K192955 · Penumbra, Inc. · HCG · Nov 20, 2019 · Neurology
Device Facts
Record ID
K192955
Device Name
Penumbra LP Coil System
Applicant
Penumbra, Inc.
Product Code
HCG · Neurology
Decision Date
Nov 20, 2019
Decision
SESE
Submission Type
Special
Regulation
21 CFR 882.5950
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Penumbra LP Coil System is indicated for the embolization of: - Intracranial aneurysms - Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae - Arterial and venous embolizations in the peripheral vasculature
Device Story
The Penumbra LP Coil System is a neurovascular embolization device used by physicians trained in interventional radiology. It consists of a coil implant, a detachment pusher, and a sheath, with a separate detachment handle. The device is delivered endovascularly to the target site (intracranial or peripheral vasculature). The physician navigates the coil through a microcatheter to the site of the aneurysm or vascular abnormality. Once positioned, the detachment handle is used to release the coil from the pusher, permanently occluding the vessel or aneurysm. This process facilitates embolization to treat neurovascular abnormalities or peripheral vascular conditions. The device is a line extension of the Penumbra Smart Coil System, utilizing existing delivery components.
Clinical Evidence
Bench testing only. No clinical data provided. Design verification testing included dimensional/visual inspection, friction testing, fatigue resistance, simulated use flow model testing in tortuous neurovascular anatomy, and distal system tensile testing. All tests met established specifications.
Technological Characteristics
Materials: 92% Platinum/8% Tungsten, 90% Platinum/10% Iridium, Stainless Steel, Adhesive, Titanium, Polymer. Delivery: Endovascular via microcatheter. Sterilization: Ethylene Oxide (EO) per ISO 11135 and ISO 10993-7 (SAL ≥ 10^-6). Configuration: Coil implant with detachment pusher and sheath. Dimensions: Primary diameter 0.0135 in, secondary diameter 1-8 mm, length 1-60 cm. Single-use.
Indications for Use
Indicated for embolization of intracranial aneurysms, neurovascular abnormalities (arteriovenous malformations, arteriovenous fistulae), and peripheral arterial/venous embolizations in patients requiring neurovascular or peripheral vascular intervention.
Regulatory Classification
Identification
A neurovascular embolization device is an intravascular implant intended to permanently occlude blood flow to cerebral aneurysms and cerebral ateriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in other vascular applications are also not included in this classification, see § 870.3300.
Special Controls
*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 882.1(e).
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services-USA. To the right, there is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION".
November 20, 2019
Penumbra, Inc. Anush Puvvada Regulatory Specialist III One Penumbra Place Alameda, California 94502
Re: K192955
Trade/Device Name: Penumbra LP Coil System Regulation Number: 21 CFR 882.5950 Regulation Name: Neurovascular Embolization Device Regulatory Class: Class II Product Code: HCG, KRD Dated: October 18, 2019 Received: October 21, 2019
Dear Anush Puvvada:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Xiaolin Zheng, Ph.D. Director (Acting) DHT5A: Division of Neurosurgical, Neurointerventional and Neurodiagnostic Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K192955
Device Name Penumbra LP Coil System
Indications for Use (Describe)
The Penumbra LP Coil System is indicated for the embolization of:
- · Intracranial aneurysms
- · Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae
- · Arterial and venous embolizations in the peripheral vasculature
Type of Use (Select one or both, as applicable)
| <div> <span> <span style="font-size:16px">☒</span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|--------------------------------------------------------------------------------------------------------------------|
| <div> <span> <span style="font-size:16px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "Penumbra" in a red font. To the right of the word is a red circle with a white "P" inside. The "P" is stylized with a white line extending from the top of the "P" to the left, connecting to the end of the word "Penumbra."
#### 1 510(k) Summary
(as required by 21 CFR 807.92)
Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra Inc. is providing the summary of Substantial Equivalence for the subject Penumbra LP Coil System (herein after referred to as LP Coil).
