K231264 · Doctus Global · EIH · Jun 30, 2023 · Dental
Device Facts
Record ID
K231264
Device Name
Easy On Universal Glaze & Stain
Applicant
Doctus Global
Product Code
EIH · Dental
Decision Date
Jun 30, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6660
Device Class
Class 2
Indications for Use
Easy On Universal Glaze & Stain is suitable for the esthetic individualization of monolithic or minimally reduced dental crown and bridge constructions made of zirconium dioxide or lithium disilicate.
Device Story
Easy On Universal Glaze & Stain is a dental ceramic finishing system consisting of paste-type stains, glazes, and liquids. Used by dental professionals to individualize the appearance of dental prostheses (crowns, bridges, inlays, onlays) made of zirconium dioxide or lithium disilicate. The system includes various shades and translucency levels to replicate natural tooth and gum tissue characteristics (incisal translucency, mamelons, halos). The paste is applied to the ceramic surface using a brush; the material is then fired at low temperatures to integrate surface textures and colors. The device provides aesthetic enhancements and surface finishing for dental restorations. It is intended for use in a dental laboratory or clinical setting by trained dental personnel.
Clinical Evidence
No clinical data. Bench testing only. Performance testing conducted per ISO 6872:2015 (flexural strength >= 50 MPa, chemical solubility < 100 µg/cm², thermal expansion, glass transition temperature). Biocompatibility testing performed per ISO 10993-1, -3, -5, -6, -11, and ISO 7405.
Technological Characteristics
Feldspar ceramic (glass oxide) paste and liquid system. Form factor: pastes and liquids. Connectivity: none. Sterilization: non-sterile. Complies with ISO 6872:2015 (Type 1, class 1b). Flexural strength >= 50 MPa. Thermal expansion coefficient (12.0±0.5) x10^-6K^-1. Glass transition temperature (560±20)°C.
Indications for Use
Indicated for esthetic individualization of monolithic or minimally reduced dental crown and bridge constructions made of zirconium dioxide or lithium disilicate.
Regulatory Classification
Identification
Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it.
June 30, 2023
Doctus Global % Bokyeong Kim Senior Researcher GMS Consulting 4th Floor, Digital Cube, 34, Sangamsan-ro Seoul, Mapo-gu 03909 SOUTH KOREA
Re: K231264
Trade/Device Name: Easy On Universal Glaze & Stain Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder For Clinical Use Regulatory Class: Class II Product Code: EIH Dated: April 28, 2023 Received: May 2, 2023
Dear Bokyeong Kim:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA
has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael E. Adjodha -S
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K231264
Device Name Easy On Universal Glaze & Stain
Indications for Use (Describe)
Easy On Universal Glaze & Stain is suitable for the esthetic individualization of monolithic or minimally reduced dental crown and bridge constructions made of zirconium dioxide or lithium disilicate.
Type of Use (Select one or both, as applicable)
|--|--|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo for a company called "Doctus G". The logo features a stylized house shape with a window inside a curved shape, suggesting a smile. The text "Doctus G" is written in a bold, modern font below the house shape. Underneath the company name is the tagline "Making Your Happy Smile".
# 510(k) Summary
[As Required by 21 CFR 807.92]
#### 1. Date Prepared [21 CFR 807.92(a)(a)]
April 28, 2023
#### 2. Submitter's Information [21 CFR 807.92(a)(1)]
| ● | Name of Sponsor:<br>- Address: | GMS Consulting<br>4th Floor, Digital Cube, 34, Sangamsan-ro, Mapo-gu, Seoul,<br>Republic of Korea,03909 |
|---|-------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| ● | Contact Name:<br>- Telephone No.:<br>- Email Address: | Beom-jin Choi, R&D Principal Researcher<br>+82-10-8989-5741<br>doctus2021@naver.com |
| ● | Name of Manufacturer:<br>- Address: | Doctus Global<br>RN.313, 51, Samwon-ro, Deogyang-gu, Goyang-si, |
Gyeonggi-do, Republic of Korea
#### 3. Trade Name, Common Name, Classification [21 CFR 807.92(a)(2)]
| Regulation Name | Porcelain Powder for Clinical Use |
|-----------------------------|-----------------------------------|
| Product Name | Easy On Universal Glaze & Stain |
| Model Name | Easy On Universal Stain |
| Regulation Number | 872.6660 |
| Classification Product Code | EIH |
| Device Class | 2 |
| 510k Review Panel | Dental |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Doctus G. The logo features a stylized letter 'D' with a small house icon inside, suggesting a connection to home or real estate. Below the symbol, the name "Doctus G" is written in a bold, modern font, with the 'G' being slightly larger and in a different color. The tagline "Making Your Happy Smile" is present below the name.
