K203214 · Shenzhen Kaiyan Medical CO , Ltd. · OLP · Jul 22, 2021 · General, Plastic Surgery
Device Facts
Record ID
K203214
Device Name
DemarkQ WOW, DemarkQ POP
Applicant
Shenzhen Kaiyan Medical CO , Ltd.
Product Code
OLP · General, Plastic Surgery
Decision Date
Jul 22, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DemarkQ is an Over-the-Counter (OTC) device intended for treatment of mild to moderate inflammatory acne.
Device Story
DemarkQ (models WOW and POP) is an OTC LED therapy device for treating mild to moderate inflammatory acne. Device uses 630nm red and 415nm blue LEDs to emit light energy simultaneously. Operated by a single key; no user settings or adjustments required. Includes main LED treatment module, USB power cord, and goggles. Used by patients in home environment. Treatment duration is 3 minutes per session; frequency is daily. Device provides light-based therapy to skin; intended to reduce acne inflammation. Benefits include non-invasive, self-administered acne management.
Clinical Evidence
Bench testing only. Evaluated for safety and performance per IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, IEC 60601-2-57, IEC 60601-1-6, IEC 62304, ISO 10993-5 (cytotoxicity), and ISO 10993-10 (irritation/sensitization).
Technological Characteristics
LED-based light therapy device. Wavelengths: 630nm (red) and 415nm (blue). Power: Lithium battery (3.7Vdc) or USB adapter (100-240Vac input). Form factor: Panel device. Standards: IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2, IEC 60601-2-57, IEC 62471, ISO 10993-5, ISO 10993-10. Software: Complies with IEC 62304.
Indications for Use
Indicated for the treatment of mild to moderate inflammatory acne in patients using the device as an Over-the-Counter (OTC) product.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
LED Light Therapy Device, model: KN-7000C (K180900)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services - USA. To the right, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" in a larger font size and "ADMINISTRATION" in a smaller font size below it. The logo is clean and professional, reflecting the organization's role in regulating food and drugs.
July 22, 2021
Shenzhen Kaiyan Medical Co Ltd Alain Dijkstra CEO 40A Fuxin Road Fuyong Subdistrict BaoAn District Shenzhen, Guangdong 518000 China
Re: K203214
Trade/Device Name: DemarkQ WOW, DemarkQ POP Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OLP Dated: June 11, 2021 Received: June 15, 2021
Dear Alain Dijkstra:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K203214
Device Name
DemarkQ, Model: DemarkQ WOW (PB-B), DemarkQ POP (SJ-72)
Indications for Use (Describe)
The DemarkQ is an Over-the-Counter (OTC) device intended for treatment of mild to moderate inflammatory acne.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary K203214
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirement of 21 CFR 807.92.
#### 1. Submitter's Information
Company: SHENZHEN KAIYAN MEDICAL CO LTD Establishment Registration Number: 3011644607 Address:40a Fuxin Road Fuyong Subdistrict Baoan District Shenzhen Guangdong, CHINA Phone: 0755-82129361 Fax No: 0755-25024651 Contact Person (including title): Alain Dijkstra(CEO) Email: alaindijkstra@kaiyanmedical.com
#### Application Correspondent:
Contact Person: Alain Dijkstra SHENZHEN KAIYAN MEDICAL CO LTD Address: 40A Fuxin Road Fuyong Subdistrict BaoAn District Shenzhen, Guangdong, China Tel: +86 755 82129361 Fax: +86 755 25024651 Email: regulation@kaiyanmedical.com
#### 2. Subject Device Information
Type of 510(k): Traditional Common Name: WOW LED therapy lamp / POP LED therapy lamp Classification Name: Over-The-Counter Powered Light Based Laser For Acne Trade Name: DemarkQ, Model: DemarkQ WOW(PB-B), DemarkQ POP(SJ-72) Review Panel: General & Plastic Surgery Product Code: OLP Regulation Number: 878.4810 Regulation Class: 2
#### 3. Predicate Device Information
{4}------------------------------------------------
| Sponsor: | SHENZHEN KAIYAN MEDICAL CO LTD |
|-----------------|--------------------------------------------------------|
| Subject Device: | DemarkQ, Model: DemarkQ WOW(PB-B), DemarkQ POP (SJ-72) |
| Document Name: | FDA 510(k) Submission Report |
| | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 |
|-----------------------|------------------------|----------------------|----------------------------------------------|
| Sponsor | LED Technologies, Inc. | Medtek Skincare, LLC | Uvbiotek, LLC |
| Device Name and Model | dpl® SpotLite | Poly Clear | LED Light Therapy Device, model:<br>KN-7000C |
| 510(k) Number | K183118 | K183708 | K180900 |
| Product Code | OLP | OLP | OLP |
| Regulation Number | 878.4810 | 878.4810 | 878.4810 |
| Regulation Class | II | II | II |
#### 4. Device Description
The DemarkQ, Model: DemarkQ WOW(PB-B), DemarkQ POP(SJ-72) are over-the counter light emitting diode (LED) panel device, that emits energy for use in dermatology for the treatment of mild to moderate inflammatory acne. The devices use two types of LEDs: 630nm red and 415nm blue. The device operated only by one key, both wavelengths will be emitted at the same time when it be turned on. The treatment time is controlled by the user. There are no user settings or adjustments required.
