K192829 · Barco N.V. · PSN · Dec 13, 2019 · General, Plastic Surgery
Device Facts
Record ID
K192829
Device Name
Demetra BDEM-01 Dermatoscope
Applicant
Barco N.V.
Product Code
PSN · General, Plastic Surgery
Decision Date
Dec 13, 2019
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4580
Device Class
Class 2
Indications for Use
The Barco Digital Dermatoscope is a non-invasive skin imaging system, which acquires multispectral and white light dermoscopic images and clinical photographs of the skin. These can be stored, retrieved, displayed and reviewed by trained medical practitioners.
Device Story
Demetra BDEM-01 is a portable, battery-powered handheld dermatoscope; captures multispectral and white light dermoscopic images, clinical close-ups (10-15 cm), and clinical overviews (35-55 cm). Device utilizes white light and multispectral LEDs (amber, red, green, cyan, blue, etc.) for image acquisition. Hardware includes an integrated AMOLED capacitive touchscreen for visualization. System integrates with cloud-based software and web applications for image storage, retrieval, and report generation. Used in professional clinical environments by medical practitioners; facilitates documentation and visualization of skin lesions. Benefits include improved imaging workflow and digital documentation of skin conditions.
Clinical Evidence
Bench testing only. Testing included IEC 62471 photobiological safety for multispectral LEDs, usability engineering for the handheld form factor, biocompatibility evaluation per ISO 10993, and review of sample images to validate acquisition performance. No clinical data provided.
Technological Characteristics
Handheld, battery-powered dermatoscope; 3840x2160 CMOS sensor; AMOLED capacitive touchscreen. Illumination: white light and multispectral LEDs (amber, deep red, green, cyan, blue, royal blue, far red, red-orange). Connectivity: wireless (FCC compliant). Biocompatibility: ISO 10993 compliant. Electrical safety: IEC 60601 compliant. Software: cloud-based with web application interface.
Indications for Use
Indicated for trained medical practitioners to acquire, store, retrieve, display, and review non-invasive multispectral and white light dermoscopic images and clinical photographs of the skin.
Regulatory Classification
Identification
A surgical lamp (including a fixture) is a device intended to be used to provide visible illumination of the surgical field or the patient.
Special Controls
*Classification.* Class II (special controls). The device, when it is an operating room lamp, a surgical instrument light, a surgical floor standing light, an endoscopic surgical light, a surgical light connector, a ceiling mounted surgical light, a surgical light carrier, surgical light accessories, a surgical lamp, a remote illuminator, or an incandescent surgical lamp, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym along with the full name of the agency on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name written out to the right of it.
December 13, 2019
Barco NV Julie Vandecandelaere Regulatory Affairs Officer President Kennedypark 35 Kortrijk, 8500 BE
Re: K192829
Trade/Device Name: Demetra BDEM-01 Regulation Number: 21 CFR 878.4580 Regulation Name: Surgical Lamp Regulatory Class: Class II Product Code: PSN Dated: September 30, 2019 Received: October 2, 2019
Dear Julie Vandecandelaere:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Purva Pandya Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K192829
Device Name Demetra BDEM-01
Indications for Use (Describe)
The Barco Digital Dermatoscope is a non-invasive skin imaging system, which acquires multispectral and white light dermoscopic images and clinical photographs of the skin. These can be stored, retrieved, displayed and reviewed by trained medical practitioners.