# 4. Identification of Predicate Device(s) [21 CFR 807.92(a)(3)]
The identified predicate devices within this submission are shown as follow:
Predicate Device #1
- 510(k) Number: K193588
- Applicant: Dental Direkt GmbH ●
- Regulation Name: Porcelain Powder for Clinical Use
- Proprietary Name: DD Contrast ●
- Classification Product Code: EIH
- Regulation Number: 21 CFR 872.6660
The predicate device has not been subject to a design-related recall
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Doctus G. The logo features a stylized letter 'D' with a small house icon inside, suggesting a connection to home or real estate. Below the symbol, the name "Doctus G" is written in a modern font, with the 'G' being larger and in a different color. The tagline "Making Your Happy Smile" is present below the name.
#### 5. Description of the Device [21 CFR 807.92(a)(4)]
#### 5.1. General
Doctus Global offers two kinds of Easy On Universal Glaze & Stain:
- V Easy On Universal Stain for artificial teeth
- > Easy On Universal Gum Stain for dental prosthesis, artificial gum tissue portion.
- V Easy Liners for Porcelain powder build-up process.
- V Easy Add On for Correcting and Repair process.
- > Easy On Crystal Glaze for glazing Lithium disilicate CAD/CAM block crown and bridge
- V Build Up Liquid for porcelain powder build up process
Both systems and other items are a useful finishing system made of paste type Stain & Glaze for dental ceramics.
Fluorescent layering materials for staining and glazing. With Easy On Universal Stain & Gum Stain, you can easily and quickly achieve high quality dental ceramic prosthesis.
Easy On Universal Stain consists of different types of self-glazing colors with various levels of translucency, each Unique replicate natural tooth shade and gum tissue colors and structures, such as incisal translucency, mamelons, detail expressions, halos, and gum tissue.
Over paint Easy On Universal Stains we layer Easy On Universal Structures. With Easy On Universal Structure you can create tooth surface textures found in natural enamel area and gum tissues, in light and thin thicknesses of 0.1 mm - 0.3 mm. With easy applying and high stability. individual surface structures can easily be integrated into the paste with small brush. The low firing temperature does not change the details of the surface, so reliable and aesthetic results are achieved with one or two firings.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo for a company called "Doctus G". The logo features a stylized house shape with a window inside a teal circle. Below the circle, the company name "Doctus G" is written in a bold, gray font. Underneath the company name is the tagline "Making Your Happy Smile" in a smaller font.
#### 5.2. Device composition
- * The components of the kit below table are sold individually or in kits. {1) ~ 2}}
- 1) Easy On Universal Stain Full Kit
Image /page/6/Picture/5 description: The image shows a kit with various colors and tools. There are 16 small containers with different colors, including light brown, red, light blue, enamel, red brown, pink, dark blue, clear, dark brown, orange, gray, white, ivory, and glaze. There is also a small bottle labeled "EASTON" and a thin brush.
- Composition
| <br>Easy On Universal<br>Glaze | <br>Stain-Shade A | <br>Stain-Shade B | <br>Stain-Shade C |
|--------------------------------|-----------------------|---------------------|----------------------|
| <br>Stain-Shade D | <br>Stain-Light Brown | <br>Stain-Red Brown | <br>Stain-Dark Brown |
| <br>Structure Stain-Enamel | <br>Stain-White | <br>Stain-Ivory | <br>Stain-Pink |
510(k) Summary
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows a logo for a company called "Doctus G". The logo features a stylized letter "D" with a window inside, suggesting a connection to homes or buildings. Below the "D" is the company name, "Doctus G", in a modern, sans-serif font. The tagline "Making Your Happy Smile" is present below the company name.
