K231999 · Wipak OY · FRG · Sep 7, 2023 · General Hospital
Device Facts
Record ID
K231999
Device Name
Steriking® LT-Blueline Pouches with Tyvek®
Applicant
Wipak OY
Product Code
FRG · General Hospital
Decision Date
Sep 7, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
Steriking® LT-Blueline Pouches with Tyvek® are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERRAD® Sterilizer Systems. The device is intended to allow sterilization of enclosed devices and to maintain sterility of the enclosed devices until used up to 1 years post sterilization. The products are for single use only. The Steriking® LT-Blueline Pouches with Tyvek are intended for use in the following STERRAD® Sterilization Cycles: NX Standard NX Advanced 100NX Duo 100NX Flex 100NX Standard
Device Story
Steriking® LT-Blueline Pouches with Tyvek® are single-use sterilization packaging for medical devices; constructed from Tyvek® and BOPET/PE plastic film; heat-sealed on three sides with an integrated adhesive strip on the fourth side for self-sealing by healthcare workers without external heat-sealing equipment. Used in clinical settings to contain devices during STERRAD® hydrogen peroxide (H2O2) gas plasma sterilization cycles. The pouch allows sterilant penetration while maintaining a microbial barrier post-sterilization. Healthcare providers use the pouches to ensure device sterility until point-of-use. The device benefits patients by providing a validated, sterile barrier for surgical instruments, reducing infection risk.
Clinical Evidence
No clinical data. Bench testing only. Testing included dye migration (ASTM F1929), tensile strength (ASTM F88), cytotoxicity (ISO 10993-5), microbial aerosol challenge (AAMI TIR12), and sterilization penetration (6-log reduction of Geobacillus stearothermophilus). Accelerated aging (3 years) and 365-day sterility maintenance studies confirmed package integrity and sterility maintenance post-sterilization.
Technological Characteristics
Materials: Tyvek® backing and BOPET/PE transparent film. Construction: Heat-sealed on three sides, self-sealing adhesive strip on the fourth. Sterilization: H2O2 gas plasma (STERRAD®). Standards: ASTM F1929, ASTM F88, ISO 11607, ISO 10993. Connectivity: None. Software: None.
Indications for Use
Indicated for healthcare workers to enclose medical devices for sterilization in STERRAD® Sterilizer Systems (NX Standard, NX Advanced, 100NX Duo, 100NX Flex, 100NX Standard). Supports single-use, terminal sterilization of devices with specific weight, lumen count (max 1), lumen ID (1-2mm), and lumen length (up to 850mm) limits. Maintains sterility for up to 1 year post-sterilization.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Steriking® LT-Blueline Pouches with Tyvek® (K221377)
Submission Summary (Full Text)
{0}------------------------------------------------
September 7, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left side of the logo is the Department of Health & Human Services logo. To the right of that is a blue square with the letters FDA in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Wipak OY % Amanda Singleton Consultant Compliance Systems International LLC 39 Lockhart Circle Amherst, New York 14228
Re: K231999
Trade/Device Name: Steriking® LT-Blueline Pouches with Tyvek® Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG Dated: August 11, 2023 Received: August 11, 2023
## Dear Amanda Singleton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Digitally signed
by Eileen Cadel
-S
Eileen
Cadel -S Date:
for
2023.09.07
13:49:47 -04'00'
Colin O'Neill, M.B.E.
Assistant Director
DHT6B: Division of Spinal Devices
OHT6: Office of Orthopedic Devices
Office of Product Evaluation and Quality
Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
### K231999
#### Device Name
Steriking® LT-Blueline Pouches with Tyvek®
#### Indications for Use (Describe)
Steriking® LT-Blueline Pouches with Tyvek® are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERRAD® Sterilizer Systems. The device is intended to allow sterilization of enclosed devices and to maintain sterility of the enclosed devices until used up to 1 years post sterilization. The products are for single use only. The Steriking® LT-Blueline Pouches with Tyvek are intended for use in the following STERRAD® Sterilization Cycles:
NX Standard NX Advanced 100NX Duo 100NX Flex 100NX Standard
Max weights: NX Standard Cycle: For pouch sizes 90 x 200 mm, 90 x 250 mm, 130 x 380 mm, .094 lbs. For pouch sizes 190 x 330 mm, 250 x 400 mm, 300 x 450 mm, 1.194 pounds.
