DVAS is an extra-oral source x-ray system to be used by trained dentists and dental technicians as an extra-oral x-ray source for producing diagnostic dental radiographic examination and diagnosis of teeth, jaw, and other oral structures using intra-oral image receptors. It is intended for both adult and pediatric subjects.
Device Story
DVAS is an extra-oral dental X-ray system; generates X-ray beams for diagnostic imaging of teeth, jaw, and oral structures. Used in dental clinics/hospitals by dentists or dental technicians. Device utilizes a cone for X-ray exposure; compatible with intra-oral image receptors (digital sensors, CR plates, or film). Output consists of anatomic X-ray images viewed via PC monitor or chemical film. Assists clinicians in diagnosing dental caries, periodontal diseases, root fractures, and other pathologies. Benefits include improved diagnostic accuracy for oral conditions.
Clinical Evidence
Bench testing only. No clinical data provided. Verification and validation testing confirmed conformance to intended use, user needs, and safety requirements. Compliance with IEC 60601-1, 60601-1-2, 60601-1-3, 60601-1-6, 60601-2-65, 62366, and 21 CFR 1020.30/31 verified.
Indicated for adult and pediatric patients requiring diagnostic dental radiographic examination of teeth, jaw, and oral structures. Used by trained dentists and dental technicians.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 8, 2023
GENORAY CO., Ltd % Kaitlynn Min Business Development GENORAY America Inc. 1220 N Simon Circle, Unit B ANAHEIM, CA 92806
Re: K232085
Trade/Device Name: DVAS (DVAS-M, DVAS-W) Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral Source X-Ray System Regulatory Class: Class II Product Code: EHD Dated: June 20, 2023 Received: November 3, 2023
Dear Kaitlynn Min:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Lu Jiang
Lu Jiang, Ph.D. Assistant Director Diagnostic X-Ray Systems Team DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
Submission Number (if known)
K232085
Device Name
DVAS (DVAS-M, DVAS-W)
#### Indications for Use (Describe)
DVAS is an extra-oral source x-ray system to be used by trained dentists and dental technicians as an extra-oral x-ray source for producing diagnostic dental radiographic examination and diagnosis of teeth, jaw, and other oral structures using intra-oral image receptors. It is intended for both adult and pediatric subjects.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Exhibit 5 K232085 510(k) Summary
This 510(k) summary information is prepared in accordance with 21 CFR 807.92
## 1. Date of Summary Preparation [21 CFR 807.92(a) (1)]
: Jun. 20, 2023
# 2. Administrative Information [21 CFR 807.92(a) (1)]
| 510(k) Submitter | GENORAY Co.,Ltd |
|------------------------|-------------------------------------------------------|
| | Address: 512, 560, Dunchon-daero, Jungwon-gu, |
| | Seongnam-si, Gyeonggi-Do, Korea |
| | Telephone No.: +82-31-5178-5500 |
| | Fax: +82-31-5178-5599 |
| | Contact Person: Inyoung Kim (iykim@genoray.com) |
| Official Correspondent | GENORAY America Inc. |
| | Address: 1220N Simon Circle, Anaheim, B, CA 92806 USA |
| | Telephone No.: +1-855-436-6729 |
| | Fax: +1-714-786-8919 |
| | Contact Person: Kaitlynn Min |
| | (kaitlynn@genorayamerica.com) |
# 3. Device Information [21 CFR 807.92(a) (2)]
| Trade / Device Name | DVAS (DVAS-M, DVAS-W) |
|---------------------------|-------------------------------|
| Common or Usual Name | Dental X-ray system |
| Classification Name | Extraoral source X-ray system |
| Classification Regulation | 21 CFR 872.1800 |
| Class of Device | Class II |
| Panel | Radiology |
| Product Code | EHD |
{4}------------------------------------------------
# 4. Predicate Device Information [21 CFR 807.92(a) (3)]
| * Predicate Device | |
|---------------------------|-----------------------------------|
| Name of Device | RIX 70 DC (K182206) |
| Manufacturer | Trident s.r.l., Italy |
| Common or Usual Name | Extraoral source X-ray system |
| Classification Name | Unit, X-ray, Extraoral with Timer |
| Classification Regulation | 21 CFR 872.1800 |
| Class of Device | Class II |
| Panel | Radiology |
| Product Code | EHD |
| * Reference Device | |
|---------------------------|-----------------------------------|
| Name of Device | PORT-X IV (K172810) |
| Manufacturer | GENORAY Co., Ltd |
| Common or Usual Name | Extraoral source X-ray system |
| Classification Name | Unit, X-ray, Extraoral with Timer |
| Classification Regulation | 21 CFR 872.1800 |
| Class of Device | Class II |
| Panel | Radiology |
| Product Code | EHD |
# 5. Description of the Device [21 CFR 807.92(a) (4)]
DVAS, the extra-oral source x-ray system for dentistry is the standard X-ray equipment used to acquire intraoral images of patients during dental diagnosis such as dental caries, periodontal diseases, dental root fracture, and other oral and dental pathologies. DVAS is used to provide anatomic X-ray images of a patient at hospitals or dental clinics using a cone for X-ray exposure.
The doctor or dentist can check the acquired X-ray images with chemical films or a PC monitor.
| No. | Item | Specification |
|-----|--------------------------------|-------------------------|
| 1 | Pixel Size | ≤ 40 μm |
| 2 | Active area size | ≥ 19 mm x 26 mm(Size 0) |
| 3 | Resolution | ≥ 10(lm/mm) |
| 4 | Digital sensor plate thickness | ≤ 19 μm * 19 μm |
| 5 | Interface | USB |
{5}------------------------------------------------
DVAS can be used Digital I/O Sensor, Image plate (CR), Film as a Image receptor. The image receptors are not part of this submission. And the minimum integration requirements for intra-oral detectors when used with this device are:
## 6. Indications for use [21 CFR 807.92(a) (5)]
DVAS is an extra-oral source x-ray system to be used by trained dentists and dental technicians as an extra-oral x-ray source for producing diagnostic dental radiographic examination and diagnosis of teeth, jaw, and other oral structures using intra-oral image receptors. It is intended for both adult and pediatric subjects.
