PORT-X IV is a portable X-ray system to be used by trained dentists and dental technicians as a mobile, extraoral x-ray source for producing diagnostic x-ray images using intraoral image receptors. It is intended for both adult and pediatric subjects.
Device Story
PORT-X IV is a portable, battery-powered, handheld extraoral X-ray system. Operated by dentists or dental technicians in clinical settings, it generates X-rays to produce diagnostic images on intraoral receptors. The device features a stationary anode X-ray tube (70 kV, 2 mA) and an LCD control panel for setting exposure parameters (0.05–1.6 sec). The operator triggers exposure via a hand switch. The system includes a rechargeable battery and recharging unit. By providing a mobile source of ionizing radiation, it enables point-of-care dental imaging, facilitating diagnosis of dental conditions. Safety is managed through compliance with radiation control standards (21 CFR 1020.30/31) and electrical safety standards (IEC 60601-1).
Clinical Evidence
Bench testing and clinical evaluation performed. Bench testing confirmed compliance with 21 CFR 1020.30/31 (focal spot to skin distance >18 cm, minimum HVL 1.5 mm) and IEC 60601-1 (loading factor accuracy, air kerma reproducibility). Software validation and EMC testing (IEC 60601-1-2) were completed. All results met essential performance requirements.
Technological Characteristics
Portable extraoral X-ray system; stationary anode tube (70 kV, 2 mA); 0.4 mm focal spot; 12.5° target angle; 2.2 mmAl total filtration; rechargeable battery; LCD display; hand switch. Complies with IEC 60601-1, IEC 60601-1-3, IEC 60601-2-65, and 21 CFR 1020.30/31.
Indications for Use
Indicated for adult and pediatric patients requiring diagnostic X-ray imaging using intraoral receptors; to be operated by trained dentists or dental technicians.
Regulatory Classification
Identification
An extraoral source x-ray system is an AC-powered device that produces x-rays and is intended for dental radiographic examination and diagnosis of diseases of the teeth, jaw, and oral structures. The x-ray source (a tube) is located outside the mouth. This generic type of device may include patient and equipment supports and component parts.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
GENORAY Co., Ltd. % Mr. Chris Yoon Business Development Manager Genoray America Inc. 147 E. Bristol Lane ORANGE CA 92865
March 7, 2018
Re: K172810
Trade/Device Name: PORT-X IV Regulation Number: 21 CFR 872.1800 Regulation Name: Extraoral source x-ray system Regulatory Class: II Product Code: EHD Dated: January 9, 2018 Received: February 12, 2018
Dear Mr. Yoon:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice
(https://www.fda.gov/MedicalDevices/DeviceRegulationandGuidance/) and CDRH Learn (http://www.fda.gov/Training/CDRHLearn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (http://www.fda.gov/DICE) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Robert Sauer
for Robert Ochs, Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K172810
Device Name PORT-X IV
Indications for Use (Describe)
PORT-X IV is a portable X-ray system to be used by trained dental technicians as a mobile, extra oral x-ray source for producing diagnostic x-ray images using intra oral image receptors. It is intended for both adult and pediativ subjects.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------|
| <div> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
| X Prescription Use (Part 21 CFR 801 Subpart D)
|_ | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### Exhibit 5 510(k) Summary
Date of Summary Preparation: Aug. 11, 2017
| 1. Submitter and US Official Correspondent | |
|---------------------------------------------------------------|-------------------------------------------------------------------------------|
| Submitter : | GENORAY Co., Ltd. |
| Address: | 512, 560, Dunchon-daero, Jungwon-gu,<br>Seongnam-si, Gyeonggi-do,13230, KOREA |
| Telephone No.: | +82-31-740-4100 |
| Fax: | +82-31-737-8018 |
| Official Correspondent (U.S): Kaitlynn Min - Business Manager | |
| Correspondent: | GENORAY America Inc. |
|----------------|-------------------------------------------|
| Address: | 147 E. Bristol Lane, Orange, CA 92865 USA |
| Telephone No.: | +1-855-436-6729 |
| Fax: | +1-714-786-8919 |
| Email: | kaitlynn @ genorayamerica.com |
## 2. Establishment Registration Number 3005843418
### 3. Device Information
| Trade/Device Name: | PORT-X IV |
|----------------------|-----------------------------------------|
| Regulation Name: | Portable X-ray System |
| Classification Name: | Extraoral Source X-ray System |
| Product Code: | EHD |
| Device Class: | Class II per regulation 21 CFR 872.1800 |
- 4. Predicate Device(Equivalent Legally Marketed Device)
| Manufacturer: | Metabiomed, Inc. |
|-----------------|--------------------------------------------------------------------------|
| Device Name: | REXTAR X |
| 510(k) Number: | K132041 (Decision Date – Jul 31, 2013) |
| Classification: | Extraoral Source X-ray System<br>Class II per regulation 21 CFR 872.1800 |
- 5. Description of the Device
The portable X-Ray system, PORT-X IV, is intended to be used by trained dentists and dental technicians as an extraoral x-ray source for producing diagnostic x-ray images using intraoral image receptors. It is the X-ray equipment and uses it to diagnose the patient to
{4}------------------------------------------------
acquire images. It also includes accessories, which are battery, recharging unit and hand switch.
