K230989 · Nu Vasive, Incorporated · OLO · May 5, 2023 · Neurology
Device Facts
Record ID
K230989
Device Name
Rod Registration Frame
Applicant
Nu Vasive, Incorporated
Product Code
OLO · Neurology
Decision Date
May 5, 2023
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.4560
Device Class
Class 2
Indications for Use
The surgical instruments are specifically designed for use with the Pulse System, which enables computer-assisted navigation for spinal surgical procedures in which the use of stereotactic surgery may be appropriate. When used with a Pulse Navigation system, the patient reference hardware is intended to provide a reference to a rigid anatomical structure that can be identified relative to the acquired image of the anatomy.
Device Story
Rod Registration Frame is patient reference hardware for Pulse Navigation system; used in spinal surgery. Device provides rigid anatomical reference by attaching to NuVasive pedicle screw systems via bent rod component. Surgeon inserts rod through screw system; attaches Patient Reference Array via poker chip proximal connection. Enables non-fiducial registration of patient anatomy to pre-acquired images. Used in OR by surgeons to facilitate computer-assisted navigation; improves accuracy of surgical instrument placement relative to patient anatomy. No radiographic markers present. Device is non-sterile.
Clinical Evidence
No clinical studies were conducted. Substantial equivalence is supported by non-clinical bench testing, including system-level accuracy testing, cadaver verification, design validation, tolerance analysis, and usability validation, using the same methodologies and acceptance criteria as the predicate device (K210574).
Technological Characteristics
Material: 17-4 PH Stainless Steel (ASTM A564). Form factor: Bent rod with distal fixation for pedicle screw systems and proximal poker chip connection for Patient Reference Array. Non-sterile. Non-fiducial registration only. No radiographic markers.
Indications for Use
Indicated for use with the Pulse System for computer-assisted navigation during spinal surgical procedures where stereotactic surgery is appropriate. Intended for patients undergoing spinal surgery requiring rigid anatomical reference for navigation.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
May 5, 2023
NuVasive, Incorporated Leah Andre Regulatory Affairs Specialist 7475 Lusk Blvd San Diego, California 92121
Re: K230989
Trade/Device Name: Rod Registration Frame Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: April 5, 2023 Received: April 6, 2023
Dear Leah Andre:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali-S
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair And Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230989
Device Name Rod Registration Frame
Indications for Use (Describe)
The surgical instruments are specifically designed for use with the Pulse System, which enables computer-assisted navigation for spinal surgical procedures in which the use of stereotactic surgery may be appropriate.
When used with a Pulse Navigation system, the patient reference hardware is intended to provide a reference to a rigid anatomical structure that can be identified relative to the acquired image of the anatomy.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| <div> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the Nuvasive logo. The logo consists of a stylized leaf shape in purple and gray on the left, followed by the word "NUVASIVE" in gray. The leaf shape is partially purple and partially gray, with the purple portion on the upper left and the gray portion on the lower left. The word "NUVASIVE" is in a sans-serif font and is slightly blurred.
## 510k Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807 and in particular 21 CFR §807.87, the following summary of information is provided:
#### A. Submitted by:
Leah Andre Regulatory Affairs Specialist NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (865) 724-9780 Date Prepared: April 5, 2023
#### B. Device Name
| Trade or Proprietary Name: | NuVasive® Rod Registration Frame |
|----------------------------|----------------------------------|
| Common Name: | Stereotaxic Instrument |
| Classification Name: | Stereotaxic Instrument |
| Device Class: | Class II |
| Regulation Number: | 21 CFR § 888.4560 |
| Product Code: | OLO |
#### C. Predicate Devices
The subject NuVasive Rod Registration Frame is substantially equivalent to the primary predicate device NuVasive Spinous Process Clamp cleared in K210574. Additional reference device includes NuVasive Hip Pin (K210574).
