CoolSeal Trinity (30 cm shaft), CoolSeal Trinity (37 cm shaft), CoolSeal Trinity (44 cm shaft)
K211579 · Bolder Surgical, LLC · GEI · Sep 7, 2021 · General, Plastic Surgery
Device Facts
Record ID
K211579
Device Name
CoolSeal Trinity (30 cm shaft), CoolSeal Trinity (37 cm shaft), CoolSeal Trinity (44 cm shaft)
Applicant
Bolder Surgical, LLC
Product Code
GEI · General, Plastic Surgery
Decision Date
Sep 7, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The CoolSeal™ Trinity is a bipolar electrosurgical instrument intended for use in minimally invasive or open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The CoolSeal™ Trinity can be used on vessels (arteries, veins, and vascular bundles) up to and including 7 mm in diameter. It is indicated for use in general surgery procedures including urologic, vascular, and gynecologic. It is indicated for use in adult and pediatric populations (infants, children, and adolescents). Procedures may include, but are not limited to, Nissen fundoplication, colectomy, cholecystectomy, adhesiolysis, hysterectomy, oophorectomy, etc. The CoolSeal™ Trinity has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the CoolSeal™ Trinity for these procedures. The device is contraindicated for use in ENT procedures.
Device Story
CoolSeal Trinity is a bipolar electrosurgical instrument for vessel sealing and division. Input: vascular structures or tissue bundles interposed between device jaws. Operation: surgeon-actuated hand lever closes jaws to grasp tissue; RF energy from CoolSeal generator creates seal; surgeon-actuated trigger deploys blade to divide tissue. Used in OR by surgeons for minimally invasive or open procedures. Output: sealed and divided tissue. Benefits: provides hemostasis in various surgical procedures. Device features 5 mm diameter shaft in 30 cm, 37 cm, and 44 cm lengths. Bilateral jaw design facilitates tissue dissection and plane separation.
Clinical Evidence
No human clinical data. Evidence includes: ex-vivo porcine vessel burst pressure testing; in-vivo porcine model thermal spread comparison (histological assessment of seal dimensions); and chronic animal study (21-day survival) confirming hemostasis and healing. Biocompatibility per ISO 10993-1. Electrical safety/EMC per ANSI/AAMI/IEC 60601-1, 60601-2-2, and 60601-1-2.
Indicated for ligation and division of vessels (arteries, veins, vascular bundles up to 7 mm), tissue bundles, and lymphatics in minimally invasive or open general surgery (urologic, vascular, gynecologic). Applicable to adult and pediatric populations (infants, children, adolescents). Contraindicated for ENT procedures and not indicated for tubal sterilization/coagulation.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 7, 2021
Bolder Surgical, LLC. Nicholas Wong Regulatory Affairs Manager 331 S. 104th Street, Suite 200 Louisville. Colorado 80027
## Re: K211579
Trade/Device Name: CoolSeal Trinity (30 cm shaft), CoolSeal Trinity (37 cm shaft), CoolSeal Trinity (44 cm shaft) Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical Cutting and Coagulation Device and Accessories Regulatory Class: Class II Product Code: GEI Dated: August 6, 2021 Received: August 9, 2021
Dear Nicholas Wong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4. Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
4-2
510(k) Number (if known) K211579
Device Name CoolSeal Trinity
### Indications for Use (Describe)
The CoolSeal™ Trinity is a bipolar electrosurgical instrument intended for use in minimally invasive or open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The CoolSeal™ Trinity can be used on vessels (arteries, veins, and vascular bundles) up to and including 7 mm in diameter. It is indicated for use in general surgery procedures including urologic, vascular, and gynecologic. It is indicated for use in adult and pediatric populations (infants, children, and adolescents). Procedures may include, but are not limited to, Nissen fundoplication, colectomy, cholecystectomy, adhesiolysis, hysterectomy, etc. The CoolSeal™ Trinity has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the CoolSeal™ Trinity for these procedures. The device is contraindicated for use in ENT procedures.
Type of Use (Select one or both, as applicable)
| <div> <span style="font-size:10pt"> <b> ☑ </b> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------------------|
| <div> <span style="font-size:10pt"> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the text 'K211579'. The text is in a sans-serif font and is black. The background is white.
