K230448 · Deepx Health, LLC · PSN · May 17, 2023 · General, Plastic Surgery
Device Facts
Record ID
K230448
Device Name
DeepX DermoSight Dermatoscope
Applicant
Deepx Health, LLC
Product Code
PSN · General, Plastic Surgery
Decision Date
May 17, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4580
Device Class
Class 2
Indications for Use
The DeepX DermoSight Dermatoscope is a non-invasive skin imaging system, which acquires white light dermatoscopic images and clinical photographs of the skin. These can be stored, retrieved, displayed and reviewed by trained medical practitioners.
Device Story
Hand-held dermatoscope (camera) with LED light source; acquires white light dermatoscopic images (contact mode) and clinical photographs (15cm distance); connects via USB to PC workstation; images/data transmitted to cloud-based software application; used in professional medical environments; enables storage, retrieval, and review of skin images/patient data by medical practitioners; supports clinical documentation and workflow; aids in visual assessment of skin lesions.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified through hardware/software validation, usability evaluation, biocompatibility testing (ISO 10993), and electrical/electromagnetic safety testing (IEC 60601).
Technological Characteristics
Hand-held camera with 5MP CMOS sensor (2592 x 1944); white light LED illumination; USB connectivity to PC; cloud-based software for storage/retrieval; contact-mode dermatoscopy and non-contact clinical photography; biocompatible nose piece/spacer (ISO 10993); IEC 60601-1/1-2 compliant.
Indications for Use
Indicated for trained medical practitioners to acquire, store, retrieve, display, and review non-invasive white light dermatoscopic images and clinical photographs of the skin.
Regulatory Classification
Identification
A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
May 17, 2023
DeepX Health LLC % Howard Schrayer Official Correspondent 8 Lookout Hilton Head Island, South Carolina 29928
Re: K230448
Trade/Device Name: DeepX DermoSight Dermatoscope Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical lamp Regulatory Class: Class II Product Code: PSN Dated: February 20, 2023 Received: February 21, 2023
Dear Howard Schrayer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Carr -S
Jessica Carr, Ph.D. Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K230488
Device Name DeepX DermoSight Dermatoscope
Indications for Use (Describe)
The DeepX DermoSight Dermatoscope is a non-invasive skin imaging system, which acquires white light dermatoscopic images and clinical photographs of the skin. These can be stored, retrieved, displayed and reviewed by trained medical practitioners.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------|
| <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------|
|-----------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Exhibit E - 510(k) Summary
| Contact: | Howard Schrayer<br>DeepX Health LLC<br>2101 L Street, NW Suite 300<br>Washington, DC 20037<br><br>Telephone: 609-273-7350<br>hs.ss@lucidmedical.net |
|-----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | February 20, 2023 |
| Device Trade Name: | DermoSight Dermatoscope |
| Manufacturer: | DeepX Health LLC<br>2101 L Street, NW Suite 300<br>Washington, DC 20037 |
| Common Name: | Light Based Imaging - Surgical Lamp |
| Classification: | Class II |
| Product Code: | PSN |
| Regulation: | 21 CFR 878.4580 |
| Predicate Devices:<br>Primary Predicate | Demetra BDEM-01<br>Surgical Lamp<br>[510(k) K192829] |
| Reference Predicate | Visiomed AG<br>MicroDERM® Model/Version 3.5<br>[510(k) K040171] |
{4}------------------------------------------------
#### Indications for Use:
The DeepX DermoSight Dermatoscope is a non-invasive skin imaging system. which acquires white light dermatoscopic images and clinical photographs of the skin. These can be stored, retrieved, displayed and reviewed by trained medical practitioners.
#### Device Description:
The DeepX Health DermoSight Dermatoscope is designed to acquire images of the skin and optimize the imaging storage and documentation workflow. The DeepX Health DermoSight Dermatoscope system consists of a hardware device and a software application. The hardware device is a hand-held, dermatoscope (camera) for acquiring and transmitting images of the skin. The hand-held, portable unit includes an LED light-source, a digital camera and a USB cable to transfer images to a standard PC workstation. The device acquires optical dermatoscopic images in a contact mode (acquired with the device in contact with the skin). In addition, the device can also acquire a clinical close-up image, when it is positioned 15cm from the skin.
The stand-alone software application is cloud-based software with a related web application. The cloud software allows for storage and retrieval of the acquired images and patient data for review by medical professionals.
DeepX Health DermoSight Dermatoscope software is a stand-alone application to be accessed via a standard computer connected to the hospital/clinic server or directly to the internet.
The software displays the patient demographic and identification information (e.g., date of image acquisition and anatomic location of photographed site). The images and identification data may be output to the computer used for image acquisition or downloaded to an alternate computer or storage device identified by the user.