| Image: Stain-Red | Image: Stain-Orange | Image: Stain-Light Blue | Image: Stain-Dark Blue |
|-------------------|------------------------------|--------------------------------|------------------------|
| Stain-Red | Stain-Orange | Stain-Light Blue | Stain-Dark Blue |
| Image: Stain-Gray | Image: Structure Stain-Clear | Easy On Universal Glaze Liquid | |
| Stain-Gray | Structure Stain-Clear | Easy On Universal Glaze Liquid | |
| Image: Brush | | | |
| Brush | | | |
| No. | Name | Description |
|-----|----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Easy On<br>Universal Glaze | A transparent glaze layer is formed on the outer surface of the ceramic<br>Inlay, Onlay, Crown, Crown & Bridge and Zirconia dental prosthesis |
| 2 | Stain-Shade A | Among the basic shade of teeth, the stain that expresses the basic<br>colors of the A |
| 3 | Stain-Shade B | Among the basic shade of teeth, the stain that expresses the basic<br>colors of the B |
| 4 | Stain-Shade C | Among the basic shade of teeth, the stain that expresses the basic<br>colors of the C |
| 5 | Stain-Shade D | Among the basic shade of teeth, the stain that expresses the basic<br>colors of the D |
| 6 | Stain-Light Brown | Stain that expresses bright brown color on the exterior of dental<br>prosthesis |
| 7 | Stain- Red Brown | Stain that expresses red brown color on the exterior of dental prosthesis |
| 8 | Stain- Dark Brown | Stain that expresses dark brown color on the exterior of dental |
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image is a logo for Doctus G. The logo features a stylized house inside of a teal-colored shape. The text "Doctus G" is written in a bold font below the shape. There is also some smaller text below the main text that says "Making Your Happy Smile."
| | | prosthesis |
|----|-----------------------------------|-----------------------------------------------------------------------------------------------------|
| 9 | Structure Stain-<br>Enamel | Stain that expresses the color and transparent of the Enamel layer |
| 10 | Stain-White | Stain that expresses the lightest part of the surface of dental prosthesis |
| 11 | Stain-Ivory | Stain that expresses lighter part of the surface of dental prosthesis |
| 12 | Stain-Pink | Stain representation of gingiva and cervical part of the dental prosthesis |
| 13 | Stain-Red | Stain representation of gingiva and red color part of the dental<br>prosthesis |
| 14 | Stain-Orange | Stain representation of occlusal and cervical part of the dental<br>prosthesis |
| 15 | Stain-Light Blue | Stain that expresses the enamel layer of the anterior and posterior tooth |
| 16 | Stain-Dark Blue | Stain that expresses the enamel layer of the anterior and posterior tooth |
| 17 | Stain-Gray | Stain that expresses the enamel layer of the anterior and posterior tooth |
| 18 | Structure<br>Stain-Clear | Stain for representing the most transparent layer of the overall<br>appearance of dental prosthesis |
| 19 | Easy On Universal<br>Glaze Liquid | If the viscosity of the product increases during use, dilute it |
| 20 | Brush | Painting instrument |
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image shows a logo for a company called "Doctus G". The logo features a stylized house with a window inside a rounded square, with swooshes underneath. The text "Doctus G" is written in a bold, sans-serif font, with the "G" slightly offset and larger than the rest of the letters. Below the company name is the slogan "Making Your Happy Smile".
- 2) Easy On Universal Gum Stain Full Kit
Image /page/9/Picture/3 description: The image shows a kit with various components arranged in a black tray. There are 13 small containers with different colored substances labeled with codes such as G1, G2, G3, G4, G5, G6, G7, G51, G52, G53, G54, G55, Dark Red, and Purple. Additionally, there is a bottle labeled "EASY ON", a container labeled "Glaze", a white rectangular card with the "EASY ON" logo, and a thin brush.
- Composition
| Image: Easy On Universal Glaze | Gum Stain-G1 | Gum Stain-G2 | Gum Stain-G3 |
|--------------------------------|------------------|-------------------------|-------------------------|
| Gum Stain-G4 | Gum Stain-G5 | Gum Stain-G6 | Gum Stain-G7 |
| Gum Stain-Dark Red | Gum Stain-Purple | Gum Structure Stain-GS1 | Gum Structure Stain-GS2 |
{10}------------------------------------------------
Image /page/10/Picture/0 description: The image shows a logo for a company called "Doctus G". The logo features a stylized letter "D" with a window inside, suggesting a connection to homes or buildings. Below the "D" is the company name, "Doctus G", in a modern, sans-serif font. The tagline "Making Your Happy Smile" is present below the company name.
