Mallya Injection Pen Adapter (Mallya® for Solostar®)
K222689 · Biocorp Production · QOG · Dec 5, 2022 · General Hospital
Device Facts
Record ID
K222689
Device Name
Mallya Injection Pen Adapter (Mallya® for Solostar®)
Applicant
Biocorp Production
Product Code
QOG · General Hospital
Decision Date
Dec 5, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5860
Device Class
Class 2
Indications for Use
The Mallya Injection Pen Adapter is indicated for the capture and wireless transmission of dosing information from compatible reusable and disposable pen injectors.
Device Story
Mallya is an add-on accessory for SoloStar insulin pen injectors; clips onto pen body and button. Device senses rotation of pen's dosing knob to calculate dose increments; detects proximity of knob to confirm dose completion. Captures dialed dose, time, and date of injection. Uses Bluetooth Low Energy (BLE) to transmit data to a mobile device/app. Used by patients in home or clinical settings. Provides LED and audible feedback to user. Does not impact drug delivery mechanism. Benefits patient by automating logging of insulin injections, reducing manual entry errors, and improving adherence tracking.
Clinical Evidence
Bench testing only. No clinical data. Testing included dose accuracy, recording/transmission accuracy, prime differentiation, complete dose notification, incomplete injection identification, physical interaction forces, biocompatibility (ISO 10993-1), electromagnetic compatibility (IEC 60601-1), cybersecurity, and human factors validation.
Technological Characteristics
Two-part plastic assembly (base and button cap). Sensing via rotational tracking of pen dose knob. Connectivity: Bluetooth Low Energy (BLE). Power: Rechargeable battery (non-replaceable). Dimensions: Base ~6x3x2 cm, Button ~2.5x2 cm. Biocompatibility: ISO 10993-1 (permanent contact with intact skin). Software: Embedded firmware for data processing and transmission.
Indications for Use
Indicated for diabetes patients using compatible SoloStar injection pens (Lantus, Toujeo, Admelog, Apidra, Soliqua 100/33) to capture and wirelessly transmit insulin dosing information.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
December 5, 2022
Biocorp Production % Lee Leichter President P/L Biomedical 5800 West State Road 80 - Lot 219 Labelle, Florida 33935
Re: K222689
Trade/Device Name: Mallya Injection Pen Adapter (Mallya® for Solostar®) Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: Class II Product Code: QOG Dated: August 29, 2022 Received: September 6, 2022
Dear Lee Leichter:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Alan M.
Stevens -S3
CAPT Alan M. Stevens Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
K222689
Device Name
Mallya Injection Pen Adapter
### Indications for Use (Describe)
The Mallya Injection Pen Adapter is indicated for the capture and wireless transmission of dosing information from compatible reusable and disposable pen injectors. The following Injection pens are compatible.
| MALLYA MODEL | INSULIN BRAND NAME | MOLECULE NAME | MOLECULE CONCENTRATION |
|----------------------------------------------------------|--------------------|------------------------------|-------------------------|
| Mallya designed for<br>Solostar® SANOFI<br>injection pen | Lantus | GLARGINE | 100 IU/mL |
| | Toujeo | GLARGINE | 300 IU/mL |
| | Admelog | LISPRO | 100 IU/mL |
| | Apidra | GLULISINE | 100 IU/mL |
| | Soliqua 100/33 | GLARGINE AND<br>LIXISENATIDE | 100 IU/mL<br>+33 mcg/mL |
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
BIOCORP 510 (k)
Mallya Injection Pen Adapter
{3}------------------------------------------------
## Summary of Safety and Effectiveness (510(k) Summary)
- 1. Prepared Bv: Alexia Garin 2. Date Prepared: Dec 5, 2022 Mallya Injection Pen Adapter (Mallya® for Solostar®) 3. Trade/Proprietary Name: 4. Common/Usual Name: Injection Data Capture Device 5. Classification Name: Piston Syringe 21 CFR 880.5860 Product Code: QOG 6. Predicate Device: K160629, Companion Medical, InPen 7. Manufacturing Site: Biocorp Production
France
- 8. Registration Number:
- 9. Substantial Equivalence: The Mallya device is substantially equivalent to the InPen (K160629). The InPen is classified as a Piston Syringe and indicated for the delivery Humalog and Novolog. Although the InPen is a full reusable Injection Pen, it contains electronics, a non-replaceable battery and a Bluetooth radio that facilitates capture and transmission of dosing information to a standalone software device. This function is the functionality added by the Mallya as an accessory to SoloStar® injection pens that are not so equipped.
ZI Lavaur La Béchade 63500 Issoire -
- 10. Technological Characteristics: The Mallya Injection Pen Adapter has the same technological characteristics as the predicate device to capture and transmit dosing information. The information is captured through a translation of the rotation of the dosing knob of the pen into a value indicating the dose increment dialed. The Mallya senses the proximity of the dose knob at the end of the delivery to indicate that the dose has been completed. Both the Mallya and the predicate device use Low Energy Bluetooth (BLE) communication technology and protocol to pair and transmit the information to a mobile device with a compatible app.
- 11. Physical description: The device is made of two parts to be assembled onto a pen injector. The Mallya base part (about 6 x 3 x 2 cm) which clips around the body of the injection pen and the Mallya button (about 2.5 x Ø2 cm) that slips over the injection pen button. A USB cable, necessary to charge the MALLYA device, and a reset key are also provided in the package.
