The Pulse System is a medical device comprised of Pulse NVM5, Pulse LessRay, and Pulse Navigation. Pulse NVM5 is intended for intraoperative neurophysiologic monitoring during spinal surgery, neck dissections, thoracic surgeries, and upper and lower extremities. The device provides information directly to the surgeon, to help assess a patient's neurophysiologic status. The Pulse NVM5 provides this information by electrically stimulating nerves via electrodes located on surgical accessories and monitoring electromyography (EMG), motor evoked potential (MEP) or somatosensory evoked potential (SSEP) responses of nerves. The System also integrates Bendini® software used to locate spinal implant instrumentation for the placement of spinal rods. · XLIF (Detection) – The XLIF (Detection) function allows the surgeon to locate and evaluate spinal nerves, and is used as a nerve avoidance tool. · Basic & Dynamic Screw Test - The Screw Test functions allow the surgeon to locate and evaluate spinal nerves by providing proximity information before, during or after bone preparation and placement of bone screws. · Free Run EMG – The Free Run EMG function identifies spontaneous EMG activity of spinal nerves by continually displaying a live stream waveform of any mechanically induced myotome contractions. · Twitch Test (Train of Four) – The Twitch Test Function allows the surgeon to assess moderate degrees of neuromuscular block in effect by evaluating muscle contraction following a train of four stimulation pulses. · MEP - Transcranial or lumbar (i.e., conus in region of L1-L2) stimulation techniques for motor evoked potentials are used to assess for acute dysfunction in axonal conduction of the corticospinal tract and peripheral nerves. The MEP function provides an adjunctive method to allow the surgeon to monitor spinal cord and motor pathway integrity during procedures with a risk of surgically induced motor injury. · SSEP = The SSEP function allows the surgeon to assess sensory spinal cord function in surgical procedures during which the spinal cord is at risk. • Remote Access - The remote monitoring and local wireless control provides real-time capabilities to the Pulse System · Bendini - The Bendini Spinal Rod Bending function is used to locate spinal implant system instrumentation (screws. hooks) to determine their relative location to one another to generate bend instructions to shape a spinal rod. A surgeon is able to use those instructions and bend a rod using the Bendini Bender, a mechanical rod bender. Pulse LessRay is intended for use in any application where a fluoroscope is incorporated to aid treatment of disease. Pulse Navigation is intended as an intraoperative image-guided localization system in either open or minimally-invasive spinal surgical procedures. Instruments and implants tracked by a passive marker sensor system are virtually displayed on a patient's 3D radiographic image data. The system enables computer-assisted navigation for spinal surgical procedures in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure can be identified relative to the acquired image of the anatomy. This may include the following spinal implant procedures: - Pedicle Screw Placement (cervical, thoracic, lumbar) - Iliosacral screw placement
Device Story
Pulse System integrates three modules: Pulse NVM5 (neuro-monitoring), Pulse LessRay (fluoroscopic image enhancement), and Pulse Navigation (image-guided surgery). NVM5 uses electrical stimulation via surgical electrodes to monitor EMG, MEP, and SSEP responses, providing real-time nerve proximity and neuromuscular block data to surgeons. Bendini software calculates spinal rod bending instructions based on implant locations. LessRay interfaces with fluoroscopes to enhance low-dose images via noise reduction and provides C-arm tracking. Pulse Navigation uses an infrared camera to track passive markers on instruments, superimposing them onto 3D radiographic images (CT/Cone Beam CT) for trajectory guidance during spinal procedures. Used in ORs by surgeons; provides visual/auditory feedback to assist in nerve avoidance, implant placement, and rod shaping. Benefits include reduced radiation exposure via LessRay, improved surgical accuracy through navigation, and real-time monitoring of neural integrity to prevent injury.
Clinical Evidence
Bench testing only. Verification and validation included tracking accuracy per ASTM F2554-10, 3D navigation registration/tracking error verification, navigation software validation, cadaver validation for pedicle screw placement, and electrical safety/EMC testing per IEC 60601. No clinical data presented.
