K200332 · GM Dos Reis Industria e Comercio Ltda. · HTN · Oct 29, 2020 · Orthopedic
Device Facts
Record ID
K200332
Device Name
EXPERT - Joint Fixation System
Applicant
GM Dos Reis Industria e Comercio Ltda.
Product Code
HTN · Orthopedic
Decision Date
Oct 29, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EXPERT - Joint Fixation System is intended as an adjunct in fracture repair involving metaphyseal and periarticular bone fragments where screws are not indicated, and as an adjunct in external and intramedullary fixation systems involving plates and nails, with fracture braces and casting. The Mini EXPERT, EXPERT Knotless and EXPERT Knotless Dual are intended to provide fixation during the healing process following: Mini EXPERT When used for fixation of bone-to-bone, the Mini EXPERT is intended as fixation posts, distribution bridges, or for distributing suture tension over areas of ligament or tendon repair, such as: - Syndesmotic trauma, such as fixation of dorsal distal radioulnar ligament (DRUL) disruptions; - Tarsometatarsal (TMT) injury, such as fixation of foot soft tissue separations due to a Lisfranc injury (Midfoot Reconstruction); and - Carpal Metacarpal (CMC) joint arthroplasty as an adjunct in the healing process of the reconstruction of the ligament at the base of the thumb metacarpal by providing stabilization between the base of the first and second metacarpal when the trapezium has been excised due to osteoarthritis. EXPERT Knotless and EXPERT Knotless Dual Syndesmotic trauma, such as fixation of syndesmosis disruptions) in connection with Weber B and C ankle fractures.
Device Story
Joint fixation system; provides bone-to-bone fixation or suture tension distribution for ligament/tendon repair. Components: titanium alloy plates (ASTM F136), UHMWPE surgical sutures/meshes, stainless steel guidewires (ASTM F138). Used in orthopedic surgery; operated by qualified surgeons. System acts as adjunct to fracture repair or external/intramedullary fixation. Surgeon places device using compatible instruments (drill bits, guides, pliers) to stabilize bone fragments or soft tissue during healing. Benefits: provides mechanical stabilization to support tissue/bone healing process.
Clinical Evidence
No clinical data were included in this submission. Performance demonstrated through mechanical testing.
Technological Characteristics
Materials: Titanium Alloy (ASTM F136) plates, UHMWPE surgical sutures/meshes, Stainless Steel (ASTM F138) guidewires. Sterilization: Ethylene oxide (ISO 11135). Shelf life: 3 years. System includes plates, sutures, and guidewires. Mechanical fixation principle.
Indications for Use
Indicated for patients requiring bone fixation during healing of metaphyseal/periarticular fractures, ligament/tendon repairs (DRUL, Lisfranc, CMC arthroplasty), and syndesmosis disruptions associated with Weber B/C ankle fractures.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
Pedimax II - Pedicular Screw Spinal System - GM dos Reis Industria E Comercio Ltda (K180626)
Mini and Micro Fragments Reconstruction System - GMReis - GM dos Reis Industria E Comercio Ltda (K182718)
Submission Summary (Full Text)
{0}------------------------------------------------
October 29, 2020
GM dos Reis Industria e Comercio Ltda % Graziela Brum Regulatory Affairs Specialist PR Servicos Regulatorios Administrativos Ltda Rua Alice Alem Saadi, 855/2402 Ribeirao Preto, 14096-570 Brazil
Re: K200332
Trade/Device Name: EXPERT - Joint Fixation System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HTN Dated: September 29, 2020 Received: October 2, 2020
Dear Graziela Brum:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K200332
#### Indications for Use (Describe)
EXPERT - Joint Fixation System
The EXPERT - Joint Fixation System is intended as an adjunct in fracture repair involving metaphyseal and periarticular bone fragments where screws are not indicated, and as an adjunct in external and intramedullary fixation systems involving plates and nails, with fracture braces and casting.
The Mini EXPERT, EXPERT Knotless and EXPERT Knotless Dual are intended to provide fixation during the healing process following:
Mini EXPERT
When used for fixation of bone, the Mini EXPERT is intended as fixation posts, distribution bridges, or for distributing suture tension over areas of ligament or tendon repair, such as:
- Syndesmotic trauma, such as fixation of dorsal distal radioulnar ligament (DRUL) disruptions;
- Tarsometatarsal (TMT) injury, such as fixation of foot soft tissue separations due to a Lisfranc injury (Midfoot Reconstruction); and
- Carpal Metacarpal (CMC) joint arthroplasty as an adjunct in the healing process of the ligament at the base of the thumb metacarpal by providing stabilization between the first and second metacarpal when the trapezium has been excised due to osteoarthritis.
