K131576 · Ge Medical Systems, LLC · JAK · Aug 30, 2013 · Radiology
Device Facts
Record ID
K131576
Device Name
OPTIMA CT660
Applicant
Ge Medical Systems, LLC
Product Code
JAK · Radiology
Decision Date
Aug 30, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The Optima CT660 is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories. This device may include data and image processing to produce images in a variety of transaxial and reformatted planes. Further the images can be post processed to produce additional imaging planes or analysis results. The system is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages. The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.
Device Story
Optima CT660 v2 is a multi-slice CT scanner consisting of a gantry, patient table, operator console, and computer. It acquires X-ray transmission data at various angles/planes via a rotating gantry (up to 0.35s/rotation) and reconstructs cross-sectional images. Operates in Axial, Cine, Helical, Cardiac, and Gated modes. Used in clinical settings by trained professionals for diagnostic imaging. New features include Organ Dose Modulation (ODM) for superficial tissue dose reduction, Ultra Kernel for improved spatial resolution, and High Pitch Helical (HPH) scanning. Output consists of transaxial and reformatted images used by radiologists/clinicians for diagnosis and therapy planning. System is an evolutionary modification of the predicate, maintaining fundamental operating principles while updating hardware for regulatory compliance (IEC 60601-1 Ed. 3, RoHS) and cost efficiency.
Clinical Evidence
Non-clinical bench testing included phantom-based image quality acceptance testing per IEC 61223-3-5 and specific validation of new features (ODM, Ultra Kernel, HPH). Clinical evidence consisted of a review of 47 subject clinical exams representing various acquisition modes and body regions. Two independent radiologists assessed diagnostic quality using a 5-point Likert scale, confirming diagnostic results were obtained for all exams.
Technological Characteristics
Multi-slice CT scanner (up to 64 slices/row). Components: gantry, table, console, computer, PDU. Materials compliant with ES 60601-1/IEC 60601-1. Features: Organ Dose Modulation (mA modulation), Ultra Kernel (reconstruction kernel), High Pitch Helical (up to 61.25mm/rotation). Compliance: 21 CFR Subchapter J, NEMA XR-25, IEC 60601-1 Ed. 3, IEC 60601-2-44 Ed. 3.0. Software: Moderate level of concern.
Indications for Use
Indicated for head, whole body, cardiac, and vascular X-ray CT imaging in patients of all ages. Used for diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
K131576
Page 1 of 6
# 510(k) Summary of Safety and Effectiveness
This 510(k) summary of Safety and Effectiveness information is submitted in accordance with the requirement of 21 CFR Part 807.87(h).
In accordance with 21 CFR 807.92 the following summary of information is provided:
| Date: | May 30, 2013 |
|-------------------------------|-------------------------------------------------------------------------------------------------------------|
| Submitter: | GE Healthcare GE Healthcare (GE Medical Systems, LLC)<br>3000 N. Grandview Blvd., W-1140 Waukesha, WI 53188 |
| Primary Contact Person: | Tomohiro Ito |
| | Regulatory Affairs Leader, MI&CT |
| | GE Healthcare (GE Healthcare Japan Corporation) |
| | Tel: +81-42-585-5383 |
| | e-mail: tomohiro.ito@ge.com |
| Secondary Contact<br>Persons: | Helen Peng |
| | Regulatory Affairs Manager, MI&CT |
| | GE Healthcare (GE Medical Systems, LLC) |
| | Tel: 262-548-5091 |
| | Fax: 262-364-2506 |
| | e-mail: hong.peng@ge.com<br>AUG 3 0 2013 |
| | John Jaeckle |
| | Chief Regulatory Affairs Strategist |
| | GE Healthcare |
| | Tel: 262-424-9547 |
| | Fax: 262-364-2506 |
| | e-mail: John.Jaeckle@ge.com |
| Product Identification: | Optima CT660 |
| Device Trade Name: | Optima CT660 |
| Common/Usual Name: | Computed Tomography X-ray System |
| Classification Names: | Computed Tomography X-ray System per 21CFR 892.1750 |
| Product Code: | 90-JAK |
:
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the General Electric (GE) logo. The logo consists of the letters "GE" intertwined within a circular border. The letters are stylized and connected, giving the logo a classic and recognizable appearance. The logo is in black and white.