NX Advanced Cycle: For pouch sizes 90 x 200 mm, 90 x 250 mm, 130 x 380 mm, .096 lbs. For pouch sizes 190 x 330 mm, 250 x 400 mm, 300 x 450 mm, 1.224 pounds.
100NX Duo Cycle: For pouch sizes 90 x 200 mm. 90 x 250 mm. 130 x 380 mm. .116 lbs. For pouch sizes 190 x 330 mm. 250 x 400 mm, 300 x 450 mm, 1.222 pounds.
100NX Flex Cycle: For pouch sizes 90 x 200 mm, 90 x 250 mm, 130 x 380 mm, .202 lbs. For pouch sizes 190 x 330 mm, 250 x 400 mm, 300 x 450 mm, 1.220 pounds.
100NX Standard Cycle: For pouch sizes 90 x 200 mm. 90 x 250 mm. 130 x 380 mm. .202 lbs. For pouch sizes 190 x 330 mm. 250 x 400 mm. 300 x 450 mm. 4.478 pounds.
Max Count of lumens: 1 lumen max for All cycles
Max ID of lumens:
NX Standard Cycle: 1 mm ID for all pouch sizes
NX Advanced Cycle: 2 mm ID for all pouch sizes
{3}------------------------------------------------
100NX Cycle: 1 mm ID for all pouch sizes
100NX Flex Cycle: 1 mm ID for all pouch sizes
100NX Standard Cycle: 2 mm ID for all pouch sizes
Max lumen lengths: NX Standard Cycle: For pouch sizes 90 x 200 mm. 90 x 250 mm. 130 x 380 mm. 125 mm. For pouch sizes 190 x 330 mm. 250 x 400 mm, 300 x 450 mm, 250 mm.
NX Advanced Cycle: For pouch sizes 90 x 200 mm. 90 x 250 mm. 130 x 380 mm. 125 mm. For pouch sizes 190 x 330 mm. 250 x 400 mm, 300 x 450 mm, 250 mm.
100NX Duo Cycle: 850 mm for all pouch sizes
100NX Flex Cycle: 850 mm for all pouch sizes
100NX Standard Cycle:
For pouch sizes 90 x 200 mm, 90 x 250 mm, 130 x 380 mm, 125 mm. For pouch sizes 190 x 330 mm, 250 x 400 mm, 300 x 450 mm, 250 mm.
Pouch Sizes: 300 x 450 mm 250 x 400 mm 190 x 330 mm 130 x 380 mm 130 x 270 mm 90 x 250 mm 90 x 200 mm
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# K231999 510(k) Summary (in accordance with 21CFR807.92)
| 510K Summary Elements per<br>21CFR807.92 | Summary |
|--------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's name, address, telephone<br>number, a contact person, and the date<br>the summary was prepared | Wipak Oy<br>Wipaktie 2<br>Nastola Finland<br>Contact: Hanna Marttila<br>Phone: +358 (0)40 124 0290<br>Date prepared: September 6, 2023<br>Submitter: Amanda Singleton, Compliance Systems International, 716.440.7364 |
| Name of the device<br>including the trade or<br>proprietary name if<br>applicable | Proprietary Name: Steriking® LT-Blueline Pouches with Tyvek® |
| Common or usual name | Common Name: Peel Pouch |
| Classification name | Classification Name: Sterilization wrap |
| Classification Product Code | FRG |
| Device Classification | Class II |
| Regulation Number | 21 CFR 880.6850 |
| Identification of the legally<br>marketed device to which<br>the submitter claims<br>equivalence<br>(Primary Predicate device) | K221377 -Steriking® LT-Blueline Pouches with Tyvek® |
| Description of the device | Steriking® LT-Blueline Pouches with Tyvek® are intended to be used to contain<br>medical devices to be terminally sterilized in the STERRAD® Sterilization<br>System. The medical devices are inserted into the Pouch, sealed, and then<br>sterilized in the STERRAD ® Sterilization System. Sterilization Cycles are noted<br>below. After completion of the sterilization process, the Pouch maintains<br>sterility of the enclosed medical devices until the seal is opened. These pouches<br>are made from a plastic film and Tyvek that is heat-sealed on three sides. The<br>fourth side is left opened and will be self-sealed when used.