{6}------------------------------------------------
# 7. Substantial equivalence chart [21 CFR 807.92(a) (6)]
| | Proposed device | | Predicate device | Reference device | SE Note | |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| | DVAS-M | DVAS-W | RIX 70 DC | PORT-X IV | | |
| Manufacturer | GENORAY Co., Ltd | | Trident s.r.l., Italy | GENORAY Co., Ltd | - | |
| 510(k) No. | K232085 | | K182206 | K172810 | - | |
| Classification<br>Name | Extraoral source x-ray system | | Extraoral source x-ray system | Extraoral source x-ray system | - | |
| Product<br>Code | EHD | | EHD | EHD | - | |
| Regulation Number | 872.1800 | | 872.1800 | 872.1800 | - | |
| Class | Class II | | Class II | Class II | - | |
| Product<br>illustration | Image: DVAS-M and DVAS-W | | Image: RIX 70 DC | Image: PORT-X IV | - | |
| Indication | DVAS is an extra-oral source x-ray | | RiX70 DC X-ray Unit is designed for use in | PORT-X IV is a | Same | |
| s for use | system to be used by trained dentists<br>and dental technicians as an extra oral x-<br>ray source for producing diagnostic<br>dental radiographic examination and<br>diagnosis of teeth, jaw, and other oral<br>structures using intra oral image<br>receptors. It is intended for both adult<br>and pediatric subjects. | | dental surgery to make endo-oral x-rays for<br>diagnostic purposes. This equipment can be<br>used to produce traditional x-rays developed<br>using chemicals or, alternatively, it can be used<br>with digital x-ray sensors. | | portable X-ray system<br>to be used by trained<br>dentists and dental<br>technicians as a<br>mobile, extraoral x-ray<br>source for producing<br>diagnostic x-ray images<br>using intraoral image<br>receptors. It is intended<br>for both adult and<br>pediatric subjects. | Same |
| Patient<br>type | Adult - Child | | Adult - Child | | Adult - Child | Same |
| Mechanical | | | | | | |
| Mechanical<br>configuration | Floor-mounted<br>type | Wall-mounted<br>type | Floor-mounted<br>type | Wall-mounted<br>type | Mobile type | Same |
| Minimum<br>source to<br>skin<br>distance | 200mm | | 200mm | | - | Same |
| X-ray<br>field Size<br>(default) | 60mm round | | 60mm round | | 60mm round | Same |
{7}------------------------------------------------
{8}------------------------------------------------
| Cycle | | | | |
|--------------------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|---------|
| Target material | Tungsten | Tungsten | Tungsten | Same |
| Target angle | 12.5° | 16° | 12.5° | Similar |
| Electrical | | | | |
| Electrical power voltage | AC 100-240V | AC 100-240V | AC 100-240V | Same |
| Focal spot | 0.4mm | 0.4mm | 0.4mm | Same |
| Exposure time | 0.05 ~ 1.6s | 0.02~2.0s | 0.05 ~ 1.6s | Similar |
| Tube current | 2.0mA fixed | 7mA fixed | 2mA fixed | Similar |
| Tube voltage | 70kV fixed | 60, 65, 70kV | 70kV fixed | Similar |
| Applied standard | IEC 60601-1 | IEC 60601-1 | IEC 60601-1 | Same |
| | IEC 60601-1-2<br>IEC 60601-1-3<br>IEC 60601-2-65<br>21 CFR 1020.30<br>21 CFR 1020.31 | IEC 60601-1-2<br>IEC 60601-1-3<br>IEC 60601-2-65<br>21 CFR 1020.30<br>21 CFR 1020.31 | IEC 60601-1-2<br>IEC 60601-1-3<br>IEC 60601-2-65<br>21 CFR 1020.30<br>21 CFR 1020.31 | |
{9}------------------------------------------------
#### 8. Safety, EMC and Performance data comparison to Predicate [21 CFR 807.92(b)]
DVAS have been successfully compeleted verification and validation testing per GENORAY quality system as well engineering bench testing in support of successfully completed verification and validation testing per GENORAY quality system and this submission.
The system has been tested and is compliant with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-3. IEC 60601-1-6. IEC 60601-2-65. IEC 62366. Also. DVAS complies with all applicable 21 CFR performance standards (21 CFR 1020.30, 21 CFR 1020.31)
And Software was validated according to the FDA Guidance "Guidance for the Content of Premarket Submissions for Software Contained in Medical devices", FDA Guidance "Guidance for the content of premarket submissions for management of cyber security". Results demonstrated that all executed verification tests were passed.
DVAS can be used not only in adults but also In the pediatric population, and therefore the system has reflected pediatric information in the labeling according to the FDA Guidance "Pediatric Information for X-ray Imaging Device Premarket Notifications," dated November 28, 2017"
Non-clinical validation testing has been performed to validate that DVAS conform to the intended use, claims, user needs, effectiveness of safety measures and instructions for use.
As a results, all test results were satisfactory and the result of bench tests indicates that the new device is as safe and effective as the predicate device.
#### 9. Conclusion
In reference to the comparison information provided in the substantial equivalence chart, and the most of functions and electronic features are similar with the predicate device. We believe that the DVAS are safe and effective as predicate device, and have no new indication for use. Therefore, DVAS are substantially equivalent to predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.