## 6. Indications for use (intended use)
PORT-X IV is a Portable X-Ray System to be used by trained dentists and dental technicians as a mobile, extraoral x-ray source for producing diagnostic x-ray images using intraoral image receptors. It is intended for both adult and pediatric subjects.
| | Proposed device | Predicate device |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | PORT-X IV | REXTAR X |
| Manufacturer | GENORAY Co., Ltd. | Metabiomed, Inc. |
| 510(k) No. | - | K132041 |
| Indications<br>for use | PORT-X IV is a portable X-ray<br>system to be used by trained dentists<br>and dental technicians as a mobile,<br>extraoral x-ray source for producing<br>diagnostic x-ray images using<br>intraoral image receptors. It is<br>intended for both adult and pediatric<br>subjects. | REXTAR X is a portable X-ray<br>system to be used by trained dentists<br>and dental technicians as a mobile,<br>extraoral x-ray source for producing<br>diagnostic x-ray Images using<br>Intraoral Image receptors. It is<br>intended for both adult and pediatric<br>subjects. |
| X-ray<br>Generator | Tube Voltage : 70 kV<br>Tube Current : 2 mA | Tube Voltage : 70 kV<br>Tube Current : 2 mA |
| X-ray Controller | Exposure Time : 0.05~1.6 sec | Exposure Time : 0.01~1.30 sec |
| X-ray Tube | Stationary Anode | Stationary Anode |
| | Focus : 0.4 mm | Focus : 0.4 mm |
| | Target Angle : 12.5°<br>Total Filtration : 2.2 mmAl<br>(Inherent Filtration : 1.0 mmAl;<br>Additional Filtration : 1.2 mmAl) | Target Angle : 12°<br>Total Filtration : 1.5 mmAl<br>(Inherent Filtration : 1.0 mmAl;<br>Additional Filtration : 0.5 mmAl) |
| | | |
| Battery | Type : Rechargeable | Type : Rechargeable |
| Display | LCD panel Display | LCD panel Display |
| Dimensions and<br>Weight | Dimension(mm):140 x 173 x 254<br>Weight(kg): 1.6 kg | Dimension(mm): 146 x 155 x 139<br>Weight(kg): 2 kg |
- 7. Substantial equivalence chart
{5}------------------------------------------------
The technical characteristics of the PORT-X IV and predicate device are similar. However the range of the exposure time and the total filtration of X-rav tube are little different. Those differences does not affect the performance and safety. EMC and performance data demonstrated that the PORT-X IV is as safe and effective as the predicate device.
The PORT-X IV is substantially equivalent to the predicate device, REXTAR X(K132041).
## 8. Safety, EMC and Performance data comparison to Predicate device
- Safety, EMC and Performance Data
PORT-X IV complies with industry standards such as IEC 60601-1 Series and 21 CFR 1020.30 and 21CFR 1020.31 to minimize electrical, mechanical and radiation hazards.
- · Electrical, mechanical, environmental safety and performance testing which are mentioned in the standard IEC 60601-1, IEC 60601-1-3 and IEC 60601-2-65 were performed.
- · EMC testing was conducted in accordance with standard IEC 60601-1-2.
- · Performance testing performed according to FDA 21 CFR 1020.30, 21 CFR 1020.31 standards, Software Validation, Clinical Evaluation. All test results were satisfactory.
- Summary of Performance Testing
The performance test for the subject device, PORT-X IV and the predicate device, REXTAR X(K132041) confirmed that the focal spot to skin distance of both devices were longer than the minimum length of 18 cm in accordance with Federal Standard (21CFR 1020.30 and 31) requirements. Also minimum HVL (half-value layer) of both devices were 1.5 mm in accordance with Federal Standard (21CFR 1020.30 and 31) requirements. Accuracy of loading factors and reproducibility of AIR KERMA for both devcies also met the essential performance requirements in accordance with IEC 60601-1 (ex. <70kVp +10%).
All the test results were satisfied and the result of bench and clinical evaluation indicates that the new device is as safe and effective as the predicate device.
## 9. FDA Guidance Documents Utilized
- Radiation Safety Considerations for X-Ray Equipment Designed for Hand-Held Use, issued December 24,2008
- Guidance for the Content of Premakret Submissions for Software Contained in Medical Devices, issued Nov 28, 2017
- Pediatric Information for X-ray Imaging Device Premarket Notifications, issued May 10, 2012
- Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, issued October 2, 2014
{6}------------------------------------------------
- General Principles of Software Validation; Final Guidance for Industry and FDA Staff, issued Jun 9, 1997.
- Guidance for the Submission of 510(k)'s for Solid State X-ray Imaging Devices, Aug 6, 1999.
## 10. Conclusion
In reference to the comparison information provided in substantial equivalence chart, most of functions and electronic features are similar with predicate device. We believe that the PORT-X IV is safe and effective as predicate device, and has no new indication for use. Therefore, PORT-X IV is substantially equivalent to predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.