Technological Characteristics
System includes control unit, digital preamplifier with A/D converter, infrared camera for optical tracking, and various surgical electrodes (Ag-AgCl film/hydrogel). Connectivity via network, USB, and DICOM. Software performs image registration, signal processing (averaging, noise reduction), and navigation. Compliant with IEC 60601-1.
Indications for Use
Indicated for patients undergoing spinal surgery, neck dissections, thoracic surgeries, and upper/lower extremity procedures requiring intraoperative neurophysiologic monitoring, nerve avoidance, or image-guided localization for spinal implants (pedicle/iliosacral screws).
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date July 29, 2021. The text is written in a clear, sans-serif font. The date is formatted with the month first, followed by the day and year. The text is black against a white background.
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a seal with an abstract design, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The text is arranged in three lines, with "FDA" being the largest and boldest, followed by "U.S. FOOD & DRUG" and then "ADMINISTRATION".
NuVasive, Incorporated Emily Chung Regulatory Affairs 7475 Lusk Boulevard San Diego, California 92121
Re: K210574
Trade/Device Name: NuVasive Pulse System Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO, PDQ, ETN, GWF, HAW, IKN, OWB, LLZ, JAA Dated: June 28, 2021 Received: June 29, 2021
Dear Emily Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K210574
Device Name NuVasive Pulse System
Indications for Use (Describe)
The Pulse System is a medical device comprised of Pulse NVM5, Pulse LessRay, and Pulse Navigation.
Pulse NVM5 is intended for intraoperative neurophysiologic monitoring during spinal surgery, neck dissections, thoracic surgeries, and upper and lower extremities. The device provides information directly to the surgeon, to help assess a patient's neurophysiologic status. The Pulse NVM5 provides this information by electrically stimulating nerves via electrodes located on surgical accessories and monitoring electromyography (EMG), motor evoked potential (MEP) or somatosensory evoked potential (SSEP) responses of nerves. The System also integrates Bendini® software used to locate spinal implant instrumentation for the placement of spinal rods.
· XLIF (Detection) – The XLIF (Detection) function allows the surgeon to locate and evaluate spinal nerves, and is used as a nerve avoidance tool.
· Basic & Dynamic Screw Test - The Screw Test functions allow the surgeon to locate and evaluate spinal nerves by providing proximity information before, during or after bone preparation and placement of bone screws.
· Free Run EMG – The Free Run EMG function identifies spontaneous EMG activity of spinal nerves by continually displaying a live stream waveform of any mechanically induced myotome contractions.
· Twitch Test (Train of Four) – The Twitch Test Function allows the surgeon to assess moderate degrees of neuromuscular block in effect by evaluating muscle contraction following a train of four stimulation pulses.
· MEP - Transcranial or lumbar (i.e., conus in region of L1-L2) stimulation techniques for motor evoked potentials are used to assess for acute dysfunction in axonal conduction of the corticospinal tract and peripheral nerves. The MEP function provides an adjunctive method to allow the surgeon to monitor spinal cord and motor pathway integrity during procedures with a risk of surgically induced motor injury.
· SSEP = The SSEP function allows the surgeon to assess sensory spinal cord function in surgical procedures during which the spinal cord is at risk.
• Remote Access - The remote monitoring and local wireless control provides real-time capabilities to the Pulse System
· Bendini - The Bendini Spinal Rod Bending function is used to locate spinal implant system instrumentation (screws. hooks) to determine their relative location to one another to generate bend instructions to shape a spinal rod. A surgeon is able to use those instructions and bend a rod using the Bendini Bender, a mechanical rod bender.
Pulse LessRay is intended for use in any application where a fluoroscope is incorporated to aid treatment of disease.
Pulse Navigation is intended as an intraoperative image-guided localization system in either open or minimally-invasive spinal surgical procedures. Instruments and implants tracked by a passive marker sensor system are virtually displayed on a patient's 3D radiographic image data. The system enables computer-assisted navigation for spinal surgical procedures in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure can be identified relative to the acquired image of the anatomy.
