LightSheer HS and ET 1060nm HPs for Duet |
{6}------------------------------------------------
#### Indications for Use:
## LightSheer Duet:
The LightSheer Duet System is indicated for use in surgical, aesthetic, and cosmetic applications in the medical specialties of general and plastic surgery, and dermatology. The LightSheer Duet System is intended for use on all skin types (Fitzpatrick skin types I - VI), including tanned skin.
- The LightSheer Duet System with LightSheer ET 805nm Laser Handpiece is intended for: -
- · Treatment of benign vascular lesions, including angiomas, telangiectasia and other benign vascular lesions and leg veins
- · Treatment of pseudofolliculitis barbae (PFB)
- · Hair removal, permanent hair reduction*
- · Treatment of benign pigmented lesions
- The LightSheer Duet System with LightSheer HS 805nm Laser Handpiece is intended for: —
- · Treatment of benign vascular lesions
- · Treatment of benign pigmented lesions
- · Hair removal, and permanent hair reduction*
## LightSheer Desire:
The LightSheer Desire System is indicated for use in surgical, aesthetic, and cosmetic applications in the medical specialties of general and plastic surgery, and dermatology. The LightSheer Desire System is intended for use on all skin types (Fitzpatrick skin types I - VI), including tanned skin.
- -The LightSheer Desire System with LightSheer ET/XC 805nm Laser Handpieces are intended for:
- · Treatment of benign vascular lesions, including angiomas, telangiectasia and other benign vascular lesions and leg veins
- · Treatment of pseudofolliculitis barbae (PFB)
- · Hair removal, permanent hair reduction*
- · Treatment of benign pigmented lesions
- । The LightSheer Desire System with LightSheer HS 805nm Laser Handpiece is intended for:
- · Treatment of benign vascular lesions
- · Treatment of benign pigmented lesions
- · Hair removal, and permanent hair reduction*
- The LightSheer Desire System with LightSheer ET/XC 1060nm and LightSheer HS 1060nm — Handpieces are intended for:
{7}------------------------------------------------
- · Treatment of benign vascular lesions, including angiomas, telangiectasia, port wine stains, leg veins and other benign vascular lesions
- · Hair removal, permanent hair reduction*
- · Treatment for Pseudofolliculitis Barbae (PFB)
- · Treatment of benign pigmented lesions, including age spots, solar lentigines, cafe-au-lait spots, nevi of Ota/Ito, melasma, Becker's nevi and other benign pigmented lesions
- · Treatment of wrinkles
# LightSheer Desire Light:
The LightSheer Desire Light System is indicated for use in surgical, aesthetic, and cosmetic applications in the medical specialties of general and plastic surgery, and dermatology. The LightSheer Desire Light System is intended for use on all skin types (Fitzpatrick skin types I -VI), including tanned skin.
- The LightSheer Desire Light System with LightSheer ET/XC 805nm Laser Handpieces are l intended for:
- · Treatment of benign vascular lesions, including angiomas, telangjectasia and other benign vascular lesions and leg veins
- · Treatment of pseudofolliculitis barbae (PFB)
- · Hair removal, permanent hair reduction*
- · Treatment of benign pigmented lesions
- । The LightSheer Desire Light System with LightSheer ET/XC 1060nm Handpiece is intended for:
· Treatment of benign vascular lesions, including angiomas, telangiectasia, port wine stains, leg veins and other benign vascular lesions.
- · Hair removal, permanent hair reduction*
- · Treatment for Pseudofolliculitis Barbae (PFB)
- · Treatment of benign pigmented lesions, including age spots, solar lentigines, cafe-au-lait spots, nevi of Ota/Ito, melasma, Becker's nevi and other benign pigmented lesions
- Treatment of wrinkles
## LightSheer Infinity:
The LightSheer Infinity System is indicated for use in surgical, aesthetic, and cosmetic applications in the medical specialties of general and plastic surgery, and dermatology. The LightSheer Infinity System is intended for use on all skin types (Fitzpatrick skin types I - VI), including tanned skin.
- The LightSheer Infinity System with LightSheer ET/LR 805nm Laser Handpieces is intended l for:
{8}------------------------------------------------
- Treatment of benign vascular lesions, including angiomas, telangiectasia and other benign vascular lesions and leg veins
- · Treatment of pseudofolliculitis barbae (PFB)
- · Hair removal. permanent hair reduction*
- · Treatment of benign pigmented lesions
- The LightSheer Infinity System with LightSheer HS 805nm Laser Handpiece is intended for: l
- · Treatment of benign vascular and pigmented lesions
- Hair removal, and permanent hair reduction* .
