K153736 · Pie Medical Imaging BV · LLZ · May 27, 2016 · Radiology
Device Facts
Record ID
K153736
Device Name
3mensio Workstation
Applicant
Pie Medical Imaging BV
Product Code
LLZ · Radiology
Decision Date
May 27, 2016
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
3mensio Workstation is a software solution that is intended to provide Cardiologists, Radiologists and Clinical Specialists additional information to aid them in reading and interpreting DICOM compliant medical images of structures of the heart and vessels. 3mensio Structural Heart enables the user to: Visualize and measure (diameters, lengths, areas, volumes, angles) structures of the heart and vessels Quantify calcium (volume, density) 3mensio Vascular enables the user to: Visualize and assess stenosis, aneurisms and vascular structures Measure the dimensions of vessels (diameters, lengths, areas, volumes, angles)
Device Story
3mensio Workstation is a software-based medical imaging tool for cardiologists, radiologists, and clinical specialists. It ingests DICOM-compliant medical images (CT, MRI, XA, etc.) from various data sources and networks. The software performs segmentation of cardiovascular structures and automatic/manual centerline detection to generate 3D models from 2D scan slices. It provides visualization techniques including Volume Rendering, MPR, Curved MPR, and MIP. Users interact with the workstation to perform measurements (diameters, lengths, volumes, angles) and calcium scoring. The output—visualizations, measurements, and reports—is used by clinicians to aid in pre-operative planning, sizing for cardiovascular interventions, and postoperative evaluation. By quantifying coronary artery dimensions and calcifications, the device supports clinical diagnosis and procedural preparation, potentially improving the accuracy and reliability of cardiovascular surgical planning.
Clinical Evidence
Bench testing only. Verification and validation performed according to Design Control procedures. System requirements were verified to be implemented correctly. A test report comparing numerical results of the subject device against predicate devices was generated to demonstrate effectiveness.
Technological Characteristics
Software-based image processing system running on Microsoft Windows. Supports DICOM-compliant datasets (CT, MRI, XA, etc.). Features include segmentation, centerline detection, and various 3D reconstruction techniques (MPR, MIP, Volume Rendering). Connectivity via standard network protocols. Conforms to NEMA PS 3.1-3.20 (DICOM) and ISO 14971 (risk management).
Indications for Use
Indicated for visualization and measurement of heart and vessel structures for pre-operational planning/sizing for cardiovascular interventions/surgery, postoperative evaluation, and support of clinical diagnosis by quantifying coronary artery dimensions and calcifications (calcium scoring).
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of three human profiles facing to the right, with flowing lines representing hair or clothing.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 27, 2016
Pie Medical Imaging B.V. % Ms. Annemeik Bouts Regulatory Affairs Coordinator Philipsweg 1 Maastricht 6227 AJ THE NETHERLANDS
Re: K153736
Trade/Device Name: 3mensio Workstation Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: April 20, 2016 Received: May 2, 2016
Dear Ms. Bouts:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{1}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Robert Ochs
Robert Ochs. Ph.D. Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known)
K153736
Device Name 3mensio Workstation
Indications for Use (Describe)
3mensio Workstation enables visualization and measurement of the heart and vessels for:
- · Pre-operational planning and sizing for cardiovascular interventions and surgery
- · Postoperative evaluation
- · Support of clinical diagnosis by quantifying dimensions in coronary arteries
- · Support of clinical diagnosis by quantifying calcifications (calcium scoring) in the coronary arteries
To facilitate the above, the 3mensio Workstation provides general functionality such as:
- · Segmentation of cardiovascular structures
- · Automatic and manual centerline detection
· Visualization and image reconstruction techniques: 2D review, Volume Rendering, MPR, Curved MPR, Stretched
- CMPR, Slabbing, MIP, AIP, MinIP · Measurement and annotation tools
- Reporting tools
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| 510(k) Summary | 3mensio Workstation 8.0 | [GEN2261]v1.0 |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------|
| Submitter/Owner Name | Pie Medical Imaging BV | |
| Address | Philipsweg 1, 6227 AJ Maastricht, The Netherlands | |
| Phone Number | +31 43 32 81 328 | |
| Fax Number | +31 43 32 81 329 | |
| Contact Person: | Annemiek Bouts, Regulatory Affairs Coordinator | |
| Email Address | annemiek.bouts@pie.nl | |
| Preparation Date | 16-dec-15 | |
| Trade Name | 3mensio Structural Heart / 3mensio Vascular | |
| Common Name | 3mensio Workstation | |