#### 1.1 Sponsor/Applicant Name and Address
Penumbra, Inc. One Penumbra Place Alameda, CA 94502 USA
#### Sponsor Contact Information 1.2
Anush Puvvada Regulatory Specialist III Tel: (510) 440-5568 Fax: (510) 217-6414 E-mail: ypuvvada@penumbrainc.com
#### 1.3 Date of Preparation of 510(k) Summarv
October 18, 2019
#### Device Trade or Proprietary Name 1.4
Penumbra LP Coil System
#### Primary Device Classification 1.5
II Regulatory Class: Classification Panel: Neurology Classification Name: Neurovascular embolization device Regulation Number: 21 CFR 882.5950 Product Code: HCG
#### 1.6 Secondary Device Classification
Regulatory Class: II Classification Panel: Cardiovascular Classification Name: Vascular embolization device Regulation Number: 21 CFR 870.3300 Product Code: KRD
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Penumbra logo. The word "Penumbra" is written in a red serif font. To the right of the word is a red circle with a white "P" inside of it.
### 1.7 Predicate and Reference Devices
| 510(k)<br>Number | Clearance Date | Name of Predicate<br>Device | Name of<br>Manufacturer |
|------------------|----------------|-----------------------------|-------------------------|
| Predicate Device | | | |
| K143218 | March 18, 2015 | | |
| K151572 | July 10, 2015 | Penumbra Smart Coil | Penumbra, Inc. |
| K160832 | April 20, 2016 | | |
| Reference Device | | | |
| K173614 | April 17, 2018 | Penumbra Coil 400 | Penumbra, Inc. |
### Predicate and Reference Device Comparison 1.8
| Attributes | Predicate Device | Reference Device | Subject Devices |
|-----------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------|----------------------|
| General | | | |
| Trade name | Penumbra Smart<br>Coil | Penumbra Coil 400 | LP Coil |
| 510(k)(s) | K143218, K151572,<br>K160832 | K173614 | K192955 |
| Classification | Class II: HCG,<br>KRD | SAME | SAME |
| Indications for Use | Indicated for the<br>embolization of:<br>• Intracranial<br>aneurysms<br>• Other<br>neurovascular<br>abnormalities<br>such as<br>arteriovenous<br>malformations<br>and arteriovenous<br>fistulae<br>• Arterial and<br>venous<br>embolizations in<br>the peripheral<br>vasculature | SAME | SAME |
| Coil Materials/Construction | | | |
| Coil | 92% Platinum/8%<br>Tungsten, 90%<br>Platinum/10%<br>Iridium, Stainless | 92% Platinum/8%<br>Tungsten,<br>Adhesive,<br>Titanium, Polymer | SAME as<br>Reference |
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Penumbra, Inc., a global healthcare company focused on innovative therapies. The logo features the word "Penumbra" in a bold, red font. To the right of the name is a circular graphic, also in red, with a stylized "P" inside, created by a white line that forms the letter and extends beyond the circle's edge.
