Static deep learning artificial intelligence model
Control point error <= 3 mm
75.3% success rate
Training, validation, and testing datasets (independent of each other)
>1 (qualified DASI Dimensions operators)
AI testing dataset
>1 (qualified DASI Dimensions operators)
Cardiovascular structure measurements
Automated algorithm involving segmentation and centerline detection
Primary measurements <= 15% error in >= 95% of cases; Secondary measurements <= 20% error in >= 95% of cases
Primary measurements <= 15% error in >= 95% of cases; Secondary measurements <= 20% error in >= 95% of cases
—
—
Automatic Measurements Validation Study
>1 (clinicians)
Aortic root and valve anatomical measurements
Semi-automatic AI enhanced workflow
Annulus area/perimeter error <= 10% in >= 95% of cases; Sinus of valsalva/sinotubular junction/ascending aorta error <= 15% in >= 95% of cases; Aortic valve angle error <= 25% in >= 95% of cases
Annulus area/perimeter mean error 0.93% and -1.02%; Sinus of valsalva/sinotubular junction/ascending aorta mean difference 4.77%, 4.13%, 3.29%, 1.25%, 2.20%, 0.12%; Aortic valve angle mean difference 2.66%
—
—
Performance Validation Study (n=40)
>1 (qualified clinician truthers)
Indications for Use
DASI Dimensions is a standalone, non-invasive, clinical decision support software solution that is intended for use by cardiologists and radiologists in context of the aortic stenosis population. DASI Dimensions provides the end-user with pre-defined images and measurements (diameters, lengths, angles, areas, and perimeters) of cardiovascular structures. The clinician receiving the results the responsibility for interpreting and validating all information and making all patient treatment decisions. DASI Dimensions is not intended to replace the clinician's decision or device's instructions for use. DASI Dimensions is prescription use only.
Device Story
DASI Dimensions is a standalone, off-site clinical decision support software for pre-TAVR planning. Input: patient multiphase DICOM CTA chest images uploaded via web portal. Operation: proprietary algorithms use a static deep learning AI model to detect aortic root control points; software calculates anatomical measurements (diameters, lengths, angles, areas, perimeters). Workflow: trained DASI Simulations operators perform quality checks on AI-generated points and measurements before releasing the report to the user via portal. Output: report containing images and cardiac structure measurements. Usage: used by cardiologists/radiologists to assist in pre-procedural planning; clinicians remain responsible for interpreting results and making treatment decisions. Benefit: provides standardized, automated anatomical quantification to support clinical decision-making for TAVR candidates.
Clinical Evidence
Bench testing only; no prospective clinical trials. Validation included: AI control point detection (75.3% success rate); automatic measurement accuracy (primary measurements ≤15% error in ≥95% cases); performance validation (n=40, mean percentage error in primary outputs 0.93% to -1.02%); operator variability (ICC 0.96); and human factors/usability testing (100% success rate in critical tasks).
Technological Characteristics
Standalone software; cloud-hosted (AWS). Inputs: multiphase DICOM CTA. Processing: static deep learning AI model for control point detection; rule-based algorithms for anatomical measurements. Connectivity: web portal for data upload and report retrieval. Cybersecurity: implemented risk control measures for known vulnerabilities. Software: non-invasive, clinical decision support.
Indications for Use
Indicated for cardiologists and radiologists for pre-procedural planning of TAVR in patients with aortic stenosis.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in blue text.
DASI Simulations Lakshmi Dasi, Chief Technology Officer 5115 Parkcenter Ave Suite 205 Dublin, Ohio 43017
Re: K231324
January 8, 2024
Trade/Device Name: DASI Dimensions (V1.0) Regulation Number: 21 CFR 892.2050 Regulation Name: Medical Image Management And Processing System Regulatory Class: Class II Product Code: QIH Dated: December 13, 2023 Received: December 14, 2023
Dear Manali Shah:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrb/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Page
2
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Lamb
Jessica Lamb Assistant Director DHT8B: Division of Radiologic Imaging Devices and Electronic Products OHT8: Office of Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
#### Indications for Use
Submission Number (if known)
K231324
Device Name
DASI Dimensions (V1.0)
Indications for Use (Describe)
DASI Dimensions is a standalone, non-invasive, clinical decision support software solution that is intended for use by cardiologists and radiologists in context of the aortic stenosis population.
DASI Dimensions provides the end-user with pre-defined images and measurements (diameters, lengths, angles, areas, and perimeters) of cardiovascular structures.
