K221678 · Sprintray, Inc. · EBI · Nov 10, 2022 · Dental
Device Facts
Record ID
K221678
Device Name
SprintRay High Impact Denture Base
Applicant
Sprintray, Inc.
Product Code
EBI · Dental
Decision Date
Nov 10, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3760
Device Class
Class 2
Indications for Use
The SprintRay High Impact Denture Base resin is a light-curable polymerizable resin intended to be used for the fabrication and repair, of full and partial removable dentures and baseplates. The material is an alternative to traditional denture base material.
Device Story
Light-curable acrylate resin for 3D printing dental appliances; used by dental professionals in clinical settings. Input: digital denture base file from optical impression system. Process: CAD/CAM system designs appliance; 3D printer fabricates base; UV light cures resin. Output: finished full/partial dentures or baseplates. Clinician verifies fit, trims, and cleans appliance before patient use. Benefits: provides alternative to traditional denture base materials; enables local, chairside manufacturing of custom prosthetics.
Clinical Evidence
Bench testing only. Testing included biocompatibility (genotoxicity, cytotoxicity, acute systemic toxicity, sensitization, irritation) and mechanical performance per ISO 20795-1 (flexural strength/modulus, water sorption/solubility, stability, residual monomers, homogeneity, surface characteristics, porosity, color stability). Results met all expected performance criteria.
Technological Characteristics
Light-curable methacrylate-based resin. Additive manufacturing (3D printing) process. Cured via UV light. Non-sterile. Complies with ISO 20795-1 for physical/mechanical properties and ISO 10993-1/ISO 7405 for biocompatibility.
Indications for Use
Indicated for the fabrication and repair of full and partial removable dentures and baseplates in patients requiring dental prosthetics.
Regulatory Classification
Identification
A denture relining, repairing, or rebasing resin is a device composed of materials such as methylmethacrylate, intended to reline a denture surface that contacts tissue, to repair a fractured denture, or to form a new denture base. This device is not available for over-the-counter (OTC) use.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the FDA logo is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
November 10, 2022
SprintRay Inc. Sara Moghtadernejad Regulatory Affairs Manager 2705 Media Center Drive, Suite 100A Los Angeles, California 90065
Re: K221678
Trade/Device Name: SprintRay High Impact Denture Base Regulation Number: 21 CFR 872.3760 Regulation Name: Denture Relining, Repairing, Or Rebasing Resin Regulatory Class: Class II Product Code: EBI Dated: October 14, 2022 Received: October 17, 2022
Dear Sara Moghtadernejad:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Michael E. Adjodha -S
Michael E. Adjodha, M.ChE. Assistant Director DHT1B: Division of Dental and ENT Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
510(k) Number (if known)
K221678
Device Name
SprintRay High Impact Denture Base
Indications for Use (Describe)
The SprintRay High Impact Denture Base resin is a light-curable polymerizable resin intended to be used for the fabrication and repair, of full and partial removable dentures and baseplates. The material is an alternative to traditional denture base material.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
O Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### 510(k) SUMMARY
## K221678
### SprintRay High Impact Denture Base
Submitter: SprintRay Inc. 2705 Media Center Drive, Suite 100A Los Angeles, CA 90065
Phone: +1 (800) 914-8004
Contact Person: Sara Moghtadernejad
Date Prepared: Nov 1, 2022
Name of Device: SprintRay High Impact Denture Base
Common or Usual Name: Denture Base, Prescription
Classification Name: Denture relining, repairing, or rebasing resin
Regulatory Class: Class II
Requlation Number: 21 CFR 872.3760
Product Code: EBI
Primary Predicate Device: Dentca Denture Base II (K162044)
Reference Devices: Dentca Denture Base (K143033)
#### Device Description
Fabrication of dental prosthetics with High Impact Denture Base resin requires computer-aided design and manufacturing system that includes the following components not part of the device: oral casting impression, digital denture base file created in an optical impression system, 3D printer, and curing light equipment.
The High Impact Denture Base consists of a curable dental acrylate resin that is designed to be used in conjunction with a scanned 3D image of a patient's teeth, and 3D printer assembly, to locally manufacture out a dental appliance in dental offices based on the clinician's judgment of patient need.
High Impact Denture Base Resin is intended exclusively for professional dental work. SprintRay High Impact Denture Base Resin is offered in following shades/colors:
- Light Pink .
- Original Pink ●
- . Light Meharry
{4}------------------------------------------------
- Original Meharry .
