K231012 · Bolder Surgical, LLC · GEI · Nov 15, 2023 · General, Plastic Surgery
Device Facts
Record ID
K231012
Device Name
CoolSeal Trinity
Applicant
Bolder Surgical, LLC
Product Code
GEI · General, Plastic Surgery
Decision Date
Nov 15, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The CoolSeal™ Trinity is a bipolar electrosurgical instrument intended for use in minimally invasive or open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The CoolSeal™ Trinity can be used on vessels (arteries, veins, and vascular bundles) up to and including 7 mm in diameter. It is indicated for use in general surgery procedures including urologic, vascular, and gynecologic. It is indicated for use in adult and pediatric populations (infants, children, and adolescents). Procedures may include, but are not limited to, Nissen fundoplication, colectomy, cholecystectomy, adhesiolysis, hysterectomy, oophorectomy, etc. The CoolSeal™ Trinity has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the CoolSeal™ Trinity for these procedures. The device is contraindicated for use in ENT procedures.
Device Story
Bipolar electrosurgical instrument; 5 mm diameter shaft; 30/37/44 cm lengths. Used in OR by surgeons for vessel/tissue bundle ligation and division. Input: RF energy from compatible generator (CoolSeal or Da Vinci E-200). Operation: Surgeon grasps tissue between double-action jaws; activates RF energy via thumb button to create seal; manually actuates trigger to deploy blade for tissue division. Output: Sealed and divided tissue. Benefits: Enables hemostasis and tissue separation in minimally invasive/open procedures. Single-use, sterile, disposable.
Clinical Evidence
No clinical data. Bench testing only. Performance supported by electrical/thermal testing and ex-vivo burst pressure testing on porcine vessels to demonstrate equivalent sealing performance to the predicate.
Technological Characteristics
Bipolar electrosurgical instrument; 5 mm shaft; 190 Vpeak. Materials: stainless steel, titanium, polymers, silicone, polymer adhesives, insulating coatings. Energy: RF via compatible generator. Connectivity: wired to generator. Sterilization: Ethylene Oxide (SAL 10^-6). Form factor: Maryland-style laparoscopic sealer/divider with >360° shaft rotation.
Indications for Use
Indicated for ligation and division of vessels, tissue bundles, and lymphatics up to 7 mm in diameter in minimally invasive or open general, urologic, vascular, and gynecologic surgeries. Applicable to adult and pediatric populations (infants, children, adolescents). Contraindicated for ENT procedures and tubal sterilization/coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 15, 2023
Bolder Surgical, LLC Nicholas Wong Sr. Manager, Regulatory Affairs 331 S. 104th Street Suite 200 Louisville, Colorado 80027
Re: K231012
Trade/Device Name: CoolSeal Trinity Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting And Coagulation Device And Accessories Regulatory Class: Class II Product Code: GEI Dated: October 20, 2023 Received: October 24, 2023
Dear Nicholas Wong:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device"
{1}------------------------------------------------
(https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30. Design controls; 21 CFR 820.90. Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review. the OS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Mark
Digitally signed by
Mark Trumbore -S
Trumbore -S Date: 2023.11.15
13:42:02-05'00'
Mark Trumbore, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices
{2}------------------------------------------------
Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
See PRA Statement below.
510(k) Number (if known)
Device Name CoolSeal Trinity
#### Indications for Use (Describe)
The CoolSeal™ Trinity is a bipolar electrosurgical instrument intended for use in minimally invasive or open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The CoolSeal™ Trinity can be used on vessels (arteries, veins, and vascular bundles) up to and including 7 mm in diameter. It is indicated for use in general surgery procedures including urologic, vascular, and gynecologic. It is indicated for use in adult and pediatric populations (infants, children, and adolescents). Procedures may include, but are not limited to, Nissen fundoplication, colectomy, cholecystectomy, adhesioly, oophorectomy, etc. The CoolSeal™ Trinity has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the CoolSeal™ Trinity for these procedures. The device is contraindicated for use in ENT procedures.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
### 510(k) Summary
### SUBMITTER
Bolder Surgical, LLC. (now part of Hologic, Inc.) 331 S. 104th Street, Suite 200 Louisville, CO 80027
Phone: 720-287-7130 Fax: 720-287-7135
Contact Person: Nick Wong Sr. Manager, Regulatory Affairs
Date Prepared: April 5th, 2023
### DEVICES
Trade Name / Model #:
CoolSealTM Trinity
- . 30 cm / CSL-TR105-30
- 37 cm / CSL-TR105-37 ●
- 44 cm / CSL -TR105-44 .
| Common or Usual Name: | Bipolar Vessel Sealing System |
|-----------------------|----------------------------------------------------------------|
| Classification Name: | Electrosurgical cutting and coagulation device and accessories |
| Regulatory Class: | II |
| Produce Code: | GEI |
### PREDICATE DEVICE
Trade Name: CoolSeal™ Trinity 510(k): K211579
### DEVICE DESCRIPTION
#### CoolSeal™ Trinity:
The CoolSeal™ Trinity, a Maryland Laparoscopic Sealer, Divider, and Dissector, with a 5 mm diameter shaft, is designed for use with the CoolSeal™ Generator or any generator with the CoolSeal™ technology. The Trinity is provided sterile and is a single-use disposable instrument. The Trinity creates seals by application of radiofrequency (RF) electrosurgical energy to vascular structures (vessels and lymphatics) or tissue bundles interposed between its jaws. A blade within the instrument is surgeon-actuated to divide tissue. The double-action jaws have been designed to dissect tissue, which includes separating tissue planes and widening openings as necessary for the surgical procedure. The CoolSeal™ Trinity includes 3 shaft lengths: 30 cm, 37 cm, and 44 cm.