This may include the following spinal implant procedures:
- Pedicle Screw Placement (cervical, thoracic, lumbar)
- Iliosacral screw placement
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Nuvasive. The logo consists of a stylized leaf shape in purple and gray, followed by the word "NUVASIVE" in a light gray sans-serif font. The leaf shape is split into two parts, with the top part in purple and the bottom part in gray.
## 510(k) Summary
## K210574
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular 21 CFR §807.92, the following summary of information is provided:
#### A. Submitted by:
Manthan J. Damani Manager, Regulatory Affairs NuVasive, Incorporated 7475 Lusk Blvd. San Diego, California 92121 Telephone: (858) 909-1800
Date Prepared: July 28, 2021
#### B. Device Name
| Proprietary Name: | NuVasive® Pulse™ System |
|------------------------------|-----------------------------------------------------------------------------------------------------------|
| Common or Usual Name: | Stereotaxic Instrument;<br>Neurological Surgical Monitor;<br>Image-intensified fluoroscopic x-ray system. |
| Classification Name: | Stereotaxic Instrument |
| Device Class: | Class II |
| Regulation Number: | 21 CFR §882.4560 |
| Classification Product Code: | OLO |
| Subsequent Product Codes: | PDO, ETN, GWF, HAW, IKN, OWB, LLZ, JAA |
#### C. Predicate Devices
The subject NuVasive Pulse System is substantially equivalent to the primary predicate NuVasive Pulse System (K180038). It is also substantially equivalent to the additional predicate devices StealthStation S8 Spine Software V1.0.0 (K170011), and NuVasive Navigation.S Instruments (K200719).
#### D. Device Description
The Pulse System is a medical device consisting of Pulse NVM5, Pulse LessRay, and Pulse Navigation. The Pulse System hardware includes a control unit, as well as accompanying accessory components.
The Pulse NVM5 is a medical device that is intended for intraoperative neurological monitoring and status assessment during spinal surgery. The device provides information directly to the surgeon, to help assess a patient's neurological status. The Pulse NVM5 provides this information by electrically stimulating nerves via electrodes located on surgical accessories and monitoring electromyography (EMG), motor evoked potential (MEP) or somatosensory evoked potential (SSEP) responses of the muscle groups innervated by the nerves. Moreover, a Twitch Test
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for Nuvasive. The logo consists of a stylized leaf shape in purple and gray on the left, followed by the word "NUVASIVE" in gray. The leaf shape is partially shaded to give it a three-dimensional appearance.
("Train of Four") function is utilized to test the ability of the nerve to respond, or contract, following four stimulation pulses to determine the presence of neuromuscular block.
Additionally, the Pulse NVM5 System includes a software function that measures spinal parameters and acquires the location of spinal implants (screws, hooks) to assist the surgeon in bending spinal rods (Bendini). Lastly, the Pulse NVM5 provides Remote Access in two pathways, Local Wireless Control and Remote Monitoring.
Pulse LessRay is a software application which can be interfaced to a fluoroscope with a video cable. The images produced by the fluoroscope are transmitted to a frame grabber in the computer running LessRay where the images are enhanced and then displayed. When used in connection with the low dose and/or pulse setting on the fluoroscope, the user can improve the quality (clarity, contrast, noise level, and usability') of a noisy (low-quality) image. Using this system, much of the graininess of low radiation dose images can be eliminated. This allows for greater utility of low dose imaging." LessRay provides the additional feature of being able to interface LessRay with a tracking system in order to aid the C-arm technician in positioning the fluoroscope between the various views of the patient necessary for the intervention. LessRay with Tracking ensures that the fluoroscope is centered over the correct anatomy prior to taking any additional x-ray images.
Pulse Navigation is a stereotactic surgical application intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. It is intended for intraoperative image-guided localization which allows for surgical instruments to be tracked in three dimensional space. The device provides real-time information directly to the surgeon, enabling the surgeon to evaluate the instrument depth and trajectory for computer-assisted navigation during spine surgery. Instruments are tracked in three dimensional space with an Infrared (IR) Camera, being virtually displayed and superimposed on registered radiographic images. Radiographic images are in the form of 3D intraoperative scan (CT or Cone Beam CT).
