K231210 · Modern Dental Laboratory (Dg) Co., Ltd. · NHA · Mar 5, 2024 · Dental
Device Facts
Record ID
K231210
Device Name
DGA Abutment
Applicant
Modern Dental Laboratory (Dg) Co., Ltd.
Product Code
NHA · Dental
Decision Date
Mar 5, 2024
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3630
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K231210 · Mar 5, 2024
DGA Abutment
Modern Dental Laboratory (Dg) Co., Ltd.
Published clinical literature (Woods, Terry O., Jana G. Delfino, and Sunder Rajan. 'Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices.')
The sponsor used published literature to perform a non-clinical worst-case MRI review to evaluate the safety of metallic devices in the MRI environment, addressing magnetically induced displacement force and torque.
MR compatibility; Literature review; Safety assessment
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices; Literature review / Scientific rationale; Follow-up/Duration: Not applicable; Study Period: Not applicable
Not applicable; Sample Size: Not applicable; Number of Sites: Not applicable
Not applicable for this study
Magnetically induced displacement force and torque
Indications for Use
DGA Abutment is a patient-specific prosthetic component directly connected to the endosseous dental implant (details as below) and is intended for use as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.
Device Story
DGA Abutment is a patient-specific prosthetic component for endosseous dental implants; provides support for single or multiple tooth prostheses. Device consists of titanium (Ti-6Al-4V ELI) abutment and fixation screw. Used in dental clinics; operated by dental professionals. Input: patient-specific anatomical data/implant platform specifications. Transformation: CNC machining of titanium stock to create custom abutment geometry. Output: physical abutment component. Healthcare provider uses output to restore patient chewing function and aesthetics; abutment is screw-retained to implant and prosthesis is cemented to abutment. Benefits: restoration of masticatory function and aesthetic profile for edentulous patients.
Clinical Evidence
Bench testing only. No clinical data. Testing included static and fatigue testing per ISO 14801:2016, biocompatibility testing (cytotoxicity, sensitization, irritation) per ISO 10993, and non-clinical MRI compatibility assessment.
Indicated for partially or fully edentulous patients requiring single or multiple tooth prostheses in the maxilla or mandible, supported by compatible Straumann Tissue Level (ST) dental implants.
Regulatory Classification
Identification
An endosseous dental implant abutment is a premanufactured prosthetic component directly connected to the endosseous dental implant and is intended for use as an aid in prosthetic rehabilitation.
Special Controls
*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 5, 2024
Modern Dental Laboratory (DG) Co., Ltd. Robin Liu Compliance Manager Room 102 & 1102 & Floor 4-10, Block 1 Modern Dental Industrial Park NO. 7 Nantou, Songshan Lake Hi-Tech Industrial Zone Dongguan, Guangdong 523000 CHINA
Re: K231210
Trade/Device Name: DGA Abutment Regulation Number: 21 CFR 872.3630 Regulation Name: Endosseous Dental Implant Abutment Regulatory Class: Class II Product Code: NHA Dated: February 5, 2024 Received: February 5, 2024
Dear Robin Liu:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
{2}------------------------------------------------
3
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Andrew I. Steen -S
Andrew I. Steen Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
Submission Number (if known)
### K231210
Device Name
DGA Abutment
Indications for Use (Describe)
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 5. 510(k) Summary
# 5.1 Submitter Information
| Company Name: | Modern Dental Laboratory (DG) Co.,Ltd. |
|------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Company Address: | Room 102&1102&Floor 4-10,Block 1 Modern Dental<br>Industrial Park, No.7 Nantou, Songshan Lake Hi-Tech<br>Industrial Zone 523000 Dongguan,Guangdong People's<br>Republic of China |
| Company Phone: | 0769-22899211 |
| Contact Person: | Robin Liu |
| Email: | Robin.liu@moderndentallab.com |
| Date Prepared | March 5,2024 |
# 5.2 Device Identification
| Device Model Name: | DGA Abutment |
|----------------------|------------------------------------|
| Classification Name: | Endosseous Dental Implant Abutment |
| Regulation Number: | 872.3630 |
| Product Code: | NHA |
| Class | II |
| Panel | Dental |
# 5.3 Predicated Devices
| Primary Predicate Device: | K213961 |
|---------------------------|------------------|
| Reference Device: | K151295, K142813 |
# 5.4 Device Description
DGA Abutment is endosseous implant abutment which is placed into a corresponding dental implant to provide support for a prosthetic restoration. DGA Abutment is made of titanium grade Ti-6Al-4V ELI (meets ASTM Standard F136). The DGA Abutment is mounted on the implant with a screw. Each abutment is supplied with two identical screws which are used for:
- a. For fixing into the endosseous implant.