The clinician receiving the results the responsibility for interpreting and validating all information and making all patient treatment decisions.
DASI Dimensions is not intended to replace the clinician's decision or device's instructions for use.
DASI Dimensions is prescription use only.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
510(k) #: K231324
510(k) Summary
Prepared on: 2024-01-02
### Contact Details
21 CFR 807.92(a)(1)
| Applicant Name | DASI Simulations | |
|------------------------------------|-------------------------------------------------------------|--------------|
| Applicant Address | 5115 Parkcenter Ave Suite 205 Dublin OH 43017 United States | |
| Applicant Contact Telephone | (909) 247-5143 | |
| Applicant Contact | Ms. Manali Shah | |
| Applicant Contact Email | manali.shah@dasisim.com | |
| Correspondent Name | DASI Simulations | |
| Correspondent Address | 5115 Parkcenter Ave Suite 205 Dublin OH 43017 United States | |
| Correspondent Contact Telephone | (770)335-8953 | |
| Correspondent Contact | Dr. Lakshmi Dasi | |
| Correspondent Contact Email | lakshmi.dasi@dasisim.com | |
| Device Name | 21 CFR 807.92(a)(2) | |
| Device Trade Name | DASI Dimensions (V1.0) | |
| Common Name | Medical image management and processing system | |
| Classification Name | Automated Radiological Image Processing Software | |
| Regulation Number | 892.2050 | |
| Product Code | QIH | |
| Legally Marketed Predicate Devices | 21 CFR 807.92(a)(3) | |
| Predicate # | Predicate Trade Name (Primary Predicate is listed first) | Product Code |
| K153736 | 3mensio Workstation | LLZ |
| Device Description Summary | 21 CFR 807.92(a)(4) | |
DASI Dimensions is an image post-processing software system intended for clinical decision support in the context of pre-procedural planning of Transcatheter Aortic Valve Replacement (TAVR) procedures. The software provides users with a report of generated dimensions of cardiac structures. DASI Dimensions software is not operated by physicians.
DASI Dimensions is an off-site software system that receives cardiologists input via upload of patient multiphase DICOM computed tomography angiography (CTA) chest image files through DASI Simulations web portal. The processed report is then available for viewing and downloading. The report is generated using proprietary algorithms that (a) detect key aortic root control points with the assistance of a static deep learning artificial intelligence (Al) model and (b) calculate anatomical measurements relevant for pre-TAVR evaluation. DASI Simulations engineers perform quality checks at both steps before releasing the report to the end user via the portal.
{4}------------------------------------------------
DASI Dimensions does not contact with the patient, nor does it control any life sustaining devices. The information provided by DASI Dimensions is not diagnostic, nor does it determine recommended medical care.
The cardiologists and/or radiologists receiving the responsibility for interpreting and validating all information and making all patient treatment decisions.
DASI Dimensions:
· Supports quantification of cardiac structures for pre-procedural planning of aotic stenosis patients in consideration for potential TAVR procedures.
· Provides the measurement of different structures of the heart, e.g., aortic root, aortic valve.
# Intended Use/Indications for Use
DASI Dimensions is a standalone, non-invasive, clinical decision support software solution that is intended for use by cardiologists and radiologists in context of the aortic stenosis population.
DASI Dimensions provides the end-user with pre-defined images and measurements (diameters, lengths, angles, areas, and perimeters) of cardiovascular structures.
The clinician receiving the responsibility for interpreting and validating all information and making all patient treatment decisions.
DASI Dimensions is not intended to replace the clinician's decision or device's instructions for use.
DASI Dimensions is prescription use only.
# Indications for Use Comparison
DASI Dimensions and the predicate device have similar indications for use and have the same intended use. Both devices provide the end-user with reports for pre-operative planning for cardiovascular interventions with images and measurements. The difference in the two indications is the support of certain modules (dimensions in coronary arteries, quantifications and certain visualization techniques such as volume rendering, MPR etc.) by the predicate device that are not supported by DASI Dimensions.
# Technological Comparison
A comparison of the technological characteristics of the subject device is presented below:
1. Target Area: The predicate device's target areas include the heart, ascending aorta and peripheral vasculature. The subject device's target areas are cardiac structures specific to Transcatheter Aortic Valve Replacement such as aortic valve, sinus of valsalva, sinotubular junction and ascending aorta.