| | Gum Structure Stain-GS3 | Gum Structure Stain-GS4 | Gum Structure Stain-GS5 | Easy On Universal Glaze Liquid |
|-----|--------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|-------------------------|--------------------------------|
| | Brush | | | |
| | Pallet | | | |
| No. | Name | Description | | |
| 1 | Easy On Universal Glaze | A transparent glaze layer is formed on the outer surface of the ceramic Inlay, Onlay, Crown, Crown & Bridge and Zirconia dental prosthesis | | |
| 2 | Gum Stain-G1 | | | |
| 3 | Gum Stain-G2 | | | |
| 4 | Gum Stain-G3 | Stains for the gingiva part of the dental prosthesis. Becomes darker as it increases by 1 to 7 | | |
| 5 | Gum Stain-G4 | | | |
| 6 | Gum Stain-G5 | | | |
| 7 | Gum Stain-G6 | | | |
| 8 | Gum Stain-G7 | | | |
| 9 | Gum Stain-Dark Red | Stain for expressing the dark red color of the attached gingiva part of dental prosthesis | | |
| 10 | Gum Stain-Purple | Stain for expressing the purple color of the attached gingiva part of dental prosthesis | | |
| 11 | Gum Structure Stain-GS1 | | | |
| 12 | Gum Structure Stain-GS2 | Stains for the gingiva part outer detail texture expression of the dental prosthesis. Becomes darker as it increases by 1 to 5 | | |
| 13 | Gum Structure Stain-GS3 | | | |
| 14 | Gum Structure Stain-GS4 | | | |
| 15 | Gum Structure Stain-GS5 | | | |
| 16 | Easy On Universal Glaze Liquid | If the viscosity of the product increases during use, dilute it | | |
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows a logo for a company called "Doctus G". The logo features a stylized letter "D" with a small house icon inside, suggesting a connection to homes or real estate. Below the "D" is the company name "Doctus G" in a bold font, and underneath that is the tagline "Making Your Happy Smile" in a smaller font. The logo uses a teal color scheme, giving it a modern and clean appearance.
| 17 | Brush | Painting instrument |
|----|--------|-----------------------------------------|
| 18 | Pallet | Products for mixing and diluting stains |
* The items below are sold individually only. {3) ~ 6)}
3) Easy Liner
| Image: Easy Liner Light | Image: Easy Liner Dark | Image: Easy Liner Pink |
|-------------------------|------------------------|------------------------|
| Easy Liner Light | Easy Liner Dark | Easy Liner Pink |
| No. | Name | Description |
|-----|------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Easy Liner Light | Materials for increasing the combined strength and shear<br>strength of Zirconia copping and Ceramic Powder in the<br>process of firing and completing by adding Ceramic Powder to<br>a substructure made of bright Zirconia framework |
| 2 | Easy Liner Dark | Materials for increasing the combined strength and shear<br>strength of Zirconia copping and Ceramic Powder in the<br>process of firing and completing by adding Ceramic Powder to<br>a substructure made of bright Zirconia framework |
| 3 | Easy Liner Pink | Materials for increasing the combined strength and shear<br>strength of Zirconia copping and Ceramic Gum color Powder<br>in the process of firing and completing by adding Ceramic Gum<br>Powder to a substructure made of bright Zirconia framework |
#### 4) Easy Add On
| <br>Easy Add On Enamel | <br>Easy Add On Clear |
|------------------------|-----------------------|
|------------------------|-----------------------|
{12}------------------------------------------------
Image /page/12/Picture/0 description: The image shows the logo for Doctus G. The logo features a stylized letter 'D' with a small house icon inside, suggesting a connection to home or real estate. Below the symbol, the name "Doctus G" is written in a modern font, with the 'G' being larger and in a different color. The tagline "Making Your Happy Smile" is present below the name.
| No. | Name | Description |
|-----|-----------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Easy Add On<br>Enamel | Materials for complete the finished dental prosthesis, that<br>complements the appearance of the enamel translucent<br>layer on the outer surface of the artificial tooth |
| 2 | Easy Add On<br>Clear | Materials for complete the finished dental prosthesis, that<br>complements the appearance of the translucent layer on<br>the outer surface of the artificial tooth |
5) Easy On Crystal Glaze, Stain-Khaki, Structure Stain Sky Blue
| Image: Easy On Crystal Glaze | Image: Stain-Khaki | Image: Structure Stain - Sky Blue |
|------------------------------|--------------------|-----------------------------------|
| Easy On Crystal Glaze | Stain-Khaki | Structure Stain - Sky Blue |
| No. | Name | Description |
|-----|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1 | Easy On Crystal<br>Glaze | Materials for during the crystallization process by heat<br>treatment of lithium disilicate ceramic prosthesis, the<br>surface of the dental prosthesis is glazed |
| 2 | Stain-Khaki | Stain that expresses khaki color on the exterior of<br>dental prosthesis |
| 3 | Structure Stain<br>-Sky Blue | Stain that expresses the light blue color and<br>transparent of the enamel layer |
{13}------------------------------------------------
Image /page/13/Picture/0 description: The image is a logo for Doctus G. The logo features a stylized house shape inside of a rounded square, with a teal swoosh underneath. The text "Doctus G" is written in a bold font below the house shape. Underneath the text is the phrase "Making Your Happy Smile".
