GenX-CR is a digital radiographic scanner for dental diagnostics intended for use by dentists and other qualified professionals. This device is used to create and display digital images by scanning intraoral X-ray images stored in an Image Plate (or Phosphore Storage Plate).
Device Story
GenX-CR is a digital dental radiographic scanner; inputs are intraoral X-ray images captured on reusable phosphor storage plates (PSP). Device scans PSP to acquire digital images; processes data; displays images on integrated touch screen and external computer monitors. Used in dental clinics by dentists or qualified professionals. Scanned images transmitted via Ethernet to computer/network for use in PortView or other diagnostic software. Enables digital viewing, storage, and deletion of images from PSP. Benefits include compact design for space-constrained environments and reduced operating costs via reusable plates.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including verification and validation of software, electrical safety (IEC 60601-1), EMC (IEC 60601-1-2), and performance standards (IEC 60601-2-65).
Technological Characteristics
Digital PSP scanner; uses reusable phosphor storage plates. Connectivity via Ethernet. Laser safety Class 1 (EN60825-1:2014). Compliant with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-65, and 62366. Image output: TIFF/Raw. Pixel sizes: 25um/50um. 16-bit grayscale.
Indications for Use
Indicated for dental diagnostics in patients requiring intraoral X-ray imaging. Intended for use by dentists and qualified professionals.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
September 13, 2023
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.
Genoray Co., Ltd. % Ms. Kaitlynn Min Business Development GENORAY America Inc. 1220 N Simon Circle, Unit B ANAHEIM CA 92806
Re: K232158
Trade/Device Name: GenX-CR Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: Class II Product Code: MUH Dated: July 19, 2023 Received: July 20, 2023
Dear Ms. Min:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiological Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
Device Name
Indications for Use (Describe)
GenX-CR is a digital radiographic scanner for dental diagnostics intended for use by dentists and other qualified professionals.
This device is used to create and display digital images by scanning intraoral X-ray images stored in an Image Plate (or Phosphore Storage Plate).
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:12px">☑ Prescription Use (Part 21 CER 801 Subpart D)</span> </div> | <div> <span style="font-size:12px">☐ Over-The-Counter Use (21 CER 801 Subpart C)</span> </div> |
|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
|-------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) summary information is prepared in accordance with 21 CFR 807.92
# 1. Date of Summary Preparation [21 CFR 807.92(a) (1)]_
: Jul. 14, 2023
# 2. Administrative Information [21 CFR 807.92(a) (1)]
| 510(k) Submitter | GENORAY Co.,Ltd<br>Address: 512, 560, Dunchon-daero, Jungwon-gu,<br>Seongnam-si, Gyeonggi-Do, Korea<br>Telephone No.: +82-31-5178-5500<br>Fax: +82-31-5178-5599<br>Contact Person: Inyoung Kim (iykim@genoray.com) |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Official Correspondent | GENORAY America Inc.<br>Address: 1220N Simon Circle, B, Anaheim, CA 92806 USA<br>Telephone No.: +1-855-436-6729<br>Fax: +1-714-786-8919<br>Contact Person: Kaitlynn Min (kaitlynn@genorayamerica.com) |
### 3. Device Information [21 CFR 807.92(a) (2)]
| Trade / Device Name | GenX-CR |
|---------------------------|-------------------------------|
| Requlation Name | Extraoral source X-ray system |
| Classification Regulation | 21 CFR 872.1800 |
| Class of Device | Class II |
| Panel | Radiology |
| Product Code | MUH |
## 4. Predicate Device Information [21 CFR 807.92(a) (3)]
| * Predicate Device | |
|---------------------------|-------------------------------|
| Name of Device | CRUXCAN(CRX-1000) |
| Manufacturer | CRUXELL Corp. |
| Regulation Name | Extraoral source X-ray system |
| Classification Regulation | 21 CFR 872.1800 |
| Class of Device | Class II |
| Panel | Radiology |
| Product Code | MUH |
{4}------------------------------------------------
#### 5. Description of the Device [21 CFR 807.92(a) (4)]
GenX-CR scans a reusable phosphor storage plate (hereinafter referred to as "PSP") instead of an analog film to acquire high-quality digital radiographic images, digitally processes them, and displays images on the equipment's touch display screen and computer screen. It's a scanner. *PSP is also known as IP (Image Plate)
After scanning the PSP, the image is saved in the device's internal memory, the scanned image can be checked in advance through the touch display, and the image stored in the PSP can be deleted and the PSP can be ejected. Scanned images can be directly transmitted to a computer or network via an Ethernet cable and are used in PortView, a dental diagnostic software package, and other diagnostic software.