| Predicate Device | K110227 - Optima CT660 |
|------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer:<br>/Design Location: | GE Healthcare Japan Corporation<br>7-127 Asahigaoka, 4-chome, Hino-shi Tokyo, 191-8503,<br>Japan |
| Manufacturing location<br>(s): | GE Healthcare Japan Corporation<br>7-127 Asahigaoka, 4-chome, Hino-shi Tokyo, 191-8503,<br>Japan<br>GE Medical Systems, LLC<br>3000 N. Grandview Blvd. Waukesha, WI 53188, USA<br>GE Hangwei Medical Systems, Co, Ltd<br>No.1, YongChang Street, Beijing Economic & Technical<br>Development Area, Beijing PR, Beijing, 100176, China |
| Distributor: | GE Medical Systems, LLC<br>3000 N. Grandview Blvd. Waukesha, WI 53188, USA |
# Marketed Devices:
The Optima CT660 v2 is of comparable type and substantially equivalent to its predicate device, GE Healthcare's currently marketed Computed Tomography X-ray System Optima CT660 (K110227). In addition, the system has similar indications for use as other GE Computed Tomography X-ray Systems and identical intended use and indications as the predicate device. The system is labeled as Optima CT660. The system has completed all design controls activities including risk management, verification and validation testing per GE's quality management system and complies with the same standards as the predicate device.
### Device Description:
The Optima CT660 v2 CT system is composed of a gantry, patient table, operator console, computer, power distribution unit (PDU), and interconnecting cables. The system includes image acquisition hardware, image acquisition, reconstruction software, associated accessories, and connections/interfaces to accessories.
The current system configuration/package names are: Optima CT660, Optima CT660 Pro, Optima CT660 s, Optima CT660 sPro, Optima CT660SE, Optima CT660 FREEdom.
The system generates images through the computer reconstruction of data acquired at different angles and planes of the rotating gantry. The gantry can rotate at up to 0.35 seconds per rotation, and can acquire up to 64 slices/rows of data per rotation with a
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined and enclosed within a circular border. The design is simple and recognizable, representing the company's brand identity.
maximum total collimation coverage of 40mm in the z direction. The system can be operated in Axial, Cine, Helical, Cardiac and Gated acquisition modes.
The system is designed as an evolutionary modification to the Optima CT660 CT Scanner System (K110227). Most of hardware is identical to the predicate system, however some has undergone changes due to the need to meet IEC60601-1 Ed. 3 and RoHS regulation and also for reducing costs while maintaining performance, compliance, and specifications. The software has been updated in accordance with GE's quality management system procedures that incorporate the software development life cvcle to introduce the new features discussed in this submittal. Software updates were also made for quality (bug) fixes and to meet the new requirements of IEC 60601-2-44 Ed. 3.0.
The Optima CT660 v2 uses virtually the same materials and identical operating principle as our existing marketed product, except in the case of using the compensatory ROHS compliant materials. The image chain components (tube, collimator, detector, DAS) are virtually identical to the predicate device.
The changes do not affect the intended use, the indications for use, patient population nor fundamental operating principles of the currently commercially available predicate system and are the identical or similar to other GE CT systems and features previously cleared.
The Optima CT660 v2 CT system is intended to be a head and whole body CT system incorporating the same basic fundamental operating principles and the same indications for use as the predicate device. Materials and construction are equivalent to our existing marketed products which are compliant with ES 60601-1. IEC 60601-1 and associated collateral and particular standards, 21 CFR Subchapter J, and NEMA XR-25. The accompanying documents also contain the information in support if IEC61223-3-5 and IEC61223-2-6 for acceptance and constancy testing. All changes have been tested and certified by a NRTL and continue to meet all applicable IEC/UL safety standards.