Image /page/3/Picture/1 description: The image shows the logo for Bolder Surgical. The word "BOLDER" is written in large, bold, blue letters. The letter "O" in "BOLDER" has a small, black, diamond shape inside of it. Below the word "BOLDER" is the word "SURGICAL" written in smaller, thinner, gray letters.
## 510(k) Summarv
## SUBMITTER
Bolder Surgical, LLC. 331 S. 104th Street, Suite 200 Louisville, CO 80027
Phone: 720-287-7130 Fax: 720-287-7135
Contact Person: Nick Wong Regulatory Affairs Manager
Date Prepared: May 19th, 2021
## DEVICES
Trade Name / Model #:
CoolSealTM Trinity
- o 30 cm / CSL-TR105-30
- 37 cm / CSL-TR105-37
- 44 cm / CSL-TR105-44
| Common or Usual Name: | Bipolar Vessel Sealing System |
|-----------------------|----------------------------------------------------------------|
| Classification Name: | Electrosurgical cutting and coagulation device and accessories |
| Regulatory Class: | II |
| Produce Code: | GEI |
## PREDICATE DEVICES
Trade Name: CoolSealTM Trinity 510(k): K203183
Trade Name: LigaSure™ Maryland 510(k) K170869
## DEVICE DESCRIPTION
## CoolSeal™ Trinity:
The CoolSeal™ Trinity, a Maryland Laparoscopic Sealer, Divider, and Dissector, with a 5 mm diameter shaft is designed for use with the CoolSeal™ Generator or any generator with the CoolSeal™ technology. The Trinity is provided sterile and is a single-use disposable instrument. The Trinity creates seals by application of radiofrequency (RF) electrosurgical energy to vascular structures (vessels and lymphatics) or tissue bundles interposed between its jaws. A blade within the instrument is surgeon-actuated to divide tissue. The double action jaws have been designed to dissect tissue, which includes separating tissue planes and widening openings as necessary for the surgical procedure. The CoolSeal™ Trinity includes 3 different shaft lengths: 30 cm, 37 cm, and 44 cm.
{4}------------------------------------------------
# INDICATIONS FOR USE
# CoolSeal™ Trinity:
The CoolSeal™ Trinity is a bipolar electrosurgical instrument intended for use in minimally invasive or open surgical procedures where ligation and division of vessels, tissue bundles, and lymphatics is desired. The CoolSeal™ Trinity can be used on vessels (arteries, veins, and vascular bundles) up to and including 7 mm in diameter. It is indicated for use in general surgery procedures including urologic, vascular, and gynecologic. It is indicated for use in adult and pediatric populations (infants, children, and adolescents). Procedures may include, but are not limited to, Nissen fundoplication, colectomy, cholecystectomy, adhesiolysis, hysterectomy, oophorectomy, etc. The CoolSeal™ Trinity has not been shown to be effective for tubal sterilization or tubal coagulation for sterilization procedures. Do not use the CoolSeal™ Trinity for these procedures. The device is contraindicated for use in ENT procedures.