#### D. Device Description
The subject NuVasive Rod Registration Frame is introducing modifications based on the design of existing patient reference hardware, the Spinous Process Clamp (Pulse System K210574). The Rod Registration Frame was designed to offer surgeons more flexibility during the registration process of Pulse Navigation by providing an additional patient reference hardware design option for attaching the existing Patient Reference Array to the patient. Despite the changes introduced to predicate Spinous Process Clamp (K210574), the subject device Rod Registration Frame is substantially equivalent to the predicate as demonstrated by verification and validation testing performed using well established and previously cleared test methods.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Nuvasive. The logo consists of a stylized leaf shape in purple and gray, followed by the word "NUVASIVE" in gray. The leaf shape is partially purple and partially gray, with the purple portion on top and the gray portion on the bottom.
#### E. Indications for Use
The surgical instruments are specifically designed for use with the Pulse System, which enables computer-assisted navigation for spinal surgical procedures in which the use of stereotactic surgery may be appropriate.
When used with a Pulse Navigation system, the patient reference hardware is intended to provide a reference to a rigid anatomical structure that can be identified relative to the acquired image of the anatomy.
#### Technological Comparison F.
The introduction of design change in patient fixation hardware introduces updated distal and proximal connection features and removal of radiographic markers (as the Rod Registration Frame is only used for non-fiducial registration).
The following table describes the summary comparison of technological characteristics of the subject device with the predicate devices:
| Characteristics | Subject Device | K210574 Predicate Device | K210574 Reference Device |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The surgical instruments are<br>specifically designed for use<br>with the Pulse System, which<br>enables computer-assisted<br>navigation for spinal surgical<br>procedures in which the use<br>of stereotactic surgery may<br>be appropriate. | The surgical instruments are<br>specifically designed for use<br>with the Pulse System, which<br>enables computer-assisted<br>navigation for spinal surgical<br>procedures in which the use<br>of stereotactic surgery may<br>be appropriate. | The surgical instruments are<br>specifically designed for use<br>with the Pulse System, which<br>enables computer-assisted<br>navigation for spinal surgical<br>procedures in which the use<br>of stereotactic surgery may<br>be appropriate. |
| | When used with a Pulse<br>Navigation system, the patient<br>reference hardware is intended<br>to provide a reference to a rigid<br>anatomical structure that can be<br>identified relative to the<br>acquired image of the anatomy. | When used with a Pulse<br>Navigation system, the patient<br>reference hardware is intended<br>to provide a reference to a rigid<br>anatomical structure that can be<br>identified relative to the<br>acquired image of the anatomy. | When used with a Pulse<br>Navigation system, the patient<br>reference hardware is intended<br>to provide a reference to a rigid<br>anatomical structure that can be<br>identified relative to the<br>acquired image of the anatomy. |
| Product Code | OLO | OLO | OLO |
| 21 CFR<br>Classification | §882.4560 | §882.4560 | §882.4560 |
| Material | 17-4 PH Stainless Steel<br>conforming to ASTM A564 | 17-4 PH Stainless Steel and 465<br>Stainless Steel conforming to<br>ASTM A564 and Nitronic 60<br>conforming to ASTM A267 | 316L SS per ASTM F138 and<br>PPSU Radel R-5100 |
| Principle of<br>Operation | Patient reference hardware that<br>provides a reference to a rigid<br>anatomical structure. | Patient reference hardware that<br>provides a reference to a rigid<br>anatomical structure. | Patient reference hardware that<br>provides a reference to a rigid<br>anatomical structure. |
| Device<br>Construct<br>Design | Patient reference hardware<br>•<br>option for attaching<br>Patient Reference Array to<br>patient to acquire accurate<br>patient registrations | Patient reference hardware<br>•<br>option for attaching Patient<br>Reference Array to patient<br>to acquire accurate patient | Patient reference hardware<br>•<br>option for attaching<br>Patient Reference Array to<br>patient to acquire accurate<br>patient registrations |