### Substantial Equivalence and Predicate Devices:
The device was shown to be substantially equivalent to a previously cleared light based imaging devices, specifically including the Demetra BDEM-01 system [510(k) K192829] predicate and the Visiomed AG MicroDERM® Model/Version 3.5 [510(k) K040171].
{5}------------------------------------------------
### Performance Testing:
The following documentation was submitted in the 510(k).
Hardware Requirements Level of Concern Statement Software Description Software Architecture Specification User Manual and Instructions for Use Software Design Specification Risk Analysis Traceability Matrix Software Validation Report Usability Evaluation Report Software Development and Lifecycle Plan Unresolved Anomalies Cybersecurity Electromagnetic Compatibility and Safety Biocompatibility Data
#### Predicate Comparison Table
| Manufacturer | DeepX Health LLC | Demetra |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name | DeepX Health<br>DermoSight Dermatoscope | Demetra BDEM-01 |
| 510(k) Number | Subject Device - TBD | K192829 |
| Regulatory Class | Class II | Class II |
| Regulation | 21 CFR 878.4580 | 21 CFR 878.4580 |
| Type of Device/<br>Product Code /<br>Regulation | Light Based Imaging<br>Surgical Lamp<br>PSN<br>21 CFR 878.4580 | Light Based Imaging<br>Surgical Lamp<br>PSN<br>21 CFR 878.4580 |
| Indications for<br>Use | The DeepX DermoSight<br>Dermatoscope is a non-invasive<br>skin imaging system, which<br>acquires white light<br>dermatoscopic images and<br>clinical photographs of the skin.<br>These can be stored, retrieved,<br>displayed and reviewed by<br>trained medical practitioners. | The Barco DermoSight<br>Dermatoscope is a noninvasive<br>skin imaging<br>system, which acquires<br>multispectral and white light<br>dermatoscopic images and<br>clinical photographs of the skin.<br>These can be stored, retrieved,<br>displayed and reviewed by<br>trained medical practitioners. |
| Intended Use<br>(From Regulation) | A surgical lamp (including a<br>fixture) is a device intended to<br>be used to provide visible | A surgical lamp (including a<br>fixture) is a device intended to<br>be used to provide visible |
| | illumination of the surgical field<br>or the patient.<br>Emission and collection of light<br>to create an image for medical<br>purposes | illumination of the surgical field<br>or the patient.<br>Emission and collection of light<br>to create an image for medical<br>purposes |
| Intended Users | Trained medical practitioners | Trained medical practitioners |
| Mechanism of Action | Acquire and display of digital<br>video images | Acquire and display of digital<br>video images |
| Form of Hardware<br>Component of<br>Device | The hardware device is a<br>portable, battery powered<br>medical device for acquiring and<br>visualizing images of the skin. | The hardware device is a<br>portable, battery powered<br>medical device for acquiring and<br>visualizing images of the skin. |
| DermoSight<br>Camera<br>Resolution | 5mp 2592 x 1944<br>CMOS sensor | 8mp 3840 x 2160<br>CMOS sensor |
| Light Source | White light LEDs | White light LEDs<br>Multispectral LEDs |
| Data transfer | Camera to PC via USB cable | FCC compliant wireless |
| Software location | Cloud based | Cloud based |
| Functions | Image acquisition<br>Image transfer<br>Image and patient data storage<br>and retrieval<br>Image display and comparison | Image acquisition<br>Image transfer<br>Image and patient data storage<br>and retrieval<br>Image display and comparison |
| Use environment | Professional medical<br>environment | Professional medical<br>environment |
| Hardware<br>Location | PC inserted between image<br>source (camera) and internet<br>connection | PC inserted between image<br>source (camera) and internet<br>connection |
| Body contact | Unbroken skin | Unbroken skin |
| User interface | Proprietary software for entry<br>and retrieval of patient<br>identification, lesion location and<br>images | Proprietary software for entry<br>and retrieval of patient<br>identification, lesion location and<br>images |
| Biocompatibility | ISO 10993 for tissue contact<br>nose piece and spacer | ISO 10993 for tissue contact<br>nose piece |
| Environmental<br>compatibility | IEC 60601-1-2:2014 and EN<br>60601-1-2:2015 compliant | IEC 60601-1-2:2014 and EN<br>60601-1-2:2015 compliant |
| Electrical safety | IEC 60601 compliant | IEC 60601 compliant |
{6}------------------------------------------------
{7}------------------------------------------------
### Conclusion
The DeeX Health DermoSight Dermatoscope system is substantially equivalent to the previously cleared Demetra BDEM-01 Image Processing System with respect to intended use, general technological characteristics and performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.