K221377 · Wipak OY · FRG · Dec 22, 2022 · General Hospital
Device Facts
Record ID
K221377
Device Name
SteriKing LT-Blueline Pouches with Tyvek
Applicant
Wipak OY
Product Code
FRG · General Hospital
Decision Date
Dec 22, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6850
Device Class
Class 2
Indications for Use
The Steriking® LT-Blueline Pouches with Tyvek® are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERRAD® Sterilizer Systems. The device is intended to allow sterilization of enclosed devices and to maintain sterility of the enclosed devices until used up to 3 years post sterilization. The products are for single use only.
Device Story
Steriking® LT-Blueline Pouches with Tyvek® are single-use sterilization packaging for medical devices. Constructed from plastic film (BOPET/PE) and Tyvek® heat-sealed on three sides; the fourth side is sealed by the user. Used in healthcare facilities to enclose instruments prior to processing in STERRAD® hydrogen peroxide gas plasma sterilization systems. After sterilization, the pouch maintains the sterility of the enclosed devices until use. The device provides a microbial barrier and allows sterilant penetration. Healthcare workers use the pouches to organize and protect instruments during and after the sterilization cycle. Benefits include maintaining instrument sterility for up to 3 years post-sterilization.
Clinical Evidence
Bench testing only. No clinical data. Testing included: Biocompatibility (ISO 10993-5), Sterilant Penetration (AAMI TIR12:2010), Package Integrity (ASTM F1929-15), Microbial Barrier Properties (ANSI/AAMI ST79:2017), Seal Strength (ASTM F88), and Shelf-Life/Accelerated Aging (ASTM F1980-16). All tests passed, confirming 6-log reduction of Geobacillus stearothermophilus and maintenance of sterility.
Technological Characteristics
Materials: Tyvek® backing, BOPET/PE transparent film. Construction: Heat-sealed pouch. Sterilization: Hydrogen peroxide gas plasma (STERRAD®). Standards: ISO 11607, ASTM F1929, ASTM F88, ISO 10993-5, AAMI TIR12, ANSI/AAMI ST79. Connectivity: None. Software: None.
Indications for Use
Indicated for healthcare workers to enclose medical devices for sterilization in STERRAD® Sterilizer Systems (NX Standard, NX Advanced, 100NX Duo, 100NX Flex, 100NX Standard). Maintains sterility for up to 3 years. Single use only.
Regulatory Classification
Identification
A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.
Predicate Devices
Sterilization Pouch/Roll Made With Tyvek® (K180672)
Submission Summary (Full Text)
{0}------------------------------------------------
December 22, 2022
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
Wipak OY % Amanda Singleton Consultant Compliance Systems International 53 Assembly Drive, Unit 149 Mendon, New York 14506
Re: K221377
Trade/Device Name: SteriKing LT-Blueline Pouches with Tyvek Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: FRG Dated: November 29, 2022 Received: November 29, 2022
Dear Amanda Singleton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Anne D. Talley -S
Colin O'Neill, M.B.E. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
#### K221377
Device Name SteriKing® LT Blueline Pouch with Tyvek®
Indications for Use (Describe)
The SteriKing® LT Blueline with Tyvek® pouches are intended to provide health care workers with an effective method to enclose devices intended for sterilization in the STERRAD® Sterilizer Systems.
The device is intended to allow sterilization of enclosed devices and to maintain sterility of the enclosed devices until used up to 3 years post sterilization. The products are for single use only.