<br><br>Steriking® LT-Blueline Pouches with Tyvek® are intended to allow sterilization<br>of enclosed devices and to maintain sterility of the enclosed devices until used<br>up to 1 year post sterilization. The products are for single use only.<br><br>The pouches are constructed from Tyvek®/plastic films. The self-sealed<br>pouches are self sealed prior to processing in the STERRAD® Sterilization<br>Systems.<br><br>Sterilization Systems:<br><br>STERRAD® Cycles:<br>NX Standard<br>NX Advanced<br>100NX Duo<br>100NX Flex<br>100NX Standard |
| Indications for Use | Steriking® LT-Blueline Pouches with Tyvek® are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERRAD® Sterilizer Systems. The device is intended to allow sterilization of enclosed devices and to maintain sterility of the enclosed devices until used up to 1 years post sterilization. The products are for single use only. The Steriking® LT-Blueline Pouches with Tyvek are intended for use in the following STERRAD® Sterilization Cycles:<br>NX Standard<br>NX Advanced<br>100NX Duo<br>100NX Flex<br>100NX Standard<br><br>Max weights:<br>NX Standard Cycle:<br>For pouch sizes 90 x 200 mm, 90 x 250 mm, 130 x 270 mm, 130 x 380 mm, .094 lbs. For pouch sizes 190 x 330 mm, 250 x 400 mm, 300 x 450 mm, 1.194 pounds.<br><br>NX Advanced Cycle:<br>For pouch sizes 90 x 200 mm, 90 x 250 mm, 130 x 270 mm, 130 x 380 mm, .096 lbs. For pouch sizes 190 x 330 mm, 250 x 400 mm, 300 x 450 mm, 1.224 pounds.<br><br>100NX Duo Cycle:<br>For pouch sizes 90 x 200 mm, 90 x 250 mm, 130 x 270 mm, 130 x 380 mm, .116 lbs. For pouch sizes 190 x 330 mm, 250 x 400 mm, 300 x 450 mm, 1.222 pounds.<br><br>100NX Flex Cycle:<br>For pouch sizes 90 x 200 mm, 90 x 250 mm, 130 x 270 mm, 130 x 380 mm, .202 lbs. For pouch sizes 190 x 330 mm, 250 x 400 mm, 300 x 450 mm, 1.220 pounds.<br><br>100NX Standard Cycle:<br>For pouch sizes 90 x 200 mm, 90 x 250 mm, 130 x 270 mm, 130 x 380 mm, .202 lbs. For pouch sizes 190 x 330 mm, 250 x 400 mm, 300 x 450 mm, 4.478 pounds.<br><br>Max Count of lumens:<br>1 lumen max for All cycles<br><br>Max ID of lumens:<br>NX Standard Cycle:<br>1 mm ID for all pouch sizes<br><br>NX Advanced Cycle:<br>2 mm ID for all pouch sizes<br><br>100NX Standard Cycle:<br>1 mm ID for all pouch sizes |
{5}------------------------------------------------
{6}------------------------------------------------
| Technological Characteristics Comparison Table | | | |
|------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Submission Device –<br>K231999<br>Steriking® LT-Blueline Pouches<br>with Tyvek® | Comparison | Primary Predicate Device –<br>K221377<br>Steriking® LT-Blueline Pouches<br>with Tyvek® |
| Device Classification | Class II | Same | Class II |
| Classification Name | Sterilization wrap | Same | Sterilization wrap |
| Regulation Name | 21 CFR 880.6850 | Same | 21 CFR 880.6850 |
| Product Code | FRG | Same | FRG |
| Indications for Use | Steriking® LT-Blueline Pouches<br>with Tyvek® are intended to<br>provide health care workers with<br>an effective method to enclose<br>devices intended for sterilization<br>in the STERRAD® | Different | The Steriking® LT-Blueline<br>Pouches with Tyvek are<br>intended to provide health care<br>workers with an effective<br>method to enclose devices<br>intended for sterilization in the |
{7}------------------------------------------------