The reason for this submission is to update indications for the Pulse Navigation application and to introduce design modifications to hardware and software components of the Navigation application.
#### E. Indications for Use
The Pulse System is a medical device comprised of Pulse NVM5, Pulse LessRay, and Pulse Navigation.
Pulse NVM5 is intended for intraoperative neurophysiologic monitoring spinal surgery, neck dissections, thoracic surgeries, and upper and lower extremities. The device provides information directly to the surgeon, to help assess a patient's neurophysiologic status. The Pulse NVM5 provides this information by electrically stimulating nerves via electrodes located on surgical accessories and monitoring electromyography (EMG), motor evoked potential (MEP) or somatosensory evoked potential (SSEP) responses of nerves. The System also integrates Bendini® software used to locate spinal implant instrumentation for the placement of spinal rods.
- XLIF (Detection) The XLIF (Detection) function allows the surgeon to locate and evaluate ● spinal nerves, and is used as a nerve avoidance tool.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for Nuvasive. The logo consists of a stylized purple and gray leaf-like shape on the left, followed by the word "NUVASIVE" in gray, sans-serif font. The leaf shape is partially purple and partially gray, with the purple portion appearing to be on top of the gray portion.
- . Basic & Dynamic Screw Test - The Screw Test functions allow the surgeon to locate and evaluate spinal nerves by providing proximity information before, during or after bone preparation and placement of bone screws.
- Free Run EMG The Free Run EMG function identifies spontaneous EMG activity of spinal . nerves by continually displaying a live stream waveform of any mechanically induced myotome contractions.
- . Twitch Test (Train of Four) - The Twitch Test Function allows the surgeon to assess moderate degrees of neuromuscular block in effect by evaluating muscle contraction following a train of four stimulation pulses.
- . MEP - Transcranial or lumbar (i.e., conus in region of L1-L2) stimulation techniques for motor evoked potentials are used to assess for acute dysfunction in axonal conduction of the corticospinal tract and peripheral nerves. The MEP function provides an adjunctive method to allow the surgeon to monitor spinal cord and motor pathway integrity during procedures with a risk of surgically induced motor injury.
- . SSEP - The SSEP function allows the surgeon to assess sensory spinal cord function in surgical procedures during which the spinal cord is at risk.
- Remote Access The remote monitoring and local wireless control provides real-time . capabilities to the Pulse System.
- Bendini The Bendini Spinal Rod Bending function is used to locate spinal implant system . instrumentation (screws, hooks) to determine their relative location to one another to generate bend instructions to shape a spinal rod. A surgeon is able to use those instructions and bend a rod using the Bendini Bender, a mechanical rod bender.
Pulse LessRay is intended for use in any application where a fluoroscope is incorporated to aid in diagnosis and treatment of disease.
Pulse Navigation is intended as an intraoperative image-guided localization system in either open or minimally-invasive spinal surgical procedures. Instruments and implants tracked by a passive marker sensor system are virtually displayed on a patient's 3D radiographic image data. The system enables computer-assisted navigation for spinal surgical procedures in which the use of stereotactic surgery may be appropriate and where a reference to a rigid anatomical structure can be identified relative to the acquired image of the anatomy.
This may include the following spinal implant procedures:
- Pedicle Screw Placement (cervical, thoracic, lumbar) l
- 트 Iliosacral screw placement
#### Technological Characteristics F.
As was established in this submission, the subject Pulse System is substantially equivalent to other predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent technological characteristics to its predicate device through comparison in areas including design, intended use, material composition, and functions.
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Nuvasive. The logo consists of a purple and gray abstract shape resembling a stylized leaf or curved shape on the left. To the right of the shape, the word "NUVASIVE" is written in gray, sans-serif font.
