K230284 · Penumbra, Inc. · HCG · Oct 25, 2023 · Neurology
Device Facts
Record ID
K230284
Device Name
Penumbra LP Coil System
Applicant
Penumbra, Inc.
Product Code
HCG · Neurology
Decision Date
Oct 25, 2023
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5950
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Penumbra LP Coil System is indicated for the embolization of: - Intracranial aneurysms - Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae - Arterial and venous embolizations in the peripheral vasculature
Device Story
The Penumbra LP Coil System is a neurovascular embolization device used by physicians to occlude blood flow in aneurysms, arteriovenous malformations, and fistulae. The system consists of a platinum embolization coil attached to a composite delivery pusher, a radiopaque positioning marker, and a detachment handle. The physician navigates the coil through a microcatheter to the target site under fluoroscopic guidance. Once positioned, the coil is detached from the pusher using the detachment handle. The device functions by physically obstructing blood flow within the target vessel or aneurysm. It is intended for single use in clinical settings. The system is provided sterile and is compatible with standard neurointerventional techniques. By inducing thrombosis and mechanical occlusion, the device helps prevent rupture or abnormal blood flow in neurovascular and peripheral vascular structures.
Clinical Evidence
No clinical or animal studies were conducted. Substantial equivalence was demonstrated through bench-top testing, including dimensional/visual inspection, friction testing, fatigue resistance, simulated use in anatomical models, radiopacity, torsional resistance, corrosion testing, coil stiffness, and tensile testing. All tests passed. Biocompatibility was verified per ISO 10993-1.
Technological Characteristics
Materials: Platinum/Tungsten coil, adhesive, titanium, polymer, plastic/stainless steel handle. Energy: None (mechanical). Dimensions: Coil length 1-70 cm, primary diameter 0.0115-0.0135 in., secondary diameter 1-10 mm. Sterilization: Ethylene Oxide (EO) per ISO 11135, SAL ≥10^-6. Connectivity: None. Software: None.
Indications for Use
Indicated for patients requiring embolization of intracranial aneurysms, neurovascular abnormalities (arteriovenous malformations, arteriovenous fistulae), and arterial/venous embolizations in the peripheral vasculature.
Regulatory Classification
Identification
A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).
{0}------------------------------------------------
October 25, 2023
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
Penumbra, Inc. Samyukta Rangachari Regulatory Specialist II One Penumbra Place Alameda, California 94502
Re: K230284
Trade/Device Name: Penumbra LP Coil System Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular Embolization Device Regulatory Class: Class II Product Code: HCG, KRD Dated: September 26, 2023 Received: September 28, 2023
Dear Samyukta Rangachari:
We have reviewed your section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (the Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database available at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Additional information about changes that may require a new premarket notification are provided in the FDA guidance documents entitled "Deciding When to Submit a 510(k) for a Change to an Existing Device" (https://www.fda.gov/media/99812/download) and "Deciding When to Submit a 510(k) for a Software Change to an Existing Device" (https://www.fda.gov/media/99785/download).
{1}------------------------------------------------
Your device is also subject to, among other requirements, the Quality System (QS) regulation (21 CFR Part 820), which includes, but is not limited to, 21 CFR 820.30, Design controls; 21 CFR 820.90, Nonconforming product; and 21 CFR 820.100, Corrective and preventive action. Please note that regardless of whether a change requires premarket review, the QS regulation requires device manufacturers to review and approve changes to device design and production (21 CFR 820.30 and 21 CFR 820.70) and document changes and approvals in the device master record (21 CFR 820.181).
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR Part 803) for devices or postmarketing safety reporting (21 CFR Part 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR Part 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR Parts 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Naira Muradyan -S
Naira Muradyan, Ph.D. Assistant Director DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K230284
Device Name Penumbra LP Coil System
Indications for Use (Describe)
The Penumbra LP Coil System is indicated for the embolization of:
- · Intracranial aneurysms
- · Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae
- · Arterial and venous embolizations in the peripheral vasculature
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Penumbra logo. The word "Penumbra" is written in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern.