GenX-CR is composed of mainly consists of PSP Scanner, PSP tray, power adapter and cable, Ethernet cable, PSP transfer box, PSP, PSP Protective Cover, Hygenic Bag, Dental PSP scanner control system software (Version: V2.2)
GenX-CR contains firmware that is part of the system. And we hereby certify that the level of concern for this software(PortView) and firmware(GenX-OP) is of Moderate level of concern. And It has not been cleared with other predicate devices but is the initial used in GenX-CR
Its compact design allows installation in space-constrained locations and minimizeds operating costs by using a low-cost, reusable PSP.
#### 6. Indications for use [21 CFR 807.92(a) (5)]
GenX-CR is a digital radiographic scanner for dental diagnostics intended for use by dentists and other qualified professionals.
This device is used to create and display digital images by scanning intraoral X-ray images stored in an Image Plate (or Phosphor Storage Plate).
{5}------------------------------------------------
# 7. Substantial equivalence chart [21 CFR 807.92(a) (6)]
| | Proposed device | Predicate device | SE Note |
|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|---------|
| | GenX-CR | CRUXCAN(CRX-1000) | |
| Manufacturer | GENORAY Co., Ltd | CRUXELL Corp. | - |
| 510(k) No. | K232158 | K183637 | - |
| Regulation Name | Extraoral source X-ray system | Extraoral source X-ray system | - |
| Classification name | System, X-ray, Extraoral Source, Digital | System, X-ray, Extraoral Source, Digital | - |
| Common Name | Dental PSP Scanner | Dental PSP Scanner | - |
| Product Code | MUH | MUH | - |
| Regulation Number | 872.1800 | 872.1800 | - |
| Class | Class II | Class II | - |
| Product illustraion | Image: GenX-CR | Image: CRUXCAN(CRX-1000) | - |
| Indications for use | GenX-CR is a digital radiographic scanner<br>for dental diagnostics intended for use by | The CRUXCAN (CRX-1000) is indicated for<br>capturing, digitization and processing of | Same |
| | | | |
| | dentists and other qualified professionals.<br>This device is used to create and display<br>digital images by scanning intraoral X-ray<br>images stored in an Image Plate (or<br>Phosphore Storage Plate). | intra-oral x-ray images stored in imaging<br>plate recording media. | |
| Image File Format | TIFF / Raw Format | TIFF / Raw Format | Same |
| Power Supply | 50/60 Hz, 100-240V~ | 50/60 Hz, 100-240V~ | Same |
| X-ray Absorber | Imaging Plate | Imaging Plate | Same |
| Image plate Size | Size 0 : 22 x 31mm<br>Size 1 : 24 x 40 mm<br>Size 2 : 31 x 41mm<br>Size 3 : 27 x 54 mm | Size 0 : 22 x 31mm<br>Size 1 : 24 x 40 mm<br>Size 2 : 31 x 41mm<br>Size 3 : 27 x 54 mm | Same |
| Image Pixel Size | 25 um (High Resolution)<br>50 um (Standard Resolution) | 25 um<br>50 um | Same |
| Gray Scale level | 16 bit | 8 bit / 16 bit | Same |
| Resolution | 12 lp/mm @ 25um | 14.0 lp/mm @ 25um | Similar |
| Imaging Plate<br>Performance<br>- DQE<br>- MTF | DQE at 10% efficiency: 2.8 lp/mm<br>MTF at 3lp/mm: 35% | DQE at 10% efficiency: 2.8 lp/mm<br>MTF at 3lp/mm: 35% | Same |
| Laser safety<br>classification | Class 1 laser product EN60825-1:2014 | Class 1 laser product EN60825-1:2014 | Same |
{6}------------------------------------------------
{7}------------------------------------------------
#### 8. Safety, EMC and Performance data comparison to Predicate [21 CFR 807.92(b)]
GenX-CR has been successfully completed verification and validation testing per GENORAY quality system and engineering bench testing in support of successfully completed verification and validation testing per GENORAY quality system and this submission. The system has been tested and is compliant with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3, IEC 60601-1-6, IEC 60601-2-65, IEC 62366.
And Software was validated according to the FDA Guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical devices", FDA Guidance "Guidance for the content of premarket submissions for management of cyber security". Results demonstrated that all executed verification tests were passed.
Non-clinical validation testing has been performed to validate that GenX-CR conforms to the intended use, claims, user needs, effectiveness of safety measures, and instructions for use. Clinical testing was not necessary for the subject device, to demonstrate substantial equivalence.
As a result, all test results were satisfactory, and the result of bench tests indicates that the new device is as safe and effective as the predicate device.
#### 9. Conclusion
In reference to the comparison information provided in the substantial equivalence chart, and the most of functions and electronic features are similar to the predicate device. We believe that the GenX-CR is safe and effective as a predicate device and has no new indication for use. Therefore, GenX-CR is substantially equivalent to a predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.