The pmd clean acne is intended for over-the-counter use to treat mild to moderate inflammatory acne.
Device Story
Handheld LED phototherapy device for OTC use; treats mild to moderate inflammatory acne. Device features blue light (415nm ± 5nm) LEDs in a treatment window to irradiate facial skin; reverse side includes soft silicone bristles for facial cleansing prior to light treatment. Powered by internal rechargeable Li-ion battery. Includes protective goggles to prevent ocular exposure. User operates device by applying blue light to face for 3-minute intervals per area. Therapeutic effect achieved via blue light irradiation. Benefits include reduction of inflammatory acne lesions. Device is intended for home use.
Clinical Evidence
Bench testing only. No clinical data provided. Compliance demonstrated with electrical safety (ANSI AAMI ES 60601-1, IEC 60601-1-11, IEC 60601-2-57), EMC (IEC 60601-1-2), photobiological safety (IEC 62471), battery safety (IEC 62133-2), and biocompatibility (ISO 10993-5, -10, -23). Software verification and validation performed per FDA guidance.
Technological Characteristics
Handheld device; materials: silicone, plastic, glass. Light source: blue LED (415nm ± 5nm). Irradiance: 70mW/cm2 ± 10%. Power: 3.7V 1000mAh Li-ion battery. Dimensions: 175.2 x 56 x 37mm. Weight: 135g. Standards: IEC 60601-1, IEC 60601-1-11, IEC 60601-2-57, IEC 62471, ISO 10993. Software-controlled operation.
Indications for Use
Indicated for over-the-counter treatment of mild to moderate inflammatory acne in patients.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters 'FDA' in a blue square, followed by the words 'U.S. FOOD & DRUG ADMINISTRATION' in blue text.
October 27, 2023
Dongguan Hunter Electronic Technology Co., Ltd. % Tracv Che Registration Engineer Feiying Drug & Medical Consulting Technical Service Group Rn 2401 Zhenye International Business Center No. 3101-90, Oianhai Road Shenzhen, Guangdong 518052 China
Re: K230490
Trade/Device Name: PMD Clean Acne. Model: 4004-CABLUE-NA, 4004-GREY-NA, 4004-YELLOW-NA. 4004-PSTO-NA. 4004-LPUR-NA. 4004-CABLUE-NA-INT. 4004-GREY-INT, 4004-YELLOW-INT, 4004-PSTO-INT, 4004-LPUR-INT Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: OLP Dated: February 23, 2023 Received: February 23, 2023
Dear Tracy Che:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{1}------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product: and 21 CFR 820.100. Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations. please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn).
{2}------------------------------------------------
Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-devices/deviceadvice-comprehensive-regulatory-assistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Tanisha L. Tanisha L. Hithe
Tanisha L. -s
Canisha L. -s Hithe -S 2023.10.27
00:05:32 -0 00:05:32 -04'00' Tanisha Hithe Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230490
#### Device Name
pmd clean acne, Model: 4004-CABLUE-NA, 4004-GREY-NA, 4004-YELLOW-NA, 4004-PSTO-NA, 4004-CABLUE-NA-INT, 4004-GREY-INT, 4004-YELLOW-INT, 4004-PSTO-INT, 4004-LPUR-INT
Indications for Use (Describe)
The pmd clean acne is intended for over-the-counter use to treat mild to moderate inflammatory acne.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510 (k) Summary K230490
This "510(k) Summary" of 510(k) safety and effectiveness information is submitted in accordance with requirements of Title 21, CFR Section 807.92.
# (1) Applicant information:
| 510(k) owner's name: | Dongguan Hunter Electronic Technology Co., Ltd. |
|---------------------------|-----------------------------------------------------------------------------------------------|
| Address: | Room 301, Building 6, 74 Qiaodong Road, Tangxia Town, Dongguan<br>City, Guangdong, P.R. China |
| Contact person: | Hunter Ye |
| Phone number: | 13554760590 |
| Email: | 13554760590@139.com |
| Date of summary prepared: | 2023-9-22 |
## (2) Reason for the submission
New device, there were no prior submissions for the device.