6) Easy On Build Up Liquid, Easy On Universal Glaze Liquid (50)
Image /page/13/Picture/3 description: The image shows two bottles of Easy On liquid. The bottle on the left is labeled "Easy On Build Up Liquid". The bottle on the right is labeled "Easy On Universal Glaze Liquid (50)". Both bottles are white and have a white cap.
| No. | Name | Description |
|-----|----------------------------------------|----------------------------------------------------------------------------|
| 1 | Easy On<br>Build Up Liquid | Liquid material by mixing dental porcelain powder when<br>build up process |
| 2 | Easy On Universal<br>Glaze Liquid (50) | If the viscosity of the product increases during use, dilute<br>it |
# 6. Indications for Use [21 CFR 807.92(a)(5)]
Easy On Universal Glaze & Stain is suitable for the esthetic individualization of monolithic or minimally reduced dental crown and bridge constructions made of zirconium dioxide or lithium disilicate.
{14}------------------------------------------------
Image /page/14/Picture/0 description: The image contains a logo for a company called "Doctus G". The logo features a stylized house icon with a window inside a rounded square, which is encircled by a teal crescent shape. Below the icon, the company name "Doctus G" is written in a bold, sans-serif font, with the "G" being slightly larger and in a different style. The tagline "Making Your Happy Smile" is present below the company name.
#### 7 . Technological Characteristics (Equivalence to Predicate Device) [21 CFR 807.92(a)(6)]
There are no significant differences in the technological characteristics of these devices compared to the predicate device which adversely affect safety or effectiveness. A table comparing the key features of the subject device and the predicate device is provided below table.
| | Proposed Device | Predicate Device | Note |
|----------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|
| K Number | - | K193588 | |
| Manufacturer | Doctus Global | Dental Direkt GmbH | |
| Product name | Easy On Universal Glaze &<br>Stain | DD Contrast | |
| Product Code | EIH | EIH | Same |
| Regulation<br>Number | 21 CFR 872.6660 | 21 CFR 872.6660 | Same |
| 510(k) Review<br>Panel | Dental | Dental | Same |
| Indications for<br>Use | Easy On Universal Glaze &<br>Stain is suitable for the<br>esthetic individualization of<br>monolithic or minimally<br>reduced dental crown and<br>bridge constructions made<br>of zirconium dioxide or<br>lithium disilicate. | The DD contrast®<br>veneering ceramics system<br>is suitable for the esthetic<br>individualization of<br>monolithic or minimally<br>reduced dental crown and<br>bridge constructions made<br>of zirconium dioxide or<br>lithium disilicate. | Same |
| Product for<br>veneering /<br>individualization<br>of dental<br>frameworks | Yes | Yes | Same |
| Framework<br>material | Zirconium dioxide and<br>lithium disilicate | Zirconium dioxide and<br>lithium disilicate | Same |
| Minimal amount<br>of fires | 1 | 1 | Same |
| For all sizes of<br>dental<br>frameworks | Yes | Yes | Same |
| Classification<br>acc. to ISO<br>6872 | Type 1, class 1b | Type 1, class 1b | Same |
| Material | Feldspar ceramics<br>(Glass Oxide) | Feldspar ceramics | Same |
| Form | Pastes and liquids | Pastes and liquids | Same |
| Shades | Different tooth and | Different tooth and | Same |
#### [Table. Comparison of Proposed Device to Predicate Device]
510(k) Summary
{15}------------------------------------------------
Image /page/15/Picture/0 description: The image is a logo for a company called "Doctus G". The logo features a stylized house with a window inside of a circle. The circle is teal and has a swoosh design around it. The text "Doctus G" is written in a bold font below the circle, and the tagline "Making Your Happy Smile" is written in a smaller font below the company name.