The SteriKing® LT Blueline with Tyvek® pouches are intended for use in the following STERRAD® Sterilization Cycles:
NX Standard NX Advanced 100NX Duo 100NX Flex 100NX Standard
Max weights:
NX Standard Cycle:
75 mm x 200 mm, 100 mm x 250 mm x 300 mm, 160 mm x 440 mm x 600 mm . 088 pounds 205 mm x 390 mm, 250 mm x 390 mm, 250 mm x 500 mm: 1.382 pounds
NX Advanced Cycle:
75 mm x 200 mm, 100 mm x 250 mm x 300 mm, 160 mm x 440 mm x 600 mm . 084 pounds 205 mm x 390 mm, 250 mm x 390 mm, 250 mm x 500 mm: 1.393 pounds
100NX Duo Cycle: 75 mm x 200 mm, 100 mm x 250 mm x 300 mm, 160 mm x 440 mm x 600 mm : 104 pounds 205 mm x 390 mm, 250 mm x 390 mm, 250 mm x 500 mm: 1.400 pounds
100NX Flex Cycle: 75 mm x 200 mm, 100 mm x 250 mm x 300 mm, 160 mm x 440 mm, 160 mm x 600 mm: .158 pounds 205 mm x 390 mm, 250 mm x 390 mm, 250 mm x 500 mm: 1.400 pounds
100NX Standard Cycle: 75 mm x 200 mm, 100 mm x 250 mm x 300 mm, 160 mm x 440 mm x 600 mm: .158 pounds 205 mm x 390 mm, 250 mm x 390 mm, 250 mm x 500 mm: 4.476 pounds
Max lumens:
1 mm min ID and 1 max lumen for all cycles
NX Standard Cycle: For pouch sizes 250x500mm, 205x390mm is 125 mm max length. Max length is 50 mm for all other pouch sizes.
{3}------------------------------------------------
NX Advanced Cycle: For pouch sizes 250x500mm, 205x390mm is 125 mm max length is 50 mm for all other pouch sizes.
100NX Duo Cycle: For pouch sizes 250x500mm, 250x390mm is 875 mm max length. Max length is 850 mm for all other pouch sizes.
100NX Flex Cycle: Max length is 850 mm for all pouch sizes.
100NX Standard Cycle: For pouch sizes 250x500mm, 205x390mm is 125 mm max length, Max length is 50 mm for all other pouch sizes.
Pouch Sizes: 250 mm x 500 mm 250 mm x 390 mm 205 mm x 390 mm 160 mm x 600 mm 160 mm x 440 mm 150 mm x 300 mm 100 mm x 250 mm 75 mm x 200 mm
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
|X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
## 510K Summary (in accordance with 21CFR807.92) -
| 510K Summary Elements per<br>21CFR807.92 | Summary |
|---------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter's name, address, telephone number, a contact person, and the date the summary was prepared | Wipak Oy<br>Wipaktie 2<br>Nastola Finland<br>Contact: Hanna Marttila Phone: +358 (0)40 124 0290<br>Date prepared: December 21, 2022<br>Consultant: Amanda Singleton, CSI LLC. Phone: 716.4407364 |
| Name of the device including the trade or proprietary name if applicable | Proprietary Name: Steriking® LT-Blueline Pouches with Tyvek® |
| Common or usual name | Common Name: Peel Pouch |
| Classification name | Classification Name: Sterilization wrap |
| Classification Product Code<br>Device Classification<br>Regulation Number | FRG<br>Class II<br>21 CFR 880.6850 |
| Identification of the legally marketed device to which the submitter claims equivalence<br>(Predicate device) | K180672, Sterilization Pouch/Roll Made With Tyvek® |
| Description of the device | Steriking® LT-Blueline Pouches with Tyvek® are intended to be used to contain medical devices to be terminally sterilized in the STERRAD® Sterilization System. The medical devices are inserted into the Pouch, sealed, and then sterilized in the STERRAD® Sterilization System. Sterilization Cycles are noted below. After completion of the sterilization process, the Pouch maintains sterility of the enclosed medical devices until the seal is opened. These pouches are made from a plastic film and Tyvek that is heat-sealed on three sides. The fourth side is left opened and will be heat-sealed when used.<br><br>Steriking® LT-Blueline Pouches with Tyvek® are intended to allow sterilization of enclosed devices and to maintain sterility of the enclosed devices until used up to 3 years post sterilization. The products are for single use only.<br><br>The pouches are constructed from Tyvek®/plastic films. The heat-sealed pouches are heat sealed prior to processing in the STERRAD® Sterilization Systems.<br><br>Sterilization Systems:<br><br>STERRAD® Cycles:<br>NX Standard<br>NX Advanced<br>100NX Duo<br>100NX Flex<br>100NX Standard |
