SKEEPER is an electronic stethoscope that collects, fitters, and amplifies the sound of a person's heart, lungs, and abdomen. It is intended for use by professional users in clinical environments or by lay users in non-clinical environments. It is not intended for self-diagnosis.
Device Story
SKEEPER (SM-300) is an electronic stethoscope collecting body sounds (heart, lung, abdomen) via a MEMS microphone diaphragm. The device converts vibrations into electrical signals, which are filtered and amplified based on selected frequency modes (Heart: 30-600Hz; Lung: 50-2000Hz; Abdomen: 30-800Hz). Output is transmitted via Bluetooth to the SKEEPER PRO APP on a mobile platform for earphone/speaker playback. The app displays phonocardiograph waveforms, stores/plays back recordings, and calculates heart rate (BPM). Used by clinicians or lay users; intended for physical assessment. Benefits include visualization and digital storage of auscultation data to assist clinical decision-making.
Clinical Evidence
No clinical data provided; substantial equivalence supported by bench testing, electrical safety (IEC 60601-1, IEC 60601-1-11, IEC 60601-1-2), biocompatibility (ISO 10993 series), software verification/validation (IEC 62304), and wireless coexistence (ANSI IEEE C63.27).
Technological Characteristics
Electronic stethoscope with MEMS microphone; rechargeable Li-ion battery; Bluetooth connectivity. Frequency modes: Heart (30-600Hz), Lung (50-2000Hz), Abdomen (30-800Hz). Software-based signal processing for filtering and amplification. Biocompatible materials per ISO 10993. Form factor: handheld device (SM-300) paired with mobile app.
Indications for Use
Indicated for professional and lay users to collect, filter, and amplify heart, lung, and abdominal sounds in clinical and non-clinical environments. Not for self-diagnosis.
Regulatory Classification
Identification
Manual stethoscope: A mechanical device used to project the sounds associated with the heart, arteries, and veins and other internal organs. Electronic stethoscope: An electrically amplified device used to project the sounds associated with the heart, arteries, and veins and other internal organs.
Special Controls
*Classification.* Class II (special controls). The device, when it is a lung sound monitor, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
3M LITTMANN ELECTRONIC STETHOSCOPE, MODEL 3200 (K083903)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 2, 2023
Smartsound Corporation Jungho Lee Representative 171, Yangjaecheon-ro, Gangnam-gu Seoul. 06302 Korea, South
Re: K230613
Trade/Device Name: SKEEPER Regulation Number: 21 CFR 870.1875 Regulation Name: Stethoscope Regulatory Class: Class II Product Code: DOD Dated: March 4, 2023 Received: March 6, 2023
Dear Jungho Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
> Sincerely, Stephen C. Browning -S
LCDR Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics, and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230613
Device Name SKEEPER
Indications for Use (Describe)
SKEEPER is an electronic stethoscope that collects, fitters, and amplifies the sound of a person's heart, lungs, and abdomen. It is intended for use by professional users in clinical environments or by lay users in non-clinical environments. It is not intended for self-diagnosis.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
Prescription Use (Part 21 CFR 801 Subpart D)
|X | Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
## K230613
#### 1. Submitter Information
| Manufacturer: | SMARTSOUND CORPORATION | | |
|-----------------|------------------------------------------------------------|--|--|
| | 171, Yangjaecheon-ro, Gangnam-gu, Seoul, Republic of Korea | | |
| Contact Person: | Jungho Lee / Representative | | |
| Tel: | +82-2-575-2252 | | |
| Fax: | +82-2-575-2201 | | |
| e-mail: | jhojholee@ismartsound.com | | |
| Date Prepared: | March 04, 2023 | | |
#### 2. Device Information
| Trade Name: | SKEEPER |
|--------------------|------------------------|
| Model Number: | SM-300 and SKEEPER APP |
| Common Name: | Electronic Stethoscope |
| Product Code: | DQD |
| Regulation Number: | 21 CFR 870.1875 |
| Class: | 2 |
#### 3. Predicate Device Information
| Predicate Device: | K200776 (Eko CORE) |
|-------------------|----------------------------------------------------------|
| Reference Device: | K083903 (3M LITTMANN ELECTRONIC STETHOSCOPE, MODEL 3200) |
#### Description of the Device 4.
SKEEPER is an electronic stethoscope used to collect and measure the sounds inside the body, and it consists of an electronic stethoscope (SM-300) and a mobile application (SKEEPER PRO APP).