- b. For dental laboratory use during fabrication of restoration.
The screw is also made of titanium grade Ti-6AI-4V ELI (meets ASTM Standard F136).
| Brand | System | Platform | Implant body diameter | 510(k) number |
|-----------|----------------------|----------|-----------------------|---------------|
| | Tissue Level<br>(ST) | RN | 4.1mm,4.8mm | K122855 |
| Straumann | | WN | 4.8mm | K122855 |
### DGA Abutment is compatible with the following implant
{5}------------------------------------------------
| Abutment post<br>height (length<br>above collar/<br>gingival height) | Angulation | Abutment<br>diameter | Wall thickness | Gingival height |
|----------------------------------------------------------------------|---------------|----------------------|---------------------|-------------------|
| 4-9 mm<br>(±0.2mm) | 0°<br>(±0.1°) | 4-7 mm<br>(±0.03mm) | ≥0.4mm<br>(±0.05mm) | 0-7mm<br>(±0.2mm) |
The range of design parameters of DGA Abutment please refer to the following sheet.
# 5.5 Indication for Use
DGA Abutment is a patient-specific prosthetic component directly connected to the endosseous dental implant (details as below) and is intended for use as a support for single or multiple tooth prostheses in the maxilla or mandible of a partially or fully edentulous patient.
| Brand | System | Platform | Implant body diameter | 510(k) number |
|-----------|-------------------|----------|-----------------------|---------------|
| Straumann | Tissue Level (ST) | RN | 4.1mm,4.8mm | K122855 |
| | | WN | 4.8mm | K122855 |
# 5.6 Technological Characteristics
- . Performance Specification
The DGA Abutment is designed to serve as a connection component between a dental implant and a dental prosthesis and to remain for indefinitely in the oral cavity. As such it must fulfil the following -
- a. Made with a biocompatible material suitable for permanent mucosal contact.
- Provide an aesthetic profile from the implant up to the dental prosthesis. b.
- Provide sufficient strength to withstand masticatory forces. C.
- . Mechanical Properties
The fatigue testing of the DGA Abutment was conducted in accordance with ISO 14801: 2016 Dentistry -- Implants -- Dynamic loading test for endosseous dental implants and FDA guidance " Class II Special Controls Guidance Document: Rootform Endosseous Dental Implants and Endosseous Dental Implant Abutments", the finished assembled implant,abutment and screw systems were used to tested.
#### . Material Properties
The DGA Abutment is made of titanium grade Ti-6AI-4V which meets ASTM F136, Standard Specification for Wrought Titanium-6Aluminum-4Vanadium ELI Alloy for Surgical Implant Applications.To further confirm finished DGA Abutment's safety, some biocompatibility testing were conducted as per ISO 10993 series of standards, such as cytotoxicity, sensitization and irritation, these tests suggest that DGA Abutment has no cytotoxicity, no sensitization and no irritation and is safety for patient.
{6}------------------------------------------------
#### . Manufacturing Process
The DGA Abutment should be machined with computer numerical control equipment and tungsten carbide cutting tools ,and water soluble coolants should be employed during the machining process. Finished DGA Abutments should be thoroughly rinsed with clean water to remove the coolants.
DGA Abutment is delivered non-sterile for single use and must be cleaned and sterilized prio to use.The sterilization was carried out according to ISO 17665-1:2006. The recommended sterilization method is provided as below:
| Method | Moist heat sterilization |
|---------------|--------------------------|
| Cycle | Pre-vacuum |
| Temperature | 132°C (270° F) |
| Exposure time | 4 min |
| Drying Time | 20 min |
#### ● Mode of Action
The mechanism of action is exactly same with the primary predicate device and reference device, and supports a determination of substantial equivalence.The infrastructure of the DGA Abutment is connected with the implant using screw-retained, and the superstructure of the DGA Abutment is connected with the dental prosthesis by cemented.The DGA Abutment is surgically placed in the maxillary and mandibular arches for supporting the dental restorations to restore patient's chewing function.
#### . MR Compatibility
Non-clinical worst-case MRI review was performed to evaluate the metallic devices in the MRI environment using scientific rationale and published literature (i.e., Woods, Terry O., Jana G. Delfino, and Sunder Rajan. "Assessment of Magnetically Induced Displacement Force and Torque on Metal Alloys Used in Medical Devices." Journal of Testing and Evaluation 49.2 (2019): 783-795), based on the entire system to include all variations (all compatible implant bodies, dental abutments, and fixation screws) and material compositions. The rationale addressed parameters per the FDA Guidance "Testing and Labeling Medical Devices for Safety in the Magnetic Resonance (MR) Environment", including magnetically induced displacement force and torque.