{5}------------------------------------------------
| Intended use of the device<br>Technological<br>characteristics<br>compared to predicate<br>(as<br>follows) | The Steriking® LT-Blueline Pouches with Tyvek® are intended to provide<br>health care workers with an effective method to enclose devices intended for<br>sterilization in the STERRAD® Sterilizer Systems.<br><br>The device is intended to allow sterilization of enclosed devices and to<br>maintain sterility of the enclosed devices until used up to 3 years post<br>sterilization.<br><br>See below as follows: | | |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Discussion of nonclinical<br>tests submitted, referenced,<br>or utilized in the premarket<br>notification submission for a<br>determination of substantial<br>equivalence | Submission Device -<br>Steriking® LT-Blueline Pouches<br>with Tyvek® | SE Determination | Predicate Device -<br>K180672 -<br>Sterilization Pouch/Roll Made<br>With Tyvek |
| Indications for Use | The SteriKing® LT Blueline with<br>Tyvek® pouches are intended to<br>provide health care workers<br>with an effective method to<br>enclose devices intended for<br>sterilization in the STERRAD®<br>Sterilizer Systems.<br><br>The device is intended to allow<br>sterilization of enclosed devices<br>and to maintain sterility of the<br>enclosed devices until used up<br>to 3 years post sterilization. The<br>products are for single use only.<br><br>The SteriKing® LT Blueline with<br>Tyvek® pouches are intended for<br>use in the following STERRAD®<br>Sterilization Cycles:<br><br>NX Standard<br>NX Advanced<br>100NX Duo<br>100NX Flex<br>100NX Standard<br><br>Max weights:<br><b>NX Standard Cycle:</b><br>75 mm x 200 mm, 100 mm x 250<br>mm, 150 mm x 300 mm, 160<br>mm x 440 mm, 160 mm x 600<br>mm: .088 pounds | Similar | The Sterilization<br>Pouch/Roll Made with<br>Tyvek® are intended to<br>provide health care<br>workers with an effective<br>method to enclose<br>devices intended for<br>sterilization in the<br>STERRAD® 100S Sterilizer.<br><br>The device is intended to<br>allow sterilization of<br>enclosed devices and also<br>to maintain sterility of the<br>enclosed devices until<br>used up to 3 years post<br>sterilization. The pouches<br>and rolls are printed with<br>a chemical indicator bar<br>which is a process<br>indicator (ISO 11140-<br>1:2005) that changes from<br>red to blue (or lighter)<br>when exposed to<br>hydrogen peroxide vapor<br>during processing in the<br>STERRAD® Sterilizer. The<br>Sterilization Pouch/Roll<br>Made with Tyvek® is<br>offered in the follow 5<br>types: Self-sealing<br>sterilization pouches<br>Sterilization pouches, Flat<br>Sterilization pouches,<br>Gusseted Sterilization |
{6}------------------------------------------------
| 390 mm, 250 mm x 500 mm:<br>1.382 pounds | rolls, Flat Sterilization<br>rolls, Gusseted |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|
| NX Advanced Cycle:<br>75 mm x 200 mm, 100 mm x 250<br>mm, 150 mm x 300 mm, 160<br>mm x 440 mm, 160 mm x 600<br>mm: .084 pounds | |
| 205 mm x 390 mm, 250 mm x<br>390 mm, 250 mm x 500 mm:<br>1.393 pounds | |
| 100NX Duo Cycle:<br>75 mm x 200 mm, 100 mm x 250<br>mm, 150 mm x 300 mm, 160<br>mm x 440 mm, 160 mm x 600<br>mm: .104 pounds | |
| 205 mm x 390 mm, 250 mm x<br>390 mm, 250 mm x 500 mm:<br>1.400 pounds | |
| 100NX Flex Cycle:<br>75 mm x 200 mm, 100 mm x 250<br>mm, 150 mm x 300 mm, 160<br>mm x 440 mm, 160 mm x 600<br>mm: .158 pounds | |
| 205 mm x 390 mm, 250 mm x<br>390 mm, 250 mm x 500 mm:<br>1.400 pounds | |