When the diaphragm at the bottom of the SM-300 is lightly attached to the body part to be measured, such as the heart, lungs, or abdomen, the diaphragm vibrates due to the sound inside the human body and generates a sound. This sound is collected using a microphone and converted into an electrical signal. After that, the electrical signal is filtered and amplified for each frequency band set according to the measurement site, amplified, and output to earphones or transmitted to the SKEEPER APP installed on the mobile platform, and then the sound is output the earphone or speaker of the corresponding platform.
The SKEEPER PRO APP graphs the waveform of the sound received by the SM-300. It also stores and plays back sound data and analyzes heart rate.
## SMARTSOUND
{4}------------------------------------------------
#### ട. Indication for Use
SKEEPER is an electronic stethoscope that collects, filters, and amplifies the sound of a person's heart, lungs, and abdomen. It is intended for use by professional users in clinical environments or by lay users in non-clinical environments. It is not intended for self-diagnosis.
#### 6. Technological Characteristics
| | Subject Device | Predicate Device | Reference Device |
|------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | K230613 | K200776 | K083903 |
| Proprietary Name | SKEEPER,<br>Model SM-300 and<br>SKEEPER APP | Eko CORE | 3M LITTMANN<br>ELECTRONIC<br>STETHOSCOPE,<br>MODEL 3200 |
| Manufacturer | SMARTSOUND<br>CORPORATION | Eko Devices, Inc. | 3M Health Care |
| Regulatory Class | Class II | Class II | Class II |
| Regulation Number | 21 CFR 870.1875 | 21 CFR 870.1875 | 21 CFR 870.1875 |
| Product Code | DQD | DQD | DQD |
| Indication for Use | SKEEPER is an<br>electronic stethoscope<br>that collects, filters, and<br>amplifies the sound of a<br>person's heart, lungs,<br>and abdomen. It is<br>intended for use by<br>professional users in<br>clinical environments or<br>by lay users in non-<br>clinical environments. It<br>is not intended for self-<br>diagnosis. | The Eko CORE is an<br>electronic stethoscope<br>that enables<br>amplification, filtering,<br>and transmission of<br>auscultation sound data<br>(heart, lungs, bowel,<br>arteries, and veins),<br>whereby a clinician at<br>one location on network<br>can listen to the<br>auscultation sounds of<br>a patient on site or at a<br>different location on the<br>network. Eko CORE is<br>intended for use on<br>pediatric and adult<br>patients. The Eko<br>CORE is intended to be<br>used by professional<br>users in a clinical<br>environment or by lay<br>users in a nonclinical<br>environment. The<br>device is not intended<br>for self-diagnosis | The 3MTM Littmann<br>Electronic Stethoscope,<br>Model 3200 is intended<br>for medical diagnostic<br>purposes only. It may<br>be used for the<br>detection and<br>amplification of sounds<br>from the heart, lungs,<br>arteries, veins, and<br>other internal organs<br>with the use of a<br>selective frequency. It<br>can be used on any<br>person undergoing a<br>physical assessment. |
| Signal Input Method | Microphone<br>(MEMS) | Microphone<br>(MEMS) | Piezoelectric Sensor |
| Frequency response<br>range | Heart mode:<br>30-600Hz<br>(emphasizes 50 Hz<br>~ 300 Hz)<br>Lung mode:<br>50-2000Hz<br>(emphasizes 100<br>Hz ~ 1200 Hz)<br>Abdomen Mode:<br>●<br>30-800Hz | 20Hz to 2KHz | Bell:<br>20-1000 Hz<br>(emphasizes 20-<br>200 Hz)<br>Diaphragm<br>20-2000 Hz<br>(emphasizes 100-<br>500 Hz)<br>Extended Range<br>20-2000 Hz |
| | (emphasizes 60 Hz<br>~ 400 Hz) | | (emphasizes 50-<br>500 Hz) |
| Sound Amplification | Yes | Yes | Yes |
| Volume Control | Yes | Yes | Yes |
| Record and Playback<br>Sounds | Yes | Yes | Yes |
| Save & Share<br>Recordings | Yes | Yes | Yes |
| Power Source | Rechargeable Lithium-<br>ion Battery | Rechargeable Lithium-<br>ion Battery | Single AA, NiMH<br>(rechargeable) or<br>Lithium batteries may<br>be used |