The device's components include the main device which containing the LED treatment module, USB power cord and goggles.
#### 5. Intended Use / Indications for Use
The DemarkQ is an Over-the-Counter (OTC) device intended for treatment of mild to moderate inflammatory acne.
#### 6. Test Summary
The DemarkQ, Model: DemarkQ WOW(PB-B), DemarkQ POP(SJ-72) have been evaluated the safety and performance by lab bench testing as following:
- IEC 60601-1:2005/(R)2012 and A1:2012. Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance
- IEC 60601-1-11 Edition 2.0 2015-01, Medical electrical equipment -- part 1-11: General requirements for basic safety and essential performance – Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment.
- IEC 60601-1-2 Edition 4.0 2014-02 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests.
{5}------------------------------------------------
- IEC 60601-2-57 Edition 1.0 2011-01 Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
- IEC 60601-1-6 Edition 3.1 2013-10 Medical electrical equipment - Part 1-6: General requirements for basic safety and essential performance - Collateral standard: Usability
- IEC 62304 Edition 1.1 2015-06, Medical device software - Software life cycle processes
- ISO 10993-5 Third edition, Biological evaluation of medical devices - Part 5: Tests for in vitro cytotoxicity
- ISO 10993-10 Third Edition, Biological evaluation of medical devices - Part 10: Tests for irritation and skin sensitization
{6}------------------------------------------------
#### 7. Comparison to predicate device and conclusion
Compare with predicate device, the subject is very similar in design principle, intended use, indications for use, functions, material and the applicable standards. The differences between subject device do not raise and new questions of safety or effectiveness.
| Elements<br>of<br>Comparison | Subject Device | | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 | Remark |
|---------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|--------------|
| Company | SHENZHEN KAIYAN<br>MEDICAL CO LTD | | LED Technologies,<br>Inc. | Medtek Skincare, LLC | Uvbiotek, LLC | -- |
| Trade Name | DemarkQ | DemarkQ | dpl SpotLite | Poly Clear | LED Light Therapy<br>Device | -- |
| Model | DemarkQ<br>WOW (PB-<br>B) | DemarkQ<br>POP(SJ-72) | -- | KN-7000C | - | -- |
| Classification<br>Name | Over-The-Counter Powered<br>Light Based Laser For Acne | | Over-The-Counter<br>Powered Light<br>Based Laser For<br>Acne | Over-The-Counter<br>Powered Light Based<br>Laser For Acne | Over-The- Counter<br>Light Based Laser for<br>Acne | SE |
| 510(k) Number | Applying | | K183118 | K183708 | K180900 | -- |
| Product Code | OLP | | OLP | OLP | OLP | SE |
| Intended Use /<br>Indications for<br>Use | The DemarkQ is an Over-<br>the-Counter (OTC) device<br>intended for treatment of<br>mild to moderate<br>inflammatory acne. | | The dpl® SpotLite is<br>an Over-the-Counter<br>(OTC) device<br>intended for<br>treatment of mild to<br>moderate<br>inflammatory acne. | The Poly Clear<br>combination of Red<br>(633nm±10nm) and<br>Blue (417nm±10nm) is<br>intended to emit<br>energy in the red, blue<br>regions of the light<br>spectrum to treat<br>dermatological<br>conditions, specifically<br>indicated to treat mild<br>to moderate acne<br>vulgaris. | The LED Light Therapy<br>Device is indicated for<br>the treatment of mild to<br>moderate inflammatory<br>acne | SE |