| Attributes | Predicate Device | Reference Device | Subject Devices |
|-----------------------------------------------------|----------------------------|-----------------------------|----------------------|
| | Steel, Adhesive, | | |
| | Titanium, Polymer | | |
| Coil Dimensions/Shape | | | |
| Coil Secondary<br>Shape | Complex, Helical | Complex, Helical,<br>Finish | Complex, Finish |
| Coil Length | 1-60 cm | SAME | SAME |
| Coil Primary<br>Diameter | 0.0105-0.0130 in. | 0.020 in. | 0.0135 in. |
| Coil Secondary<br>Diameter | 1-18 mm | 2-40 mm | 1-8 mm |
| Other | | | |
| Sterilization | EO | SAME | SAME |
| Sterilization<br>assurance level<br>(SAL) | $\geq$ 10-6 | SAME | SAME |
| Use | Single Use | SAME | SAME |
| Shelf-life (coil<br>implant &<br>detachment pusher) | 5 Years | 8 Years | SAME as<br>Predicate |
| Shelf-life<br>(detachment handle) | 5 Years | 3 Years | SAME as<br>Predicate |
| Device Packaging<br>Materials and<br>Dimensions | As specified in<br>K143218 | As described in<br>K173614 | SAME as<br>Predicate |
#### Device Description 1.9
The LP Coil consists of a Coil Implant attached to a Detachment Pusher, both contained within a Sheath. The Detachment Pusher comprises a shaft with a radiopaque positioning marker, a Distal Detachment Tip (DDT), and a pull wire. A Detachment Handle (packaged separately) is used to detach the Coil Implant from the Detachment Pusher. The LP Coil is designed for endovascular embolization in the neuro and peripheral vasculature. Intended users for this device are physicians who have received appropriate training in interventional radiology. The LP Coil is a line extension to the Penumbra Smart Coil System (hereinafter referred to as Smart Coil) that uses the existing Sheath, Detachment Pusher, and Detachment Handle from the Smart Coil. This line extension includes two configurations:
- Ruby LP .
- Finish LP ●
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Penumbra, Inc., a global healthcare company focused on innovative therapies. The logo features the word "Penumbra" in a bold, red font. To the right of the name is a circular graphic, also in red, with a stylized "P" inside, connected to the wordmark by a thin line.
## 1.10 Indications For Use
The Penumbra LP Coil System is indicated for the embolization of:
- Intracranial aneurysms .
- . Other neurovascular abnormalities such as arteriovenous malformations and arteriovenous fistulae
- Arterial and venous embolizations in the peripheral vasculature .
### 1.11 Summary of Non-Clinical Data
Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, a summary of information regarding Substantial Equivalence of the device is as follows.
Included in this section are descriptions of the design control testing performed on the subject LP Coil based on a risk analysis of any technological changes compared to the predicate device. Design Verification (Bench-Top Testing and Sterilization Testing) was performed on the subject device as part of the design control activities. The subject device met all established requirements. No new biocompatibility testing was performed due to the same materials of construction between the subject and predicate devices.
## 1.11.1 Bench-top Testing
Design Verification testing was conducted to evaluate the physical and mechanical properties of the subject devices and demonstrate substantial equivalence to predicate. The following tests were performed, and all tests passed:
| Attribute | Specification | Results |
|-------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|
| Dimensional/Visual<br>Inspection | Confirm the dimensions of the units meet all product<br>specifications. | Pass |
| Friction Testing | Push/pull friction acceptable through a 0.0165 in. ID<br>microcatheter | Pass |
| Fatigue Resistance | The Coil Implant retains its secondary shape after<br>being cycled into/out of the 0.0165" ID<br>microcatheter 5 times | Pass |
| Simulated Use Flow<br>Model Testing | Simulated use testing with accessory devices in an<br>anatomical model which simulated the tortuosity of<br>the neurovasculature. Devices were delivered<br>through the model to evaluate the effectiveness of<br>the devices to embolize targeted vasculature. | Pass |
| Distal System Tensile<br>Test | Minimum per specification | Pass |
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the word "Penumbra" in a red sans-serif font. To the right of the word is a red circle with a white "P" inside. The "P" is stylized with a line extending from the top of the "P" to the left, connecting to the word "Penumbra."
# 1.11.2 Sterilization Testing
The LP Coil System was tested to be sterile using identical acceptance criteria and testing methods as the predicate device. Testing was accordance with ISO 11135 and ISO 10993-7.
# 1.12 Summary of Substantial Equivalence
The subject LP Coil is substantially equivalent to the predicate device Smart Coil. The subject device has identical intended use as the predicate device. The subject and the predicate devices differ slightly in regard to minor technological variations, while maintaining the same fundamental scientific technology. However, these differences do not raise different questions of safety and effectiveness.
The device testing described in the 510(k) Summary demonstrate the subject devices are substantially equivalent to the predicate device in regard to operating principle, fundamental technology and device performance.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.