EXPERT Knotless and EXPERT Knotless Dual
Syndesmotic trauma, such as fixation of syndesmosis disruptions) in connection with Weber B and C ankle fractures.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------|---------------------------------------------|
|----------------------------------------------|---------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## 510(k) Summary
ADMINISTRATIVE INFORMATION
| Labels | Values |
|---------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Sponsor/Manufacturer Name | GM dos Reis Indústria e Comércio Ltda Avenida Pierre Simon de La Place 600 Campinas, São Paulo, Brazil 13069-320 Telephone: +55 (19) 3765-9900 |
| Prepared by | Janine Treter and Graziela Brum Passarini Regulatory Affairs PR Serviços Regulatórios Administrativos Ltda E-Mail: graziela@rapassarini.com.br Telephone +55 (16) 3421-8488 |
| Submission date | 29/Sep/2020 |
DEVICE NAME AND CLASSIFICATION
| Labels | Values |
|---------------------------|-----------------------------------------------------------------------------|
| Trade/ Proprietary Name | EXPERT - Joint Fixation System |
| Common Name | Washer, Bolt Nut |
| Classification Name | Single/multiple component metallic bone fixation appliances and accessories |
| Product Code | HTN |
| Classification Regulation | 21 CFR 888.3030, Class II |
| Review Panel | Orthopedic |
PREDICATE DEVICE INFORMATION
| Labels | Values |
|-------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Predicate Devices | K090107 - Mini TightRope - Arthrex, Inc. K061925- Mini TightRopeTM Repair Kit - Arthrex, Inc. K043248 - Arthrex TightRope™ Syndesmosis Devices, Arthrex, Inc. K111032 - CMC Cable FIX - Instratek, Inc. |
| Reference Devices | K100006 - Polyester Non-absorbable Surgical Suture, Polyblend Non- absorbable Surgical Suture, Silk Non-absorbable Surgical Suture, Polypropylene Surgical Suture, Nylon Non-absorbable Surgical Suture, Riverpoint Medical K180626 - Pedimax II - Pedicular Screw Spinal System - GM dos Reis Industria E Comercio Ltda K182718 - Mini and Micro Fragments Reconstruction System - GMReis - GM dos Reis Industria E Comercio Ltda |
{4}------------------------------------------------
K200332
#### INDICATIONS FOR USE
The EXPERT – Joint Fixation System is intended as an adjunct in fracture repair involving metaphyseal and periarticular bone fragments where screws are not indicated, and as an adjunct in external and intramedullary fixation systems involving plates and nails, with fracture braces and casting.
The Mini EXPERT, EXPERT Knotless and EXPERT Knotless Dual are intended to provide fixation during the healing process following:
#### Mini EXPERT
When used for fixation of bone-to-bone, the Mini EXPERT is intended as fixation posts, distribution bridges, or for distributing suture tension over areas of ligament or tendon repair, such as:
- Syndesmotic trauma, such as fixation of dorsal distal radioulnar ligament (DRUL) disruptions;
- Tarsometatarsal (TMT) injury, such as fixation of foot soft tissue separations due to a Lisfranc injury (Midfoot Reconstruction); and
- Carpal Metacarpal (CMC) joint arthroplasty as an adjunct in the healing process of the reconstruction of the ligament at the base of the thumb metacarpal by providing stabilization between the base of the first and second metacarpal when the trapezium has been excised due to osteoarthritis.
EXPERT Knotless and EXPERT Knotless Dual
Syndesmotic trauma, such as fixation of syndesmosis disruptions) in connection with Weber B and C ankle fractures.
## DEVICE DESCRIPTION
The EXPERT – Joint Fixation System is available in three models. Each model is composed of different medical devices (i.g. plates, surgical suture) assembled together for the purpose of use as a set. Each set is composed of plates from different sizes and shapes made of Titanium Alloy (ASTM F136) associated with non-absorbable surgical sutures and/or surgical meshes made of an Ultra-High Molecular Weight Polyethylene (UHMWPE).
The metallic components are manufactured by GM dos Reis Indústria e Comércio Ltda and the nonabsorbable Surgical Sutures and Surgical Suture Meshes are purchased in bulk from River Point Medical (K100006).
Each EXPERT model comes with a transient use guidewire accessory in the same package. The guidewire is made of Stainless Steel (ASTM F138) to aid in the EXPERT - Joint Fixation System placement.
In order to promote a correct placement of EXPERT – Joint Fixation System, GMReis has also available a range of instruments (class I exempt) to serve the surgeon such as drill bits, drill guides, plier, among others. GMReis recommends the use of these instruments in order to ensure the compatibility with the subject devices and promote the success of the procedure.
{5}------------------------------------------------
The devices must only be used by qualified surgeons mastering the surgical technique, having been trained and qualified in orthopedic surgeries.
## EQUIVALENCE TO MARKETED DEVICE
The subject device is substantially equivalent in indications and design principles to the following predicate devices:
K090107 - Mini TightRope - Arthrex, Inc. K061925- Mini TightRope™ Repair Kit - Arthrex, Inc.. K043248 - Arthrex TightRope™ Syndesmosis Devices, Arthrex, Inc. K111032 - CMC Cable FIX - Instratek, Inc.
The subject device and the predicate devices have equivalent intended use and equivalent technological characteristics. The subject and predicate devices are all manufactured from identical or equivalent materials and share equivalent design characteristics. The subject and predicate devices encompass equivalent physical dimensions and are to be sterilized by identical method. Any differnce in the technological characteristics do not raise new issues of safety or efficacy.
Biocompatibility of the titanium alloy plates are supported by the reference devices K180626 and K182718. Biocompatibility of the UHMWPE surgical sutures are supported by the reference devices K100006. Biocompatibility of the stainless steel of the guide wire are supported for the following test reports: Cytotoxicity, Sensitization, Intracutaneous Reactivity, Acute Systemic Toxicity and Materialmediated Pyrogenicity Testing according the FDA guidance entitled Use of International Standard ISO 10993-1,"Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process" dated June 16, 2016 according to the device categorization and contacting profile. Pyrogenicity monitoring according to AAMI/ANSI ST72 meeting pyrogen limit specifications.
The subject devices are provided sterile and have shelf life of 3 years. Ethylene oxide sterilization validation was performed according to ISO 11135.
The performance of the subject devices compared with the predicate devices are demonstrated through mechanical testing. No clinical data were included in this submission.
## CONCLUSION
The documentation submitted in this premarket notification demonstrates that the subject devices have comparable features and performance and, therefore, are substantially equivalent to the identified predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.