K230397 · GM Dos Reis Industria e Comercio Ltda. · HWC · Jun 12, 2023 · Orthopedic
Device Facts
Record ID
K230397
Device Name
Cut Screw - Percutaneous Compression Screw
Applicant
GM Dos Reis Industria e Comercio Ltda.
Product Code
HWC · Orthopedic
Decision Date
Jun 12, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Cut Screw - Percutaneous Compression Screw are single use devices indicated for fixing and stabilizing the bones of the mid foot, metatarsal and phalanges of the foot using an appropriate relation bone-screw size.
Device Story
Cut Screw is a metallic bone fixation fastener used for bone synthesis in the midfoot, metatarsals, and phalanges. The device features a cannulated design (central hole) allowing passage over a guide wire, enabling surgeons to verify positioning via radiological imaging before final placement. It is designed for percutaneous insertion to provide anatomical reduction, stable fixation, and preservation of blood supply. The screw includes a hexagonal connection for a wrench. It is intended for use by qualified orthopedic surgeons in a clinical/surgical setting. The device is provided in various diameters to accommodate patient anatomy and is available either sterile (Ethylene Oxide) or non-sterile (requiring end-user cleaning and steam sterilization). By providing stable fixation, the device facilitates bone healing and stabilization.
Clinical Evidence
Bench testing only. No clinical data included. Mechanical performance validated per ASTM F543 (torsional properties, driving torque, axial pullout strength). Biocompatibility supported by reference device K182718. Pyrogenicity monitoring performed per AAMI/ANSI ST72.
Indicated for fixing and stabilizing bones of the midfoot, metatarsals, and phalanges in patients requiring orthopedic bone fixation.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the date June 12, 2023. The text is written in a clear, serif font. The date is presented in the month-day-year format. The text is black against a white background.
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size below that.
GM Dos Reis Industria e Comercio Ltda Guilherme Esteves Pontes Regulatory Affairs Analyst Avenida Pierre Simon de La Place 600 Campinas, SP 13069-320 Brazil
Re: K230397
Trade/Device Name: Cut Screw - Percutaneous Compression Screw Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth Or Threaded Metallic Bone Fixation Fastener Regulatory Class: Class II Product Code: HWC Dated: May 11, 2023 Received: May 11, 2023
Dear Guilherme Esteves Pontes:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Tejen D. Soni -S 2023.06.12 16:49:16 -04'00'
For Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230397
Device Name
Cut Screw - Percutaneous Compression Screw
Indications for Use (Describe)
Cut Screw - Percutaneous Compression Screw are single use devices indicated for fixing and stabilizing the bones of the mid foot, metatarsal and phalanges of the foot using an appropriate relation bone-screw size.
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the logo for GMREIS. The logo consists of a circular graphic on the left and the text "GMREIS" on the right. The graphic is a combination of blue and red shapes, forming a stylized circle. Below the text "GMREIS" is the phrase "Qualidade para Vida" in a smaller font.
#### 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92.
## I. Submitter:
GM Dos Reis Industria e Comercio Ltda Avenida Pierre Simon de La Place 600 Campinas, São Paulo, Brazil 13069-320 Guilherme Esteves Pontes Regulatory Affairs Analyst Telephone: +55 (19) 3765-9900 Email: guilherme.qualidade@gmreis.com.br Date prepared: February 06, 2023
#### II. Device Name:
| Trade Name: | Cut Screw - Percutaneous Compression Screw |
|----------------------|----------------------------------------------------|
| Common Name: | Screw, Fixation, Bone |
| Classification Name: | Smooth or threaded metallic bone fixation fastener |
| Device Class: | II |
| Product Codes: | HWC |
| Regulation Number: | 21 CFR 888.3040 |
#### III. Predicate Devices:
Legally marketed devices to which we are claiming "Substantial Equivalence" are the following:
Nexis® osteosynthesis compressive screws (K143229) (Primary predicate device).
Mini and Micro Fragments Reconstruction System - GMReis (GMReis, K182718) (Reference device).