{5}------------------------------------------------
ਨੀ ਹੋਵੀ (
OMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATED DEVICES
able 1 below presents the comparison of technological characteristics between the two identified predic
| Table 1. presents the subject Device compared to its two predicates, CoolSeal Trinity (K203183) and LigaSure Maryland (K170869)Description | CoolSeal™ Trinity<br>(Subject Device) | CoolSeal™ Trinity<br>(Primary Predicate/K203183) | LigaSure Maryland (Secondary Predicate/K170869) |
|--------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The CoolSeal™ Trinity is a bipolar<br>electrosurgical instrument intended for<br>use in minimally invasive or open<br>surgical procedures where ligation and<br>division of vessels, tissue bundles, and<br>lymphatics is desired. The CoolSeal™<br>Trinity can be used on vessels (arteries,<br>veins, and vascular bundles) up to and<br>including 7 mm in diameter. It is<br>indicated for use in general surgery<br>procedures including urologic, vascular,<br>and gynecologic. It is indicated for use in<br>adult and pediatric populations (infants,<br>children, and adolescents). Procedures<br>may include, but are not limited to,<br>Nissen fundoplication, colectomy,<br>cholecystectomy, adhesiolysis,<br>hysterectomy, oophorectomy, etc. | The CoolSeal™ Trinity is a bipolar<br>electrosurgical instrument intended for<br>use in minimally invasive or open<br>surgical procedures where ligation and<br>division of vessels, tissue bundles, and<br>lymphatics is desired. The CoolSeal™<br>Trinity can be used on arteries up to and<br>including 6 mm, veins, and vascular<br>bundles up to and including 7 mm in<br>diameter. It is indicated for use in general<br>surgery procedures including urologic,<br>vascular, and gynecologic. It is indicated<br>for use in adult and pediatric populations<br>(infants, children, and adolescents). | The LigaSure Sealer/Divider is a bipolar<br>electrosurgical instrument intended for<br>use in minimally invasive or open<br>surgical procedures where ligation and<br>division of vessels, tissue bundles, and<br>lymphatics is desired. The LigaSure<br>Sealer/ Divider can be used on vessels<br>(arteries, veins, and vascular bundles) up<br>to and including 7 mm in diameter. It is<br>indicated for use in general surgery<br>procedures and such surgical specialties<br>as urologic, vascular, thoracic, and<br>gynecologic. Procedures may include, but<br>are not limited to, Nissen fundoplication,<br>colectomy, cholecystectomy,<br>adhesiolysis, hysterectomy,<br>oophorectomy, etc. |
| | The CoolSeal™ Trinity has not been<br>shown to be effective for tubal<br>sterilization or tubal coagulation for<br>sterilization procedures. Do not use the<br>CoolSeal™ Trinity for these procedures.<br>The device is contraindicated for use in<br>ENT procedures. | The CoolSeal™ Trinity has not been<br>shown to be effective for tubal<br>sterilization or tubal coagulation for<br>sterilization procedures. Do not use the<br>CoolSeal™ Trinity for these procedures.<br>The device is contraindicated for use in<br>ENT procedures. | The LigaSure system has not been shown<br>to be effective for tubal sterilization or<br>tubal coagulation for sterilization<br>procedures. Do not use the LigaSure<br>system for these procedures. |
| Where used<br>(environment) | Operating Room | | Operating Room |
{6}------------------------------------------------
| ਖੱ | |
|--------|-----|
| பப | C |
| 1<br>l | |
| | 기 이 |
| 0- | |
| | |
| a o | |
| Description | CoolSeal™ Trinity<br>(Subject Device) | CoolSeal™ Trinity<br>(Primary Predicate/K203183) | LigaSure Maryland (Secondary<br>Predicate/K170869) |
|--------------------------------------------|-------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------|
| Intended User | Surgeons | Surgeons | Surgeons |
| Anatomical<br>Sites | Vessels, tissue bundles, and lymphatics | Vessels, tissue bundles, and lymphatics | Vessels, tissue bundles, and lymphatics |
| Anatomical Size | Arteries, veins, and vascular bundles up<br>to and including 7 mm in diameter | Arteries up to and including 6 mm,<br>veins, and vascular bundles up to and<br>including 7 mm in diameter | Arteries, veins, and vascular bundles up<br>to and including 7 mm in diameter |
| Patient<br>Population | Adult and pediatric populations (infants, children, and adolescents). | | Not specified |
| Power Source | Bipolar energy platform | Bipolar energy platform | Bipolar energy platform |
| Primary<br>Functions | Grasp, Dissect, Seal, Divide | Grasp, Dissect, Seal, Divide | Grasp, Dissect, Seal, Divide |