| Classification: | Classification Name: Image Processing System<br>Classification Panel: Radiology<br>Regulation Class: Class II<br>Regulation number: 21 CFR 892.2050<br>Classification Product Code: LLZ | |
| Predicate Devices | 3mensio Workstation (K120367)<br>Philips Intellispace Cardiovascular (K153022) | |
| Device Description | The 3mensio Workstation software allows cardiologists, radiologist and clinical<br>specialists to select patient studies from various data sources, view them, and<br>process the images with the help of a comprehensive set of tools. 3mensio<br>Workstation works with all major medical image formats and can access multiple<br>data stores and across networks.<br><br>3mensio Workstation can combine 2D scan slices into comprehensive 3D models<br>of the patient, and can display supporting ultrasound and X-ray Angio data. The<br>software accurately represents different types of tissue, making it easier to<br>diagnose anomalies in scans.<br><br>3mensio Workstation contains two modules, 3mensio Structural Heart<br>(containing submodule cCTA), and 3mensio Vascular which can be marketed in<br>combination or as separate solutions:<br>3mensio Structural Heart enables assessment and measurement of differen<br>structures of the heart, e.g. aortic valve, mitral valve and ventricles. It provides<br>analysis of the feasibility of a transapical, transfemoral or subclavian approach to<br>structures for replacement or repair procedures. 3mensio cCTA enables assessment and measurement of the coronary<br>arteries and can help in the quantification of the dimensions of the<br>coronary arteries and of the calcified plaques present in coronary<br>arteries. 3mensio Vascular enables assessment of vessels and can help identify<br>calcifications, aneurysms and other anomalies to quickly and reliably prepare for<br>various types of vascular procedures. | |
| The results can be displayed, printed or saved in a variety of formats to a hard disk, network, PACS system or CD. Results can also be printed as a hardcopy. | | |
| Intended Use | 3mensio Workstation is a software solution that is intended to provide Cardiologists, Radiologists and Clinical Specialists additional information to aid them in reading and interpreting DICOM compliant medical images of structures of the heart and vessels.<br>3mensio Structural Heart enables the user to:<br>Visualize and measure (diameters, lengths, areas, volumes, angles) structures of the heart and vessels Quantify calcium (volume, density) 3mensio Vascular enables the user to: Visualize and assess stenosis, aneurisms and vascular structures Measure the dimensions of vessels (diameters, lengths, areas, volumes, angles) | |
| Indications for Use | 3mensio Workstation enables visualization and measurement of structures of the heart and vessels for:<br>Pre-operational planning and sizing for cardiovascular interventions and surgery Postoperative evaluation Support of clinical diagnosis by quantifying dimensions in coronary arteries Support of clinical diagnosis by quantifying calcifications (calcium scoring) in the coronary arteries To facilitate the above, the 3mensio Workstation provides general functionality such as: Segmentation of cardiovascular structures Automatic and manual centerline detection | |
{4}------------------------------------------------
- Measurement and annotation tools ●
- . Reporting tools
Technological Characteristics Comparison
A comparison of the technological characteristics of the predicate and subject device is given the table below.
#### New Device Predicate Device Predicate Device 3mensio Workstation Device name 3mensio Workstation Intellispace Cardiovascular Pie Medical Imaging Manufacturer Pie Medical Imaging Philips K120367 510(k) number K153022 Intended Use and Indications for Use Intended use 3mensio Workstation is a software 3mensio Workstation is a software Philips IntelliSpace Cardiovascular solution that is intended to provide solution that is intended to provide software product is an integrated cardiologists, radiologists and clinical cardiologists, radiologists and clinical multimodality image and information specialists additional information to aid specialists additional information to aid system designed to perform the them in reading and interpreting them in reading and interpreting DICOM necessary functions required for DICOM compliant medical images of compliant medical images of structures of import, export, storage, archival, structures of the heart and vessels. the heart and vessels. review, analysis, quantification, reporting and database management 3mensio Structural Heart enables the 3mensio Structural Heart enables the user of digital medical images. user to: to: · Visualize and measure (diameters, · Visualize and measure (diameters, lengths, areas, volumes, angles) lengths, areas, volumes, angles) structures of the heart and vessels structures of the heart and vessels · Quantify calcium (volume, density) · Quantify calcium (volume, density) 3mensio Vascular enables the user to: 3mensio Vascular enables the user to: · Visualize and assess stenosis, · Visualize and assess stenosis, aneurisms aneurisms and vascular structures and vascular structures · Measure the dimensions of vessels · Measure the dimensions of vessels (diameters, lengths, areas, volumes, (diameters, lengths, areas, volumes,
{5}------------------------------------------------
