K231820 · Biocorp Production · QOG · Jan 12, 2024 · General Hospital
Device Facts
Record ID
K231820
Device Name
SoloSmart Injection Pen Adapter (SoloSmart®)
Applicant
Biocorp Production
Product Code
QOG · General Hospital
Decision Date
Jan 12, 2024
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5860
Device Class
Class 2
Indications for Use
The SoloSmart Injection Pen Adapter is indicated for the capture and wireless transmission of dosing information from compatible reusable and disposable pen injectors.
Device Story
SoloSmart is an injection pen adapter that attaches to compatible insulin pens; it captures dosing information by sensing the rotation of the pen's dosing mechanism during administration. The device uses Bluetooth Low Energy (BLE) to transmit the dialed dose, date, and time of injection to a mobile application. It is intended for patient use in the same environments as their injection pens. The device provides visual feedback via LED. By automating the logging of injection data, it assists patients in tracking their insulin therapy, potentially improving adherence and clinical decision-making by providing accurate records to healthcare providers.
Clinical Evidence
Bench testing only. Testing included dose accuracy (ISO 11608-1), recording/transmission accuracy, prime differentiation, incomplete injection identification, and physical interaction forces. Biocompatibility testing followed ISO 10993-1. Electrical safety and EMC testing followed IEC 60601-1 and collateral standards. Human factors validation confirmed usability of the form factor and instructions.
Technological Characteristics
Cap-style adapter (21.4mm diameter, 23.3mm height) for pen injectors. Sensing principle: rotation of dosing mechanism. Connectivity: Bluetooth Low Energy (BLE). Power: Non-replaceable rechargeable battery (USB charged). Standards: ISO 11608-1 (dose accuracy), ISO 10993-1 (biocompatibility), IEC 60601-1 (electrical safety), IEC 60601-1-11 (home healthcare).
Indications for Use
Indicated for diabetes patients using compatible reusable and disposable pen injectors (Sanofi SoloStar) to capture and wirelessly transmit dosing information.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right of this symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
January 12, 2024
Biocorp Production Alexia Garin Vice President & Regulatory Affairs Zi Lavaur -La Béchade Issoire, 63500, France
Re: K231820
Trade/Device Name: SoloSmart Injection Pen Adapter (SoloSmart®) Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: QOG Dated: December 20, 2023 Received: December 20, 2023
Dear Alexia Garin:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Shruti N. Mistry -S
Shruti Mistry, MS Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of Gastrorenal, ObGyn, General Hospital, and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231820
Device Name
SoloSmart Injection Pen Adapter (SoloSmart®)
### Indications for Use (Describe)
The SoloSmart Injection Pen Adapter is indicated for the capture and wireless transmission of dosing information from compatible reusable and disposable pen injectors.
The following Injection pens are compatible:
| DEVICE MODEL NAME | INSULIN BRAND NAME | MOLECULE NAME | MOLECULE CONCENTRATION |
|----------------------------------------------------------------|--------------------|------------------------------|------------------------|
| SoloSmart<br>designed for<br>Solostar® SANOFI<br>Injection pen | Lantus | GLARGINE | 100 IU/mL |
| | Toujeo | | 300 IU/mL |
| | Admelog | LISPRO | 100 IU/mL |
| | Apidra | GLULISINE | 100 IU/mL |
| | Soliqua 100/33 | GLARGINE AND<br>LIXISENATIDE | 100 IU/mL<br>+33mcg/mL |
Type of Use *(Select one or both, as applicable)*
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------|-----------------------------------------------------------|
|------------------------------------------------------------|-----------------------------------------------------------|
**CONTINUE ON A SEPARATE PAGE IF NEEDED.**
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary)
- 1. Prepared By: Alexia Garin
2.
- Date Prepared: December 20, 2023
- 3. Trade/Proprietary Name: SoloSmart Injection Pen Adapter (SoloSmart®)
- 4. Common/Usual Name: Iniection Data Capture Device Piston Syringe 5. Classification Name:
- 21 CFR 880.5860
- 6. Predicate Device: K222689, Biocorp Production, Mallya Injection Pen Adapter
Product Code: QOG
- 7. Biocorp Production Applicant: ZI Lavaur La Béchade 63500 Issoire - France
- Substantial Equivalence: The SoloSmart® device is substantially equivalent to the Mallya 8. (K222689). The device has the same indication for use, intended use and same fundamental scientific technology.
- 9. Technological Characteristics: The Solosmart® Injection Pen Adapter has the same technological characteristics as the predicate device to capture and transmit dosing information. The information is captured through the rotation of the dosing button of the pen into a value indicating the dose increment injected. Both the SoloSmart® and the predicate device use same Low Energy Bluetooth (BLE) communication technology and same protocol to pair and transmit the information to a mobile device with a compatible app.
- 10. Physical description: The device is made of a cap to be assembled onto a pen injector by covering the injection pen button. The overall device size is 21.4mm diameter & 23.3mm height. A USB cable, necessary to charge the SoloSmart® device is also provided in the package.