| Data transfer to<br>Compatible Computing<br>platform | Mobile and tablets<br>(SKEEPER-APP) | Yes<br>(Eko App) | Yes<br>(StethAssist™) |
| Wireless<br>Technology | Bluetooth® | Bluetooth® | Bluetooth® |
| Software Function | • Receive and store<br>the sound data<br>• Display the<br>phonocardiograph of<br>heart sound data<br>• Reply the sound data<br>• Measure the heart<br>rate (BPM) | • Receive and store<br>the sound data<br>• Display the<br>phonocardiograph of<br>heart sound data<br>• Reply the sound data | • Receive and store<br>the sound data<br>• Display the<br>phonocardiograph of<br>heart sound data<br>• Reply the sound data<br>• Measure the heart<br>rate (BPM) |
{5}------------------------------------------------
## Indication for Use
Subject, predicate, and reference devices are the same in that they collect, filter, and amplify sound inside the human body.
The subject and predicate devices are intended to be used by medical personnel and lay users in clinical and non-clinical environment.
Therefore, the indication for use of subject device is substantially the same as predicate device.
#### . Signal Input Method
The subjective device's signal input method is the same as that of the predicate device, which is different from the reference device.
#### . Frequency response range
The subjective device changes the frequency of the sound collected depending on the mode selection. A predicate device collects a wide range of sound frequencies in one mode. The reference device also has different collected frequencies depending on the mode.
The frequency band used throughout is 30 Hz to 2000 Hz for the target device and 20 Hz to 2000 Hz for the subject and reference devices. The frequencies used by the subject and reference devices are substantially equivalent because they cover the frequency bands used by the subject device.
#### . Functions
Sound Amplification, Volume Control, Record and Playback Sounds, Save & Share Recordings, Power Source, and wireless communication method are all the same for all devices.
#### . Application software
Using software to store, play, and graphically display sounds is all the same.
Predicate devices do not have the function of analyzing sound, but Reference devices do have the function of analyzing heart rate. This is the same as the function of analyzing heart rate for the Subject device.
{6}------------------------------------------------
#### Summary of Non-Clinical Data 7.
The following non-clinical data were provided in support of the substantial equivalence determination.
Electrical safety and electromagnetic compatibility(IEC 60601-1, IEC 60601-1-11, and IEC 60601-1-2), Biocompatibility(ISO 10993-1, ISO 10993-5, ISO 10993-10, ISO 10993-23), Software verification and validation (IEC 62304), Wireless coexistence (ANSI IEEE C63.27) and Bench testing
## 10. Summary of Clinical Data
Clinical Data was not required to demonstrate the substantial equivalence.
## 11. Conclusion
SKEEPER has the same intended uses and similar indications as the predicate device in that it filters and amplifies the sound of the human body. The two devices are also identical in that they send sound data to mobile applications for storage, playback, and visualization.
The difference between SKEEPER and predicate device is that SKEEPER's mobile application provides heart rate analysis, but predicate device's mobile application does not provide heart rate analysis. However, heart rate analysis is provided in the reference device's mobile application.
These functional differences do not introduce new intended uses and do not raise other concerns about safety and effectiveness when used as labeled.
Therefore, SKEEPER is substantially equivalent to a predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.