| 100NX Standard Cycle:<br>75 mm x 200 mm, 100 mm x 250<br>mm, 150 mm x 300 mm, 160<br>mm x 440 mm, 160 mm x 600<br>mm: .158 pounds | |
| 205 mm x 390 mm, 250 mm x<br>390 mm, 250 mm x 500 mm:<br>4.476 pounds | |
| Max lumens:<br>1 mm min ID and Max 1 lumen.<br>for all cycles | |
| NX Standard Cycle: For pouch<br>sizes 250x500mm, 250x390mm,<br>205x390mm is 125 mm max<br>length. Max length is 50 mm for<br>all other pouch sizes. | |
{7}------------------------------------------------
| | NX Advanced Cycle: For pouch<br>sizes 250x500mm, 250x390mm,<br>205x390mm is 125 mm max<br>length. Max length is 50 mm for<br>all other pouch sizes.<br>100NX Duo Cycle: For pouch<br>sizes 250x500mm, 250x390mm,<br>205x390mm is 875 mm max<br>length. Max length is 850 mm<br>for all other pouch sizes.<br>100NX Flex Cycle: Max length is<br>850 mm for all pouch sizes.<br>100NX Standard Cycle: For<br>pouch sizes 250x500mm,<br>250x390mm, 205x390mm is 125<br>mm max length. Max length is<br>50 mm for all other pouch sizes.<br>Pouch Sizes:<br>250 mm x 500 mm<br>250 mm x 390 mm<br>205 mm x 390 mm<br>160 mm x 600 mm<br>160 mm x 440 mm<br>150 mm x 300 mm<br>100 mm x 250 mm<br>75 mm x 200 mm | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Design | These pouches are made from a<br>plastic film and Tyvek that is<br>heat sealed on three sides. The<br>fourth side is left opened and<br>will be heat-sealed when used. | Same | These pouches are made<br>from a plastic film and<br>Tyvek that is heat sealed<br>on three sides. The fourth<br>side is left opened and will<br>be heat-sealed when<br>used. |
| Backing Material | Tyvek® | Same | Tyvek® |
| Transparent Film | BOPET/PE | Similar | PE |
| Sterilization Processes | H2O2 Sterilization Process as per:<br>STERRAD® | Similar | STERRAD® 100S |
| | NX Standard | | |
| | NX Advanced | | |
| | 100NX Duo | | |
| | 100NX Flex | | |
| | 100NX Standard | | |
| Sterilant | Achieved a 10-6 Sterility Assurance | Similar | Showed a 6 log reduction of |
| Validation | Level (SAL) of Geobacillus | | Geobacillus |
| | stearothermophilus. | | stearothermophilus |
| AAMI TIR12:2010<br>ANSI/AAMI ST79<br>ANSI/AAMI/ISO 14937:2009<br>ANSI/AAMI/ISO 17664:2017 | NX Standard<br>NX Advanced<br>100NX Duo<br>100NX Flex<br>100NX Standard | | |
| Dye Migration<br><br>ISO 11607<br>ASTM F1929<br>AAMI TIR12:2010 | No leaks detected after dye<br>migration | Same | No leaks detected after dye<br>migration |
| Seal Strength<br><br>ASTM F88<br>AAMI TIR12:2010<br>AAMI/ISO TIR16775:2014<br>ISO 11607 | > 1.5N/15mm per ASTM F88 | Same | > 1.5N/15mm per ASTM F88 |
| Microbial Barrier<br>Properties<br><br>ISO 11607-1<br>AAMI TIR12:2010<br>ANSI/AAMI ST79:2017 | Pass<br><br>Each pouch met the sterility<br>maintenance requirement as there<br>was no growth in any of the culture<br>tubes containing the stainless-steel<br>coupons at the end if the incubation<br>period. | Similar | Pass |
| Biocompatibility | Pass<br>ISO 10993-5 | Similar | Pass<br>ISO 10993-12 |
| Shelf-Life Pre-Sterilization<br><br>ASTM F1929-15<br>ASTM F88 | 5 years | Similar | 3 years |
| Maintenance of Sterility –<br>Accelerated Aging<br><br>ANSI/AAMI<br>ST8:2013/(R)2018<br>AAMI TIR12:2010<br>ANSI/AAMI ST77:2017<br>ANSI/AAMI ST79:2017<br>ISO 11607-1<br>ASTM F1980<br>ASTM F1929-15<br>ASTM F88 | 3 years | Same | 3 years |
{8}------------------------------------------------
{9}------------------------------------------------
### K221377 Page 6 of 8
| Dimensions and max load weight | | | | | | Similar | Not listed |
|--------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|-----------------------|------------------------|----------------------------|-----------|------------|
| Pouch size | NX<br>Advanced<br>Cycle | NX<br>Standard<br>Cycle | 100NX<br>Duo<br>Cycle | 100NX<br>Flex<br>Cycle | 100NX<br>Standard<br>Cycle | | |