| | | | Pouch Sizes: | | Pouch Sizes: |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sterilizer Systems. The device is<br>intended to allow sterilization of<br>enclosed devices and to<br>maintain sterility of the enclosed<br>devices until used up to 1 years<br>post sterilization. The products<br>are for single use only. The<br>Steriking® LT-Blueline Pouches<br>with Tyvek are intended for use<br>in the following STERRAD®<br>Sterilization Cycles:<br>NX Standard<br>NX Advanced<br>100NX Duo<br>100NX Flex<br>100NX Standard | STERRAD® Sterilizer Systems.<br>The device is intended to allow<br>sterilization of enclosed devices<br>and to maintain sterility of the<br>enclosed devices until used up<br>to 3 years post sterilization. The<br>products are for single use only<br>The Steriking® LT-Blueline<br>Pouches with Tyvek are<br>intended for use in the<br>following STERRAD®<br>Sterilization Cycles:<br>NX Standard<br>NX Advanced<br>100NX Duo<br>100NX Flex<br>100NX Standard | | 300 x 450 mm | | 250 mm x 390 mm |
| Max weights:<br>NX Standard Cycle:<br>For pouch sizes 90 x 200 mm, 90<br>x 250 mm, 130 x 270 mm, 130 x<br>380 mm, .094 lbs. For pouch<br>sizes 190 x 330 mm, 250 x 400<br>mm, 300 x 450 mm, 1.194<br>pounds. | Max weights:<br>NX Standard Cycle:<br>75 mm x 200 mm, 100 mm x<br>250 mm, 150 mm x 300 mm,<br>160 mm x 440 mm, 160 mm x<br>600 mm: .088 pounds<br>205 mm x 390 mm, 250 mm x<br>390 mm, 250 mm x 500 mm:<br>1.382 pounds | | 250 x 400 mm | | 205 mm x 390 mm |
| NX Advanced Cycle:<br>For pouch sizes 90 x 200 mm, 90<br>x 250 mm, 130 x 270 mm, 130 x<br>380 mm, .096 lbs. For pouch<br>sizes 190 x 330 mm, 250 x 400<br>mm, 300 x 450 mm, 1.224<br>pounds. | NX Advanced Cycle:<br>75 mm x 200 mm, 100 mm x<br>250 mm, 150 mm x 300 mm,<br>160 mm x 440 mm, 160 mm x<br>600 mm: .084 pounds<br>205 mm x 390 mm, 250 mm x<br>390 mm, 250 mm x 500 mm:<br>1.393 pounds | | 190 x 330 mm | | 160 mm x 600 mm |
| 100NX Duo Cycle:<br>For pouch sizes 90 x 200 mm, 90<br>x 250 mm, 130 x 270 mm, 130 x<br>380 mm, .116 lbs. For pouch<br>sizes 190 x 330 mm, 250 x 400<br>mm, 300 x 450 mm, 1.222<br>pounds. | 100NX Duo Cycle:<br>75 mm x 200 mm, 100 mm x<br>250 mm, 150 mm x 300 mm,<br>160 mm x 440 mm, 160 mm x<br>600 mm: .104 pounds<br>205 mm x 390 mm, 250 mm x<br>390 mm, 250 mm x 500 mm:<br>1.400 pounds | | 130 x 380 mm | | 160 mm x 440 mm |
| 100NX Flex Cycle:<br>For pouch sizes 90 x 200 mm, 90<br>x 250 mm, 130 x 270 mm, 130 x<br>380 mm, .202 lbs. For pouch<br>sizes 190 x 330 mm, 250 x 400<br>mm, 300 x 450 mm, 1.220<br>pounds. | 100NX Flex Cycle: | | 130 x 270 mm | | 150 mm x 300 mm |
| 100NX Standard Cycle:<br>For pouch sizes 90 x 200 mm, 90 | | | 90 x 250 mm | | 100 mm x 250 mm |
| 380 mm, .202 lbs. For pouch<br>sizes 190 x 330 mm, 250 x 400<br>mm, 300 x 450 mm, 4.478<br>pounds. | 75 mm x 200 mm, 100 mm x<br>250 mm, 150 mm x 300 mm,<br>160 mm x 440 mm, 160 mm x<br>600 mm: .158 pounds | | 90 x 200 mm | | 75 mm x 200 mm |
| Max Count of lumens:<br>1 lumen max for All cycles | 205 mm x 390 mm, 250 mm x<br>390 mm, 250 mm x 500 mm:<br>1.400 pounds | Design | These pouches are made from a<br>plastic film and Tyvek that is<br>heat sealed on three sides. The<br>fourth side is left opened and<br>will be self-sealed when used. | | These pouches are made from<br>a plastic film and Tyvek that is<br>heat sealed on three sides. The<br>fourth side is left opened and<br>will be heat-sealed when used. |