- 12. Mallya device is compatible with the SoloStar® injection pens available on the US market such as the insulin brand names, molecules and associated concentrations listed in the table below:
| MALLYA MODEL | INSULIN BRAND NAME | MOLECULE NAME | MOLECULE<br>CONCENTRATION |
|----------------------------------------------------------|--------------------|------------------------------|---------------------------|
| | Lantus | | 100 IU/mL |
| Mallya designed for<br>SoloStar® SANOFI<br>injection pen | Toujeo | GLARGINE | 300 IU/mL |
| | Admelog | LISPRO | 100 IU/mL |
| | Apidra | GLULISINE | 100 IU/mL |
| | Soliqua 100/33 | GLARGINE AND<br>LIXISENATIDE | 100 IU/mL<br>+33 mcg/mL |
{4}------------------------------------------------
# COMPARISONS TO PREDICATE DEVICE
| Attribute | Subject Device | Predicate | Discussion/Comments |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device<br>Name | Mallya | InPen<br>(data capture and<br>transmission technology<br>only) | The InPen functions as a pen injector in<br>addition to capturing and<br>communicating information. The Mallya<br>works only as a data capture and<br>transmission device accessory to<br>approved pen injectors |
| | Information Capture and Transmission Technology | | |
| 510(k)<br>Number | N/A | K160629 | N/A |
| Device<br>Classification | Class 2 - QOG | Class 2 - FMF | The InPen is classified as a Piston<br>Syringe and not the new Procode of<br>Injection Data Capture Device. Both are<br>within the same classification |
| Indication for<br>Use | Indicated for the capture and<br>wireless transmission of<br>dosing information from<br>compatible reusable and<br>disposable pen injectors. | The InPen is a home-use<br>reusable pen injector for<br>single-patient use | The InPen is indicated for use as an<br>injection pen and dose calculator, but<br>does not have a specific indication for<br>use related to information capture and<br>transmission |
| Single patient<br>Use | Yes | Yes | Same |
| Intended Use | Intended to be used by<br>patients in the same use<br>environment as their<br>compatible injection pen. | Intended to be used by<br>people with diabetes age 12<br>and older for the self-<br>injection of a desired dose<br>of insulin. | The InPen is intended to deliver insulin<br>whereas the Mallya will work with any<br>compatible injection pen for which it is<br>labeled. |
| Prescription<br>use | No | Yes | injection pens are for use by<br>prescription only, but a separate<br>prescription is not required for the<br>Mallya |
| User Group | User of Compatible Injection<br>Pens; includes diabetes<br>patients | Diabetes patients treated<br>with multiple daily insulin<br>injection (MDI) therapy | Same |
| User<br>Feedback | Electronic - LED (light) and<br>Audible (Beeps) | Mechanical - Audible and<br>tactile clicks per increment | Feedback to user is for injection pen<br>function |
| Wireless<br>Connectivity | Bluetooth Low Energy (BLE) | Bluetooth Low Energy<br>(BLE) | Same |
| Control or<br>impact<br>Drug<br>delivery | No | No | Same, the information recording and<br>transmission function does not impact<br>drug delivery |
| Software | Yes | Yes | The recording and transmission of<br>information is controlled by software |
| Fluid Pathway<br>Contact | None | None | Same |
| Dose<br>Recorded | Calculated based on dose<br>set | Calculated based on dose<br>set | Same |
| Information<br>Transmitted | Dialed dose, and time and<br>date of injection | Dialed dose, and time and<br>date of injection | Same |
| Mechanism for<br>Recording | Sense dose dialed through<br>rotation of dosing mechanism | Sense dose dialed through<br>rotation of dosing mechanism | Although the technology is different<br>(magnet vs. electronic rotational) |
| Attribute | Subject Device | Predicate | Discussion/Comments |
| dose dialed | during dose setting | during dose setting | encoder), both use the rotation of the<br>dose knob during dose setting to<br>determine the dose that has been<br>dialed. Accuracy of Mallya dose<br>recording has been assessed and<br>verified during Design Verification., |
| Differentiates<br>Prime vs. Dose | Yes | Yes | Same |
| Battery | Non-Replaceable;<br>Rechargeable | Non-Replaceable; Non-<br>Rechargeable | The InPen cannot be recharged. |
| Lifetime | 2 years of use | 1 year of use | The Mallya can be used for up to two<br>years |
{5}------------------------------------------------
#### 14. Performance Data
Data on the following testing were generated verifying the design of the device:
- Electromagnetic Compatibility and Electrical Safety IEC 60601-1 and appropriate ● collateral requirements from IEC 60601-2 and IEC 60601-1-11
- Bench testing on performance, using ISO 11608-1 as a quide. ●
- Dose accuracy of pen is not affected by Mallya .
- . Recording and transmission Accuracy
- . Dose Prime Differentiation
- . Complete Dose Notification
- . Incomplete Iniection Identification
- Physical Interaction with pen, including force to mount, maintain and remove Mallya
- Biocompatibility ISO 10993-1 and FDA guidance; permanent contact with intact skin ●
- Lifetime The product met test criteria to verify it will operate/function for 2 years (Use ● Lifetime) following storage for up to two years (Shelf Life).
- Cybersecurity Testing and Software Verification and Validation per FDA Guidelines ●
- Human Factors HF Validation per FDA guidance ●
#### 15. Conclusion
The data presented supports that the Mallya is substantially equivalent to the predicate for specific indications and technological characteristics of the predicate device with regards to the data recording and transmission.
### 16. Contact person:
Alexia Garin, pms@biocorp.fr +33 473 557050
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.