# (3) Proprietary name of the device
| Trade name/model: | pmd clean acne, Model: 4004-CABLUE-NA, 4004-GREY-NA,<br>4004-YELLOW-NA, 4004-PSTO-NA, 4004-LPUR-NA,<br>4004-CABLUE-NA-INT, 4004-GREY-INT, 4004-YELLOW-INT,<br>4004-PSTO-INT, 4004-LPUR-INT |
|--------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common name: | Over-The-Counter Powered Light Based Laser For Acne |
| Regulation number: | 21 CFR 878.4810 |
| Product code: | OLP |
| Review panel: | General & Plastic Surgery |
| Regulation class: | Class II |
## (4) Predicate device
| Sponsor | Li-Tek Electronics Technologies |
|-----------------------|---------------------------------------|
| Device Name and Model | LED Phototherapy Device. Model:PL-120 |
| 510(k) Number | K162098 |
| Product Code | OLP, OHS |
| Regulation Number | 21 CFR 878.4810 |
| Regulation Class | II |
# (5) Description/ Design of device:
The pmd clean acne adopts light emitting diodes (LED) in the blue (415nm ± 5nm) spectrum to
{5}------------------------------------------------
irradiate on the face to realize its therapeutic effect. The pmd clean acne adopts handheld design, LEDs are contained in the blue light treatment window, and on its reverse side, the soft silicone bristols can be used to clean the face before the blue light treatment. The product is provided with internal battery and it can be charged by external adaptor. To prevent irradiation of LED lights to eyes during the treatment, pmd clean acne is equipped with goggles which blocks light energy from LEDs.
# (6) Indications for use:
The pmd clean acne is intended for over-the-counter use to treat mild to moderate inflammatory acne.
# (7) Materials
| Component<br>name | Material of Component | Body Contact Category | Contact Duration |
|-------------------|--------------------------|----------------------------------------|--------------------|
| pmd clean<br>acne | Silicone, plastic, glass | Surface-contacting device: Intact skin | Less than 24 hours |
We have tested the device for biocompatibility by a reliable third-party lab. For details, please refer to "Biocompatibility Discussion".
| Item | Subject device | Predicate device | Remark |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------|--------|
| Trade name | pmd clean acne, Model:<br>4004-CABLUE-NA,<br>4004-GREY-NA,<br>4004-YELLOW-NA,<br>4004-PSTO-NA,<br>4004-LPUR-NA,<br>4004-CABLUE-NA-INT,<br>4004-GREY-INT,<br>4004-YELLOW-INT,<br>4004-PSTO-INT,<br>4004-LPUR-INT | LED Phototherapy Device.<br>Model:PL-120 | / |
| 510 (k) number | Pending | K162098 | / |
| Manufacturer | Dongguan Hunter Electronic<br>Technology Co., Ltd. | Li-Tek Electronics<br>Technologies | / |
| Regulation number | 21 CFR 878.4810 | 21 CFR 878.4810 | Same |
| Classification name | Over-The-Counter Powered<br>Light Based Laser For Acne | Over-The-Counter<br>Powered Light Based Laser<br>For Acne | Same |
# (8) Technological characteristics and substantial equivalence:
{6}------------------------------------------------
| Product code | OLP | OLP, OHS | Same |
|------------------------------------------------------|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|
| Class | II | II | Same |
| Indications for use/<br>Intended use | The pmd clean acne is<br>intended for over-the- counter<br>use to treat mild to moderate<br>inflammatory acne. | The red light is intended<br>for the treatment of periorbital wrinkles, and<br>the blue light is intended<br>for the treatment of the<br>mild to moderate<br>inflammatory acne. | Similar, the<br>indications for<br>use of the<br>subject device<br>is within that of<br>the predicate<br>device. |
| Location for use | Face | Face | Same |
| OTC or prescription | OTC | OTC | Same |
| <b>Basic unit characteristics</b> | | | |
| Power supply | 3.7V 1000mAh Li-ion Battery | 3.7V 1050mAh Li battery | Similar |