#### Reverse Engineering Analysis ●
Dimensional analysis and reverse engineering of the implant-to-abutment connection platform were performed, including an assessment of maximum and minimum dimensions of critical design aspects, tolerances, and cross-sectional images of the submission device and compatible OEM implant body, OEM abutment, and OEM fixation screw. The testing demonstrated implant to abutment compatibility and has established substantial equivalency of the proposed device with predicate
{7}------------------------------------------------
devices.Static and fatigue testing according to ISO 14801 were performed on worst case scenario. The results of the above tests have met the criteria of the standards and demonstrated the substantial equivalence with the primary predicate.
# 5.7 Risk Management
The risks to health have been identified, identified risks generally associated with the use of the DGA Abutment. The identifying approach of risk is specified as per the standard ISO 14971: 2019 Medical devices - Application of risk management to medical deices and ISO TR 24971: 2020 Medical devices - Guidance on the application of ISO 14971.
{8}------------------------------------------------
# 5.8 Substantial Equivalence Comparison
| Elements | Subject Device | Primary Predicate Device | Reference Device | Reference Device | Explanation |
|----------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | DGA Abutment | TruAbutment DS | Biodenta Customised<br>Abutment -Titanium | Duplex Abutment | |
| 510(k)<br>Number | New device | K213961 | K151295 | K142813 | |
| Manufactur<br>er | Modern Dental<br>Laboratory (DG) Co.,<br>Ltd. | TruAbutment Inc. | Biodenta Swiss AG | Biogenesis Co., Ltd. | |
| Regulation<br>Number | 21 CFR 872.3630 | 21 CFR 872.3630 | 21 CFR 872.3630 | 21 CFR 872.3630 | Same |
| Classificati<br>on Name | Endosseous Dental<br>Implant Abutment | Endosseous Dental Implant<br>Abutment | Endosseous Dental<br>Implant Abutment | Endosseous Dental Implant<br>Abutment | Same |
| Product<br>Code | NHA | NHA | NHA | NHA | Same |
| Device<br>Class | Class II | Class II | Class II | Class II | Same |
| Indication<br>for Use | DGA Abutment is a<br>patient-specific<br>prosthetic component<br>directly connected to<br>the endosseous dental<br>implant (details as<br>below) and is intended<br>for use as a support for | TruAbutment DS is a<br>patient-specific CAD/CAM<br>abutment, directly connected to<br>endosseous dental implants and<br>is intended for use as an aid in<br>prosthetic rehabilitation. It is<br>compatible with the following | The Biodenta<br>Customized Abutment-<br>Titanium is intended for<br>use with dental<br>implants as a support<br>for single or multiple<br>tooth prostheses in the | The Biogenesis Implant<br>System - Kisses is indicated<br>for use in partially or fully<br>edentulous mandibles and<br>maxillae, in support of single<br>or multiple -unit restorations<br>including; cemented retained, | Same. Slightly<br>difference in<br>description, however<br>the content is<br>substantial<br>equivalence. |
| | single or multiple tooth<br>prostheses in the<br>maxilla or mandible of a<br>partially or fully<br>edentulous patient.<br>Brand: Straumann<br>System: Tissue Level<br>(ST)<br>Platform & Implant body<br>diameter: RN (4.1mm,<br>4.8mm), WN (4.8mm)<br>510(k) number:<br>K122855 | systems:<br>• Straumann Tissue Level Implant<br>(K122855, K202942): 4.1(RN),<br>4.8(RN), 6.5(WN) mm | maxilla or mandible of a<br>partially or fully<br>edentulous patient. | screw retained,or<br>overdenture restorations, and<br>terminal or intermediate<br>abutment support for fixed<br>bridgework. The Biogenesis<br>Implant System - Kisses is for<br>single and two stage surgical<br>procedures.It is for delayed<br>loading | The material of subject<br>device is same as that<br>of predicate device<br>and reference device<br>K151295. |
| Abutment<br>and Screw<br>material | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti-6Al-4V ELI | Ti Gr.4 | |
| Abutment<br>fixation | Screw-retained | Screw-retained | Screw-retained | Screw-retained | Same |
| Abutment<br>Angulation | 0°<br>(±0.1°) | Max 25° | Max 30° | straight | the subject's allowed<br>angulation is smaller<br>than that of predicate<br>/reference( K151295)<br>device, and is same as<br>that of reference<br>device K142813. |
| Abutment<br>Diameter | 4-7 mm<br>(±0.03mm) | 5.2-8.0 mm | 4.5-15 mm | 4.0-6.5mm | Belong to patient -<br>specific slightly |
| | | | | | difference in |
| | | | | | parameter in diameter,<br>however it doesn't<br>affect the substantial<br>equivalence. |
