DD Contrast

K193588 · Dental Direkt GmbH · EIH · Apr 16, 2020 · Dental

Device Facts

Record IDK193588
Device NameDD Contrast
ApplicantDental Direkt GmbH
Product CodeEIH · Dental
Decision DateApr 16, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.6660
Device ClassClass 2

Indications for Use

The DD contrast® veneering ceramics system is suitable for the esthetic individualization of monolithic or minimally reduced dental crown and bridge constructions made of zirconium dioxide or lithium disilicate.

Device Story

DD contrast is a veneering ceramic system used by dental technicians to individualize dental restorations. The system consists of pre-mixed ceramic pastes and liquids applied to sintered zirconium dioxide or lithium disilicate crown and bridge frameworks. The pastes allow for coloring, texturing, and glazing of the restoration surface. After application, the restoration is fired in a standard dental ceramic furnace. The device enables the creation of thin shade and glaze layers or structure-building areas to achieve desired tooth morphology and esthetics. It is used in a dental laboratory setting to provide prostheses for patients with missing or damaged teeth.

Clinical Evidence

No clinical data. Substantial equivalence is supported by non-clinical bench testing, including biocompatibility (ISO 10993-1, ISO 7405) and physical property testing (ISO 6872).

Technological Characteristics

Feldspar ceramic paste and liquid system. Main components: SiO2, Al2O3, B2O3, K2O, Na2O. Classified as Type 1, class 1b per ISO 6872. Flexural strength ≥ 50 MPa. Applied to zirconia or lithium disilicate frameworks and fired in a standard dental furnace.

Indications for Use

Indicated for the esthetic individualization of monolithic or minimally reduced dental crown and bridge constructions made of zirconium dioxide or lithium disilicate.