#### 1 510(k) Summary K230284
(as required by 21 CFR 807.92)
Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra, Inc. is providing the summary of Substantial Equivalence for the subject Penumbra LP Coil System (herein after referred to as LP Coil).
#### Sponsor/Applicant Name and Address 1.1
Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA
#### 1.2 Sponsor Contact Information
Samyukta Rangachari Regulatory Affairs Specialist Tel: (763) 742-9306 Fax: (510) 217-6414 E-mail: srangachari@penumbrainc.com
#### 1.3 Date of Preparation of 510(k) Summary
October 25, 2023
#### 1.4 Device Trade or Proprietary Name
Penumbra LP Coil System
#### Primary Device Classification 1.5
Regulatory Class: II Classification Panel: Neurology Classification Name: Neurovascular embolization device Regulation Number: 21 CFR 882.5950 Product Code: HCG
#### Secondary Device Classification 1.6
Regulatory Class: II Classification Panel: Cardiovascular Classification Name: Vascular embolization device Regulation Number: 21 CFR 870.3300 Product Code: KRD
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The "P" is stylized with a horizontal line running through it.
### 1.7 Predicate Device
| 510(k) Number | Device Name |
|---------------|-------------------------|
| K192955 | Penumbra LP Coil System |
#### 1.8 Comparison of Technological Characteristics with the Predicate Device
| Attributes | Predicate Device | Subject<br>Device |
|--------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|
| General | | |
| Trade Name | Penumbra LP Coil System | Penumbra LP<br>Coil System |
| 510(k) Number | K192955 | K230284 |
| Classification | Class II: HCG, KRD | SAME |
| Indications for Use | The Penumbra LP Coil System is<br>indicated for the embolization of:<br>• Intracranial aneurysms<br>• Other neurovascular abnormalities<br>such as arteriovenous malformations<br>and arteriovenous fistulae<br>• Arterial and venous embolizations<br>in the peripheral vasculature | SAME |
| Materials | | |
| Coil | Platinum/Tungsten, Adhesive,<br>Titanium, Polymer | SAME |
| Introducer Sheath | Polymer | Equivalent |
| Detachment Handle | Plastic, Stainless Steel | SAME |
| Dimensions/Shape | | |
| Coil Secondary Shape | Complex, Finish | Complex, Finish,<br>Helical, Hybrid |
| Coil Length | 1-60 cm | 1-70 cm |
| Coil Primary Diameter | 0.0135 in. | 0.0115 in.-<br>0.0135 in. |
| Coil Secondary<br>Diameter | 1-8 mm | 1-10 mm |
| Detachment Pusher<br>Length | 185 cm | SAME |
| Introducer Sheath Length | 125 cm max | 126 cm max |
| Other | | |
| Sterilization | EO | SAME |
| Sterilization Assurance<br>Level (SAL) | ≥10-6 | SAME |
| Attributes | Predicate Device | Subject<br>Device |
| Use | Single Use | SAME |
| Shelf-life (coil implant &<br>detachment pusher) | 5 Years | 1 year |
| Shelf-life (detachment<br>handle) | 5 Years | SAME |
| Device Packaging<br>Materials and Dimensions | As specified in<br>K192955 | SAME |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the Penumbra company logo. The word "Penumbra" is written in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The logo is simple and modern.
#### Device Description 1.9
The Penumbra LP Coil System is comprised of the Penumbra LP Coil, a platinum embolization coil attached to a composite delivery pusher with a radiopaque positioning marker and the Penumbra LP Coil Detachment Handle.