| Max ID of lumens: | 100NX Standard Cycle: | Backing Material | Tyvek® | No Change | Tyvek® |
| NX Standard Cycle:<br>1 mm ID for all pouch sizes | 75 mm x 200 mm, 100 mm x<br>250 mm, 150 mm x 300 mm,<br>160 mm x 440 mm, 160 mm x<br>600 mm: .158 pounds | Transparent Film | BOPET/PE | No Change | BOPET/PE |
| NX Advanced Cycle:<br>2 mm ID for all pouch sizes | 205 mm x 390 mm, 250 mm x<br>390 mm, 250 mm x 500 mm:<br>4.476 pounds | Sterilization Processes | H2O2 Sterilization Process as per:<br>STERRAD®<br>NX Standard<br>NX Advanced<br>100NX Duo<br>100NX Flex<br>100NX Standard | No Change | H2O2 Sterilization Process as per:<br>STERRAD®<br>NX Standard<br>NX Advanced<br>100NX Duo<br>100NX Flex<br>100NX Standard |
| 100NX Cycle:<br>1 mm ID for all pouch sizes | Max lumens:<br>1 mm min ID and Max 1 lumen<br>for all cycles | Sterilant<br>Validation | Achieved a 10-6 Sterility Assurance<br>Level (SAL) of Geobacillus<br>stearothermophilus.<br>NX Standard<br>NX Advanced<br>100NX Duo<br>100NX Flex<br>100NX Standard | No Change | Achieved a 10-6 Sterility Assurance<br>Level (SAL) of Geobacillus<br>stearothermophilus.<br>NX Standard<br>NX Advanced<br>100NX Duo<br>100NX Flex<br>100NX Standard |
| 100NX Flex Cycle:<br>1 mm ID for all pouch sizes | NX Standard Cycle: For pouch<br>sizes 250x500mm, 250x390mm<br>205x390mm is 125 mm max<br>length. Max length is 50 mm for<br>all other pouch sizes. | Package Integrity | Pass | No Change | Pass |
| 100NX Standard Cycle:<br>2 mm ID for all pouch sizes | NX Advanced Cycle: For pouch<br>sizes 250x500mm, 250x390mm<br>205x390mm is 125 mm max<br>length. Max length is 50 mm for<br>all other pouch sizes. | Biocompatibility | Pass | No Change | Pass |
| Max lumen lengths:<br>NX Standard Cycle:<br>For pouch sizes 90 x 200 mm, 90<br>x 250 mm, 130 x 270 mm, 130 x<br>380 mm, 125 mm. For pouch<br>sizes 190 x 330 mm, 250 x 400<br>mm, 300 x 450 mm, 250 mm. | 100NX Duo Cycle: For pouch<br>sizes 250x500mm, 250x390mm<br>205x390mm is 875 mm max<br>length. Max length is 850 mm<br>for all other pouch sizes. | Shelf-Life Pre-Sterilization | 3 years | Different | 5 years |
| NX Advanced Cycle:<br>For pouch sizes 90 x 200 mm, 90<br>x 250 mm, 130 x 270 mm, 130 x<br>380 mm, 125 mm. For pouch<br>sizes 190 x 330 mm, 250 x 400<br>mm, 300 x 450 mm, 250 mm. | 100NX Flex Cycle: Max length<br>is 850 mm for all pouch sizes. | Maintenance of Sterility | 1 year | Different | 3 years |
| 100NX Duo Cycle:<br>850 mm for all pouch sizes | 100NX Standard Cycle: For<br>pouch sizes 250x500mm,<br>250x390mm, 205x390mm is<br>125 mm max length. Max<br>length is 50 mm for all other<br>pouch sizes. | | | | |
| 100NX Flex Cycle:<br>850 mm for all pouch sizes | Pouch Sizes:<br>250 mm x 500 mm | | | | |
| 100NX Standard Cycle:<br>For pouch sizes 90 x 200 mm, 90<br>x 250 mm, 130 x 270 mm, 130 x<br>380 mm, 125 mm. For pouch<br>sizes 190 x 330 mm, 250 x 400<br>mm, 300 x 450 mm, 250 mm. | | | | | |
{8}------------------------------------------------
{9}------------------------------------------------
{10}------------------------------------------------
| Max weights | | | | | | Different | | | | | |
|----------------------|-------------------------------------------------------------------------------------|-------------------------|-----------------------------------------------------------------------------------------------------------|------------------------|----------------------------|--------------|-------------------------|-------------------------|-----------------------|------------------------|----------------------------|