2. Functionality: Both the subject and predicate device enable linear (diameter), angular and areal measurements of cardiac structures with annotations. The predicate device has additional tools for volume measurements and scoring of calcification in addition to the functionality described above.
3. Interpretation of Images: The outputs of both the subject and predicate devices are interpreted by Health Care Professionals (HCP).
4. Technical Method: Both the subject and predicate devices use an automated algorithm involving segmentation and centerline detection of cardiac structures.
5. Software Architecture: Both the subject and predicate devices have multiphase DICOM data provided by clinical users as input, while the reports are prepared by qualified operators and made available digitally.
#### Non-Clinical and/or Clinical Tests Summary & Conclusions 21 CFR 807.92(b)
Non-clinical bench performance testing includes the relevant information described in the FDA guidance document "Recommended Content and Format of Non-Clinical Bench performance Testing Information in Premarket Submissions". The subject device has been evaluated and verified in accordance with specifications and applicable performance standards through software verification and validation testing.
Software verification and validation testing confirmed that the software requirements fulfilled the predefined acceptance criteria. The reference standard was derived from 2 qualified truthing each CTA, whose measurements were averaged for each case. If there
## 21 CFR 807.92(a)(5)
## 21 CFR 807.92(a)(5)
21 CFR 807.92(a)(6)
{5}------------------------------------------------
was a significant variance between the initial two truthers, an adjudicator was involved. To ensure the uniqueness of all data used in the studies, each case UD (unique identifer) was monitored. Relevant studies incorporated data with various distributions of ethnicity, gender, and age. All data was collected through the DASI Simulations secure server, AWS, through internal workstations operated by trained DASI Dimensions operators.
Software verification and validation testing were conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices".
As described in the DASI Dimensions Software Verification and Validation Plan (TP131), the following testing was completed:
(i) Software Verification and Validation: Software vere done to verify that design inputs described in the software requirements specifications can be traced to design outputs including software design documents, test procedures, and system software validation activities were performed to ensure that the design outputs meet design inputs using a complete DASI Dimensions system software release.
a. Traceability Matrix (RMF112): Generated to trace each functional requirement (design input) through implementation (design output). Functional Requirements testing (TR142 and TR143) were performed to ensure that each design input can be traced to the design output. The results showed successful performance of the functional testing.
b. Al Control Point Detection Validation Study (W112 and TR132) trained and tested the Al model using training, validation, and testing datasets. Control point deviation has e clinically significant impact on the final outputs when > 3 mm of error is introduced thus an acceptance criteria of ≤ 3 mm was used to determine success of an individual control point. In the Al testing dataset the Al generated control points were compared to manual control points generated by qualified DASI Dimensions operators (ground truth), error resulted in a success rate of 75.3% of points. All testing, validation, and training cases were independent of each other and were not used in any other validation studies.
c. Automatic Measurements Validation Study (TP133 and TR133) was performed to verify the accuracy of the automatic measurement process of DASI Dimensions. The automatic measurement outputs generated using control points were compared to the measurements generated manually by clinicians using current standard of care methods. The primary measurements showed ≤15% error in ≤95% of cases thus satisfying the acceptance criteria. The secondary measurements showed ≤20% error in ≤95% of the cases thus satisfying the acceptance criteria. The automatic measurement with the manual measurements generated by clinicians with no statistically significant differences between the two measurement groups.
d. Performance Validation Study (TP134 and TR134) was used as validation of the overall performance of the device. A cohort of patient CT images (n=40), representative of the general population of patients with tricuspid aortic stenosis, were processed through the DASI Dimensions semi-automatic Al enhanced workflow by trained DASI Simulations operators where outputs were compared to qualified clinician truther generated control points performed at an 85.3% success rate achieving the acceptance criteria of ≥75%. The mean percentage error in primary outputs was 0.93% (Cl: positive 8.65%, negative -6.80%) and -1.02% (Cl: positive 3.49%, nespectively for the annulus area and perimeter, satisfying the acceptance criteria of ≤10% error in ≥95% of cases. The mean percentage difference in secondary outputs was 4.77% (Cl: positive 11.26%, negative -1.72%), 4.13% (Cl: positive 11.61%, negative -3.35%), 3.29% (C: 8.83%, negative - 2.25%) respectively for the left-, ight- and noncoronary sinus of valsalva diameters, 1.25% (Cl: positive -6.30%) and 2.20% (Cl: positive 8.46%, negative -4.06%) respectively for sinotubular junction maximum and minimum diameters, and 0.12% (Cl: positive -5.05%) for the ascending aorta maximum diameter, satisfying the acceptance criteria of ≤15% of cases. The mean percentage difference in the tertiary output between DASI Dimensions and clinician truthing was 2.66% (C.: positive - 19.66%) for the aortic valve angle, satisfying the acceptance criteria of ≤25% of cases. No trends were observed between the aotic valve annulus error and patient or scan characteristics. Scans were independent of those used for Al training or any other validation study.