| | needed for navigated<br>surgery<br>Distal Connection<br>Feature/Patient Fixation<br>Method: Bent Rod<br>component utilized for<br>patient fixation by<br>insertion through<br>NuVasive pedicle screw<br>systems<br>Proximal Connection<br>Feature: Poker Chip<br>component utilized for<br>fixation to Extension Arm<br>to facilitate connection to<br>Patient Reference Array<br>No radiographic markers<br>due to use of non-fiducial<br>(Non-fiducial) registration<br>only | registrations needed for<br>navigated surgery<br>Distal Connection<br>Feature/Patient Fixation<br>Method: Clamp<br>component utilized for<br>patient fixation by<br>actuating jaw features to<br>clamp directly to spinous<br>process<br>Proximal Connection<br>Feature: PRA hub<br>component utilized for<br>fixation to Extension Arm<br>with Articulating Adapter<br>to facilitate connection to<br>Patient Reference Array.<br>Can also be attached<br>directly to Patient<br>Reference Array.<br>Radiographic markers used<br>specifically for fiducial<br>registration only | needed for navigated<br>surgery<br>Distal Connection<br>Feature/Patient Fixation<br>Method: Sharp pin<br>component utilized for<br>patient fixation by fixation<br>into pelvis<br>Proximal Connection<br>Feature: PRA hub<br>component utilized for<br>fixation to Extension Arm<br>with Articulating Adapter<br>to facilitate connection to<br>Patient Reference Array.<br>Can also be attached<br>directly to Patient<br>Reference Array<br>May or may not have<br>radiographic markers used<br>specifically for fiducial<br>registration only |
| Offered Lengths | Short and Long | Short and Long | Short and Long |
| Offered<br>Diameters | Various; compatible with<br>NuVasive Screw Systems | N/A | N/A |
| Registration<br>Compatibility | Non-fiducial Registration | Fiducial and Non-fiducial<br>Registration | Fiducial and Non-fiducial<br>Registration |
| Method of<br>Attachment | Inserted through NuVasive<br>screw systems | Clamps directly to spinous<br>process | Fixated into pelvis |
| Region of<br>Anatomy | Cervical, Thoracic and Lumbar<br>regions of the spine | Cervical, Thoracic and Lumbar<br>regions of the spine | Lumbar and Sacral regions of<br>the spine |
| Performance<br>Data/<br>Testing/Analysis | System Level Accuracy<br>Testing<br>Cadaver Verification<br>Design Validation<br>Tolerance Analysis<br>Usability Validation | System Level Accuracy<br>Testing<br>Cadaver Verification<br>Design Validation<br>Tolerance Analysis<br>Usability Validation | System Level Accuracy<br>Testing<br>Cadaver Verification<br>Design Validation<br>Tolerance Analysis<br>Usability Validation |
| Accessory<br>Instrumentation | Extension Arm which holds<br>Patient Reference Array | Extension Arm<br>with Articulating Adapter<br>and/or Patient Reference Array | Extension Arm<br>with Articulating Adapter<br>and/or Patient Reference Array |
| Sterilization | Non-sterile | Non-sterile | Sterile (EO) |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Nuvasive. The logo consists of a stylized leaf shape on the left, with the word "NUVASIVE" in gray letters to the right of the leaf. The leaf is split into two colors, with the top half being a dark purple and the bottom half being a light gray.
#### G. Performance Data
Non-clinical testing was performed using the same methodology and acceptance criteria as the previously cleared Spinous Process Clamp (K210574), demonstrating that the subject device's performance is substantially equivalent to the legally marketed predicate device. The following testing was performed:
- . System Level Accuracy Testing
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for Nuvasive. The logo consists of a stylized leaf shape in purple and gray, followed by the word "NUVASIVE" in gray sans-serif font. The leaf shape is on the left side of the logo, and the word "NUVASIVE" is on the right.
- Cadaver Verification ●
- Design Validation
- . Tolerance Analysis
- Usability Validation ●
The results demonstrate that the subject Rod Registration Frame meets the same criteria as the predicate devices, and the subject device was therefore found to be substantially equivalent to the predicate. No clinical studies were conducted.
#### H. Conclusion
Results of non-clinical testing demonstrate that the subject device Rod Registration frame is substantially equivalent to the predicate device cleared in K210574.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.