| Abutment<br>Post Height<br>(length above collar/<br>gingival height) | 4-9 mm<br>(±0.2mm) | 4.0-7.0 mm | 5.0 - 12.3 mm | 4.0-7.0mm | Belong to patient<br>-specific,slightly<br>difference in<br>parameter in height,<br>however it doesn't<br>affect the substantial<br>equivalence. |
| Wall<br>thickness | Min 0.4mm<br>(±0.05mm) | Min 0.4mm | Unknown | unknown | Same as predicate<br>device |
| | | | | | Belong to patient<br>-specific,slightly |
| Gingival<br>height | 0-7mm<br>(±0.2mm) | Unknown | Unknown | 1.0-7.0mm | difference in<br>parameter in height,<br>however it doesn't<br>affect the substantial<br>equivalence. |
| Installation<br>procedure | 1. The final DGA<br>Abutment is attached to<br>the implant using the<br>corresponding dental<br>screw.<br>2. The dental screw is | a. After osseointegration has<br>been successfully achieved for<br>the fixture, design an appropriate<br>CAD/CAM custom abutment<br>suitable to the oral environment.<br>b. Reproduce the configuration | An abutment screw<br>provided by Biodenta is<br>used to secure the<br>abutment to the<br>implant, whereas the<br>prosthesis is custom | It's not available from public<br>database. | Subject device is<br>substantial<br>equivalence to primary<br>predicate and<br>reference device<br>(K151295) . Slightly |
| | tightened in the | of the oral cavity of the patient | made and | | difference in |
| | recommended torque | through impression and send the | cement-retained on the | | description, however |
| | 3. In order to obtain the | impression to a dental laboratory | abutment. The | | the content is |
| | recommended torque, a | to fabricate the final prosthesis. | abutment screw should | | substantial |
| | dental torque wrench | Once the final prosthesis | be tightened with a | | equivalence. |
| | with a suitable screw | fabrication is completed, deliver it | manual torque wrench | | |
| | driver must be used in | to the patient to improve their | according to the | | |
| | accordance with the | masticating and aesthetic | implant manufacturer's | | |
| | relevant manufacturer's | functions. | specifications. | | |
| | instructions | When affixing an abutment | | | |
| | 4. The screw channel | onto a fixture, the recommended | | | |
| | must always be sealed | torque value is equivalent to the | | | |
| | after the abutment is | manufacturers' value for the | | | |
| | attached to the implant. | respective implant system. The | | | |
| | | torque value should be | | | |
| | | determined based on the clinical | | | |
| | | assessment of the bone quality, | | | |
| | | fixture dimensions, and | | | |
| | | prosthesis type, etc. | | | |
| CAD/CAM<br>Processing | Milled at Modern<br>Dental Laboratory<br>under QSR control. | All digitally designed abutments<br>and/or coping for use with the<br>TruAbutment DS abutments are<br>intended to be sent to a<br>TruAbutment-validated milling<br>center for manufacture. | Milled in Biodenta<br>milling center under<br>QSR control. | It's not available from public<br>database. | Subject device is<br>substantial<br>equivalence to primary<br>predicate and<br>reference device<br>(K151295) |
| Compatible<br>Implant | Straumann Tissue | Straumann Tissue Level Implant | Dentsply, AstraTech<br>OsseoSpeed | Biogenesis Implant System -<br>Kisses | Subject device and<br>primary predicate |
| System | Level Implant | (K122855, K202942): 4.1(RN), | Biomet 3i, Certain | device both are | |
| | (K122855): 4.1(RN), | 4.8(RN), 6.5(WN) mm | Internal | compatible with | |
| | 4.8(RN) ,4.8(WN) mm | | Zimmer, Screw Vent | Straumann Tissue | |
| | | | Nobel Biocare,Nobel<br>Replace<br>Nobel Biocare,Nobel<br>Active<br>Straumann, Bone Level<br>BioHorizons,Internal<br>Osstem, GS/TS<br>Biodenta, Bone Level<br>and Tapered | Level Implant | |
| Sterilizatio<br>n | Non-sterile | Non-sterile | Non-Sterile | Same | |
| | | | Non-Sterile | | |
Page 5 of 10
{9}------------------------------------------------
Page 6 of 10
{10}------------------------------------------------
Page 7 of 10
{11}------------------------------------------------
Page 8 of 10
{12}------------------------------------------------
{13}------------------------------------------------
# 5.9 Substantial Equivalence Conclusion
The DGA Abutment is substantially equivalent to the primary predicate device and the reference devices in terms of indication for use, technical characteristics and function. Although there are slightly differences in size parameter of devices, these differences are very minor not affecting substantial equivalence.
Therefore, the subject device, the predicate device and the reference devices are substantial equivalence.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.