| | New Device | Predicate Device | Predicate Device |
|-------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device name | 3mensio Workstation | 3mensio Workstation | Intellispace Cardiovascular |
| Manufacturer | Pie Medical Imaging | Pie Medical Imaging | Philips |
| 510(k) number | - | K120367 | K153022 |
| | angles) | angles) | |
| Indications<br>for Use | 3mensio Workstation enables<br>visualization and measurement of<br>structures of the heart and vessels for:<br>• Pre-operational planning and sizing<br>for cardiovascular interventions and<br>surgery<br>• Postoperative evaluation<br>• Support of clinical diagnosis by<br>quantifying dimensions in coronary<br>arteries<br>• Support of clinical diagnosis by<br>quantifying calcified plaques (calcium<br>scoring) in the coronary arteries<br><br>To facilitate the above, the 3mensio<br>Workstation provides general<br>functionality such as:<br>• Segmentation of cardiovascular<br>structures<br>• Automatic and manual centerline<br>detection<br>• Visualization and image<br>reconstruction techniques: 2D review,<br>Volume Rendering, MPR, Curved<br>MPR, Stretched CMRP, Slabbing,<br>MIP, AIP, MinIP<br>• Measurement and annotation tools<br>• Reporting tools | 3mensio Workstation enables visualization<br>and measurement of structures of the heart<br>and vessels for:<br>• Pre-operational planning and sizing for<br>cardiovascular interventions and<br>surgery<br>• Postoperative evaluation<br><br>To facilitate the above, the 3mensio<br>Workstation provides general functionality<br>such as:<br>• Segmentation of cardiovascular<br>structures<br>• Automatic and manual centerline<br>detection<br>• Visualization and image<br>reconstruction techniques: 2D review,<br>Volume Rendering, MPR, Curved<br>MPR, Stretched CMRP, Slabbing,<br>MIP, AIP, MinIP<br>• Measurement and annotation tools<br>• Reporting tools | Philips IntelliSpace Cardiovascular<br>software product is an integrated<br>multimodality image and information<br>system designed to perform the<br>necessary functions required for<br>import, export, storage, archival,<br>review, analysis, quantification,<br>reporting and database management<br>of digital cardiovascular images,<br>waveforms<br>and data related to cardiology.<br><br>Philips IntelliSpace Cardiovascular<br>offers support for third party<br>applications in order to enable the use<br>of commercially available tools and<br>specified applications for analysis,<br>quantification and reporting. It allows<br>multiple users fast access to, and<br>exchange of specific and/or multiple<br>cardiology exams.<br><br>Philips IntelliSpace Cardiovascular<br>software runs on standard information<br>technology hardware and software,<br>utilizing the standard information<br>technology operating systems and<br>user interface.<br><br>Communication and data exchange<br>are done using standard protocols.<br><br>Philips IntelliSpace<br>Cardiovascular will also be made<br>available for use on specified<br>Cardiovascular Monitoring Systems,<br>which use suitable hardware<br>components.<br><br>The modular design allows<br>configurability to tailor the image<br>import, archive and communications<br>solution<br>to one's particular budgetary and<br>performance needs. The number of<br>modalities and reporting and/or<br>viewing sites can be configured per<br>system. |
| | Technological Characteristics | | |
| Data type | • CT data in DICOM format (vendor<br>independent) | • CT data in DICOM format (vendor<br>independent) | • Any image modality data in<br>DICOM format (vendor<br>independent) |
| Input of Patient<br>Data | • Manual through keyboard/mouse<br>• Command line interface | • Manual through keyboard/mouse<br>• Command line interface | • Manual through<br>keyboard/mouse<br>• Command line interface |
| Study list image<br>functionality | • Exporting<br>• Deleting<br>• Anonymizing (no automatic deletion<br>of original patient data)<br>• Search | • Exporting<br>• Deleting<br>• Anonymizing (no automatic deletion<br>of original patient data)<br>• Search | Full patient/study list support |
| Centreline<br>extraction | • Realign orthogonal MPRs<br>• Segmentation toolset:<br>- Automatic segmentation<br>- Automatic centreline<br>- Manual centreline | • Realign orthogonal MPRs<br>• Segmentation toolset:<br>- Automatic segmentation<br>- Automatic centreline<br>- Manual centreline | (Features not mentioned since they<br>are not relevant for the equivalence) |
| | New Device | Predicate Device | Predicate Device |
| Device name | 3mensio Workstation | 3mensio Workstation | Intellispace Cardiovascular…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.