Regulatory Classification

Identification

Porcelain powder for clinical use is a device consisting of a mixture of kaolin, felspar, quartz, or other substances intended for use in the production of artificial teeth in fixed or removable dentures, of jacket crowns, facings, and veneers. The device is used in prosthetic dentistry by heating the powder mixture to a high temperature in an oven to produce a hard prosthesis with a glass-like finish.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. April 16, 2020 Dental Direkt GmbH Patrick Berz Regulatory Affairs Manager Industriezentrum 106-108 Spenge, 32139 Germany Re: K193588 Trade/Device Name: DD Contrast Regulation Number: 21 CFR 872.6660 Regulation Name: Porcelain Powder for Clinical Use Regulatory Class: Class II Product Code: EIH Dated: January 15, 2020 Received: January 17, 2020 Dear Patrick Berz: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's {1}------------------------------------------------ requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, for Srinivas "Nandu" Nandkumar, PhD Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) K193588 Device Name DD Contrast Indications for Use (Describe) The DD contrast® veneering ceramics system is suitable for the esthetic individualization of monolithic or minimally reduced dental crown and bridge constructions made of zirconium dioxide or lithium disilicate. | Type of Use (Select one or both, as applicable) | | |-------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------| | <div> <span> <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> | CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*** The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff *PRAStaff@fda.hhs.gov* "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image contains the logo for Dental Direkt. The logo consists of a stylized cross made up of gray and dark gray squares, with one orange square in the center. To the right of the cross are the words "Dental" in light gray and "Direkt" in dark gray, stacked on top of each other. Dental Direkt GmbH I Industriezentrum 106 – 108 I D-32139 Spenge ## 510(k) Summary | Submitter of 510(k) | Dental Direkt GmbH<br>Industriezentrum 106-108<br>32139 Spenge / Germany | |--------------------------------|----------------------------------------------------------------------------------------------------------------------------| | Contact Person | Mr. Uwe Greitens, CEO<br>Phone:+49 5225 86319-0<br>Fax: +49 5225 86319-99<br>u.greitens@dentaldirekt.de | | Regulatory Correspondent | Mr. Patrick Berz, Manager Regulatory Affairs<br>Phone:+49 5225 86319-0<br>Fax: +49 5225 86319-99<br>p.berz@dentaldirekt.de | | Establishment Registration No. | 3008347275 | | Date Prepared | 2020/04/16 | | Trade Name of Device | DD contrast | | Common Name | Powder, Porcelain | | Classification Name | Porcelain Powder for clinical use | | Product Code | EIH | | Regulation Number | 21 CFR 872.6660 | | Classification | Class II | | 510(k) Number | K193588 | {4}------------------------------------------------ | Primary Predicate Device | K111743: Jensen Industries:<br>InSync Ceramic System (Model name: "MiYO – liquid<br>ceramic") | |-------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for Use | The DD contrast® veneering ceramics system is<br>suitable for the esthetic individualization of monolithic<br>or minimally reduced dental crown and bridge<br>constructions made of zirconium dioxide or lithium<br>disilicate. | | Device Description | DD contrast veneering ceramics are dental ceramics<br>for the esthetic individualization of crowns and<br>bridges made from zirconia or lithium disilicate. The<br>DD contrast group consists of various ceramic<br>veneering pastes (already pre-mixed) for coloring,<br>texturing and glazing as well as liquids for adjusting<br>the textures of the paste variants.<br>The DD contrast products are applied to the sintered<br>zirconium dioxide or lithium disilicate restoration and<br>fired in a standard furnace for veneering ceramics.<br>Due to the different consistencies, the pastes can be<br>used to achieve thin shade and glaze layers as well<br>as structure-building areas on the crowns.<br>The veneering ceramics are of type I, class 1b)<br>according to ISO 6872 (FDA Recognition Number 4-<br>251). | | Technological Characteristics | The technological characteristics of the new product<br>are the same as those of the legally marketed device.<br>Both products are feldspar ceramics fired onto a<br>dental ceramic framework made of zirconium dioxide<br>or lithium disilicate. With this veneering technique it is<br>possible to design a crown or bridge to individual<br>patient requirements. | | Material composition | Feldspar ceramics, main components: SiO2, Al2O3,<br>B2O3, K2O, Na2O | {5}------------------------------------------------ | Discussion of<br>Tests Performed | Clinical Tests<br>Dental Direkt GmbH did not conduct, nor rely upon, clinical tests to determine substantial equivalence as dental ceramics that fall under FDA product code EIH have a long history of safe and effective use in the US.<br><br>Non-Clinical Tests<br>Non-clinical testing was performed in order to validate the product against the company's specified design requirements according to the following standards:<br>ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process (FDA Recognition #4-212) ISO 7405, Dentistry - Evaluation of biocompatibility of medical devices used in dentistry (FDA Recognition #2-258) The DD contrast products are proven to be biocompatible. ISO 6872, Dentistry - Ceramic materials (FDA Recognition #4-251) | |----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Substancial<br>Equivalence | Material<br>The composition of DD contrast corresponds to that of the predicate device.<br><br>Physical Properties | DD contrast has physical properties comparable to those of predicates. {6}------------------------------------------------ ## Technical comparison | Feature | Predicate Device | New Device | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Trade name | InSync Ceramic System | DD contrast | | Model name | MiYO Esthetic System Kit | DD contrast | | Indications | | | | Indications for Use<br>Statement | "InSync Ceramic System":<br>The pressable ceramic<br>pellets are pressed onto<br>zirconia frames by dental<br>technicians to fabricate full<br>ceramic crowns and the<br>ceramic layering porcelain<br>and liquids are used to<br>build up the pressed<br>ceramic to final tooth<br>morphology and shade.<br>The ceramic layering<br>porcelain and liquids are<br>also used in building<br>ceramic crowns and<br>bridges on titanium and<br>titanium alloy substructures.<br>Both applications are to<br>provise protheses for<br>missing / damaged teeth.<br>"MiYO Esthetic System Kit":<br>A Type 1, class 1b ceramic<br>for the coloring, enhancing,<br>and glazing of the following<br>restoration types: Zirconia<br>and Lithium Disilicate. | The DD contrast®<br>veneering ceramics<br>system is suitable for the<br>esthetic individualization of<br>monolithic or minimally<br>reduced dental crown and<br>bridge constructions made<br>of zirconium dioxide or<br>lithium disilicate. | | Comment on Indications for Use Statement:<br>The indication of the predicate ("InSync Ceramic<br>System") is comparable to that of the new device, even if<br>it is written in a very general way and includes different<br>veneering ceramics. The relevant parts of the indication<br>describe "layering porcelain and liquids" which are<br>applied to a pressed ceramic veneer which has been<br>previously pressed on a zirconium framework. Even<br>better, however, is the comparability of the corresponding<br>model variant (“MiYO Esthetic System Kit”) resulting from<br>this group, whose indication is identical to that of the new<br>product. | | | | Product for veneering /<br>individualization of dental<br>frameworks | Yes | Yes | {7}------------------------------------------------ | Feature | Predicate Device | New Device | |---------------------------------------|---------------------------------------------|---------------------------------------------| | Framework material | Zirconium dioxide and<br>lithium disilicate | Zirconium dioxide and<br>lithium disilicate | | Minimal amount of fires | 2 | 1 | | For all sizes of dental<br>frameworks | Yes | Yes | | Classification acc. to ISO<br>6872 | Type 1, class 1b | Type 1, class 1b | {8}------------------------------------------------ | Feature | Predicate Device | New Device | |-------------------------------|----------------------------------------------|----------------------------------------------| | Technological Characteristics | | | | Material | Feldspar ceramics | Feldspar ceramics | | Form | Pastes and liquids | Pastes and liquids | | Shades | Different tooth and individualization shades | Different tooth and individualization shades | | Tested according to ISO 6872 | Yes | Yes | | Flexural strength [MPa] | $\ge$ 50 | $\ge$ 50 | ## Substantial Equivalence Conclusion Dental Direkt GmbH believes that the new product DD contrast is substantially equivalent to the predicate device when used as instructed by knowledgeable and trained dental personnel. There is no change of the intended use of the new device and no difference in fundamental scientific technology. The new device is made from the same materials as the predicate device. The new device was verified in accordance with the abovementioned FDA recognized standards. Dental Direkt GmbH therefore believes that DD contrast is substantially equivalent to the legally marketed predicate device.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...