The modified system has been developed under the same GE quality system and has completed all design controls, including risk management, verification and validation.
#### Intended Use:
The system is intended to be used for head, whole body Computed Tomography applications.
### Indications for Use:
The Optima CT660 is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes. including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions. These images may be obtained either with or without contrast. This device may include signal
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'G' and 'E' intertwined within a circular border. The letters are stylized and appear to be hand-drawn or calligraphic, giving the logo a classic and recognizable look.
analysis and display equipment, patient and equipment supports, components and accessories.
This device may include data and image processing to produce images in a variety of transaxial and reformatted planes. Further the images can be post processed to produce additional imaging planes or analysis results.
The system is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages.
The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.
### Technology:
The Optima CT660 v2 employs the same fundamental scientific technology as its predicate device and other marketed CT systems.
The modification for this 510(k) review is the introduction of the three new optional
features: Organ Dose Modulation (ODM); Ultra Kernel (UK); and High Pitch Helical (HPH).
### Organ Dose Modulation
This feature enables prescription of X-Ray tube current modulation to control radiation dose to the anterior body surface where some sensitive organs such as breasts. thyroids, or eyes are located creating a virtual shield. It is a modification to our existing Automatic Exposure Control (AutomA/SmartmA) where the mA modulation now can be reduced over the user prescribed section of the rotations in consideration of more radiosensitive superficial organs/tissues ..
#### High Pitch Helical
Enables scanning at 30.62mm/rotation or 61.25mm/rotation for 20 and 40mm collimation respectively. This feature introduces a new higher pitch helical scan acquisition parameter that meets GE's existing helical image quality specifications.
#### Ultra Kernel
This is a new reconstruction kernel for filtered back projection that is designed for applications where improved detail is desired.
### Potential Adverse Effects on Health:
Potential electrical, mechanical and radiation hazards are identified in a risk management including hazard analysis and controlled by:
System verification and validation to ensure performance to specifications, Federal Regulations, and user requirements.
{4}------------------------------------------------
GE Healthcare
Image /page/4/Picture/1 description: The image shows the General Electric (GE) logo. The logo consists of the letters 'GE' intertwined within a circle. The letters and the circle are in black, contrasting with the white background. The logo is a well-known symbol representing the multinational conglomerate.
510/k} Premarket Notification Submission for Optima CT660 v2
Adherence and certification to industry and international standards. (UL/CSA and . IEC).
- Compliance to applicable CDRH 21CFR subchapter J requirements. .
- Compliance to NEMA XR-25 .
The device is designed and manufactured under the Quality System Regulations of 21CFR820.
# Determination of Substantial Equivalence:
The Optima CT660 v2 is a modified device based on the hardware and software platform of the predicate device. It was designed and is manufactured under GE's quality system that meet the Quality System Regulations of 21CFR 820 and ISO 13485. All the changes were fully verified and validated to the acceptance criteria per GE Healthcare's design control procedures under our quality system before the modified device was commercially introduced in applicable countries. In addition the Optima CT660 v2 has been successfully tested to demonstrate compliance with IEC 60601-1 (edition 3) and its associated collateral and particular standards. 21CFR Subchapter J. and NEMR XR-25. The following quality assurance measures were applied to the development of the system:
- . Risk Analysis
- Required Reviews .
- . Design Reviews
- . Testing on unit level (Module Verification)
- Integration testing (System Verification) .
- . Performance testing (Verification)
- . Safety testing (Verification)
- . Simulated use testing (Validation)
All changes were verified and validated on the bench, and the testing did not reveal any new questions of safety or effectiveness. GE believes the Optima CT660 v2 is of comparable type and substantially equivalent to our currently marketed system: Optima CT660 (K110227). The substantial equivalence was also based on software documentation for a "Moderate" level of concern device.