- 11. Indications for Use: The SoloSmart Injection Pen Adapter is indicated for the capture and wireless transmission of dosing information from compatible reusable and disposable pen injectors. The following Injection pens are compatible:
| DEVICE MODEL | INSULIN BRAND<br>NAME | MOLECULE NAME | MOLECULE<br>CONCENTRATION |
|--------------------------------------------------------------------|-----------------------|------------------------------|------------------------------|
| SoloSmart®<br>designed for<br>SoloStar®<br>SANOFI<br>injection pen | Lantus | GLARGINE | 100 IU/mL |
| | Toujeo | GLARGINE | 300 IU/mL |
| | Admelog | LISPRO | 100 IU/mL |
| | Apidra | GLULISINE | 100 IU/mL |
| | Soliqua 100/33 | GLARGINE AND<br>LIXISENATIDE | GLARGINE AND<br>LIXISENATIDE |
{4}------------------------------------------------
#### 12. Substantial Equivalence discussion:
| Attribute | Subject Device | Predicate | Discussion/Comments |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| Device Name | SoloSmart® | Mallya | - |
| | | Data Capture and Transmission Technology | |
| 510(k) Number | K231820 | K222689 | - |
| Device Classification | Class 2 - QOG | Class 2 - QOG | |
| Indication for Use | Indicated for the<br>capture and wireless<br>transmission of dosing<br>information from<br>compatible reusable<br>and disposable pen<br>injectors. | Indicated for the<br>capture and wireless<br>transmission of dosing<br>information from<br>compatible reusable<br>and disposable pen<br>injectors. | |
| Single patient Use | Yes | Yes | |
| Reusable Device | Yes | Yes | |
| Intended Use | Intended to be used<br>by patients in the<br>same use<br>environment as their<br>compatible injection<br>pen. | Intended to be used<br>by patients in the<br>same use<br>environment as their<br>compatible injection<br>pen. | same |
| Prescription use | No | No | |
| User Group | User of Compatible<br>Injection Pens;<br>includes diabetes<br>patients | User of Compatible<br>Injection Pens;<br>includes diabetes<br>patients | |
| User Feedback | Electronic - LED<br>(light) | Electronic - LED<br>(light) and Audible<br>(Beeps) | Both provide visible<br>feedback (light).<br>Removal of audible<br>feedback was<br>confirmed as<br>unnecessary in a HF<br>Validation study |
| Wireless<br>Connectivity | Bluetooth Low Energy<br>(BLE) | Bluetooth Low Energy<br>(BLE) | Same connectivity BLE<br>protocol |
| Control or impact<br>Drug delivery | No | No | |
| Fluid Pathway<br>Contact | None | None | same |
| Software controlled | Yes | Yes | same |
| Dose Recorded | Calculated based on<br>dose set | Calculated based on<br>dose set | Change from two-part<br>to one part device |
| Information<br>Transmitted | Dialed dose, and time<br>and date of injection | Dialed dose, and time<br>and date of injection | necessitated and<br>enabled changes to the |
| Mechanism for<br>Recording dose<br>dialed | Sense dose dialed<br>through rotation of<br>dosing mechanism<br>during dose<br>administration | Sense dose dialed<br>through rotation of<br>dosing mechanism<br>during dose setting | hardware and software<br>to capture, record and<br>transmit the dose<br>delivered, rather than |
| Differentiates Prime<br>vs. Dose | Yes | Yes | only the dose that was<br>set (dialled). |
| Battery | Non-Replaceable;<br>Rechargeable | Non-Replaceable;<br>Rechargeable | Same type of battery<br>with same design input<br>specification regarding<br>battery |
| Lifetime | Up to 3 years of use<br>from manufacturing<br>date | 2 years of use | Extended lifetime |
{5}------------------------------------------------
#### 13. Performance Data
Data on the following testing were generated verifying the design of the device:
- . Bench testing on performance, using ISO 11608-1 as a guide.
- Dose accuracy of pen is not affected by SoloSmart® .
- . Recording and transmission Accuracy
- Dose Prime Differentiation .
- . Complete Dose Notification
- . Incomplete Injection Identification
- Physical Interaction with pen, including force to mount, maintain and remove . SoloSmart®
- Biocompatibility - ISO 10993-1 and FDA quidance; permanent contact with intact skin
- Lifetime The product met test criteria to verify it will operate/function for 2 years (Use . Lifetime) following storage for up to two years (Shelf Life).
- Electrical safety, EMC and Radiocommunication - IEC 60601-1 and appropriate collateral requirements from IEC 60601-2 and IEC 60601-1-11
- Cybersecurity Testing and Software Verification and Validation per FDA Guidelines
#### 14. Human factor validation
Biocorp performed a HF Validation study to confirm the usability of SoloSmart device form factor and associated instructions for use per FDA guidance.
#### 15. Conclusion
Biocorp as concluded that the evidences documented during design control activities supports that the SoloSmart is substantially equivalent to the predicate for specific indications and technological characteristics of the predicate device with regards to the data recording and transmission.
- 16. Contact person: Alexia Garin, pms@biocorp.fr +33 473 557050
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.