| | Proposed Device | Predicate Device | Note |
|------------------------|--------------------------|--------------------------|------|
| | individualization shades | individualization shades | |
| Physical<br>properties | Conform to ISO 6872:2015 | Conform to ISO 6872:2015 | Same |
| Flexural strength | $\ge$ 50 MPa | $\ge$ 50 MPa | Same |
| Single use | Yes | Yes | Same |
| Sterility | Non-sterile | Non-sterile | Same |
Doctus Global believes that the new product Easy On Universal Glaze & Stain is substantially equivalent to the predicate device when used as instructed by knowledgeable and trained dental personnel. There is no change of the intended use of the new device and no difference in fundamental scientific technology. The new device is made from the same materials as the predicate device. The new device was verified in accordance with the abovementioned FDA recognized standards.
In conclusion, Easy On Universal Glaze & Stain is substantially equivalent to the legally marketed predicate device.
#### 8. Non-Clinical Test Summary [21 CFR 807.92(b)(1)]
The following data were provided in support of the substantial equivalence determination:
- 1) Bio-compatibility
Bio-compatibility test is performed according to below:
- FDA Guidance document Use of International Standard ISO 10993-1, Biological evaluation of . medical devices- Part 1: Evaluation and testing within a risk management process
- . ISO 10993-1:2018 Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
- . ISO 10993-3:2014 Biological evaluation of medical devices - Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity
- ISO 10993-5:2009 Biological evaluation of medical devices Part 5: Tests for in vitro ● cytotoxicity
- ISO 10993-6:2016 Biological evaluation of medical devices — Part 6: Tests for local effects after implantation
- . ISO 10993-11:2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
- ISO 7405:2018 Dentistry - Evaluation of biocompatibility of medical devices used in dentistry
2) Performance Test
510(k) Summary
13 / 15
{16}------------------------------------------------
Image /page/16/Picture/0 description: The image shows a logo for a company called "Doctus G". The logo features a stylized letter "D" with a small house icon inside, surrounded by a crescent shape. Below the "Doctus G" text, there is a tagline that reads "Making Your Happy Smile".
The performance tests were performed according to ISO 6872:2015 Dentistry - Ceramic materials, and the test results showed that the proposed device meets the requirements specified in the standard.
| ISO 6872 Cl. | Test item | Requirements |
|--------------|-----------------------|---------------------------------------------------------|
| 5.1 | Uniformity | The color shall be uniformly dispersed |
| 5.2 | Extraneous materials | It shall be free from extraneous materials |
| 5.3 | Mixing and | It shall neither from lumps, nor granules during mixed, |
| | condensation | and it shall neither crack, nor crumble during drying. |
| 7.2 | Radioactivity | Uranium238 shall be less than 1.0 Bq/g |
| 7.3 | Flexural strength | it shall not be less than 50MPa |
| 7.4 | Linear thermal | It shall be within (12.0±0.5) x10-6K-1 |
| | expansion coefficient | |
| 7.5 | Glass transition | It shall be within (560±20)°C |
| | temperature | |
| 7.6 | Chemical solubility | The mass loss shall be less than 100 µg/cm² |
The requirements for other test items are as follows.
| Test item | Requirements |
|------------|---------------------------------------------------------------------------------------------------------------------------|
| Appearance | When tested by visual inspection, it shall be free form extraneous materials and<br>suitable for use. |
| Package | When tested by visual inspection, it shall be no damage, extraneous materials<br>and the other harmful defects for usage. |
| Weight | When measured the weight using a balance, the tolerance shall be withing<br>±5%. |
#### 9. Clinical Test Summary [21 CFR 807.92(b)(2)]
Clinical testing is not included in this submission.
#### 10.Substantial Equivalence [21 CFR 807.92(b)(1) and 807.92]
There are no significant differences between proposed device and the predicate devices that would adversely affect the use of the product. It is substantially equivalent to these devices in indications for use and technology characteristics.
{17}------------------------------------------------
Image /page/17/Picture/0 description: The image shows the logo for Doctus G. The logo features a stylized letter 'D' with a small house icon inside, suggesting a connection to home or real estate. Below the symbol, the name "Doctus G" is written in a bold, modern font, with the 'G' being slightly larger and in a different color. The tagline "Making Your Happy Smile" is present below the name.
## 11.Conclusion [21 CFR 807.92(b)(3)]
In according with the Federal Food & Drug and cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification, concludes that the Easy On Universal Glaze & Stain is substantially equivalent in safety and effectiveness to the predicate device as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.