## Summary of Additional Testing
- Non-Clinical Testing
In addition to the verification and validation testing successfully completed as required by GE Healthcare's quality system, additional engineering testing was performed to provide the requisite data to substantiate performance claims, safety and efficacy, and ultimately substantial equivalence.
These tests include the objective image quality acceptance testing conducted using phantoms and performed in accordance IEC 61223-3-5 as well as the following specific tests performed on the new features:
{5}------------------------------------------------
KB1576
Page le of le
For Organ Dose Modulation, tests on phantoms confirmed that ODM enables reducing radiation dose in the superficial tissues while maintaining diagnostic image quality. . Tests were also conducted on phantoms that confirmed the ODM enables equivalent mean CT number accuracy and uniformity.
For Ultra Kernel, phantoms were used to test and demonstrate that Ultra kernel improves visual spatial resolution while maintaining pixel noise standard deviation and level of image artifacts.
For High Helical Pitch, performance tests on phantoms were conducted and demonstrated that the helical pitch 1.531 meets GE's image quality specifications for lower pitch acquisitions.
### Sample Clinical Images
47 subject clinical exams representing various acquisition modes and body regions were collected and reviewed. The exams acquired used either one or more of the 3 new features or were acquired using the system without the new features at the physician's discretion. An assessment of diagnostic quality using a 5 point Likert scale was performed by two independent radiologists.
The image quality assessment on the sample clinical CT exam images obtained on the GE Healthcare Optima CT660 2.0 scanner with the three new features demonstrated that diagnostic results (and actual diagnoses) were obtained for all 47 subject exams. This sample data was representative of a wide range of anatomical coverage and patient indications and serves to help demonstrate the modified Optima CT660 2.0 scanner with the new features continues to perform as intended and in the substantially equivalent manner to the unmodified predicate device.
### Conclusion:
The Optima CT660 v2 does not raise any new potential safety risks and performs as well as devices currently on the market. Based on the conformance to standards, development under our quality system, engineering testing, and sample clinical images provided, GE Healthcare believes that the Optima CT660 v2 is as safe and effective, and performs in a substantially equivalent manner to the predicate device, Optima CT600 (K110227).
{6}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle faces right and is positioned within a circular border. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the border of the circle.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 30, 2013
GE Medical Systems, LLC % Ms. Helen Peng Regulatory Affairs Manager 3000 N. Grandview Blvd. WAUKESHA WI 53188
Re: K131576 Trade/Device Name: Optima CT660 V2 Regulation Number: 21 CFR 892.1750 Regulation Name: Commuter Tomography X-ray System Regulatory Class: Class II Product Code: JAK Dated: August 16, 2013 Received: August 19, 2013
Dear Ms. Peng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device. subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and fisting (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
forth in the quality systems (QS) regulation (21 CFR Part,820); and if applicable, the electronic product radiation control provisions (Scctions 531-542 of the Act); 21 CFR 1000-1050.
{7}------------------------------------------------
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known):
Device Name: Optima CT660 V2
Indications for Use:
The Optima CT660 is intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data taken at different angles and planes, including Axial, Cine, Helical (Volumetric), Cardiac, and Gated acquisitions. These images may be obtained either with or without contrast. This device may include signal analysis and display equipment, patient and equipment supports, components and accessories.
This device may include data and image processing to produce images in a variety of transaxial and reformatted planes. Further the images can be post processed to produce additional imaging planes or analysis results.
The system is indicated for head, whole body, cardiac and vascular X-ray Computed Tomography applications in patients of all ages.
The device output is a valuable medical tool for the diagnosis of disease, trauma, or abnormality and for planning, guiding, and monitoring therapy.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
\$m_{h,p}\$).
(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health
> 510(k) K131576
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.