| Mechanism of<br>Grasping | Hand actuated lever allows user to open or close | Hand actuated lever allows user to open or close | Hand actuated lever allows user to open<br>or close |
| Mechanism of<br>Dissection<br>(Separation) | Bilateral jaw allows the user to separate planes of tissue | Unilateral jaw allows the user to separate planes of tissue | Unilateral jaw allows the user to separate<br>planes of tissue |
| Mechanism of<br>Action (Sealing) | Seal is created by application of RF energy to structures interposed between the jaws<br>of the instrument. | Seal is created by application of RF energy to structures interposed between the jaws<br>of the instrument. | Seal is created by the application of RF<br>energy to structures interposed between<br>the jaws of the instrument. |
| Seal Activation | Button on the sealer instrument handle activated by thumb | Button on the sealer instrument handle activated by thumb | Button on the sealer instrument handle<br>activated by full compression of the lever<br>or optional footswitch pedal |
| Knife Activation | Cutting trigger – non-energized | Cutting trigger – non-energized | Cutting trigger – non-energized |
| Automatic<br>sealing cycle | Yes | Yes | Yes |
| Rated Voltage<br>(Vpeak) | 190 Vpeak | | 288 Vpeak |
| Shaft Diameter | 5 mm | 5 mm | 5 mm |
| Shaft Length | 30 cm, 37 cm, 44 cm | 30 cm, 37 cm, 44 cm | 23 cm, 37 cm, 44 cm |
| Description | CoolSeal™ Trinity<br>(Subject Device) | CoolSeal™ Trinity<br>(Primary Predicate/K203183) | LigaSure Maryland (Secondary<br>Predicate/K170869) |
| Shaft Rotation | | >360° | 159°-359° |
| Seal Length | | 19 mm | 20 mm |
| How Supplied | | Single-use disposable | Single-use disposable |
| Tissue Contact<br>Materials | | Stainless steel, titanium, polymers, silicone, polymer adhesives, and insulating<br>coatings | Stainless steel, titanium, polymers,<br>silicone, polymer adhesives, and<br>insulating coatings |
| Surgical<br>Approach | | Open or laparoscopic | Open or laparoscopic |
| Sterilization | | Ethylene Oxide | Ethylene Oxide |
| Sterility<br>Assurance Level | | 10-6 | 10-6 |
{7}------------------------------------------------
# BOLDER
{8}------------------------------------------------
Image /page/8/Picture/2 description: The image shows the logo for Bolder Surgical. The word "BOLDER" is in large, bold, blue letters, with a stylized "O" that has a black triangle in the upper left quadrant. Below "BOLDER" is the word "SURGICAL" in smaller, thinner, blue letters.
# PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
# Biocompatibility Testing
CoolSeal™ Trinity:
The biocompatibility evaluation for the CoolSeal™ Trinity was conducted in accordance with ISO 10993-1, Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process," as recognized by FDA.
# Electrical Safety and Electromagnetic Compatibility (EMC)
The CoolSeal™ system complies with relevant clauses of the ANSI/AAMI/IEC 60601-1 and ANSI/AAMI/IEC 60601-2-2 standards for electrical safety and ANSI/AAMI/IEC 60601-1-2 standard for EMC.
# Mechanical and Functional Testing
Mechanical, electrical, and functional testing was carried out to verify that the proposed device performed as expected.
# Ex-vivo and In-vivo Vessel Burst Pressure
CoolSeal™ Trinity:
Ex-vivo burst pressure testing of excised fresh porcine blood vessels was conducted on the subject device to demonstrate effective bipolar electrosurgical vessel sealing performance.
## In-vivo Thermal Spread Comparison
## CoolSeal™ Trinity:
Open laparotomy was performed in a porcine model using the subject device and predicate device. Each device was used to seal vessels and tissue bundles. Samples were excised for threedimensional histological assessments (e.g., length, width, and depth) to quantify thermal spread in seals created by both devices. This study demonstrated that the subject device is as safe and effective as the predicate device.
## Chronic Animal Study
## CoolSeal™ Trinity:
Chronic animal studies were conducted to assess safety and performance of bipolar vessel sealing with the subject device over the minimum 21-day survival period. All animals survived without any complications. All vessel sealing effects on tissue maintained chronic hemostasis and healed as anticipated.
## CONCLUSIONS
Based on a review of performance data, comparison of the device classification, intended use, operating principles, technological characteristics, sterility, and biocompatibility, the subject device is safe, as effective, and performs as well as the legally marketed predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.