#### Table 1 – Comparison of Technical Characteristics
| Specification/<br>Property | Predicate Device<br>NuVasive Pulse System<br>(K180038) | Predicate Device<br>StealthStation S8 Spine Software v1.0.0<br>(K170011) | Subject Device<br>Pulse System | |
|---------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use /<br>Indications for Use | The Pulse System is a medical device comprised of Pulse NVM5,<br>Pulse LessRay, and Pulse Navigation.<br><br>The Pulse NVM5 is a medical device that is intended for<br>intraoperative neurophysiologic monitoring during spinal<br>surgery, neck dissections, thoracic surgeries, and upper and<br>lower extremities. The device provides information directly to<br>the surgeon, to help assess a patient's neurophysiologic<br>status. Pulse NVM5 provides this information by electrically<br>stimulating nerves via electrodes located on surgical accessories<br>and monitoring electromyography (EMG), transcranial or<br>lumbar motor evoked potential (MEP) or somatosensory<br>evoked potential (SSEP) responses of nerves. The System also<br>integrates Bendini™ software used to locate spinal implant<br>instrumentation for the placement of spinal rods.<br><br>• XLIF (Detection) – The XLIF (Detection) function allows<br>the surgeon to locate and evaluate spinal nerves, and is used<br>as a nerve avoidance tool.<br>• Basic & Dynamic Screw Test - The Screw Test functions<br>allow the surgeon to locate and evaluate spinal nerves by<br>providing proximity information before, during or after bone<br>preparation and placement of bone screws.<br>• Free Run EMG – The Free Run EMG function identifies<br>spontaneous EMG activity of spinal nerves by continually<br>displaying a live stream waveform of any mechanically<br>induced myotome contractions.<br>• Twitch Test (Train of Four) – The Twitch Test Function<br>allows the surgeon to assess moderate degrees of<br>neuromuscular block in effect by evaluating muscle<br>contraction following a train of four stimulation pulses.<br>• MEP - Transcranial or lumbar (i.e., conus in region of L1-L2)<br>stimulation techniques for motor evoked potentials are used<br>to assess for acute dysfunction in axonal conduction of the<br>corticospinal tract and peripheral nerves. The MEP function<br>provides an adjunctive method to allow the surgeon to<br>monitor spinal cord and motor pathway integrity during<br>procedures with a risk of surgically induced motor injury.<br>• SSEP - The SSEP function allows the surgeon to assess<br>sensory spinal cord function in surgical procedures during<br>which the spinal cord is at risk. | The StealthStation System is intended as an<br>aid for precisely locating anatomical<br>structures in either open or percutaneous<br>procedures.<br><br>The StealthStation® System, with<br>StealthStation Spine Software, is intended as<br>an aid for precisely locating anatomical<br>structures in either open or percutaneous<br>neurosurgical and orthopedic procedures.<br>Their use is indicated for any medical<br>condition in which the use of stereotactic<br>surgery may be appropriate, and where<br>reference to a rigid anatomical structure, such<br>as the spine or pelvis, can be identified<br>relative to images of the anatomy.<br><br>This can include the following spinal implant<br>procedures, such as:<br>• Pedicle Screw Placement<br>• Iliosacral Screw Placement<br>• Interbody Device Placement | The Pulse System is a medical device comprised of Pulse NVM5,<br>Pulse LessRay, and Pulse Navigation.<br><br>Pulse NVM5 is a medical device that is intended for<br>intraoperative neurophysiologic monitoring during spinal<br>surgery, neck dissections, thoracic surgeries, and upper and<br>lower extremities. The device provides information directly to<br>the surgeon, to help assess a patient's neurophysiologic<br>status. Pulse NVM5 provides this information by electrically<br>stimulating nerves via electrodes located on surgical accessories<br>and monitoring electromyography (EMG), transcranial or<br>lumbar motor evoked potential (MEP) or somatosensory<br>evoked potential (SSEP) responses of nerves. The System also<br>integrates Bendini™ software used to locate spinal implant<br>instrumentation for the placement of spinal rods.<br><br>• XLIF (Detection) – The XLIF (Detection) function allows<br>the surgeon to locate and evaluate spinal nerves, and is used<br>as a nerve avoidance tool.