- l The LightSheer Infinity System with LightSheer ET/LR 1060nm and LightSheer HS 1060nm Handpieces is intended for:
- · Treatment of benign vascular lesions, including angiomas, hemangiomas, telangiectasia, port wine stains, leg veins and other benign vascular lesions.
- · Hair removal, permanent hair reduction
- · Treatment for Pseudofolliculitis Barbae (PFB)
- · Treatment of benign pigmented lesions, including age spots, solar lentigines, cafe-au-lait spots, nevi of Ota/Ito, melasma, Becker's nevi and other benign pigmented lesions
- Treatment of wrinkles
*Permanent hair reduction is defined as the long-term, stable reduction in the number of hairs re-growing when measured at 6, 9, and 12 months after the completion of a treatment regime.
## Device Description:
The Lumenis LightSheer Family of Pulse Diode Array Laser Systems consists of four laser consoles that can be used with up to four different types of handpieces. TThe four laser consoles are: LightSheer Duet, LightSheer Infinity, LightSheer Desire, and LightSheer Desire Light. The laser consoles provide (1) a graphical user interface and software for control of the system. (2) the needed electronics to control and power the accessories. (3) handpiece connection port(s), (4) a vacuum pump (most models) and (5) cooling system.
Handpieces can be divided into two categories, depending on the technology applied to the skin during treatment: high speed (HS) Handpieces with vacuum technology (HIT™) or ET, XC and LR Handpieces with cooling technology (ChillTip), available with different sized tips. Some handpieces are universal, supporting different ChillTip sizes within a single handpiece. In addition, these handpieces either deliver pulsed diode laser light with wavelengths ranging from 790 - 950nm (805nm nominal) or 1040-1080nm (1060nm nominal), depending on the diode array type. The same LightSheer Handpieces are compatible with multiple systems (such as the LightSheer Duet, LightSheer Infinity, LightSheer Desire, and LightSheer Desire Light), in which case they differ only in terms of the mechanical connection to the console.
{9}------------------------------------------------
#### Reason for Submission:
The purpose of this 510(k) is to add the 1060nm handpieces to the LightSheer Desire and LightSheer Desire Light Systems. These handpieces have been previously cleared for use with other systems in the LightSheer Family of Pulse Diode Array Laser Systems. For completeness, this submission also includes a description of all four systems in the LightSheer Family of Pulsed Diode Array Laser Systems.
# Substantial Equivalence:
The intended use and indications for use of the modified LightSheer Desire Laser and LightSheer Desire Light are identical to the intended use and indications for use of its predicate device LightSheer Duet. In addition, the same technological characteristics and principles of operation apply for these systems. The modifications introduced to the subject LightSheer Desire Laser and LightSheer Desire Light involve adding compatibility to LightSheer1060nm Handpieces that have been previously been cleared for use with the LightSheer Duet System. In addition, there have also been some modifications made to the systems since their last clearance. These modifications have undergone a contemporaneous regulatory analysis that concluded that the cumulative modifications could not significantly affect safety or effectiveness. Performance testing was also conducted in order to demonstrate the performance of the modified LightSheer Family of Pulsed Diode Array Laser Systems and to verify that no different questions of safety and effectiveness have been raised due to the modifications introduced. The following activities were performed:
- Risk analysis per ISO 14971. .
- . Electrical and laser safety and electromagnetic compatibility testing as required to conform to performance standards as follows:
- o 60601-1: Medical electrical equipment Part 1: General requirements for basic safety and essential performance
- 60601-1-2: Medical electrical equipment Part 1-2: General requirements for o basic safety and essential performance- Collateral standard: Electromagnetic compatibility- requirements and tests
- 60825-1: Safety of laser products -Part 1: Equipment classification and o requirements
- 60601-2-22: Medical electrical equipment- part 2: particular requirements for the o safety of diagnostic and therapeutic laser equipment.
- Software verification and validation ●
Consequently, it is Lumenis' belief that the LightSheer Family of Pulsed Diode Array Laser Systems is substantially equivalent to its legally marketed predicates.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.