The coil/delivery pusher is packaged separately from the Penumbra LP Coil Detachment Handle. Penumbra LP Coil includes the following coil configurations:
- Ruby Coil LP ●
- Packing Coil LP ●
- PTC Coil LP ●
- Helical Coil LP ●
- Hybrid Coil LP ●
### 1.10 Indications For Use
The Penumbra LP Coil System is indicated for the embolization of:
- . Intracranial aneurysms
- Other neurovascular abnormalities such as arteriovenous malformations and . arteriovenous fistulae
- Arterial and venous embolizations in the peripheral vasculature .
### 1.11 Summary of Non-Clinical Data
Included in this section are summary descriptions of the testing performed on the subject Penumbra LP Coil System based on the risk analysis of changes compared to the predicate device.
### 1.11.1 Bench-top Testing
Design Verification testing was conducted to evaluate the physical and mechanical properties of the subject devices and demonstrate substantial equivalence to predicate. The following tests were performed, and all tests passed:
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the Penumbra company logo. The logo consists of the word "Penumbra" in a red, sans-serif font, followed by a red circle with a white "P" inside. The "P" is stylized with a horizontal line extending from the top of the letter to the edge of the circle. There is a registered trademark symbol in the bottom right corner of the logo.
| Attribute | Specification | Results |
|----------------------------------|------------------------------------------------------------------------------------------------------|---------|
| Dimensional/Visual<br>Inspection | Confirm the dimensions of the test units meet all<br>product specifications. | Pass |
| Friction Testing | Confirm that the friction within a microcatheter is<br>acceptable. | Pass |
| Fatigue Resistance<br>Testing | Confirm that the coil implant retains its shape after<br>being cycled into/out of the microcatheter. | Pass |
| Simulated Use Testing | Simulated use testing in an anatomical model and<br>post-detachment dimensional inspection. | Pass |
| Radiopacity Testing | Confirm fluoroscopic visibility of test units. | Pass |
| Torsional Resistance<br>Testing | Confirm torsional resistance of test units. | Pass |
| Corrosion Testing | Confirm that there is no visible corrosion after<br>testing. | Pass |
| Coil Stiffness Testing | Confirm that test units meet product specifications<br>related to coil stiffness. | Pass |
| Tensile Testing | Confirm tensile strength of coil implant and<br>Detachment Pusher joints. | Pass |
Table 1: Bench-Top Testing Summary
## 1.11.2 Biocompatibility Testing
Non-clinical testing determined the Penumbra LP Coil System to be biocompatible according to the requirements of ISO 10993-1. Some biocompatibility testing of the predicate Penumbra LP Coil System was deemed applicable considering the same materials.
The following tests were performed for the Detachment Pusher of the subject Penumbra LP Coil System:
- . Cytotoxicity
- Hemocompatibility .
The following tests were performed for the Introducer Sheath of the subject Penumbra LP Coil System:
- Cytotoxicity ●
- Sensitization ●
- Irritation ●
- Systemic Toxicity .
- Hemocompatibility ●
- Material-Mediated Pyrogenicity ●
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Penumbra" in a red, sans-serif font. To the right of the word is a red circle with a white "P" inside. The "P" is stylized with a line extending from the top of the "P" to the edge of the circle.
### 1.11.3 Sterilization Testing
The Penumbra LP Coil System was tested to be sterile using identical acceptance criteria and testing methods as the predicate device in accordance with ISO 11135 and ISO 10993-7.
### 1.12 Performance Data – Animal, Clinical
No animal or clinical studies were conducted because bench testing was determined sufficient for verification and validation purposes.
### 1.13 Summary of Substantial Equivalence
The subject Penumbra LP Coil System is substantially equivalent to the predicate device Penumbra LP Coil System. The subject device has identical intended use as the predicate device. The subject and the predicate devices differ with minor technological variations, while maintaining the same fundamental scientific technology. However, these differences do not raise different questions of safety and effectiveness.
The device testing described in the 510(k) Summary demonstrates the subject device is substantially equivalent to the predicate device with regards to operating principle, fundamental technology, and device performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.