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for GMREIS. The logo consists of a circular graphic on the left, with a blue outer ring and a red semi-circular shape inside. To the right of the graphic is the text "GMREIS" in large, bold, blue letters. Below the text is the phrase "Qualidade para Vida" in a smaller, lighter font.
EXPERT - Joint Fixation System - GMReis (GMReis, K200332) (Reference Device).
### IV. Device Description:
The purpose of this submission is to obtain marketing clearance for the Cut Screw -Percutaneous Compression Screw.
The Cut Screw - Percutaneous Compression Screw for bone synthesis, with a hole in the center in order to allow the passage of the guide wire, thus facilitating its placement, as the surgeon is able to verify the positioning of the screw through the radiological image of the guide wire before its placement. In addition, the Cut Screw -Percutaneous Compression Screw, provides anatomical reduction, stable fixation, and preservation of blood supply by using a percutaneous incision.
The screws have a hexagonal connection for a wrench. They are found in a variety of diameters to meet the range of anatomies of the patients, are presented in Titanium Alloy according to the standard ASTM F136, it could be provided in sterile condition sterilized by Ethylene Oxide or non-sterile condition to end user and must be cleaned, and steam sterilized before use.
The devices must only be used by qualified surgeons mastering the surgical technique, having been trained and qualified in orthopedic surgeries.
#### V. Statement of Indications for Use of the Device:
Cut Screw - Percutaneous Compression Screw are single use devices indicated for fixing and stabilizing the bones of the mid foot, metatarsal and phalanges of the foot using an appropriate relation bone-screw size.
### VI. Comparison of Technological Characteristics with The Predicate Device:
The subject device is substantially equivalent in indications and design principles to the following predicate device:
K143229 - Nexis® osteosynthesis compressive screws - Novastep
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for GMREIS. The logo consists of a circular graphic on the left and the text "GMREIS" on the right. Below the text is the phrase "Qualidade para Vida". The circular graphic is blue and red, and it has a white outline.
The subject and predicate devices have equivalent intended use and equivalent technological characteristics. The subject and predicate devices are all manufactured from identical materials and share equivalent design characteristics. The subject and predicate devices encompass equivalent physical dimensions. Any difference in the technological characteristics do not raise new issues of safety or efficacy.
Biocompatibility of the subject devices are supported by the reference device K182718. The performance of the subject device screws was compared with the specific guidance "Orthopedic Non-Spinal Metallic Bone Screws and Washers – Performance Criteria for Safety and Performance Based Pathway" and are demonstrated through mechanical testing according to ASTM F543. No clinical data were included in this submission. The subject devices could be provided sterile by Ethylene Oxide method or could be provided non-sterile to be sterilized at the end-user and have no expiration date defined. Both sterilization method was validated according to the applicable standards. Pyrogenicity monitoring according to AAMI/ANSI ST72 meeting pyrogen limit specifications.
#### VII. Performance Data:
The Non-clinical tests performed and the Recognized Consensus Standards to which the Cut Screw - Percutaneous Compression Screw have been tested are specifically the following:
- Torsional Properties of Metallic Bone Screws per ASTM F543 .
- Driving Torque of Medical Bone Screws per ASTM F543 .
- Axial Pullout Strength of Medical Bone Screws per ASTM F543 .
The test results demonstrate that the subject device presents safety and efficacy in terms of mechanical properties.
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for GMREIS. The logo consists of a circular graphic on the left and the text "GMREIS" on the right. Below the text is the phrase "Qualidade para Vida". The circular graphic is made up of blue, red, and gray shapes.
### VIII. Conclusions:
As was established in this submission, the subject Cut Screw - Percutaneous Compression Screw are substantially equivalent to the predicate devices cleared by the FDA for commercial distribution in the United States. The subject device was shown to be substantially equivalent and have the same technological characteristics, intended use, indications for use, material composition, anatomical region, multiple sizes, and basic design features compared to its predicate devices. Any differences between the Cut Screw - Percutaneous Compression Screw and the predicate devices are considered minor and do not raise different questions of safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.