| Pouch size | NX<br>Advanced<br>Cycle | NX<br>Standard<br>Cycle | 100NX<br>Duo<br>Cycle | 100NX<br>Flex<br>Cycle | 100NX<br>Standard<br>Cycle | Pouch size | NX<br>Advanced<br>Cycle | NX<br>Standard<br>Cycle | 100NX<br>Duo<br>Cycle | 100NX<br>Flex<br>Cycle | 100NX<br>Standard<br>Cycle |
| 300 x 450 mm | 1.224 | 1.194 | 1.222 | 1.220 | 4.748 | 250 x 500 mm | 1.393 | 1.382 | 1.400 | 1.400 | 4.476 |
| 250 x 400 mm | 1.224 | 1.194 | 1.222 | 1.220 | 4.748 | 250 x 390 mm | 1.393 | 1.382 | 1.400 | 1.400 | 4.476 |
| 190 x 330 mm | 1.224 | 1.194 | 1.222 | 1.220 | 4.748 | 205 x 390 mm | 1.393 | 1.382 | 1.400 | 1.400 | 4.476 |
| 130 x 380 mm | 0.096 | 0.094 | 0.116 | 0.202 | 0.202 | 160 x 600 mm | 0.084 | 0.088 | 0.104 | 0.158 | 0.158 |
| 130 x 270 mm | 0.096 | 0.094 | 0.116 | 0.202 | 0.202 | 160 x 440 mm | 0.084 | 0.088 | 0.104 | 0.158 | 0.158 |
| 90 x 250 mm | 0.096 | 0.094 | 0.116 | 0.202 | 0.202 | 150 x 300 mm | 0.084 | 0.088 | 0.104 | 0.158 | 0.158 |
| 90 x 200 mm | 0.096 | 0.094 | 0.116 | 0.202 | 0.202 | 100 x 250 mm | 0.084 | 0.088 | 0.104 | 0.158 | 0.158 |
| | | | | | | 75 x 200 mm | 0.084 | 0.088 | 0.104 | 0.158 | 0.158 |
| Max Lumen Dimensions | Max Count of lumens: | Different | 1 mm min ID Lumen and Max 1 lumen for all cycles | | | | | | | | |
| | 1 lumen max for all pouch sizes | | | | | | | | | | |
| | | | | | | | | | | | |
| | Max lumen Id: | | For NX Standard Cycle: For<br>pouch sizes 250x500mm,<br>250x390mm, 205x390mm is<br>125 mm max length. Max | | | | | | | | |
| | | | | | | | | | | | |
| | NX Standard: | | | | | | | | | | |
| | 1 mm ID for all pouch sizes | | length is 50 mm for all other<br>pouch sizes. | | | | | | | | |
| | | | | | | | | | | | |
| | NX Advanced: | | NX Advanced Cycle: For pouch | | | | | | | | |
| | 2 mm ID for all pouch sizes | | sizes 250x500mm, 250x390mm,<br>205x390mm is 125 mm max | | | | | | | | |
| | | | length. Max length is 50 mm for | | | | | | | | |
| | 100NX Duo: | | all other pouch sizes. | | | | | | | | |
| | 1 mm ID for all pouch sizes | | | | | | | | | | |
| | | | 100NX Duo Cycle: For pouch | | | | | | | | |
| | 100NX Flex: | | sizes 250x500mm, 250x390mm, | | | | | | | | |
| | 1 mm ID for all pouch sizes | | 205x390mm is 875 mm max | | | | | | | | |
| | | | length. Max length is 850 mm | | | | | | | | |
| | 100NX Standard: | | for all other pouch sizes. | | | | | | | | |
| | 2 mm ID for all pouch sizes | | | | | | | | | | |
| | | | 100NX Flex Cycle: Max length | | | | | | | | |
| | Max lumen lengths: | | is 850 mm for all pouch sizes. | | | | | | | | |
| | | | | | | | | | | | |
| | NX Standard Cycle: | | 100NX Standard Cycle: For pouch | | | | | | | | |
| | For pouch sizes 90 x 200 mm, 90<br>x 250 mm, 130 x 270 mm, 130 x<br>380 mm, 125 mm. | | sizes 250x500mm, 250x390mm, | | | | | | | | |
| | | | 205x390mm is 125 mm max | | | | | | | | |
| | For pouch sizes 190 x 330 mm,<br>250 x 400 mm, 300 x 450 mm,<br>250 mm. | | length. Max length is 50 mm for all | | | | | | | | |
| | | | other pouch sizes. | | | | | | | | |
| | | | | | | | | | | | |
| | NX Advanced Cycle: | | | | | | | | | | |
| | For pouch sizes 90 x 200 mm, 90<br>x 250 mm, 130 x 270 mm, 130 x<br…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.