The CRUXCAN (CRX-1000) is indicated for capturing, digitization and processing of intra-oral x-ray images stored in imaging plate recording media.
Device Story
CRUXCAN (CRX-1000) is a dental computed radiography system; captures, digitizes, and processes intra-oral X-ray images. Input: X-ray exposed photostimulable phosphor (PSP) imaging plates (polyester film with inorganic crystals). Operation: Scanner reads stored X-ray energy from plates via light stimulation; converts to digital image; transfers to workstation for processing/routing. Features built-in erase function and color touch-screen LCD. Used in radiological environments by qualified staff. Output: Digital X-ray images displayed via CRUXCAN TWAIN Driver and Viewer software. Assists clinicians in dental diagnostic imaging. Benefits: Provides digital conversion of conventional X-ray imaging, enabling efficient image storage, processing, and viewing.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via bench testing, including electromagnetic compatibility and safety testing per IEC 60601 standards, and software validation testing. Imaging plate performance metrics (DQE, MTF, resolution) were compared against predicate devices.
Technological Characteristics
Dental PSP scanner; uses photostimulable phosphor imaging plates. Power: 100-240V, 50-60Hz. Connectivity: TWAIN driver/viewer software. Interface: Color touch-screen LCD. Imaging performance: 25um/50um pixel size, 14.0 lp/mm resolution. Software: Windows 10 or higher. Complies with IEC 60601 for safety and EMC.
Indications for Use
Indicated for capturing, digitization, and processing of intra-oral x-ray images stored in imaging plate recording media for dental radiography.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 12, 2019
CRUXELL Corp. % Ms. Priscilla Chung Regulatory Affairs Consultant LK Consulting Group USA, Inc. 1150 Roosevelt, STE 200 IRVINE CA 92620
Re: K183637
Trade/Device Name: CRUXCAN (CRX-1000) Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: Class II Product Code: MUH Dated: December 10, 2018 Received: December 26, 2018
Dear Ms. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/CombinationProducts/GuidanceRegulatoryInformation/ucm597488.htm); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Michael D. O'Hara
Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K183637
Device Name
CRUXCAN (CRX-1000)
Indications for Use (Describe)
The CRUXCAN (CRX-1000) is indicated for capturing, digitization and processing of intra-oral x-ray images stored in imaging plate recording media.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------|--|
| <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> | |
| <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> | |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
(K183637)
This summary of 510(k) is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: Feb 5, 2019
## 1. 510K Applicant / Submitter:
CRUXELL Corp. A-405, Migun Techno World II, 187 techno 2-ro. Yuseong-gu. Daejeon, 34025 South Korea Tel : +82-42-935-2554 Fax : +82-42-931-2554
## 2. Submission Contact Person
LK Consulting Group USA, Inc. 1150 Roosevelt, STE 200, Irvine CA 92620 Priscilla Juhee Chung Phone: 714.202.5789 Fax: 714-409-3357 Email: juhee.c@lkconsultinggroup.com
## 3. Device
- . Proprietary Name: CRUXCAN (CRX-1000)
- Common Name: Dental PSP Scanner ●
- Classification: Class II (21 CFR 872.1800) ●
- Product Code: MUH .
- Regulation Name: Extraoral source x-ray system .
## 4. Predicate Devices
- 4.1. Digora Optime (K041050) by Soredex Instrumentariumn Corporation
- Common Name: Dental PSP Scanner .
- Classification: Class II (21 CFR 872.1800) .
- Product Code: MUH
- Regulation Name: Extraoral source x-ray system .
{4}------------------------------------------------
4.2. Digora Optime (K133231) by Instrunmentariuin Dental
- . Common Name: Dental PSP Scanner
- Classification: Class II (21 CFR 872.1800) .
- . Product Code: MUH
- Regulation Name: Extraoral source x-ray system .
## 5. Device Description:
The CRUXCAN (CRX-1000) is a dental computed radiography system and is intended to produce digital X-Ray images for dental radiography purposes. It consists of a scanner, reusable imaging plates and workstation software. This device scans X-Ray exposed imaging plate and produces X-Ray image in digital form. Then, digital image is transferred to workstation for further processing and routing.
The design features a built-in erase function and a color touch-screen LCD panel without physical push buttons for device operation. This device is intended to be operated in a radiological environment by qualified staff.
The Imaging Plate is a polyester film made of densely applied particles of inorganic crystals called photostimulable phosphor. It is a flexible X-ray sensor for the CR system which can be used with a conventional medical X-ray imaging system and can be employed as a substitute for the screen/film system. The phosphor layer has a function of recording an X-ray image. This photostimulable phosphor is a special luminescent material which stores X-ray energy and emits light proportional to the stored X-ray energy when stimulation energy such as visible light is irradiated onto it.
CRUXCAN TWAIN Driver and Viewer which transmit and display images are substantially equivalent to the predicate device software as well in terms of functions and principle of operations. The User interface is slightly different but the differences in design do not raise questions in device effectiveness.