| Power Supply | Power by: 2<br>x (3.7Vdc,<br>2700 mAh<br>lithium | Power by:<br>3.7Vdc,<br>1500 mAh<br>lithium | Not publicly available | 100-240V,<br>50/60Hz±2%, 300VA<br>(Mains Connected) | Adaptermodel number:<br>LXCP12-005200DEG,<br>Input: 100-240V a.c.<br>50/60Hz. | SE<br>Note 1 |
| battery,<br>10.26Wh)<br>Adapter<br>Input: 100-<br>240Vac,<br>50/60Hz<br>Adapter<br>Output: DC<br>5V, 2A<br>Main unit<br>input: DC<br>5V, 2A | battery,<br>10.26Wh<br>Adapter<br>Input: 100-<br>240Vac,<br>50/60Hz<br>Adapter<br>Output: DC<br>5V, 2A<br>Main unit<br>input: DC<br>5V, 2A | | | 0.5Amax. Output: 5V<br>d.c. 2A<br>Main unit: input: 5V d.c.<br>2A/Internal battery:<br>3.6Vd.c. 2200mAh<br>Internal battery<br>specification:<br>CR18650-22F 3.6V d.c.<br>2200mAh | | |
| Wavelengths | 630nm±10nm<br>415nm±10nm | | 415nm, 630nm | 633±10nm<br>417±10nm | Red light 633nm±10nm<br>Blue light 417nm±10nm | SE |
| Irradiance<br>source | LEDs | | LEDs | No publicly available | LEDs | SE |
| Number of<br>LEDs | 112 each<br>panel | Total 72 | Not publicly available | Red: 910<br>Blue: 910 | Red: 48<br>Blue: 48<br>Total: 96 | SE<br>Note 2 |
| Irradiance<br>(mW/cm²) | Red: 5<br>Blue: 25 | Red: 25<br>Blue: 25±5 | Not publicly available | Combo Red/Blue<br>head:<br>Red: 30m±10<br>Blue: 20m±10 | Red light 45±5<br>Blue light 25±5 | SE<br>Note 2 |
| Time<br>range<br>and frequency | 3 minutes per treatment.<br>Can be used daily. | | 3 minutes per<br>treatment | No publicly available | 3 minutes per skin<br>area. Can be used<br>daily. | SE |
| EMC | IEC 60601-1-2 | | IEC 60601-1-2 | IEC 60601-1-2 | IEC 60601-1-2 | SE |
| Safety | IEC 60601-1<br>IEC 60601-2-57<br>IEC 60601-1-11<br>IEC 62471 | | IEC 60601-1 | IEC 60601-1<br>IEC 60601-2-57<br>IEC 62471 | IEC 60601-1<br>IEC 60601-1-11<br>IEC 60601-2-57<br>IEC 62471 | SE |
| Biocompatibility | ISO 10993-1,<br>ISO 10993-5,<br>ISO 10993-10 | ISO 10993-5,<br>ISO 10993-10 | -- | ISO 10993-1,<br>ISO 10993-5,<br>ISO 10993-10 | SE | |
{7}------------------------------------------------
Sponsor: SHENZHEN KAIYAN MEDICAL CO LTD DemarkQ, Model: DemarkQ WOW(PB-B), DemarkQ POP (SJ-72) Subject Device: FDA 510(k) Submission Report Document Name:
{8}------------------------------------------------
Sponsor: SHENZHEN KAIYAN MEDICAL CO LTD DemarkQ, Model: DemarkQ WOW(PB-B), DemarkQ POP (SJ-72) Subject Device: FDA 510(k) Submission Report Document Name:
{9}------------------------------------------------
## Comparison in Detail(s):
#### Note 1:
Although the "Power Supply" is different from the predicate devices, but they all complied with the IEC 60601-1, IEC 60601-1-11 and IEC 60601-1-2 safety standards' requirements. So, these slight differences will not raise any safety or effectiveness issue.
# Note 2:
Although the "Number of LEDs" and "Irradiances" are different from the predicate devices, But they are all within the scope of comparing device parameters and they all complied with the IEC 60601-1, IEC 60601-1-2, IEC 62471 and IEC 60601-2-57 safety standards' requirements. So, these differences will not raise any safety or effectiveness issue.
## Final Conclusion:
After an analysis of the safety indications, intended uses, performance, and other technological characteristics, the Sponsor believes that no significant differences exist between the new device and the predicate devices and no new issues of safety or effectiveness are raised. Thus, the subject device DemarkQ, Model: DemarkQ WOW(PB-B), DemarkQ POP(SJ-72) are Substantial Equivalence to the predicate devices K183118, K183708 and K180900.
8. Date of the summary prepared: June 11, 2021
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.