Type of Use (Select one or both, as applicable)
| <span style="font-size:120%">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) |
|------------------------------------------------------------------------------------|
| <span style="font-size:120%">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## K192829
| 510(k) Summary (in accordance with 21 CFR 807.92) | |
|---------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 1. Company | Barco N.V.<br>Healthcare Division<br>35 President Kennedypark<br>8500 Kortrijk<br>BELGIUM |
| 2. Contact<br>person | Julie Vandecandelaere<br>Regulatory Affairs Officer<br>Tel: +32 (0)56 26 13 19<br>julie.vandecandelaere@barco.com |
| 3. Date of<br>submission | September 30 2019 |
| 4. Device<br>information | Trade name/model: Demetra BDEM-01<br>Common name: Dermatoscope / Skin Analysis System<br>Classification name: Light Based Imaging – Surgical lamp<br>Classification code: PSN<br>Regulation number: 878.4580 |
| 5. Predicate<br>device | Microderm - K040171 |
| 6. Device<br>description | The Barco Demetra BDEM-01 is designed to capture images of the skin and optimize<br>the imaging and documentation workflow.<br><br>The “Barco Demetra BDEM-01” consists of a hardware device and a software<br>application.<br><br>The hardware device is a portable, battery powered medical device for acquiring and<br>visualizing images of the skin. The device acquires multispectral optical dermoscopic<br>images in a contact mode (device in contact with the skin). In addition, the device can<br>also acquire a clinical close-up image, when it is held at up to 10 to 15 cm from the<br>skin, and a clinical overview image when it is held at a distance of approximately 35-55<br>cm from the skin.<br><br>The stand-alone software application is cloud software with a related web application.<br>The cloud software can also generate one or more reports and feedback to the device. |
| 7. Intended Use<br>of the Device | The Barco Digital Dermatoscope is a non-invasive skin imaging system, which acquires<br>multispectral and white light dermoscopic images and clinical photographs of the skin.<br>These can be stored, retrieved, displayed and reviewed by trained medical<br>practitioners.<br><br>This indication for use statement is equivalent to the one of the predicate device. The acquisition of<br>multispectral images is not a critical difference for the use of the device and does not affect the<br>safety or effectiveness of the device. |
| 8. Comparison of | |
|------------------------|---------------------|
| Trade Name/Device Name | BDEM-01 |
| | MicroDERM (K040171) |
P 1 / 3
BAF
CO
{4}------------------------------------------------
| technological<br>characteristics | Product Code | PSN | PSN (originally filed as FSS) |
|----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Regulatory Class | II | II |
| | Indications for use | The Barco Digital<br>Dermatoscope is a non-<br>invasive skin imaging<br>system, which acquires<br>multispectral and white light<br>dermoscopic images and<br>clinical photographs of the<br>skin. These can be stored,<br>retrieved, displayed and<br>reviewed by trained medical<br>practitioners. | MicroDERM is intended for<br>the acquisition and storage<br>of images of skin surfaces.<br>These images can then be<br>retrieved, printed,<br>reviewed and displayed. |
| | Illumination for | Skin | Skin |
| | Intended users | Trained medical practitioners | Physicians |
| | Where used | Professional environment | Professional environment |
| | Digital or Analog | Digital | Digital |
| | Illumination | White light LEDs (for live<br>preview mode and<br>dermoscopic images)<br>Multispectral LEDs (for<br>dermoscopic images)<br>• Amber<br>• Deep Red<br>• Green<br>• Cyan<br>• Blue<br>• Royal Blue<br>• Far Red<br>• Red-Orange<br>(Only for acquisition and storage,<br>not for visualization in the current<br>intended use) | White light LEDs |
| | Magnification | 10-50x fold<br>(1 sensor pixel = 6.383um in<br>contact) | 15-50x fold |
| | Sensor resolution | 3840 (H) × 2160 (V) CMOS<br>sensor | 752 x 582 pixel CCD |
| | Design | Camera, software and<br>integrated AMOLED<br>capacitive touchscreen | Camera, software and<br>frame grabber card |
| | Human factors | Handheld device | Handheld device + card |
| | Net weight scope | Without non-contact cone:<br>528 g<br>With non-contact cone: 552<br>g | 2150 g |
| | Biocompatibility | Biocompatible (ISO 10993) | Unknown |
| | Data connection | Wireless communication<br>FCC compliant | Model or network card |
| | Compatibility with the<br>environment and other<br>devices | IEC 60601-1-2:2014 and EN<br>60601-1-2:2015 compliant | IEC 60601 compliant |
| | Electrical Safety | IEC 60601 compliant | IEC 60601 compliant |
| 9.<br>Performance<br>testing | The bench tests mentioned below were performed to validate the device<br>characteristics that differ from the predicate device: | | |
| | Modification to device | Test performed | Criteria |
| | Multispectral LEDs<br>illumination | Per IEC 62471 | PASS |
| | Only hand-held and<br>device with less weight | Usability Engineering | PASS |
| | Biocompatibility | Biological Evaluation | PASS |
| | Image acquisition | Review of sample images | PASS |
| 10.<br>Conclusion | Demetra BDEM-01 was found to be substantially equivalent to the predicate device, due to<br>the following reasons:<br>a) Device and predicate device have the same intended use<br>b) The technological characteristics differences from the predicate device do not affect<br>safety or effectiveness<br>c) Bench testing showed that the device has similar characteristics compared to the<br>predicate device and did not reveal new issues of safety and performance. | | |
P 2 / 3
Image /page/4/Picture/4 description: The image shows the word "BARCO" in white font on a red background. The letters are bold and sans-serif. To the right of the word is a stylized graphic of a circle with two short lines extending from either side.
{5}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.