# INDICATIONS FOR USE
CoolSeal™ Trinity:
{5}------------------------------------------------
The CoolSeal™ Trinity is a bipolar electrosurgical instrument intended for use in minimally invasive or open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The CoolSeal™ Trinity can be used on vessels (arteries, veins, and vascular bundles) up to and including 7 mm in diameter. It is indicated for use in general surgery procedures including urologic, vascular, and gynecologic. It is indicated for use in adult and pediatric populations (infants, children, and adolescents). Procedures may include, but are not limited to. Nissen fundoplication, colectomy, cholecystectomy, adhesiolysis, hysterectomy, oophorectomy, etc. The CoolSeal™ Trinity has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the CoolSeal™ Trinity for these procedures. The device is contraindicated for use in ENT procedures.
### COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICES
Table 1 below compares technological characteristics between the subject and predicate devices. As Table 1 illustrates, all characteristics are the same, with the only difference being the identification of a new compatible electrosurgical generator.
| Description | CoolSeal™ Trinity<br>(Subject Device) | CoolSeal™ Trinity<br>(Predicate - K211579) |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------|
| Indications for Use | The CoolSeal™ Trinity is a bipolar electrosurgical instrument intended for use<br>in minimally invasive or open surgical procedures where ligation and division of<br>vessels, tissue bundles, and lymphatics is desired. The CoolSeal™ Trinity can be<br>used on vessels (arteries, veins, and vascular bundles) up to and including 7 mm<br>in diameter. It is indicated for use in general surgery procedures including<br>urologic, vascular, and gynecologic. It is indicated for use in adult and pediatric<br>populations (infants, children, and adolescents). Procedures may include, but are<br>not limited to, Nissen fundoplication, colectomy, cholecystectomy, adhesiolysis,<br>hysterectomy, oophorectomy, etc.<br>The CoolSeal™ Trinity has not been shown to be effective for tubal sterilization<br>or tubal coagulation for sterilization procedures. Do not use the CoolSeal™<br>Trinity for these procedures.<br>The device is contraindicated for use in ENT procedures. | |
| Where used<br>(environment) | Operating Room | |
| Intended User | Surgeons | |
| Anatomical Sites | Vessels, tissue bundles, and lymphatics | |
| Anatomical Size | Arteries, veins, and vascular bundles up to and including 7 mm in diameter | |
| Patient Population | Adult and pediatric populations (infants, children, and adolescents). | |
| Power Source | Bipolar energy platform | |
| Compatible<br>Electrosurgical<br>Generator | CoolSeal Generator<br>Da Vinci E-200 | CoolSeal Generator |
| Primary Functions | Grasp, Dissect, Seal, Divide | |
Table 1. Presents the subject device compared to the predicate device.
{6}------------------------------------------------
| Description | CoolSeal™ Trinity<br>(Subject Device) | CoolSeal™ Trinity<br>(Predicate - K211579) |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------|--------------------------------------------|
| Mechanism of<br>Grasping | Hand actuated lever allows user to open or close | |
| Mechanism of<br>Dissection<br>(Separation) | Bilateral jaw allows the user to separate planes of tissue | |
| Mechanism of<br>Action (Sealing) | Seal is created by application of RF energy to structures interposed between the<br>jaws of the instrument. | |
| Seal Activation | Button on the sealer instrument handle activated by thumb | |
| Knife Activation | Cutting trigger – non-energized | |
| Automatic sealing<br>cycle | Yes | |
| Rated Voltage<br>(Vpeak) | 190 Vpeak | |
| Shaft Diameter | 5 mm | |
| Shaft Length | 30 cm, 37 cm, 44 cm | |
| Shaft Rotation | >360° | |
| Seal Length | 19 mm | |
| How Supplied | Single-use disposable | |
| Tissue Contact<br>Materials | Stainless steel, titanium, polymers, silicone, polymer adhesives, and insulating<br>coatings | |
| Surgical Approach | Open or laparoscopic | |
| Sterilization | Ethylene Oxide | |
| Sterility Assurance<br>Level | 10-6 | |
# PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## Electrical and Thermal Testing
Electrical and thermal testing verified that the proposed subject device performed as expected when connected to the E-200 generator.
## Ex-vivo Vessel Burst Pressure
### CoolSeal™ Trinity:
Ex-vivo burst pressure testing of excised fresh porcine blood vessels was conducted on the subject device to demonstrate effective bipolar electrosurgical vessel sealing performance that is equivalent to the predicate device.
## CONCLUSIONS
Based on a review of performance data, comparison of the device classification, intended use, operating principles, and technological characteristics, the subject device is safe, as effective, and performs as well as the legally marketed predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.