| Weight | 135g | 150g | Different |
| Dimensions | 175.2 * 56 * 37mm | 187*65*51mm | Different |
| Irradiation area | 11.04cm2±5% | 30cm2±5% | Different |
| Software/<br>Firmware/<br>Microprocessor<br>Control? | Yes | Yes | Same |
| Sterilization | Not required | Not required | Same |
| <b>Output specifications</b> | | | |
| Light source | Light Emitting Diodes (LED) | Light Emitting Diodes<br>(LED) | Same |
| Wavelength | Blue: 415nm±5nm | Red: 630 ± 3nm,<br>Blue: 415 ± 3nm | Similar |
| Irradiance | Blue: 70mW/cm2±10% | Red: 80mW/cm2±10%<br>Blue: 65mW/cm2±10% | Similar |
| <b>Additional features</b> | | | |
| Treatment<br>duration | 3 minutes for each part | 3 minutes per target area | Same |
| Materials of<br>skin-contacting<br>components | Silicone, plastic, glass | ABS plastic | Different |
| Work<br>environment | Temperature: 5-30°C<br>Humidity: 10-80%RH<br>Atmospheric pressure:<br>700hPa-1060hPa | Temperature: 5 ~ 40°C<br>Humidity: 10%-80%<br>Atmospheric pressure:<br>700hPa-1060hPa | Similar |
| Biocompatibility<br>feature | All body-contacting<br>materials are complied with<br>ISO10993-5, ISO 10993-10<br>and ISO 10993-23 | ABS plastic in hand hold<br>part can be considered<br>safety | Similar |
| Electrical safety | IEC 60601-1;<br>IEC 60601-1-11;<br>IEC 60601-2-57 | IEC 60601-1;<br>IEC 60601-1-11;<br>IEC 60601-2-57 | Same |
| EMC | IEC 60601-1-2 | IEC 60601-1-2 | Same |
| Photobiological<br>safety | IEC 62471 | IEC 62471 | Same |
{7}------------------------------------------------
Dongguan Hunter Electronic Technology Co., Ltd. 510(k)s - Section 8. 510 (k) Summary
## (9) Non-clinical studies and tests performed:
Non-clinical testings have been conducted to verify that the pmd clean acne meets all design specifications which supports the conclusion that it's Substantially Equivalent (SE) to the predicate device. The testing results demonstrate that the subject device complies with the following standards:
- A ANSI AAMI ES 60601-1, Medical electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2, Medical electrical equipment Part 1-2: General requirements for basic safety A and essential performance - Collateral standard: Electromagnetic disturbances - Requirements and tests
- A IEC 60601-1-11, Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
- > IEC 60601-2-57, Medical Electrical Equipment - Part 2-57: Particular requirements for the basic safety and essential performance of non-laser light source equipment intended for therapeutic, diagnostic, monitoring and cosmetic/aesthetic use
- A IEC 62471, Photobiological safety of lamps and lamp systems
- A IEC 62133-2, Secondary cells and batteries containing alkaline or other non-acid electrolytes -Safety requirements for portable sealed secondary cells, and for batteries made from them, for use in portable applications - Part 2: Lithium systems
The device has been tested for biocompatibility, it complies with the following standards.
- A ISO 10993-5, Biological Evaluation of Medical Devices - Part 5: Tests for In Vitro Cytotoxicity
- > ISO 10993-10, Biological evaluation of medical devices - Part 10: Tests for skin sensitization
- > ISO 10993-23, Biological evaluation of medical devices - Part 23: Tests for irritation
We have also conducted:
- A Software verification and validation test according to the requirements of the FDA "Guidance for Pre Market Submissions and for Software Contained in Medical Devices"
## (10) Conclusion
{8}------------------------------------------------
Based on the above analysis and non-clinical tests performed, it can be concluded that the subject device pmd clean acne is as safe, as effective, and performs as well as the legally marketed predicate device, K162098, LED Phototherapy Device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.