## 8. Indications for Use
The CRUXCAN (CRX-1000) is indicated for capturing, digitization and processing of intraoral x-ray images stored in imaging plate recording media.
{5}------------------------------------------------
## 9. Substantial Equivalence Discussion:
| | Subject Device | Predicate Device | Reference Device | SE Note |
|---------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|
| 510k # | K183637 | K041050 | K133231 | |
| Device Name | CRUXCAN | Digora Optime | Digora Optime | |
| Manufacturer | CRUXELL Co., Ltd. | Soredex Instrumentariumn<br>Corporation | Instrunmentariuin<br>Dental | |
| Indications for Use | The CRUXCAN (CRX-1000) is indicated for<br>capturing, digitization<br>and processing of intra-oral x-ray images stored<br>in imaging plate<br>recording media. | The Digora Optime<br>imaging system is indicated<br>for capturing, digitization<br>and processing of intra oral<br>x-ray images stored in<br>imaging plate recording<br>media. | The DIGORA0<br>Optime imaging<br>system is indicated for<br>capturing, digitization<br>and processing of<br>intra oral x-ray images<br>stored in imaging<br>plate recording media. | Equivalent |
| Communication | TIFF / Raw Format | DICOM | DICOM | No issue on<br>safety and<br>effectiveness |
| Power Supply | 50 – 60 Hz<br>100 - 240 V ~ | 50 – 60 Hz<br>100 - 240 V ~ | 50 – 60 Hz<br>100 - 240 V ~ | Equivalent |
| Operating System | Windows 10 or higher | Windows NT 4.0/2000/XP | Windows 7 or 8 | Equivalent |
| X-ray Absorber | Imaging Plate | Imaging Plate | Imaging Plate | Equivalent |
| Cassette Size | Size 0 : 22 x 31mm<br>Size 1 : 24 x 40mm<br>Size 2 : 31 x 41mm<br>Size 3 : 27 x 54mm | Size 0 : 22 x 31mm<br>Size 1 : 24 x 40mm<br>Size 2 : 31 x 41mm<br>Size 3 : 27 x 54mm | Size 0 : 22 x 31mm<br>Size 1 : 24 x 40mm<br>Size 2 : 31 x 41mm<br>Size 3 : 27 x 54mm | No issue on<br>safety and<br>effectiveness |
| Pixel Size | 25 um<br>50 um | 40 um<br>64 um | 30um<br>60um | No issue on<br>safety and<br>effectiveness |
| Dynamic Range<br>(Acquisition) | 8bit /16 bit | 14 bit | 14 bit | Equivalent |
| Resolution | 14.0lp/mm @ 25um | 12.5 lp/mm @ 40um | 16.7 lp/mm@30um | No issue on<br>safety and<br>effectiveness |
| Imaging Plate<br>Performance<br>- DQE<br>- MTF<br>- Max. Resolution | DQE at 10% efficiency:<br>2.8 lp/mm<br>MTF at 3lp/mm : 35%<br>Max resolution : 25um | DQE at 10% eff : 2.4<br>lp/mm<br>MTF at 3lp/mm : 32%<br>Max resolution : 40um | DQE at 10% eff : 2.4<br>lp/mm<br>MTF at 3lp/mm : 32%<br>Max resolution : 30um | No issue on<br>safety and<br>effectiveness |
The CRUXCAN is substantially equivalent to the predicate device, Digora Optime (K041050) and Digora Optime (K133231) in indications for use, imaging principle, and physical characteristics. The CRUXCAN also has similarities in power supply, operating system, X-ray absorber and imaging plate size.
The differences of the three devices are pixel size, dynamic range, resolution, and image plate performance. As compared in the chart above, the subject device has better or substantially equivalent resolution, DQE, and MTF data. The smaller pixel size of subject device than
{6}------------------------------------------------
the predicate device offers wider dynamic range with storing image data using more bits.
The imaging plates are similar to the predicate plates and they employ the same technology and principle of operation. Despite the minor differences, the results of the performance testing provided in the submission support that the subject device is substantially equivalent to the predicate device.
## 10. Performance Tests
- . Performance studies were conducted on the device imaging plate according to the FDA guidance "Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices".
- . The CRUXCAN (CRX-1000) has been tested for Electromagnetic Compatibility and Safety in accordance with the IEC60601 standards. The devices operated within the EMI emission, susceptibility and static discharge levels specified in the IEC 60601 standards.
- . Software Validation Test
The test results of the tests performed on the subject device supported that it is substantially equivalent to the predicate device.
Clinical testing was not necessary for the subject device, in order to demonstrate substantial equivalence.
## 11. Conclusions:
Based on the information provided in this premarket notification, CRUXELL Corp. concludes that the CRUXCAN (CRX-1000) is substantially equivalent to the predicate device as described herein in.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.