- . Dark Meharry
High Impact Denture Base is designed to meet appropriate ISO standards for flexibility, sorption, and solubility to withstand prolonged use in the oral cavity. It is delivered non-sterile, and instructions are provided on cleaning the material prior to providing it to a patient. Curing is performed with a UV lamp. The appliance is then cleaned, trimmed, and verified to fit in the dental office before the patient leaves.
## Intended Use / Indications for Use
The SprintRay High Impact Denture Base resin is a light-curable polymerizable resin intended to be used for the fabrication and repair, of full and partial removable dentures and baseplates. The material is an alternative to traditional denture base material.
## Summary of Technological Characteristics
Light curing of a 3D printed acrylate resin is the technological principle for both the subject and predicate devices. The High Impact Denture Base is poured into a 3D printer, which relies on scanned images of the patient's oral cavity to produce a dental appliance. At a high level, the subject and predicate devices are based on the following same technological elements:
- Are a pourable acrylate resin
- · Are used in coniunction with 3D printers, which rely on common 3D images to define the fabricated dental appliance
- · Are cured prior to final trimming and cleaning
- · Are used for the fabrication of orthodontic and dental appliances
The following technological differences exist between the subject and predicate devices:
- · Differences in acrylate resin material
{5}------------------------------------------------
## Performance Data
The following performance data were provided in support of the substantial equivalence determination.
### Biocompatibility Testing
The biocompatibility evaluation for High Impact Denture Base was conducted in accordance with the FDA Blue Book Memorandum #G95-1 and International Standard ISO 10993-1 and ISO 7405, as recognized by FDA. The battery of testing included the following tests:
- Genotoxicity ●
- Cytotoxicity ●
- Acute Systematic Toxicity ●
- Sensitization ●
- Irritation
High Impact Denture Base is considered tissue contacting for a period longer than 30 days (a removable prosthesis).
### Bench Testing
High Impact Denture Base was tested for conformity with the industry consensus standard ISO 20795-1. The battery of testing included the following tests:
- Flexural Strength and Modulus ●
- Water Sorption and Solubility ●
- Stability
- Residual Methyl Methacrylate Monomers ●
- Homogeneity ●
- Surface Characteristics ●
- Shape Capability, Translucency, and Polishability ●
- Freedom from Porosity ●
- . Color Stability
In all instances, High Impact Denture Base functioned as intended and the outcomes were as expected.
{6}------------------------------------------------
# Equivalence to Marketed Devices
| Feature | Dentca Denture Base II<br>Predicate Device | Dentca Denture Base<br>Reference Device | SprintRay High Impact<br>Denture Base | Conclusion |
|------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Product Code | EBI | EBI | EBI | Similar |
| Regulation | 21 CFR 872.3760 | 21 CFR 872.3760 | 21 CFR 872.3760 | Similar |
| Intended Use &<br>Indications for Use | Dentca Denture Base II<br>is a light-cured resin<br>indicated for fabrication<br>and repair of full and<br>partial removable<br>dentures and<br>baseplates. The<br>material is an<br>alternative to<br>traditional heat-cured<br>and auto polymerizing<br>resins.<br><br>Fabrication of dental<br>prosthetics with Dentca<br>Denture Base II requires<br>a computer-aided<br>design and<br>manufacturing<br>(CAD/CAM) system that<br>includes the following<br>components not part of<br>the device: oral casting<br>impression, digital<br>denture base file<br>created in an optical<br>impression system,<br>stereolithographic<br>additive printer, and<br>curing light equipment. | Dentca Denture Base is<br>a light-cured resin<br>indicated for fabrication<br>and repair of full and<br>partial removable<br>dentures and<br>baseplates. The material<br>is an alternative to<br>traditional heat-cured<br>and auto polymerizing<br>resins.<br><br>Fabrication of dental<br>prosthetics with Dentca<br>Denture Base requires a<br>computer-aided design<br>and manufacturing<br>(CAD/CAM) system that<br>includes the following<br>components not part of<br>the device: oral casting<br>impression, digital<br>denture base file<br>created in an optical<br>impression system,<br>stereolithographic<br>additive printer, and<br>curing light equipment. | The SprintRay High<br>Impact Denture Base<br>resin is a light-curable<br>polymerizable resin<br>intended to be used for<br>the fabrication and<br>repair, of full and partial<br>removable dentures and<br>baseplates. The material<br>is an alternative to<br>traditional denture base<br>material. | Similar |