<br>• Basic & Dynamic Screw Test - The Screw Test functions<br>allow the surgeon to locate and evaluate spinal nerves by<br>providing proximity information before, during or after bone<br>preparation and placement of bone screws.<br>• Free Run EMG – The Free Run EMG function identifies<br>spontaneous EMG activity of spinal nerves by continually<br>displaying a live stream waveform of any mechanically<br>induced myotome contractions.<br>• Twitch Test (Train of Four) – The Twitch Test Function<br>allows the surgeon to assess moderate degrees of<br>neuromuscular block in effect by evaluating muscle<br>contraction following a train of four stimulation pulses.<br>• MEP - Transcranial or lumbar (i.e., conus in region of L1-L2)<br>stimulation techniques for motor evoked potentials are used<br>to assess for acute dysfunction in axonal conduction of the<br>corticospinal tract and peripheral nerves. The MEP function<br>provides an adjunctive method to allow the surgeon to<br>monitor spinal cord and motor pathway integrity during<br>procedures with a risk of surgically induced motor injury.<br>• SSEP – The SSEP function allows the surgeon to assess<br>sensory spinal cord function in surgical procedures during<br>which the spinal cord is at risk. | |
| Specification/<br>Property | Predicate Device<br>NuVasive Pulse System<br>(K180038) | Predicate Device<br>StealthStation S8 Spine Software v1.0.0<br>(K170011) | Subject Device<br>Pulse System | |
| | • Remote Access - The remote monitoring and local wireless<br>control provides real-time capabilities to the Pulse System | | • Remote Access - The remote monitoring and local wireless<br>control provides real-time capabilities to the Pulse System | |
| | • Bendini – The Bendini Spinal Rod Bending function is used<br>to locate spinal implant system instrumentation (screws,<br>hooks) to determine their relative location to one another to<br>generate bend instructions to shape a spinal rod. A surgeon<br>is able to use those instructions and bend a rod using the<br>Bendini Bender, a mechanical rod bender | | • Bendini – The Bendini Spinal Rod Bending function is used<br>to locate spinal implant system instrumentation (screws,<br>hooks) to determine their relative location to one another to<br>generate bend instructions to shape a spinal rod. A surgeon<br>is able to use those instructions and bend a rod using the<br>Bendini Bender, a mechanical rod bender | |
| Intended Use /<br>Indications for Use<br>(Continued) | The Pulse LessRay is intended for use in any application where<br>a fluoroscope is incorporated to aid in diagnosis and treatment<br>of disease. | | Pulse LessRay is intended for use in any application where a<br>fluoroscope is incorporated to aid in diagnosis and treatment of<br>disease. | |
| | Pulse Navigation is intended as an intraoperative image-guided<br>localization system in either open or minimally-invasive spinal<br>surgical procedures. Instruments and implants tracked by a<br>passive marker sensor system are virtually displayed on a<br>patient's 2D or 3D radiographic image data. The system<br>enables computer-assisted navigation for spinal surgical<br>procedures in which the use of stereotactic surgery may be<br>appropriate and where a reference to a rigid anatomical<br>structure can be identified relative to the acquired image of the<br>anatomy.<br>This may include the following spinal implant procedures: o<br>Pedicle Screw Placement (2D Navigation in Sacral and Lumbar<br>Spine and 3D Navigation in Sacral and Thoracolumbar Spine)<br>o Interbody Device Placement (2D and 3D Navigation in<br>Lumbar Spine via Lateral Approach) | | Pulse Navigation is intended as an intraoperative image-guided<br>localization system in either open or minimally-invasive spinal<br>surgical procedures. Instruments and implants tracked by a<br>passive marker sensor system are virtually displayed on a<br>patient's 3D radiographic image data. The system enables<br>computer-assisted navigation for spinal surgical procedures in<br>which the use of stereotactic surgery may be appropriate and<br>where a reference to a rigid anatomical structure can be<br>identified relative to the acquired image of the anatomy.<br>This may include the following spinal implant procedures:<br>o Pedicle Screw Placement (cervical, thoracic, lumbar)<br>o Iliosacral screw placement | |
| Device Class | II | II | II | |
| Product Code | OLO, PDQ, ETN, GWF, HAW, IKN, OWB, LLZ, JAA | OLO | OLO, PDQ, ETN, GWF, HAW, IKN, OWB, LLZ, JAA…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.