e. Operator Variability Study (TP136 and TR136) validated the repeatability of DASI Dimensions outputs produced by different operators. The study evaluated whether the observed inter-operator variability in the selection of control points and alteration of measurement location/contours can introduce significant difference in the DASI Dimensions measurement outputs. Using a dataset of CTAs (n = 14), comparison of the annulus area outputs from DASI Dimensions made by trained DASI Simulations operators (n = 5) showed excellent inter-operator agreement (precision) and to clinician ground truth measurements (accuracy), with an ICC of 0.96 and ≤ 10% difference from clinician measurements in ≥95% of cases Both acceptance criteria were met, thus demonstrating that the variability in DASI Dimensions outputs due to operator inputs is within acceptable limits.
f. Human Factors/Usability Validation (TP137 and TR137) was conducted to assess the DASI Dimensions operator (n =6) interactions with the Processing Stage of DASI Dimensions software system user interface to identify use errors that would or could result in serious harm to the patient or user. The Human Factors and Usability Testing is to assess the effectiveness of risk management measures. Overall, the DASIDimensions Processing Stage Human Factors evaluation showed that the DASI Dimensions Processing Stage user interface is user-friendly and reduces use-related risk. There were no instances of a system failure is defined as the inability for a participant to complete any of the critical tasks provided to them. The findings of the DASI Dimensions Processing Stage Human Factors evaluation shows that the relevant hazards have been reduced to a residual risk that is acceptable. The
{6}------------------------------------------------
100% success rate of the system and positive feedback regarding how intuitive the user interface was showed that the "Operational Hazard Use Error" (Hazard ID Section 9.0 and "Information Hazard ID Section 10.0) hazards were successfully reduced to an acceptable residual risk due to a remote probability of ever occurring.
g. Control Point Sensitivity Study (TP140 and TR140) focused on assessing the Sensitivity of the DASI Dimensions automatic measurement outputs to the error in the aortic control points coordinates. Using a dataset of CTAs (n = 14), measurements were generated using control points that were iteratived in designated directions and magnitudes and were compared with the baseline measurements which were generated using an unperturbed set of verified control points. At both 1.5 mm and 3.0mm perturbations of the control points, the resulting automatic annulus area measurements had percent errors ≤ 10%, meeting the acceptance criteria.
(ii) Cybersecurity testing (TP130 and TR130) was conducted to ensure that there were no unidentified vulnerabilities and that the appropriate risk control measures were implemented to protect from known vulnerabilities when the device of threat. The testing showed that appropriate risk control measures were implemented.
The non-clinical verfication and validation test results ed that the device meets the design requirements and intended use. During the development, potential hazards were evaluated and controlled through risk management activities. The performance testing demonstrates that the device meets all its specifications.
No prospective clinical trials were conducted in support of this Traditional 510(k). Clinical evaluation, as a set of ongoing activities conducted in the assessment and analysis of DASI Dimensions clinical safety, effectiveness and performance, are outlined in non-clinical testing summary.
The subject device has been evaluated and verified in accordance with predefications and applicable performance standards through software verification and validation testing. The non-clinication and validation test results established that the device meets its design requirements and intended use. During the development, potential hazards were evaluated and controlled through risk management activities. The performance testing demonstrates that the device meets all its specifications. The data generated from each test also supports a finding of substantial equivalence based on meeting testing requirements, special controls, and consensus standards.
No prospective clinical trials were conducted in support of this Traditional 510(k). Clinical evaluation, as a set of ongoing activities conducted in the assessment and analysis of DASI Dimensions clinical safety, effectiveness and performance, are outlined in non-clinical testing summary.
The subject device has been evaluated and verified in accordance with predefications and applicable performance standards through software verfication and validation testing. The non-clinical verification test results established that the device meets its design requirements and intended use. During the development, potential hazards were evaluated and controlled through risk management activities. The performance testing demonstrates that the device meets all its specifications. The data generated from each test also supports a finding of substantial equivalence based on meeting tequirements, special controls, and consensus standards.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.