| | weight (pounds) | | | | | | |
| 250 x 500 mm | 1.393 | 1.382 | 1.400 | 1.400 | 4.476 | | |
| 250 x 390 mm | 1.393 | 1.382 | 1.400 | 1.400 | 4.476 | | |
| 205 x 390 mm | 1.393 | 1.382 | 1.400 | 1.400 | 4.476 | | |
| 160 x 600 mm | 0.084 | 0.088 | 0.104 | 0.158 | 0.158 | | |
| 160 x 440 mm | 0.084 | 0.088 | 0.104 | 0.158 | 0.158 | | |
| 150 x 300 mm | 0.084 | 0.088 | 0.104 | 0.158 | 0.158 | | |
| 100 x 250 mm | 0.084 | 0.088 | 0.104 | 0.158 | 0.158 | | |
| 75 x 200 mm | 0.084 | 0.088 | 0.104 | 0.158 | 0.158 | | |
| Max Lumen Dimensions | 1 mm min ID Lumen and Max 1<br>lumen for all cycles<br><br>For NX Standard Cycle: For<br>pouch sizes 250x500mm,<br>250x390mm, 205x390mm is 125<br>mm max length. Max length is<br>50 mm for all other pouch sizes.<br><br>NX Advanced Cycle: For pouch<br>sizes 250x500mm, 250x390mm,<br>205x390mm is 125 mm max<br>length. Max length is 50 mm for<br>all other pouch sizes.<br><br>100NX Duo Cycle: For pouch<br>sizes 250x500mm, 250x390mm,<br>205x390mm is 875 mm max<br>length. Max length is 850 mm<br>for all other pouch sizes.<br><br>100NX Flex Cycle: Max length is<br>850 mm for all pouch sizes.<br><br>100NX Standard Cycle: For<br>pouch sizes 250x500mm,<br>250x390mm, 205x390mm is 125<br>mm max length. Max length is<br>50 mm for all other pouch sizes. | | | | | Different | Not listed |
#### Summary of Non-Clinical Testing
The Steriking® LT-Blueline Pouches with Tyvek® (submission device) has the identical intended use and indications for use as
the predicate devices. The subject device was compared to the predicate device by testing the Sterilant Penetration, Biocompatibility, Package Integrity, Material Compatibility, and Sterility Maintenance. The results of the Steriking® LT-Blueline Pouches with Tyvek® validation studies demonstrate that the sterilization pouches perform as intended. The results are summarized as follows:
- The Sterilant Penetration testing performed as described in AAMI / ANSI / ISO 14937:2009, "Sterilization of Health Care Products – General Requirements for Characterization of a Sterilizing Agent and the Development, Validation, and Routine Control of a Sterilization Process for Medical Devices". The results confirm that the 6 of 8
{10}------------------------------------------------
sterilant is able to penetrate Steriking® LT-Blueline Pouches with Tyvek® and sustain with the medical instrument inside the subject device.
- The Biocompatibility testing performed as described in ISO 10993-5, Biological evaluation of medical devices — Part 5: Tests for in vitro cytotoxicity. The testing results demonstrate the Steriking® LT-Blueline Pouches with Tyvek® did not elicit any cytotoxic effect.
- The Package Integrity, Material Compatibility, Sterility Maintenance, and Microbial Aerosol Challenge testing was performed as described in ISO 11607. The use of the aerosol challenge test to analyze the microbial barrier properties of the samples was considered to be rigorous. By exposing the sealed test samples' exterior with aerosolized spores, the permeability of the samples to microorganisms was challenged. A sample which demonstrates that all items remain sterile following this test is considered safe and effective at maintaining package integrity. The use of stainless-steel coupons was based on several considerations. First, they offer a convenient way to identify specific pieces to be sterility tested and to confirm that samples were indeed in the specified locations. Second, coupons are relatively easy to manipulate, and reduces the likelihood of the introduction of adventitious growth into the sterility test results. Finally, the coupons are comprised of metals consistent with metals used to manufacture medical instruments.