| User Population | Clinicians in dental offices | Clinicians in dental offices | Clinicians in dental offices | Similar |
| Chemical<br>Description | Methacrylate-based resin | Methacrylate-based resin | Methacrylate-based resin | Similar |
| Material Type | Light-curable Resin | Light-curable Resin | Light-curable Resin | Similar |
| Curing Method | UV Light | UV Light | UV Light | Similar |
| Product State | Liquid | Liquid | Liquid | Similar |
| Manufacturing<br>Technology Type | Additive | Additive | Additive | Similar |
| Volume provided | 1kg bottle | 1kg bottle | 1kg bottle | Similar |
| Shelf life | >1.5 years | >1.5 years | 1.5 years | Similar |
| Standards | ISO 20795-01 | ISO 20795-01 | ISO 20795-01 | Similar |
| | | | | |
| Physical and<br>Mechanical<br>Properties | Flexural Strength and<br>Modulus | Flexural Strength and<br>Modulus | Flexural Strength and<br>Modulus | Similar |
| | Water Sorption and<br>Solubility | Water Sorption and<br>Solubility | Water Sorption and<br>Solubility | |
| | Stability | Stability | Stability | |
| | Residual Methyl<br>Methacrylate Monomers | Residual Methyl<br>Methacrylate Monomers | Residual Methyl<br>Methacrylate Monomers | |
| | Homogeneity | Homogeneity | Homogeneity | |
| | Surface Characteristics | Surface Characteristics | Surface Characteristics | |
| | Shape Capability,<br>Translucency, and<br>Polishability | Shape Capability,<br>Translucency, and<br>Polishability | Shape Capability,<br>Translucency, and<br>Polishability | |
| | Freedom from Porosity | Freedom from Porosity | Freedom from Porosity | |
| | Color Stability | Color Stability | Color Stability | |
| Biocompatibility | Tested to ISO 7405, ISO-<br>10993-1 | Tested to ISO 7405, ISO-<br>10993-1 | Tested to ISO 7405, ISO-<br>10993-1 | Similar |
| | Genotoxicity (Part 3) | Genotoxicity (Part 3) | Genotoxicity (Part 3) | |
| | Cytotoxicity (Part 5) | Cytotoxicity (Part 5) | Cytotoxicity (Part 5) | |
| | Acute Systematic Toxicity<br>(Part 11) | Acute Systematic Toxicity<br>(Part 11) | Acute Systematic Toxicity<br>(Part 11) | |
| | Sensitization (Part 10) | Sensitization (Part 10) | Sensitization (Part 10) | |
| | Irritation (Part 10) | Irritation (Part 10) | Irritation (Part 10) | |
| Additive<br>Manufacturing | Testing, according to<br>FDA's guidance<br>Technical<br>Considerations for<br>Additive Manufactured<br>Medical Devices, was<br>performed and results<br>were provided in the<br>510(k). These tests<br>included evaluation of<br>all relevant properties<br>of the printed resin<br>using the permitted<br>machines. Further,<br>tests based on<br>considerations of the<br>orientation during<br>manufacturing were | Testing, according to<br>FDA's guidance<br>Technical<br>Considerations for<br>Additive Manufactured<br>Medical Devices, was<br>performed and results<br>were provided in the<br>510(k). These tests<br>included evaluation of<br>all relevant properties<br>of the printed resin<br>using the permitted<br>machines. Further,<br>tests based on<br>considerations of the<br>orientation during<br>manufacturing were | Testing, according to<br>FDA's guidance<br>Technical<br>Considerations for<br>Additive Manufactured<br>Medical Devices, was<br>performed and results<br>were provided in the<br>510(k). These tests<br>included evaluation of<br>all relevant properties<br>of the printed resin<br>using the permitted<br>machines. Further,<br>tests based on<br>considerations of the<br>orientation during<br>manufacturing were | Similar |
| Printer Device | SprintRay Pro 95 | SprintRay Pro 95 | SprintRay Pro 95 | Similar |
| Post-Cure Device | SprintRay ProCure 2 | SprintRay ProCure 2 | SprintRay ProCure 2 | Similar |
| Sterility | Non-sterile | Non-sterile | Non-sterile | Similar |
{7}------------------------------------------------
{8}------------------------------------------------
## Conclusions
The High Impact Denture Base resin is as safe and effective as its predicate and reference devices. The High Impact Denture Base has the same intended use and indication, and similar technological characteristics, and principles of operation as its predicate and reference devices. The minor technological differences between the High Impact Denture Base and its predicate and reference devices raise no new issues of safety or effectiveness. Performance data demonstrate that the High Impact Denture Base is as safe and effective as the predicate and reference devices. Thus, the High Impact Denture Base is substantially equivalent.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.