- The Microbial Aerosol Challenge study is a whole package test which challenges all aspects of the pouch in its final state. Conversely the ISO 11607 tests (Dye Migration and Seal Peel) test only one component of the pouch at time as it is not a whole package test. The Microbial Aerosol Challenge is a rigorous whole package test, and therefore the results are more comprehensive and supersede the results from Dye Migration and Seal Peel tests.
| Test Methodology | Purpose | Acceptance Criteria | Results |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ISO 10993-5 Biological Evaluation of medical<br>devices - Part 5: Tests for in vitro cytotoxicity | Biocompatibility | no reduction of cell growth<br>or cell lysis was observed | Pass |
| AAMI TIR12:2010 - Designing, testing, and<br>labeling reusable medical devices for<br>reprocessing in health care facilities | Sterilant Penetration | Demonstrating that a<br>minimum of 1.0x10^6<br>Geobacilus<br>stearothermophilus spores<br>were killed in a half-cycle (6-log reduction) | Pass |
| | | | Negative for growth |
| ASTM F1929-15: Standard test method for<br>detecting seal leaks in porous medical<br>packaging by dye migration | Package Integrity | Dye Penetration: none | Pass |
| ANSI/AAMI ST79:2017 - Comprehensive guide<br>to steam sterilization and sterility assurance<br>in health care facilities<br><br>AAMI TIR12:2010 - Designing, testing, and<br>labeling reusable medical devices for<br>reprocessing in health care facilities | Microbial Barrier Properties | A sample which<br>demonstrates that all items<br>remain sterile following<br>exposure | Pass<br>Each pouch met the<br>sterility maintenance<br>requirement as there<br>was no growth in any<br>of the culture tubes<br>containing the<br>stainless-steel<br>coupons at the end if<br>the incubation period. |
| ASTM F1929-15: Standard test method for<br>detecting seal leaks in porous medical<br>packaging by dye migration | Shelf Life Pre Sterilization | Dye Penetration: none<br>Seal peel: =/>1.5N/15mm | Pass |
{11}------------------------------------------------
| ASTM F88: Standard Test Method for Seal<br>Strength of Flexible Barrier Materials | | | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------|---------------------------------------------------------------------------------------------|------|
| ASTM F1980 – 16 Standard Guide for<br>Accelerated Aging of Sterile Barrier Systems<br>for Medical Devices | Shelf Life Post Sterilization | Dye Penetration: none<br>Seal peel: =/>1.5N/15mm | Pass |
| ASTM F1929-15: Standard test method for<br>detecting seal leaks in porous medical<br>packaging by dye migration | | | |
| ASTM F88: Standard Test Method for Seal<br>Strength of Flexible Barrier Materials | | | |
| 1. ASTM F1608, ASTM F2638 - Microbial<br>Barrier<br>2. ASTM D3776 - Basis Weight of Tyvek<br>3. ASTM D2724 - Delamination of Tyvek<br>4. TAPPI T460 - Gurly Hill porosity of Tyvek<br>5. ISO 1924-2 - Tensile Strength, MD of Tyvek<br>6. ISO 1924-2 - Elongation, MD of Tyvek<br>7. ASTM D1424 - Elmendorf Tear, MD of<br>Tyvek<br>8. ISO 534 - Thickness of Tyvek<br>9. ISO 5636-3 - Bendtsen air permeability of<br>Tyvek<br>10. TAPPI T523 - Moisture vapor transmission<br>Rate of Tyvek<br>11. ISO 2758 - Mullen, burst of Tyvek<br>12. ASTM D3420 - Spencer puncture of Tyvek | Material Compatibility | Meets specifications per<br>Technical Data Sheets.<br>Included in Bench Testing<br>Summary. | Pass |
Conclusion:
The conclusions drawn from the non-clinical tests demonstrate that the proposed device is as safe, as effective, and performs as well as or better than the predicate device K180672.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.